Clinical performance study: 18 clinical cases for phase detection; 29 contrast enhanced liver studies for image review.
2 (US board certified radiologists)
Indications for Use
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: ·Proton density (PD) (also called hydrogen density) ·Spin-lattice relaxation time (T1) ·Spin-spin relaxation time (T2) ·Flow dynamics ·Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Galan 3T is a 3T MRI system for diagnostic imaging. Input: RF signals from protons in body tissues/fluids. Operation: System uses pulse sequences to acquire spatial data; V9.0 software update adds AI/ML-based reconstruction and workflow tools (PIQE, IMC, NeuroLine+, Free Breathing Dynamic). Output: Cross-sectional images (transaxial, coronal, sagittal, oblique). Used in clinical settings by radiologists/technicians. PIQE uses deep learning to generate high-matrix images from low-matrix acquisitions, reducing scan time and ringing artifacts. IMC uses ML to correct motion artifacts. NeuroLine+ automates slice positioning. Free Breathing Dynamic uses ML for arterial phase detection. Output aids physicians in diagnosis by providing high-quality anatomical images. Benefits include reduced scan times, improved image sharpness, and standardized workflow.
Clinical Evidence
Bench testing and clinical image reviews performed. PIQE evaluated via phantom and 292 clinical scans (brain/knee); 6 radiologists scored images as clinically acceptable (Likert ≥3). NeuroLine+ validated on 13 clinical cases; 96% success in auto-positioning. IMC validated on 300 image volumes (volunteers) and 34 clinical patient volumes; radiologists confirmed improved image quality/diagnostic confidence. Free Breathing Dynamic DLR validated on 18 cases (94.4% success in phase detection) and 29 clinical liver studies; met acceptance criteria for diagnostic confidence.
Technological Characteristics
3T MRI system. Materials/hardware same as predicate. Features: Deep Learning-based reconstruction (PIQE, IMC), ML-based slice positioning (NeuroLine+), and automated workflow tools. Connectivity: 10GbE high-speed reconstruction kit. Software: IEC 62304 compliant. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS 1-5.
Indications for Use
Indicated for patients requiring diagnostic MRI imaging of head or body structures. No specific age or gender contraindications provided beyond standard MRI safety protocols.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR (K222387)
Reference Devices
Orian 1.5T scanner
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
August 30, 2023
Canon Medical Systems Corporation % Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive Tustin, California 92780
Re: K230355
Trade/Device Name: Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MR Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: August 1, 2023 Received: August 2, 2023
Dear Janine Reyes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230355
#### Device Name
Vantage Galan 3T. MRT-3020. V9.0 with AiCE Reconstruction Processing Unit for MR
#### Indications for Use (Describe)
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
·Proton density (PD) (also called hydrogen density) ·Spin-lattice relaxation time (T1) ·Spin-spin relaxation time (T2) ·Flow dynamics ·Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
# 1. CLASSIFICATION and DEVICE NAME
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|-------------------------------------------------------------------------------------|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction<br>Processing Unit for MR |
| Model Number: | MRT-3020 |
# 2. SUBMITTER'S NAME
Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
# 3. OFFICIAL CORRESPONDENT
Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance Canon Medical Systems Corporation
# 4. CONTACT PERSON, U.S. AGENT and ADDRESS
## Contact Person
Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 E-mail: jfreyes@us.medical.canon
# Official Correspondent/U.S. Agent
Paul Biggins Senior Director, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-7808 Fax: (714) 730-1310 E-mail: pbiggins@us.medical.canon
{4}------------------------------------------------
- 5. MANUFACTURING SITE Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
- 6. ESTABLISHMENT REGISTRATION 9614698
- 7. DATE PREPARED February 8, 2023
## 8. DEVICE NAME
Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MR
## 9. TRADE NAME
Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MR
## 10. Classification NAME
Magnetic Resonance Diagnostic Device (MRDD)
## 11. CLASSIFICATION PANEL
Radiology
## 12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
## 13. PRODUCT CODE
90-LNH
## 14. PREDICATE DEVICE
Predicate Device: Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR (K222387)
| System | Predicate Device |
|----------------------------------------------------------------------------------|---------------------------------|
| Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR | |
| Marketed By | Canon Medical Systems USA, Inc. |
| 510(k) Number | K222387 |
| Clearance Date | August 31, 2022 |
## 15. REASON FOR SUBMISSION
Modification of a cleared device
#### 16. SUBMISSION TYPE
Traditional 510(k) Premarket Notification
{5}------------------------------------------------
## 17. DEVICE DESCRIPTION
The Vantage Galan (Model MRT-3020) is a 3 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K222387. This system is based upon the technology and materials of previously marketed Canon Medical Systems and is intended to acquire and display crosssectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body.
## 18. SUMMARY OF CHANGE(S)
This submission is to report the following changes:
## Summary of Software Changes:
- Exsper: Exsper has been updated to support 3D encoding sequences.
- Slice shimming: When performing the multi-slice image acquisition, parameters for magnetic field homogeneity are corrected per slice.
- Iterative Motion Correction (IMC): IMC has been updated to utilize Deep Learning based methods in addition to traditional model-based methods.
- . Free Breathing Dynamic: Free Breathing Dynamic is contrast enhanced dynamic imaging technique. Utilizing Deep Learning reconstruction, Free Breathing Dynamic enables a single continuous scan to aid patients who have difficulty holding their breath.
- UTE imaging: UTE imaging has been updated to support CG-Recon to reduce the scan time while maintaining resolution and SNR.
- Precise IQ Engine (PIQE): PIQE is Deep Learning based technique that generates higher inplane matrix images from low matrix images while mitigating the ringing artifact. PIQE is targeted for brain and knee regions.
- Auto Protocol Brain Application: Auto Protocol Brain Application allows to automatically proceed with examination according to a pre-defined scenario.
- . Auto Consult Brain Package: Auto Consult performs the examinations for brain region all the way from patient setting down to the scan and check.
- . NeuroLine+: NeuroLine+ has been updated to utilize Machine Learning based method.
- . Auto Scan Assist: Auto Scan Assist takes away the variability and helps operators improve workflow with automatic standardizing workflow with automatic positioning. Planning of re-scan which is based on the result of image analysis are also available.
- DSD filter: DSD filter removes image noise while retaining the optimal smoothness and sharpness.
- Ringing correction: Ringing Correction detects areas in the image that contain truncation artifacts and then reduces or eliminates the artifacts by applying minute subpixel shifting to the pixels in those areas of the image. Ringing correction is targeted for head imaging.
## Summary of Accessory Changes:
- . Tablet UX: Tablet UX has been updated so that display movie from patient camera can be displayed on tablet, voice communication with a patient to be available and scan control to be possible.
- 10GbE High-speed reconstruction kit: This kit increases the communication speed of the AiCE reconstruction processing unit for MR.
{6}------------------------------------------------
- Ceiling camera: Position calculation method for some body regions have been updated. ●
## 19. SAFETY PARAMETERS
| Item | Subject Device:<br>Vantage Galan 3T, MRT-3020, V9.0<br>with AiCE Reconstruction<br>Processing Unit for MR | Predicate Device:<br>Vantage Galan 3T, MRT-3020, V8.0<br>with AiCE Reconstruction<br>Processing Unit for MR (K222387) | Notes |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------|
| Static field strength | 3T | 3T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter<br>display | SAR, dB/dt | SAR, dB/dt | Same |
| ii. Operating mode<br>access requirements | Allows screen access to 1st level<br>operating mode | Allows screen access to 1st level<br>operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015) | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015) | Same |
| Maximum dB/dt | 1st operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015 | 1st operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015 | Same |
| Potential emergency<br>condition and means<br>provided for shutdown | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Same |
#### 20. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K222387.
## 21. INDICATIONS FOR USE
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- . Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
*Note: No change from the previous predicate submission, K222387.
{7}------------------------------------------------
*Made For life*
## 22. SUMMARY OF DESIGN CONTROL ACTIVITIES
Risk Management activities for this modification are included in this submission. The test methods used are the same as those submitted in the previously cleared submission of the predicate device, Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR (K222387). A declaration of conformity with design controls is included in this submission.
#### 23. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):
#### LIST OF APPLICABLE STANDARDS
- ANSI AAMI ES60601-1:2005/(R)2012 ● and A1:2012
- IEC60601-1-2 (2020) ●
- IEC60601-1-6 (2010), Amd.1 (2013), ● Amd.2 (2020)
- IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015)
- IEC60825-1 (2014)
- IEC62304 (2006), Amd.1 (2015) ●
- IEC62366-1 (2020)
- NEMA MS 1 (2008)
- NEMA MS 2 (2008)
- NEMA MS 3 (2008) ●
- NEMA MS 4 (2010)
- NEMA MS 5 (2010) ●
#### 24. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrate that the system requirements have been met. Additionally, image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria.
Exsper 3D was evaluated utilizing phantom images. Testing confirmed that Exsper 3D reduced artifacts caused by unfolding error compared to conventional SPEEDER.
Slice Shim was evaluated using clinical images. Testing confirmed that images with the Slice Shim were equal to or better than those with the Standard Shim especially for off-center slices.
UTE (Ultra Short TE) CG Recon was evaluated utilizing both phantom and clinical images. Testing confirmed CG recon performs better at maintaining both image resolution and image SNR as compared to conventional grid recon when scan time is reduced.
DSD (Dynamic Shrinkage Denoise) Filter was evaluated utilizing phantom images. Testing confirmed, when compared with the conventional filter, the DSD filter has the equivalent low-contrast imaging capability and has greater denoising effect under high spatial resolution imaging conditions in which SNR decreases.
{8}------------------------------------------------
Ringing Correction was evaluated utilizing both phantom and clinical images. Testing confirmed Ringing Correction reduced ringing and met predetermined acceptance criteria.
Auto Scan Assist was evaluated utilizing clinical images. Testing confirmed the operation of slice positioning utilizing Auto Scan Assist applications NeuroLine+, ડીંક્ટિ/O/ Liver, LiverLine+, and W-SpineLine+ resulted in less time and less steps as compared to slice positioning without Auto Scan Assist.
Ceiling Camera patient orientation and patient anatomy position detection method was evaluated using clinical images. Testing confirmed the percentage of successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria. Additionally, testing confirmed the ceiling camera resulted in less or comparable patient setting time compared to conventional manual patient setting, regardless of the operator.
PIQE (Precise IQ Engine) for MR underwent performance (bench testing) using ACR phantom images. Using metrics of SNR and signal intensity profiles for ringing and sharpness, testing confirmed PIQE generates higher in-plane matrix from lower matrix image and PIQE contributes to ringing artifact reduction and increase of sharpness. The bench testing also included evaluation in typical clinical images of brain and knee, the metrics of Edge Slope Width (to evaluate image sharpness), Ringing Variable Mean (to evaluate ringing artifacts), Signal-to-Noise ratio, and Contrast Change Ratio. Comparing images with PIQE at various scaling factors to standard clinical techniques such as zeropadding interpolation and GA filter, confirmed that PIQE generates images with sharper edges while mitigating the smoothing and ringing effects and maintaining similar or better contrast and SNR.
Additionally, a randomized, blinical image review study was conducted with 6 USA board certified radiologists (3 per anatomy). Using the conventional method (i.e., matrix expansion with Fine Reconstruction and processing with GA Filter) as a reference, the images reconstructed with either the conventional method or the new method PIQE were randomized, blinded to the reviewers, and scored by the 3 reviewers per anatomy in various clinically-relevant categories (including ringing, sharpness, SNR, overall IQ, and feature conspicuity) using a modified 5-point Likert scale, where 3 or above is considered clinically acceptable. A total of 17 unique subjects, from USA, Europe and Japan, were scanned in brain or knee to provide the test data sets separately from the training data sets using Vantage Galan 3T and anatomy-appropriate coils, comprising a total of 292 scans (neuro and knee), in multiple orientations (axial, sagittal and coronal), and multiple contrast weightings (T1-/T2-/PDweighted with/without Fat saturation) within the FSE2D family of pulse sequences. The resulting reconstructions (conventional and new) were all scored at, or above, clinically acceptable by three board-certified Radiologists per anatomy. The reviewers exhibited a strong agreement at the "good" and "very good" level for all IQ metrics such as SNR, image ringing, overall IQ and feature conspicuity. In conclusion, Testing confirmed (a) PIQE generates higher spatial in-plane resolution images from lower resolution images (with the ability to triple the matrix dimensions in both in-plane directions, i.e. a factor of 9x), (b) PIQE contributes to ringing artifact reduction, denoising and increased sharpness, (c) PIQE is able to accelerate scanning by reducing the acquisition matrix only, while maintaining clinical matrix size and image quality, (d) PIQE benefits can be obtained on regular clinical protocols without requiring acquisition parameter adjustment.
{9}------------------------------------------------
NeuroLine+ underwent performance (bench) testing using clinical images. Testing of NeuroLine+ was conducted with an independent group of patients, newly collected and entirely separate from the training group. The data, comprising 13 clinical cases (1 male, 12 female, height = 165±7, weight = 68±13, BMI = 25±4, Age = 60±14, national identities include France), were acquired with the typical localizer imaging condition of Fast Field Echo 3D sequence (TR/TE = 4.0/1.4 [ms], Matrix = 128 x 128 x 128, Spatial resolution = 2.5 [mm] isotropic) on the Vantage Galan 3T System. The acceptance criteria was a successful scan alignment (i.e., offset and angle within acceptable error defined as typical interrater variability) greater than 80% of the time. To assess the accuracy of the auto-detected angle or center position of the target planes, the angle and position of the target planes were manually annotated by two experienced ARRT licensed MR technologists. For the angular error, NeuroLine+ met the acceptance criteria being similar or better as compared to the conventional method. For the autopositioning, NeuroLine+ yielded 96.0% success which met the acceptance criteria. The results support the conclusion that NeuroLine+ is a clinically acceptable option for the slice-alignment procedure in head examination.
lterative Motion Correction (IMC) underwent performance (bench testing) using clinical datasets (12) without subject motion and with mathematically simulated motion added. Using the metrics of peak SNR and structural similarity (SSIM), testing demonstrated that IMC is effective in reducing motion artifacts and met predetermined acceptance criteria.
With scanning protocols typically employed clinically for brain and cervical spine, 21 volunteers were imaged at 3T in either the 16channel or 32channel coil with various orientations and weightings using the FSE 2D family of sequences. Two acquisitions of each image type and orientation were acquired, one with subject motionless, the other with subject moving as instructed. From the two acquisitions, three reconstructions were performed. The motion free acquisition was reconstructed using typical clinical routine, and the image acquired during the subject motion was reconstructed in two ways: once without IMC applied and again with IMC applied. In total, there were 300 image volumes (100 per group). The image review using the randomized and blinded 300 image volumes was performed. Three US board certified radiologists, specializing in neuro imaging, read and scored the images in various clinically-relevant categories (such as SNR, tissue contrast, image sharpness, and diagnostic confidence) based on a pre-determined 5-point modified Likert scale, where a value of 3 or greater was considered clinically acceptable. Testing confirmed the IMC technique performs as expected, significantly reducing motion artifacts, and improving overall image quality metrics as evaluated via SNR, tissue contrast, image sharpness, and diagnostic confidence. IMC corrected images are the same as, or better than, images without IMC applied.
Another clinical image review was performed by the three US board certified radiologists using an additional 34 image volumes in brain or cervical spine acquired, at 3T, from typical clinical patients with pathology and motion. This study further confirmed the diagnostic information in IMC images was the same or better than those without IMC applied. All testing data were acquired separately and independently from the training data after the machine learning training was completed.
Free Breathing Dynamic DLR. The automatic arterial phase detection feature of Free Breathing Dynamic DLR underwent performance (bench) testing using clinical images from patients receiving clinicallyindicated contrast. Firstly, testing of the arterial phase detection was conducted with 18 clinical cases
{10}------------------------------------------------
(9 male, 9 female, age = 64±13, national identities include Japan), and yielded 94.4% success (i.e., the automatically proposed phases included the gold standard phase as manually selected by experienced radiologists) which met the acceptance criteria (greater than or equal to 80%). Additionally, a clinical image review was conducted with 2 US board certified radiologists who read and scored a total of 29 (14 male, 14 female, 1 unknown, age = 63±12, national identities include United States, France, and Japan) contrast enhanced Free Breathing Dynamic liver studies (50 slices per study, repeated for 23 dynamic phases) acquired using either Galan 3T or Orian 1.5T scanner. Based on a modified 5-point Likert scale in categories of diagnostic importance, a score of 3 or higher was considered clinically acceptable. The average of visual scores for overall IQ, feature conspicuity and diagnostic confidence met the acceptance criteria. All testing data were acquired separately and independently from the training data after the machine learning training was completed. The results support the conclusion that Free Breathing Dynamic is a clinically acceptable option for the acquisition of freebreathing contrast enhanced dynamic liver exams providing acceptable diagnostic confidence.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
## 25. SUBSTANTIAL EQUIVALENCE
Canon Medical Systems Corporation believes that the Vantage Galan 3T, MRT-3020, V9.0, Magnetic Resonance Imaging (MRI) System with AiCE Reconstruction Processing Unit for MR, is substantially equivalent to the previously cleared predicate device referenced in this submission.
Canon Medical Systems Corporation believes that the changes incorporated into the Vantage Galan 3T, MRT-3020, V9.0, Magnetic Resonance Imaging (MRI) System with AiCE Reconstruction Processing Unit for MR, are substantially equivalent to the previously cleared predicate device.
# 26. CONCLUSION
The modifications incorporated into the Vantage Galan 3T, MRT-3020, V9.0, Magnetic Resonance Imaging (MRI) System with AiCE Reconstruction Processing Unit for MR, do not change the indications for use or the intended use of the device. Based upon bench testing, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.