LiverMultiScan v5 (LMSv5) is indicated for use as a magnetic resonance diagnostic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times. LMSv5 is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced. LMSv5 provides several tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessment of selected regions includes the determination of triglyceride fat fraction in the liver (PDFF), T2*, LIC (Liver Iron Concentration) and iron corrected T1 (cT1) measurements. These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
Device Story
LMSv5 is a standalone post-processing software device for liver evaluation. It ingests DICOM 3.0 MR image datasets from compatible scanners. The device uses machine-learning-based automated liver segmentation and artifact detection to assist operators in placing ROIs on parametric maps. It calculates quantitative metrics: PDFF, T2*, LIC, and cT1. The software generates a summary report for interpretation by a clinician. Used at Perspectum’s image analysis center by trained internal operators. The output assists clinicians in diagnosing and monitoring liver health. It does not replace clinical procedures or provide automated diagnoses; it serves as an aid to clinical decision-making.
Clinical Evidence
Bench testing only. Performance validated via functional verification, phantom testing (accuracy/precision across physiological ranges), and in-vivo volunteer data to assess measurement precision across supported scanners. Inter- and intra-operator variability was evaluated. No clinical trials were required as the device is a post-processing software tool.
Technological Characteristics
Standalone software application; DICOM 3.0 compliant; platform-independent (installed on general-purpose workstations). Employs machine-learning for automated liver segmentation and artifact detection. Quantitative methods include IDEAL (for PDFF) and MOST (for LIC). No energy delivery or patient monitoring. Software level of concern: Moderate.
Indications for Use
Indicated for patients suitable for MRI who are not contraindicated. Used for noninvasive liver evaluation, including quantification of PDFF, T2*, LIC, and cT1 measurements to assist clinicians in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 6, 2022
Perspectum Ltd. % Ioan Wigley Head of Regulatory Affairs Gemini One 5520 John Smith Drive Oxford, Oxfordshire OX 2LL UNITED KINGDOM
Re: K213960
Trade/Device Name: LiverMultiScan v5 (LMSv5) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: July 26, 2022 Received: July 26, 2022
Dear Ioan Wigley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K213960
Device Name LiverMultiScan v5
#### Indications for Use (Describe)
LiverMultiScan v5 (LMSv5) is indicated for use as a magnetic resonance diagnostic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.
LMSv5 is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.
LMSv5 provides several tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessment of selected regions includes the determination of triglyceride fat fraction in the liver (PDFF), T2*, LIC (Liver Iron Concentration) and iron corrected T1 (cT1) measurements.
These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Dect
Date Prepared:
01 September 2022
K213960
## Submitter Details
Owner Address:
Perspectum Ltd Gemini One, 5520 John Smith Drive, Oxford Business Park, Oxford, OX4 2LL +44 (0) 1865 655329
| Owner/Operator Number: | 10056574 |
|------------------------------------|------------------------------------------------------------------|
| Establishment Registration Number: | 3014232555 |
| Contact Person: | loan Wigley<br>loan.Wigley@perspectum.com<br>+44 (0) 1865 655329 |
## Subject and Predicate Devices
| | Subject Device | Primary Predicate Device | Secondary Predicate Device |
|----------------------|-----------------------------------------------|-----------------------------------------------|-----------------------------------------------|
| 510(k) number | Not known | K202170 | K043271 |
| Legal Manufacturer | Perspectum Ltd. | Perspectum Ltd. | Inner Vision Biometrics Pty |
| Owner/Owner Operator | 10056574 | 10056574 | / |
| Device Name | LiverMultiScan v5.0.0 | LiverMultiScan v4.0.0 | R2-MRI Analysis System<br>(FerriScan) |
| Proprietary/Common | LiverMultiScan, LMSv5, LMS | LiverMultiScan, LMSv4, LMS | FerriScan |
| Panel | Radiology | Radiology | Radiology |
| Regulation | 892.1000 | 892.1000 | 892.1000 |
| Risk Class | Class II | Class II | Class II |
| Product Class code | LNH | LNH | LNH |
| Classification | System, Nuclear Magnetic<br>Resonance Imaging | System, Nuclear Magnetic<br>Resonance Imaging | System, Nuclear Magnetic<br>Resonance Imaging |
{5}------------------------------------------------
# Subject Device
#### General Description
LiverMultiScan v5 (LMSv5) is a standalone software device. The LiverMultiScan v5 device is to assist a trained operator with the evaluation of information from Magnetic Resonance (MR) images from a single time-point (a patient visit).
LiverMultiScan is a post-processing software device, a trained operator uses tools within the device interface to quantify liver tissue characteristics from parametric maps. LiverMultiScan v5 includes automatic processing functionality based on machine-learning to assist in the quantification of metrics during analysis, such as automatic artefact detection and automatic segmentation of the liver. A summary report from the analysis conducted is generated for interpretation by a clinician.
LiverMultiScan v5 is not intended to replace the established procedures for the assessment of a patient's liver health by a clinician, providing many opportunities for competent human intervention in the clinical care of patients.
The metrics are intended to be used as an additional diagnostic input to provide information to clinicians as part of a wider diagnostic process. It is expected that in the normal course of clinical care, patients with clinical symptoms or risk factors which may indicate liver disease. The interpreting clinician needs to take into consideration the device's limitations and accuracy during clinical interpretation.
Information gathered through existing diagnostic tests and clinical evaluation of the patient, as well as information obtained from LiverMultiScan v5 metrics, may contribute to a diagnostic decision.
LiverMultiScan v5 is not a computer-aided diagnostic device and can only present imaging information which must be interpreted by a qualified clinician. LiverMultiScan v5 is an aid to diagnosis and treatment decisions remain the responsibility of the clinician.
#### Intended Use and Indications for Use
LiverMultiScan v5 (LMSv5) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.
LMSv5 is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR Systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.
LMSv5 provides several tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessment of selected regions includes the determination of triglyceride fat fraction in the liver (PDFF), T2*, LlC (Liver Iron Concentration) and iron corrected T1 (cT1) measurements.
These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis.
#### Indications and contraindications
LiverMultiScan is indicated for use where MRI is not contraindicated.
#### Contraindications
LiverMultiScan is indicated for use where MRI is not contraindicated.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Perspectum" in a dark gray sans-serif font. To the right of the word is a circular logo divided into four colored sections: yellow, blue, green, and pink. The logo has a small "R" in a circle in the upper right corner.
#### Intended Conditions
LiverMultiScan is not intended to be used for use on any specific disease or condition, but the information provided in the report, when interpreted by a licensed physician, may benefit the clinical management, including diagnosis and monitoring of patients.
#### Standalone Software
LiverMultiScan is a post-processing software device. All operations are directly controlled by the LiverMultiScan device. LiverMultiScan does not control other firmware or software outside of the device
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Perspectum" in a dark gray sans-serif font. To the right of the word is a circular logo that is divided into four colored sections: yellow, blue, green, and pink. The logo also has a small "registered trademark" symbol in the upper right corner.
### Subject and Predicate Comparison
The following characteristics were compared between the predicate device in order to demonstrate substantial equivalence.
| Characteristi<br>c | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | LiverMultiScan v5 (LMSv5) is<br>indicated for use as a magnetic<br>resonance diagnostic device software<br>application for non-invasive liver<br>evaluation that enables the<br>generation, display and review of 2D<br>magnetic resonance medical image<br>data and pixel maps for MR relaxation<br>times.<br><br>LMSv5 is designed to utilize DICOM<br>3.0 compliant magnetic resonance<br>image datasets, acquired from<br>compatible MR Systems, to display<br>the internal structure of the abdomen<br>including the liver. Other physical<br>parameters derived from the images<br>may also be produced. | LiverMultiScan (LMSv4) is indicated<br>for use as a magnetic resonance<br>diagnostic device software<br>application for non-invasive liver<br>evaluation that enables the<br>generation, display and review of 2D<br>magnetic resonance medical image<br>data and pixel maps for MR relaxation<br>times.<br><br>LiverMultiScan (LMSv4) is designed to<br>utilize DICOM 3.0 compliant magnetic<br>resonance image datasets, acquired<br>from compatible MR systems, to<br>display the internal structure of the<br>abdomen including the liver. Other<br>physical parameters derived from the<br>images may also be produced. | For the analysis of multi-<br>slice, spin-echo MRI data<br>sets of the liver for the<br>measurement of liver R2<br>and liver iron concentration. |
| Characteristi<br>c | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
| | LMSv5 provides several tools, such as<br>automated liver segmentation and<br>region of interest (ROI) placements,<br>to be used for the assessment of<br>selected regions of an image.<br>Quantitative assessment of selected<br>regions includes the determination of<br>triglyceride fat fraction in the liver<br>(PDFF), T2*, LIC (Liver Iron<br>Concentration) and iron corrected T1<br>(cT1) measurements. | LiverMultiScan (LMSv4) provides a<br>number of tools, such as automated<br>liver segmentation and region of<br>interest (ROI) placements, to be used<br>for the assessment of selected<br>regions of an image. Quantitative<br>assessments of selected regions<br>include the determination of<br>triglyceride fat fraction in the liver<br>(PDFF), T2* and iron-corrected T1<br>(cT1) measurements. T2* may be<br>optionally computed using the DIXON<br>or LMS MOST methods. | |
| | These images and the physical<br>parameters derived from the images,<br>when interpreted by a trained<br>clinician, yield information that may<br>assist in diagnosis. | These images and the physical<br>parameters derived from the images,<br>when interpreted by a trained<br>clinician, yield information that may<br>assist in diagnosis. | |
| Indications<br>for Use | Same as intended use | Same as intended use | The R2-MRI Analysis System<br>is an accessory diagnostic<br>device to MRI scanners and<br>is intended for diagnostic<br>use to present images that<br>reflect the magnetic<br>resonance spectra for the<br>determination of iron on the<br>liver. |
| Characteristi<br>c | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
| Target<br>Population | Patients suitable to undergo an MRI<br>scan and not contra-indicated for MRI | Patients suitable to undergo an MRI<br>scan and not contra-indicated for MRI | Patients suitable to undergo<br>an MRI scan and not contra-<br>indicated for MRI |
| Device User | Trained Perspectum internal operator | Trained Perspectum internal operator | Resonance Health's trained<br>analyst |
| Report User | An interpreting clinician or healthcare<br>practitioner | An interpreting clinician or healthcare<br>practitioner | An interpreting clinician or<br>healthcare practitioner |
| Device Use<br>Environment | Installation of LiverMultiScan v5 is<br>controlled and is installed on general<br>purpose workstations that meet the<br>minimum technical requirements at<br>Perspectum's image analysis centre<br>by specialist members of staff. | Installation of LMSv4 is controlled and<br>is installed on general purpose<br>workstations that meet the minimum<br>technical requirements at<br>Perspectum's image analysis centre<br>by specialist members of staff. | Image analysis and LIC<br>reporting is performed at a<br>central ISO 13485 certified<br>Service Centre. Hosting<br>platform is resonance<br>health's internal server. |
| Clinical<br>Setting | LiverMultiScan v5 is a standalone<br>post-processing software device that<br>is intended to be installed on general<br>use workstations at Perspectum's<br>image analysis centre.<br>Operators use LMSv5 to conduct<br>quantitative analysis of liver tissue<br>characteristics to produce a report.<br>The end-users for the output from<br>the device, the report, are clinicians<br>who receive and interpret<br>LiverMultiScan (LMSv5) reports. | LMSv4 is a standalone post-<br>processing software device that is<br>intended to be installed on general<br>use workstations at Perspectum's<br>image analysis centre.<br>Operators use LMS to conduct<br>quantitative analysis of liver tissue<br>characteristics to produce a report.<br>The end-users for the output from<br>the device, the report, are clinicians<br>who receive and interpret LMSv4<br>reports. | FerriScan is a standalone<br>software tool that is<br>intended to be used by<br>Resonance Health's trained<br>analyst, FerriScan is installed<br>and runs on the Resonance<br>Health's internal server.<br>The report result is overseen<br>by the radiologist and the<br>final decision for clinical<br>management of the patient<br>is made by their treating<br>clinician. |
| Anatomical<br>Location | Abdomen, including the liver | Abdomen, including the Liver | Abdomen, including the<br>Liver. |
| Characteristi<br>c | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
| Energy | LiverMultiScan v5 is a standalone<br>software application, it does not<br>deliver, monitor or depend on energy<br>delivered to or from patients. | LMS is a standalone software<br>application, it does not deliver,<br>monitor or depend on energy<br>delivered to or from patients. | FerriScan is a standalone<br>software application. It does<br>not deliver, monitor or<br>depend on energy delivered<br>to or from patients. |
| Design:<br>Purpose | LiverMultiScan v5 is a standalone<br>software application that imports MR<br>data sets encompassing the<br>abdomen, including the liver.<br>Visualisation and display of 2D multi-slice, spin-echo MR data can be<br>analysed and quantitative metrics of<br>tissue characteristics are then<br>reported.<br><br>Datasets imported into<br>LiverMultiScan (LMSv5) are DICOM<br>3.0 compliant, reported metrics are<br>independent of the MRI equipment<br>vendor. | LMS is a standalone software<br>application that imports MR data sets<br>encompassing the abdomen,<br>including the liver. Visualisation and<br>display of 2D multi-slice, spin-echo<br>MR data can be analysed and<br>quantitative metrics of tissue<br>characteristics are then reported.<br><br>Datasets imported into LMS are<br>DICOM 3.0 compliant, reported<br>metrics are independent of the MRI<br>equipment vendor. | FerriScan is intended for:<br>• Supporting clinical<br>diagnoses about the<br>status of liver iron<br>concentration.<br>• Supporting the<br>subsequent clinical<br>decision-making<br>processes.<br>• Supporting the use in<br>clinical research trials,<br>directed at studying<br>changes in liver iron<br>concentration as a<br>result of interventions.<br>• It contains an image<br>viewer for importing<br>DICOM images,<br>browsing through<br>patient datasets,<br>viewing images and<br>performing region of<br>interest analysis |
| Characteristi | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System |
| c | | | (FerriScan) (Secondary<br>Predicate) |
| Design: Tools | Allows for the visualisation via<br>parametric maps and quantification<br>of metrics (cT1, LIC and PDFF) from<br>liver tissue and exportation of results<br>& images to a deliverable report. | Allows for the visualisation via<br>parametric maps and quantification<br>of metrics (cT1, T2* and PDFF) from<br>liver tissue and exportation of results<br>& images to a deliverable report. | Allows for the visualisation<br>of images that reflect the<br>magnetic resonance spectra<br>for the determination of<br>iron on the liver. It contains |
| | Quantification is through either full<br>segmentation of the outer liver<br>contour and liver vasculature or<br>manual placement of ROI's on the<br>parametric maps. IQR and median<br>metrics are reported from the<br>segmentation/ROI quantification.<br>automatically detects artefacts on the<br>and are delineated on the parametric<br>map computed images as well as<br>recommends slices to be used as the | Quantification is through either full<br>segmentation of the outer liver<br>contour and liver vasculature or<br>manual placement of ROI's on the<br>parametric maps. IQR and median<br>metrics are reported from the<br>segmentation/ROI quantification. | an image viewer for<br>• importing DICOM<br>images,<br>• browsing through<br>patient datasets,<br>• viewing images and<br>performing ROI analysis. |
| Design: MR<br>Relaxometry | quantitative output of the device.<br>T1, iron corrected T1 (cT1) mapping<br>and LIC | T1, iron corrected T1 (cT1) and T2*<br>mapping | R2 (Signal Decay Rate)<br>mapping |
| Design: Liver<br>Fat<br>Quantificatio<br>n | Utilizes MR images that exploit the<br>difference in resonance frequencies<br>between hydrogen nuclei in water<br>and triglyceride fat using the<br>LiverMultiScan v5 IDEAL method. | Utilizes MR images that exploit the<br>difference in resonance frequencies<br>between hydrogen nuclei in water<br>and triglyceride fat using the LMS<br>IDEAL method. | N/A |
| Design:<br>Parametric<br>Maps | Iron corrected T1 (cT1), Proton<br>Density Fat Fraction (PDFF) and LIC<br>parametric maps can be created from<br>all supported scanners. | Iron corrected T1 (cT1), T2* and<br>Proton Density Fat Fraction (PDFF)<br>parametric maps can be created from<br>all supported scanners. | FerriScan calculates the<br>signal decay rate (R2) that is<br>used to characterise iron<br>loading in the liver. It<br>produces an output report |
| Characteristic | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
| | It is possible to use the LIC maps and<br>knowledge of the measurements and<br>the scanner field strength to correct<br>for signal changes related to iron<br>deposits, producing a cT1 map. The<br>cT1 map eliminates the effects of<br>elevated iron from the T1<br>measurement (3) and standardizes<br>for the fat signal across scanner<br>manufacturers.<br>PDFF is quantified using the LMS<br>IDEAL method. Parametric maps of<br>LIC are quantified using the LMS<br>MOST method.<br><br>LMSv5 uses the measured T2* value<br>and uses them to characterise iron<br>loading in the liver which is then<br>transformed by a defined calibration<br>curve to provide a quantitative<br>measure of liver iron concentration in<br>vivo. LMSv5 presents images that<br>reflect the magnetic resonance<br>spectra for iron determination on the<br>liver. | It is possible to use the T2* and PDFF<br>maps and knowledge of the T2* and<br>PDFF measurements and the scanner<br>field strength to correct for signal<br>changes related to iron deposits,<br>producing a cT1 map. The cT1 map<br>eliminates the effects of elevated iron<br>from the T1 measurement (3) and<br>standardizes for the fat signal across<br>scanner manufacturers.<br>PDFF is quantified using the LMS<br>IDEAL method. Parametric maps of<br>T2* may optionally be computed<br>using either the three-point DIXON<br>method or the LMS MOST method. | comprising R2 which is<br>transformed by a defined<br>calibration curve into a<br>quantitative measure of liver<br>iron concentration in vivo.<br>FerriScan presents images<br>that reflect the magnetic<br>resonance spectra for iron<br>determination on the liver. |
| Design:<br>Visualisation | Numerous views within the<br>LiverMultiScan v5 interface can be<br>used to assist in analysis of Iron<br>corrected T1 (cT1), triglyceride fat<br>(also known as Proton Density Fat<br>Fraction (PDFF)) and LIC parametric | Numerous views within the LMSv4<br>interface can be used to assist in<br>analysis, Iron-corrected T1 (cT1), T2*<br>and triglyceride fat (also known as<br>Proton Density Fat Fraction (PDFF))<br>parametric maps can be created from | Visualisation of multi-slice,<br>spin-echo MRI data sets<br>encompassing the<br>abdomen.<br>Software tool calculates the<br>signal decay rate (R2) that is |
| Characteristi<br>c | LMSv5 (Subject device) | LMSv4 (Primary Predicate Device) | R2-MRI Analysis System<br>(FerriScan) (Secondary<br>Predicate) |
| | maps can be created from all<br>supported scanners.<br><br>Parametric maps displayed using the<br>LMSv5 colourmap, designed to have<br>maximum contrast on liver<br>parenchymal tissue. | all supported scanners. R2 maps can<br>also be utilised to assess the quality<br>of the map fitting.<br><br>Iron- corrected T1 (cT1) displayed<br>using LMSv4 colourmap, designed to<br>have maximum contrast on liver<br>parenchymal tissue. | used to characterise iron<br>loading in the liver, which is<br>then transformed by a<br>defined calibration curve to<br>provide a quantitative<br>measure of liver iron<br>concentration in vivo.<br><br>LIC is displayed on a map<br>and histogram within the<br>FerriScan LIC report. |
| Design:<br>Supported<br>Modalities | DICOM 3.0 compliant MR data from | DICOM 3.0 compliant MR data from | DICOM Image Format.<br>Functionality independent<br>of MRI equipment vendor. |
| | supported MRI scanners. | supported MRI scanners. | |
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Perspectum" in a bold, dark gray font. To the right of the word is a circular logo. The logo is divided into four quadrants, each with a different color: yellow, blue, green, and pink.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for Perspectum. The word "Perspectum" is written in a bold, sans-serif font in dark gray. To the right of the word is a circular graphic with four different colored sections: yellow, blue, green, and pink. There is a registered trademark symbol in the upper right corner of the graphic.
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Perspectum" in a bold, sans-serif font. To the right of the word is a circular logo that is divided into four colored sections: yellow, blue, green, and pink. The logo also has a small "R" in a circle in the upper right corner.
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "Perspectum" in a bold, dark gray font. To the right of the word is a colorful circular logo. The logo is divided into four quadrants, with each quadrant a different color: yellow, blue, green, and pink. There is a registered trademark symbol in the upper right corner of the logo.
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Perspectum" in a dark gray sans-serif font. To the right of the word is a circular logo that is divided into four colored sections: yellow, blue, green, and pink. There is a registered trademark symbol next to the logo.
{13}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "Perspectum" in a bold, dark gray font. To the right of the word is a circular logo divided into four colored sections: yellow, blue, green, and pink. A small registered trademark symbol is located in the upper right corner of the logo.
{14}------------------------------------------------
## Differences between subject and predicate devices
- The LMSv5 device offers additional quantification tools in its interface that are not available in the primary . predicate device, LMSv4. LMSv5 automatically detects artefacts on the and are delineated on the parametric map computed images as well as recommends slices to be used as the quantitative output of the device. Artefact areas are overlayed on the parametric maps to assist in quantification and interpretation.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.