The general-purpose magnetic resonance imaging (MRI) system is designed to scan any targeted area of the body, to collect, display and analyse MR images and other real-time imaging procedures. Imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic, the lumbar and the sacral sections as portions of the spinal column, imaging the pelvis and imaging the head. Outcomes related to diagnosis: Magnifico is a Magnetic Resonance (MR) system that produces cross-section images of the limbs, joints, spinal column, pelvis and head. MRI provides better soft tissue contrast than CT and can differentiate better between fat, water, muscle, and other soft tissue than CT (CT is usually better at imaging bones). These images provide information to physicians and can be useful in diagnosing a wide variety of diseases and conditions.
Device Story
Magnifico is an open-magnet MRI system used in clinical settings by radiologists and technicians to produce cross-sectional images of limbs, joints, spine, pelvis, and head. The system utilizes a 0.4T permanent NdFeB magnet to generate MR signals, which are processed by the console (PC, monitor, keyboard) to display and analyze images. The open-magnet design improves patient comfort, particularly for claustrophobic patients and children. The device includes a transparent headcoil and specialized coils for cervical and knee applications. Physicians use the resulting high-contrast soft tissue images to diagnose various diseases and conditions. The system supports bilateral hip scanning and pelvic imaging. Clinical benefit is derived from superior soft tissue differentiation compared to CT, aiding in accurate clinical decision-making.
Clinical Evidence
Clinical evidence consisted of a set of clinical images in DICOM format covering all intended anatomical regions and common sequences. Images were evaluated by a board-certified radiologist (Dr. Mark Awh) and determined to be of good diagnostic quality, comparable to images obtained from similar MRI equipment. No formal clinical trial with statistical endpoints was required; effectiveness was demonstrated through qualitative expert review of image quality.
Technological Characteristics
Permanent magnet MRI system (0.4T NdFeB, C-shape). Components: patient table, magnetic unit, console (PC/monitor), electronic box. Connectivity: DICOM. Software: EVO'21 release. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-33, ISO 14971, ISO 62304, IEC 62366, NEMA MS-1, NEMA MS-3.
Indications for Use
Indicated for patients requiring MRI imaging of the upper limbs, lower limbs, temporomandibular joint, spinal column (cervical, thoracic, lumbar, sacral), pelvis, and head for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K090433 — I_OPEN 0.4T · Beijing Wandong Medical Equipment Co., Ltd. · Mar 25, 2009
K033315 — SUPEROPEN 0.23T, MODEL NAM-P023A · Neusoft Digital Medical Systems Co., Ltd. · Oct 29, 2003
K030918 — SUPEROPEN 0.35T, MODEL NSM-P035 · Neusoft Digital Medical Systems Co., Ltd. · Apr 1, 2003
K062860 — MODIFICATION TO SUPEROPEN 0.23T, MODEL NAM-P023A · Neusoft Medical Systems Co., Ltd. · Oct 27, 2006
K130658 — ELIXBO PM545 · Xbo Medical Systems Co., Ltd. · Jul 19, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
February 4, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Esaote S.p.A. %Alberto Carcagni Regulatory Affairs Officer Via Enrico Melen 77 Genoa. Genoa 16152 ITALY
Re: K212419
Trade/Device Name: Magnifico Open, Magnifico MSK Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: July 26, 2021 Received: January 5, 2022
Dear Alberto Carcagni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K212419
#### Device Name
Magnifico (Magnifico Open and Magnifico MSK configurations)
Indications for Use (Describe)
The general-purpose magnetic resonance imaging (MRI) system is designed to scan any targeted area of the body, to collect, display and analyse MR images and other real-time imaging procedures.
Imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, caff, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic, the lumbar and the sacral sections as portions of the spinal column, imaging the head.
Outcomes related to diagnosis:
Magnifico is a Magnetic Resonance (MR) system that produces cross-section images of the limbs, joints, spinal column, pelvis and head. MRI provides better soft tissue contrast than CT and can differentiate better between fat, water, muscle, and other soft tissue than CT (CT is usually better at imaging bones). These images provide information to physicians and can be useful in diagnosing a wide variety of diseases and conditions.
| <span> <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
Tyne of Use (Select one or hoth as annlicable)
{3}------------------------------------------------
510(k) Summary Magnifico Esaote S.p.A.
# 6 - 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(a).
### Submitter Information
Esaote, S.p.A. Via E. Melen 77 Genoa 16152 Italy
Alberto Carcagnì Contact Person: Phone: +39 055 4229365 +39 055 4229424 Fax: Mobile: +39 338 7170634 alberto.carcagni@esaote.com
| Date: | July 2021 |
|-------------------------|--------------------------------------|
| Trade Name: | Magnifico |
| Classification Panel: | Radiology |
| Classification Name(s): | Magnetic Resonance Diagnostic Device |
| Product Code: | LNH |
# Predicate Device(s)
| Trade Name | Common name | Class | Product<br>code | Manufacturer | K<br>number |
|-------------|-----------------------------------------------|-------|-----------------|---------------|-------------|
| S-scan | System, nuclear magnetic<br>resonance imaging | II | LNH | ESAOTE S.P.A. | K161973 |
| G-scan Brio | System, nuclear magnetic<br>resonance imaging | II | LNH | ESAOTE S.P.A. | K180592 |
{4}------------------------------------------------
#### Technological Characteristics
### Device Description
Magnifico is a Magnetic Resonance (MR) system with two configurations:
1. Magnifico Open, "Whole body" configuration (all above listed, in intended use, anatomical regions) 2. Magnifico MSK, musculoskeletal configuration (all above listed anatomical regions, excluded head and pelvis)
which produces images of the internal structures of the patient's limbs, joints and spinal column. The system comprises four main parts:
- Patient Table
- . Magnetic Unit, containing a permanent magnet
- Console, comprising a PC, keyboard, mouse, monitor and operating table
- Electronic box
Magnifico has an open magnet that makes comfortable MRI exam for all patients, including claustrophobic patients, in particular children.
Additionally, Magnifico comes with a transparent headcoil for enhanced patient comfort.
Magnifico is similar to Esaote S-scan system, cleared via K161973
With respect to the predicate devices, Magnifico introduces improvements, a new software release and few features.
These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following:
- A new Software version (EVO'21) including the following features: .
- 1. Pelvis anatomical district
- 2. Hip scan Bilateral (instead of unilateral as for Esaote S-scan and G-scan Brio)
- . A new multi-channel technology C-spine coil for cervical application and a new coil for knee scan
- 0.4T Magnet instead of the 0.25T installed in S-scan.
# Indications for Use
The general-purpose magnetic resonance imaging (MRI) system is designed to scan any targeted area of the body, to collect, display and analyse MR images and other real-time imaging procedures. Imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic, the lumbar and the sacral sections as portions of the spinal column, imaging the pelvis and imaging the head. Outcomes related to diagnosis
Maqnifico is a Maqnetic Resonance (MR) system that produces cross-section images of the limbs, ioints, spinal column, pelvis and head. MRI provides better soft tissue contrast than CT and can differentiate better between fat, water, muscle, and other soft tissue than CT (CT is usually better at imaqing bones). These images provide information to physicians and can be useful in diagnosing a wide variety of diseases and conditions.
{5}------------------------------------------------
510(k) Summary Magnifico Esaote S.p.A.
The technological characteristics of the Magnifico systems with the addition of new 0.4T Magnet**, Pelvis anatomical district and Bilateral Hip scan, reflected in this 510(k), do not alter the scientific technology of the Magnifico systems and are substantially equivalent to those of the predicate devices. The modification are implemented in order to improve image quality, analysis and comparison of Magnetic Resonance images.
**The new magnet is always a permanent one, based on NdFeB (neodymium), C-shape, with the bed positioned across the magnet and orthogonally to it, as for the previous 0.25T Magnet.
# Summary of Non-Clinical Tests
The Magnifico has been evaluated to demonstrate substantial equivalence related to medical electrical equipment, risk management, and software verification and has been found to conform to the following medical device safety standards:
- IEC 60601-1 ●
- IEC 60601-1-2 ●
- IEC 60601-1-6
- IEC 60601-2-33
- ISO 14971
- ISO 62304
- IEC 62366
- NEMA MS-1
- NEMA MS-3
# Summary of Clinical Tests
Within this submission is included a complete set of clinical images in DICOM format, for all anatomical regions foreseen in the intended use with the sequences commonly used for these anatomies. The clinical images were acquired and found to be of good diagnostic quality by the board-certified radiologist, Dr. Mark Awh, diplomat American Board of Radiology and president of RadSource.
# Conclusion
The Non-clinical testing demonstrates that the Magnifico is safe. Clinical Testing, with clinical images quality, comparable to those obtained from similar MRI equipment, demonstrate that Magnifico is effective and performs as wellas or better than the predicate system. Magnifico is substantially equivalent to the legally marketed devices and conforms to applicable medical device safety and performance standards.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.