← Product Code [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH) · K202170

# LiverMultiScan (K202170)

_Perspectum, Ltd. · LNH · Oct 2, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K202170

## Device Facts

- **Applicant:** Perspectum, Ltd.
- **Product Code:** [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH.md)
- **Decision Date:** Oct 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

LiverMultiScan (LMSv4) is indicated for use as a magnetic resonance diagnostic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times. LiverMultiScan (LMSv4) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced. LiverMultiScan (LMSv4) provides a number of tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessments of selected regions include the determination of triglyceride fat fraction in the liver (PDFF), T2* and iron-corrected T1 (cT1) measurements. T2* may be optionally computed using the DIXON or LMS MOST methods. These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis

## Device Story

LiverMultiScan (LMSv4) is a standalone post-processing software application; imports DICOM 3.0 MR image datasets of the abdomen; enables automated liver segmentation and manual ROI placement; computes quantitative metrics including PDFF, T2*, and iron-corrected T1 (cT1). Used at Perspectum's image analysis center by trained operators; produces a PDF report for interpreting clinicians. Assists clinicians in characterizing liver tissue; provides additional diagnostic input alongside blood tests, ultrasound, and biopsy. Does not replace clinical judgment; not a computer-aided diagnostic device. Benefits patients by providing noninvasive quantitative data to aid in liver health assessment and identifying patients for further evaluation.

## Clinical Evidence

No clinical investigations were conducted. Performance was assessed via bench testing using purpose-built phantoms and in-vivo volunteer data. Bench testing evaluated accuracy of cT1, T2*, and PDFF against gold standards. Clinical performance testing assessed repeatability and reproducibility across scanners/field strengths, reporting Limits of Agreement for ROI and segmentation methods. Results confirmed measurements are highly repeatable and reproducible, meeting acceptance criteria.

## Technological Characteristics

Standalone software; non-contact/non-invasive. Operates on general-purpose workstations (Mac OS). Inputs: DICOM 3.0 MR datasets. Outputs: Parametric maps (cT1, T2*, PDFF) and PDF reports. Algorithms: Automated liver segmentation, ROI-based quantification, and parametric mapping (IDEAL, DIXON, LMS MOST). Compliant with IEC 62304, IEC 62366, DICOM 3.0, ISO 14971, and ISO 13485.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- LiverMultiScan (LMSv3) ([K190017](/device/K190017.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 2, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Perspectum LTD % Jaco Jacobs Chief Compliance Officer Gemini One, 5520 John Smith Drive Oxford, Oxfordshire OX4 2LL UNITED KINGDOM

Re: K202170

Trade/Device Name: LiverMultiScan (LMSv4) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: March 13, 2020 Received: August 3, 2020

Dear Jaco Jacobs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael D. O'Hara

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K202170

Device Name LiverMultiScan (LMSv4)

### Indications for Use (Describe)

LiverMultiScan (LMSv4) is indicated for use as a magnetic device software application for noninvasive liver evaluation that enables the generation, display and review of 2D magnetic resonance medical image data and pixel maps for MR relaxation times.

LiverMultiScan (LMSv4) is designed to utilize DICOM 3.0 compliant magnetic resonance image datasets, acquired from compatible MR systems, to display the internal structure of the abdomen including the liver. Other physical parameters derived from the images may also be produced.

LiverMultiScan (LMSv4) provides a number of tools, such as automated liver segmentation and region of interest (ROI) placements, to be used for the assessment of selected regions of an image. Quantitative assessments of selected regions include the determination of triglyceride fat fraction in the liver (PDFF), T2* and iron-corrected T1 (cT1) measurements. T2* may be optionally computed using the DIXON or LMS MOST methods.

These images and the physical parameters derived from the images, when interpreted by a trained clinician, yield information that may assist in diagnosis

| Type of Use ( <i>Select one or both, as applicable</i> )                                                                                  |  |
|-------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><span style="font-size:1.5em;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |  |
| <div style="display:flex; align-items:center;"><span style="font-size:1.5em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div>  |  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

### This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date Prepared:

21st of July 2020

# 1. Submitter Details

| Owner Address:                     | Perspectum Ltd<br>Gemini One,<br>5520 John Smith Drive,<br>Oxford Business Park,<br>Oxford,<br>OX4 2LL |
|------------------------------------|--------------------------------------------------------------------------------------------------------|
| Owner/Operator Number:             | 10056574                                                                                               |
| Establishment Registration Number: | 3014232555                                                                                             |
| Contact Person:                    | Dr Jaco Jacobs<br>jaco.jacobs@perspectum.com                                                           |

+44 (0) 1865 655329

# 2. Subject and Predicate Device

|                      | Subject Device                                | Predicate Device                              |
|----------------------|-----------------------------------------------|-----------------------------------------------|
| 510(k) number        | Not known                                     | K190017                                       |
| Legal Manufacturer   | Perspectum Ltd.                               | Perspectum Diagnostics Ltd.                   |
| Owner/Owner Operator | 10056574                                      | 10056574                                      |
| Device Name          | LiverMultiScan (LMSv4)                        | LiverMultiScan (LMSv3)                        |
| Proprietary/Common   | LiverMultiScan                                | LiverMultiScan                                |
| Panel                | Radiology                                     | Radiology                                     |
| Regulation           | 892.1000                                      | 892.1000                                      |
| Risk Class           | Class II                                      | Class II                                      |
| Product Class code   | LNH                                           | LNH                                           |
| Classification       | System, Nuclear Magnetic<br>Resonance Imaging | System, Nuclear Magnetic<br>Resonance Imaging |

# 3. Subject Device Description

### 3.1. General

LiverMultiScan is a standalone software device. The purpose of the LiverMultiScan device is to assist a trained operator with the evaluation of information from Magnetic Resonance (MR) images from a single time-point (a patient visit). LiverMultiScan is a post-processing software device, a trained operator uses tools such as automatic liver segmentation and region of interest placement upon previously acquired MR images, from which a summary report is generated. The summary report is subsequently sent to an interpreting clinician at the acquiring site.

LiverMultiScan is not intended to replace the established procedures for the assessment of a patient's liver health by an interpreting clinician, providing many opportunities for competent human intervention in the interpretation of images and information displayed.

{4}------------------------------------------------

The metrics are intended to be used as an additional diagnostic input to provide information to clinicians as part of a wider diagnostic process. It is expected that in the normal course of liver disease diagnosis, patients with clinical symptoms or risk factors which may indicate liver disease. The interpreting clinician needs to take into consideration the device's limitations and accuracy during clinical interpretation.

Liver function tests, blood tests, ultrasound scanning as well as liver biopsy are all expected to be used at the discretion of a qualified clinician in addition to information obtained from the use of LiverMultiScan metrics. The purpose of LiverMultiScan metrics is to provide imaging information to assist in characterizing tissue in the liver, in addition to existing methods for obtaining information relating to the liver. LiverMultiScan metrics are not intended to replace any existing diagnostic source of information but can be used to identify patients who may benefit most from further evaluation, including biopsy.

Information gathered through existing diagnostic tests and clinical evaluation of the patient, as well as information obtained from LiverMultiScan metrics, may contribute to a diagnostic decision.

LiverMultiScan is not a computer-aided diagnostic device and can only present imaging information which must be interpreted by a qualified clinician. LiverMultiScan is an aid to diagnosis and treatment decisions remains the responsibility of the clinician.

In consequence, the product is considered to have no adverse effect on health since the results represent only a part of the information that the user will utilize for final interpretation. In this regard, LiverMultiScan presents a moderate level of concern with respect to patient safety.

### 3.2. Sterilization and Shelf Life

LMSv4 is a standalone software device thus it is non-contact, non-invasive and non-sterile. The shelf life of LMSv4 is indefinite as long as the manufacturer continues to support the device. Both sterilization and shelf life characteristics are equivalent of the predicate device.

### 3.3. Biocompatibility

LMSv4 is a standalone software device thus it is non-contact and non-invasive. No biocompatibility testing was deemed necessary to demonstrate the safety and effectiveness of LMSv4 does not consists of materials that differ from the predicate device.

### 3.4. Software

LMSv4 was successfully validated and verified against the requirements specification and its intended use. The results from the validation and verification activities, documented in this submission, corroborate that LMSv4 meets the product requirement specifications and intended use, which is deemed to be substantially equivalent to the predicate (see section below).

Validation and verification activities were conducted in a controlled environment and in compliance with IEC 62304:2006, ISO 13485:2016 and 21 CFR 820. LMSv4 is also in compliance with the DICOM standard.

### 3.5. Electromagnetic and Electrical Safety

LMSv4 is a standalone software device, there are no electrical safety risks associated with the direct use of the LMSv4 device. No electromagnetic or electrical safety testing was deemed necessary to demonstrate the safety and effectiveness of LMSv4.

{5}------------------------------------------------

# Perspectur

# 4. Subject and Predicate Comparison

### Subject and Predicate Device Comparison 4.1.

The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence.

| Comparison of Subject and Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic                             | LMSv4 (Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LMSv3 (Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use<br>and Indications<br>for Use | "LiverMultiScan (LMSv4) is indicated for use as<br>a magnetic resonance diagnostic device<br>software application for non-invasive liver<br>evaluation that enables the generation, display<br>and review of 2D magnetic resonance medical<br>image data and pixel maps for MR relaxation<br>times.                                                                                                                                                                       | "LiverMultiScan (LMSv3) is indicated for use as<br>a magnetic resonance diagnostic device<br>software application for non-invasive liver<br>evaluation that enables the generation, display<br>and review of 2D magnetic resonance medical<br>image data and pixel maps for MR relaxation<br>times.                                                                                                                              |
|                                            | LiverMultiScan (LMSv4) is designed to utilize<br>DICOM 3.0 compliant magnetic resonance<br>image datasets, acquired from compatible MR<br>systems, to display the internal structure of the<br>abdomen including the liver. Other physical<br>parameters derived from the images may also<br>be produced.                                                                                                                                                                 | LiverMultiScan (LMSv3) is designed to utilize<br>DICOM 3.0 compliant magnetic resonance<br>image datasets, acquired from compatible MR<br>systems, to display the internal structure of<br>the abdomen including the liver. Other<br>physical parameters derived from the images<br>may also be produced.                                                                                                                        |
|                                            | LiverMultiScan (LMSv4) provides a number of<br>tools, such as automated liver segmentation<br>and region of interest (ROI) placements, to be<br>used for the assessment of selected regions of<br>an image. Quantitative assessments of<br>selected regions include the determination of<br>triglyceride fat fraction in the liver (PDFF), T2*<br>and iron-corrected T1 (cT1) measurements.                                                                               | LiverMultiScan (LMSv3) provides a number of<br>tools, such as automated liver segmentation<br>and region of interest (ROI) placements, to be<br>used for the assessment of selected regions of<br>an image. Quantitative assessments of<br>selected regions include the determination of<br>triglyceride fat fraction in the liver (PDFF), T2*<br>and iron-corrected T1 (cT1) measurements.                                      |
|                                            | T2* may be optionally computed using the<br>DIXON or LMS MOST methods.                                                                                                                                                                                                                                                                                                                                                                                                    | PDFF may optionally be computed using the<br>LMS IDEAL or three-point DIXON<br>methodology.                                                                                                                                                                                                                                                                                                                                      |
|                                            | These images and the physical parameters<br>derived from the images, when interpreted by<br>a trained clinician, yield information that may<br>assist in diagnosis."                                                                                                                                                                                                                                                                                                      | These images and the physical parameters<br>derived from the images, when interpreted by<br>a trained clinician, yield information that may<br>assist in diagnosis."                                                                                                                                                                                                                                                             |
| Target                                     | Patients suitable to undergo an MRI scan and                                                                                                                                                                                                                                                                                                                                                                                                                              | Patients suitable to undergo an MRI scan and                                                                                                                                                                                                                                                                                                                                                                                     |
| Population<br>Device User                  | not contra-indicated for MRI.<br>Trained Perspectum operator                                                                                                                                                                                                                                                                                                                                                                                                              | not contra-indicated for MRI.<br>Trained Perspectum operator                                                                                                                                                                                                                                                                                                                                                                     |
| Report User                                | An interpreting clinician or healthcare<br>practitioner.                                                                                                                                                                                                                                                                                                                                                                                                                  | An interpreting clinician or healthcare<br>practitioner.                                                                                                                                                                                                                                                                                                                                                                         |
| Device Use<br>Environment                  | Installation of LMSv4 is controlled and installed<br>on general purpose workstations at<br>Perspectum's image analysis centre, by<br>specialist members of staff.                                                                                                                                                                                                                                                                                                         | Installation of LMSv3 is controlled and<br>installed on general purpose workstations at<br>Perspectum's image analysis centre by<br>specialist members of staff.                                                                                                                                                                                                                                                                 |
| Comparison of Subject and Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Characteristic                             | LMSv4 (Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LMSv3 (Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                         |
| Clinical Setting                           | LMSv4 is a standalone software device that is<br>intended to be installed on general use<br>workstations at Perspectum's image analysis<br>centre. The intended device users will log on to<br>the workstations, access the device, and use<br>the device on general-use HD monitors.                                                                                                                                                                                     | LMSv3 is a standalone software device that is<br>intended to be installed on general use<br>workstations at Perspectum's image analysis<br>centre. The intended device users will log on<br>to the workstations, access the device, and<br>use the device on general-use HD monitors.                                                                                                                                            |
|                                            | LMSv4 is a post-processing software, the<br>intended device users are trained internal<br>Perspectum operators. Operators use LMS to<br>conduct quantitative analysis of liver tissue<br>characteristics to produce a report.                                                                                                                                                                                                                                             | LMSv3 is a post-processing software, the<br>intended device users are trained internal<br>Perspectum operators. Operators use LMS to<br>conduct quantitative analysis of liver tissue<br>characteristics to produce a report.                                                                                                                                                                                                    |
|                                            | The end-users for the output from the device,<br>the pdf report, are clinicians who receive and<br>interpret LMSv4 reports through the QAS.                                                                                                                                                                                                                                                                                                                               | The end-users for the output from the device,<br>the pdf report, are clinicians who receive and<br>interpret LMSv3 reports through the QAS.                                                                                                                                                                                                                                                                                      |
| Anatomical<br>Location                     | Abdomen, Liver                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Abdomen, Liver                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Energy<br>Considerations                   | Software only application. The device, a<br>standalone software application, does not<br>deliver, monitor or depend on energy<br>delivered to or from patients.                                                                                                                                                                                                                                                                                                           | Software only application. The device, a<br>standalone software application, does not<br>deliver, monitor or depend on energy<br>delivered to or from patients.                                                                                                                                                                                                                                                                  |
| Design: Purpose                            | LMS is a standalone software application that                                                                                                                                                                                                                                                                                                                                                                                                                             | LMS is a standalone software application that                                                                                                                                                                                                                                                                                                                                                                                    |
|                                            | imports MR data sets encompassing the<br>abdomen, including the liver. Visualisation and<br>display of 2D multi-slice, spin-echo MR data<br>can be analysed, and quantitative metrics of<br>tissue characteristics are then reported.<br>Datasets imported into LMS are DICOM 3.0<br>compliant, reported metrics are independent<br>of the MRI equipment vendor.                                                                                                          | imports MR data sets encompassing the<br>abdomen, including the liver. Visualisation<br>and display of 2D multi-slice, spin-echo MR<br>data can be analysed, and quantitative<br>metrics of tissue characteristics are then<br>reported.<br>Datasets imported into LMS are DICOM 3.0<br>compliant, reported metrics are independent                                                                                              |
| Design: Tools                              | Allows for the visualisation via parametric<br>maps and quantification of metrics (cT1, T2*<br>and PDFF) from liver tissue and exportation of<br>results & images to a deliverable report.<br>LMSv4 allows for:                                                                                                                                                                                                                                                           | of the MRI equipment vendor.<br>Allows for the visualisation via parametric<br>maps and quantification of metrics (cT1, T2*<br>and PDFF) from liver tissue and exportation of<br>results & images to a deliverable report.<br>LMSv3 allows for:                                                                                                                                                                                  |
|                                            | cT1<br>Full segmentation of the outer liver<br>contour and liver vasculature of the cT1<br>parametric map. Interquartile Range (IQR)<br>and median metrics are reported from the<br>segmentation.                                                                                                                                                                                                                                                                         | cT1<br>Full segmentation of the outer liver<br>contour and liver vasculature of the cT1<br>parametric map.<br>Interquartile Range (IQR) and median<br>metrics are reported from the<br>segmentation.                                                                                                                                                                                                                             |
| Comparison of Subject and Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Characteristic                             | LMSv4 (Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LMSv3 (Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                            | ROI placed method on the cT1 map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices.                                                                                                                                                                                                                                                                                                                                           | ROI placed method on the cT1 map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices.                                                                                                                                                                                                                                                                                                  |
|                                            | T2*  ROI placed method on the T2* map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices. T2* parametric maps are calculated from the MOST method or the three-point DIXON method.                                                                                                                                                                                                                                             | T2*  ROI placed method on the T2* map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices. T2* parametric maps are calculated from the three-point DIXON method.                                                                                                                                                                                                                       |
|                                            | PDFF  Full liver segmentation of the PDFF parametric map where IQR and median metrics are reported from the segmentation. ROI placed method on the PDFF map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices. PDFF parametric maps are calculated using the LMS IDEAL method 1                                                                                                                                               | PDFF  Full liver segmentation of the PDFF parametric map where IQR and median metrics are reported from the segmentation. ROI placed method on the PDFF map with IQR and median metrics from the placed ROI's potentially across multiple acquired slices. PDFF parametric maps can be calculated using either the LMS IDEAL method 1 or the three-point DIXON method 2                                                          |
| Design: MR<br>Relaxometry                  | T1, iron-corrected T1 (cT1) and T2* mapping.                                                                                                                                                                                                                                                                                                                                                                                                                              | T1, iron-corrected T1 (cT1) and T2* mapping.                                                                                                                                                                                                                                                                                                                                                                                     |
| Design: Liver Fat<br>Quantification        | Utilizes MR images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat using the LMS IDEAL method.                                                                                                                                                                                                                                                                                                                 | Utilizes MR images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat using either the LMS IDEAL method or three-point DIXON method.                                                                                                                                                                                                                                     |
| Design: Liver<br>Segmentation              | LMSv4 supports automatic multi-slice full liver segmentation of the cT1 and PDFF parametric map, use of this functionality is at the discretion of the operator instead or in combination with the ROI based method.<br>The cT1 segmented liver is presented in colour level window, while the rest of the cT1 image is presented in greyscale level window with ducts and liver vasculature excluded from the segmented volume.                                          | LMSv3 supports automatic multi-slice full liver segmentation of the cT1 and PDFF parametric map, use of this functionality is at the discretion of the operator instead or in combination with the ROI based method.<br>The cT1 segmented liver is presented in colour level window, while the rest of the cT1 image is presented in greyscale level window with ducts and liver vasculature excluded from the segmented volume. |
|                                            | Comparison of Subject and Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Characteristic                             | LMSv4 (Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LMSv3 (Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                         |
| Design: Regions<br>of Interest (ROI)       | Median and interquartile range measurements<br>created from a cross sectional slice of liver<br>tissue. For each parametric map, statistics<br>from multiple Regions of Interest (ROIs) –<br>potentially placed across multiple slices - are<br>summarised.                                                                                                                                                                                                               | Median and interquartile range measurements<br>created from a cross sectional slice of liver<br>tissue. For each parametric map, statistics<br>from multiple Regions of Interest (ROIs) –<br>potentially placed across multiple slices – are<br>summarised.                                                                                                                                                                      |
| Design:<br>Parametric<br>Maps              | Iron corrected T1 (cT1), T2* and Proton<br>Density Fat Fraction (PDFF) parametric maps<br>can be created from all supported scanners.<br>PDFF is quantified using the LMS IDEAL<br>method. Parametric maps of T2* may be<br>optionally be computed using either the<br>three-point DIXON method or the LMS MOST<br>method.                                                                                                                                                | Iron corrected T1 (cT1), T2* and Proton<br>Density Fat Fraction (PDFF) parametric maps<br>can be created from all supported scanners.<br>PDFF is quantified using the LMS IDEAL<br>method or the three-point DIXON method.                                                                                                                                                                                                       |
| Design:<br>Visualisation                   | Numerous views within the LMSv4 interface<br>can be used to assist in analysis, Iron-<br>corrected T1 (cT1), T2* and triglyceride fat<br>(also known as Proton Density Fat Fraction<br>(PDFF)) parametric maps can be created from<br>all supported scanners. R² maps can also be<br>utilised to assess the quality of the map fitting.<br>Iron- corrected T1 (cT1) displayed using LMSv4<br>colourmap, designed to have maximum<br>contrast on liver parenchymal tissue. | Numerous views within the LMSv4 interface<br>can be used to assist in analysis, iron<br>corrected T1 (cT1), T2* and triglyceride fat<br>(also known as Proton Density Fat Fraction<br>(PDFF)) parametric maps can be created from<br>all supported scanners.<br>Iron corrected T1 (cT1) displayed using the<br>LMSv3 colourmap, designed to have maximum<br>contrast on liver parenchymal tissue.                                |
| Design:<br>Supported<br>Modalities         | DICOM 3.0 compliant MR data from supported<br>MRI scanners.                                                                                                                                                                                                                                                                                                                                                                                                               | DICOM 3.0 compliant MR data from supported<br>MRI scanners.                                                                                                                                                                                                                                                                                                                                                                      |
| Design: Report                             | Quantified metrics and images derived from<br>the analysis conducted of liver tissue<br>characteristic on parametric maps are collated<br>into a report for evaluation and interpretation<br>by a clinician.                                                                                                                                                                                                                                                              | Quantified metrics and images derived from<br>the analysis conducted of liver tissue<br>characteristic on parametric maps are<br>collated into a report for evaluation and<br>interpretation by a clinician.                                                                                                                                                                                                                     |
| Compatibility<br>with the<br>environment   | Installation of LMSv4 is controlled and is<br>installed on general purpose workstations that<br>meet the minimum technical requirements at<br>Perspectum's image analysis centre by<br>specialist members of staff.                                                                                                                                                                                                                                                       | Installation of LMSv3 is controlled and is<br>installed on general purpose workstations<br>that meet the minimum technical<br>requirements at PD's image analysis centre by<br>specialist members of staff.                                                                                                                                                                                                                      |
| Performance                                | Device performance was assessed with<br>purpose-built phantoms and in-vivo acquired<br>data from volunteers covering a range of<br>physiological values for cT1, T2* and PDFF.                                                                                                                                                                                                                                                                                            | Device performance was assessed with<br>purpose-built phantoms and in-vivo acquired<br>data from volunteers covering a range of<br>physiological values for cT1, T2* and PDFF.                                                                                                                                                                                                                                                   |
| Supported MRI<br>Systems                   | Validated across all listed supported<br>manufacturers and field strengths.                                                                                                                                                                                                                                                                                                                                                                                               | Validated across all listed supported<br>manufacturers and field strengths.                                                                                                                                                                                                                                                                                                                                                      |
| Standards                                  | IEC 62304, IEC 62366, DICOM 3.0, ISO 14971,<br>ISO 13485                                                                                                                                                                                                                                                                                                                                                                                                                  | IEC 62304, IEC 62366, DICOM 3.0, ISO 14971,<br>ISO 13485                                                                                                                                                                                                                                                                                                                                                                         |
| Comparison of Subject and Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Characteristic                             | LMSv4 (Subject device)                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LMSv3 (Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                         |
| System/Operati<br>ng System                | Mac OS                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Mac OS                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Materials                                  | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Biocompatibility                           | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterility                                  | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Electrical Safety                          | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Mechanical<br>Safety                       | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Chemical Safety                            | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Thermal Safety                             | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |
| Radiation Safety                           | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                                                                      | Not applicable, standalone software.                                                                                                                                                                                                                                                                                                                                                                                             |

{6}------------------------------------------------

# Perspectum 6

{7}------------------------------------------------

# Perspectum 6

{8}------------------------------------------------

# Perspectum t

{9}------------------------------------------------

Table 1. Comparison of similar characteristics between the subject and predicate device.

In conclusion, the subject device does not result in any new potential safety risk when compared to the chosen predicate device and performs in accordance with its use characteristics and intended use.

## 5. Performance Testing

LMSv4 underwent performance testing under controlled conditions to corroborate that it is safe and effective when used as intended. The performance testing conducted demonstrates that LMSv4 is at least as safe and effective as the predicate device and does not introduce any new risks.

### Performance Testing - Bench 5.1.

The accuracy and precision of device measurements was assessed using purpose-built phantoms containing vials with different relaxation times corresponding to the physiological range of liver tissue values expected to be seen in-vivo. Accuracy was measured by the proximity of quantified LMSv4 values to the gold standard. The results of which are summarized below:

| Metric     | Accuracy            |                     |
|------------|---------------------|---------------------|
|            | 1.5T                | 3T                  |
| cT1        | -189.5 to -35.11 ms | -187.0 to -19.12 ms |
| T2*        | -0.68 to 0.64 ms    | -0.30 to 0.39 ms    |
| IDEAL PDFF | -3.80 to 6.08%      | -1.39 to 5.58%      |

Table 2. Pooled performance testing - bench

- The MOLLI-based T1 measurement produced by LMSv4 is consistent with the literature-reported . underestimation of ground truth T1 using MOLLI techniques 34
- . LMSv4 measurement of T2*is accurate over the expected physiological range of values
- LMSv4 measurement of PDFF is accurate over the expected physiological range of values o

### 5.2. Performance Testing - Clinical

To assess the precision of LMSv4 measurements across supported scanners in-vivo acquired volunteer data was used, volunteers participating in the performance testing were representative of the intended patient population. Inter and intra operator variability was also assessed. The results of which are summarized below:

{10}------------------------------------------------

| Metric                    | Repeatability       | Reproducibility     |
|---------------------------|---------------------|---------------------|
|                           | Limits of Agreement | Limits of Agreement |
| cT1 (ROI)                 | -43.25 to 26.77 ms  | -103 to 91.8 ms     |
| cT1 (Segmentation)        | -40.75 to 25.02 ms  | -102.3 to 93.69 ms  |
| T2* (DIXON)               | -5.21 to 6.01 ms    | -1.74 to 0.35 ms    |
| T2* (MOST)                | -3.17 to 3.25 ms    | -2.40 to 2.15 ms    |
| IDEAL PDFF (ROI)          | -1.48 to 1.42%      | -2.88 to 2.53%      |
| IDEAL PDFF (Segmentation) | -1.31 to 1.34%      | -2.94 to 2.53%      |

Table 3. Pooled performance testing - clinical

- LMSv4 measurements of cT1, T2* and PDFF are highly repeatable .
- LMSv4 measurements of cT1, T2* and PDFF are reproducible between scanners and field strengths .
- The variation introduced by operator measurements with both ROI placement and segmentation method is well . within the prescribed acceptance criteria.

### 5.3. Clinical Investigation

No clinical investigations or studies were conducted during performance testing of LMSv4.

### 6. Conclusion

The subject device is substantially equivalent to the predicate device, both regulated under regulation 21 CFR 892.1000. Substantial equivalence is based on the following observations:

- . The indications for use and intended uses of both the subject device and predicate device are substantially equivalent.
- The subject and predicate devices are both software applications which facilitate the import and visualization of • MR data sets to visualise and enable quantification of physiological characteristics in the liver to provide measurements.
- The subject and predicate devices both support tools and features to derive measurements from MR images . and parametric maps of tissue characteristics.
- The subject and predicate device facilitate the creation of a medical report containing the images and analysis . output derived from quantification of liver tissue parameters intended to be interpreted by a trained clinician.
- . Both the subject and predicate devices are designed to run on general-purpose computing hardware and intended to be used in the same environment.
- Performance testing and risk management demonstrates that the subject device performs at least as safely and . effectively as the proposed predicate device and supports the determination of substantial equivalence.

In conclusion, the subject device does not result in any new potential safety risks when compared to the chosen predicate device and performs in accordance with its use characteristics and intended use.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the word "Perspectum" in a bold, dark gray font. To the right of the word is a colorful circular logo. The logo is divided into four sections, with each section a different color: yellow, blue, green, and pink.

### References

- 1. Reeder, S. B. et al. Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): Application with fast spin-echo imaging. Magn. Reson. Med. 54, 636–644 (2005).
- 2. Dixon, W. T. Simple proton spectroscopic imaging. Radiology 153, 189–194 (1984).
- 3. Piechnik, S. K. et al. Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J. Cardiovasc. Magn. Reson. 12, 69 (2010).
- 4. Puntmann, V. O. et al. Are T1 values to characterize myocardial tissue equivalent between various sequences: comparison of MOLLI, shMOLLI, 3'5-MOLLI and SASHA. J. Cardiovasc. Magn. Reson. 15, E18 (2013).

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K202170](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K202170)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
