The OASIS MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-latice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Device Story
OASIS MRI system; 1.2 Tesla superconducting magnet gantry; non-invasive imaging. Inputs: proton magnetic resonance signals. Transformation: RF excitation pulses applied; spatial localization via magnetic field gradients (x, y, z); echoes measured; 3D matrix reconstruction via image processing. Output: cross-sectional anatomical images (transverse, coronal, sagittal, oblique, curved). Used in clinical settings by trained physicians. Output interpreted by physicians for diagnostic determination. Benefits: high-quality anatomical imaging without ionizing radiation; supports T1, T2, proton density, diffusion, and spectroscopy.
Clinical Evidence
Clinical usability evaluated via collection and analysis of clinical image examples for new features, specifically the WIT Spine Coil. No formal clinical trial with statistical endpoints reported; bench testing provided for performance verification.
Technological Characteristics
1.2 Tesla superconducting open magnet; 16-channel RF receiver; 33mT/m gradient strength; 100 T/m/sec slew rate. Standards: NEMA MS 1-5, 8; IEC 60601-1, 60601-1-2, 60601-2-33, 62304. OS: Windows 10 IoT. CPU: Xeon 3.8 GHz. Software: V7.0D.
Indications for Use
Indicated for patients requiring diagnostic imaging of the head, body, spine, or extremities using T1, T2, proton density weighted imaging, diffusion weighted imaging, MR angiography, or spectroscopy.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 21, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
Hitachi Healthcare Americas % Aaron Pierce Director, RA/QA 1959 Summit Commerce Park TWINSBURG OH 44087
Re: K202030
Trade/Device Name: OASIS MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: October 12, 2020 Received: October 14, 2020
Dear Aaron Pierce:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
2
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
Device Name OASIS MRI system Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
#### Indications for Use (Describe)
The OASIS MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-latice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
| Anatomical Region: | Head, Body, Spine, Extremities |
|--------------------|-----------------------------------------|
| Nucleus excited: | Proton |
| Diagnostic uses: | T1, T2, proton density weighted imaging |
| | Diffusion weighted imaging |
| | MR Angiography |
| | Image processing |
| | Spectroscopy |
| | Whole Body |
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 510(k) Statement or Summary
## Submitter Information
| Submitter: | Hitachi Healthcare Americas<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371 |
|-------------------|----------------------------------------------------------------------------------------|
| Contact: | Aaron Pierce |
| Telephone number: | 330-425-1313 |
| Telephone number: | 330-963-0749 |
| E-mail: | piercea@hitachihealthcare.com |
| Date: | October 12, 2020 |
## Subject Device Name
| Trade/Proprietary Name: | OASIS MRI system |
|-------------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Name: | System, Nuclear Magnetic Resonance Imaging |
| Product Code | LNH |
| Class | II |
| Panel | Radiology |
### Predicate Device Name
| Predicate Device(s): | OASIS MRI System (K192851) |
|----------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Name: | System, Nuclear Magnetic Resonance Imaging |
| Product Code | LNH |
| Class | II |
| Panel | Radiology |
### Indications for Use
The OASIS MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Anatomical Region: Head, Body, Spine, Extremities Nucleus excited: Proton T1, T2, proton density weighted imaging Diagnostic uses: Diffusion weighted imaging MR Angiography Image processing Spectroscopy Whole Body
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Hitachi logo. The word "HITACHI" is in bold, black letters on the top line. The words "Inspire the Next" are in a smaller font below the word "HITACHI."
## Device Description
### Function
The OASIS is a Magnetic Resonance Imaging System that utilizes a 1.2 Tesla superconducting maqnet in a qantry design.
### Scientific Concepts
Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiguitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.
The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.
### Physical and Performance Characteristics
MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules. In addition the OASIS MR system has the Function of measuring spectroscopy.
### Performance Evaluation
The OASIS V7.0D MRI System is equivalent to the OASIS MRI (K192851) with the following exceptions:
- Control unit changed from RFIP unit to IRCP unit ●
- RF receiver channel is expanded from eight (8) to sixteen (16) ●
- WIT Spine Coil is added to product specification ●
- . RAPID Body Coil and RAPID C-spine Coil are deleted from product specification.
- Operating System is changed from Windows 7 to Windows 10 IoT .
- CPU performance is improved from Xeon 3.5 GHz to Xeon 3.8 GHz. ●
- Application software is revised to V7.0D. ●
A rationale analysis was then conducted and the results are contained below.
| Testing Type | Rationale Analysis |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing - Bench | Performance bench testing was conducted on the applicable new features. Test data confirmed that each new feature<br>perform as intended for diagnostic use. |
| Performance Testing -<br>Clinical | Clinical image examples are provided for each applicable new feature and or coil that we judged to be sufficient to evaluate<br>clinical usability. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Hitachi logo. The word "HITACHI" is written in large, bold, black letters. Below it, in a smaller font, is the phrase "Inspire the Next" in black, except for a small red accent mark above the letter 't' in "Next."
## Device Technological Characteristics
The control and image processing hardware and the base elements of the system software are identical to the predicate device. The OASIS MRI system w/V6.0F software is substantially equivalent to the OASIS (K093044). See tables below.
The technological characteristics in regards to hardware of the OASIS MRI system w/V6.0F and the predicate are listed below:
| ITEM | OASIS (K192851) | OASIS | DIFFERENCE | |
|---------------------|-------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------|
| System | Standards Met | NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8,<br>IEC: 60601-1, 60601-1-2, 60601-2-33, 62304 | NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8,<br>IEC: 60601-1, 60601-1-2, 60601-2-33, 62304 | No |
| Magnet and Gantry | Type and Field Strength | Super-conducting open magnet, 1.2 Tesla | Super-conducting open magnet, 1.2 Tesla | No |
| | Resonant Frequency | 49.39MHz ±100 kHz | 49.39MHz ±98 kHz | See Table 1 |
| Gradient System | Gradient Strength | 33mT/m | 33mT/m | No |
| | Slew Rate | 100 T/m/sec | 100 T/m/sec | No |
| | Rise Time | 300µsec to30mT/m | 300µsec to30mT/m | No |
| | Audible Noise (MCAN) | | | |
| | Ambient | 63.0 dBA | 63.0 dBA | No |
| | Lpeak | 126.3dB | 126.3dB | No |
| | Leq | 119dBA | 119dBA | No |
| RF System | Transmitter channels | 2 | 2 | No |
| | Peak Envelop Power | 18 kW | 18 kW | No |
| | Duty Cycle | 60% (Gating max), 10% at full power | 85% (Gating max), 10% at full power | See Table 1 |
| | RF receiver channel | 8 | 16 | See Table 1 |
| System Control Unit | Unit Type | RFIP | IRCP | See Table 1 |
The hardware differences from the OASIS MRI system to the predicate device are analyzed in the table below:
#### Table 1 Hardware NSE Analysis
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see<br>section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-------------|----------------------------------|
| Device<br>Modification<br>Summary | ● Reference range of "Resonant Frequency" changed, but this is a correction of miscalculation according to the<br>conversion from field strength (Tesla) to frequency (Hz). No change to the magnet itself.<br>● Maximum duty of RF amplifier is changed from 60% to 85%. The device is changed but there is no performance<br>change as a system because the RF output is controlled by the same sequence specification as OASIS<br>(K192851).<br>● RF receiver channel is expanded from eight (8) to sixteen (16).<br>● System control unit is changed from RFIP unit to IRCP unit. The device is changed but there is no change in<br>performance except for the number of receiver channels. | | | |
| Significant<br>Changes | □Manufacturing Process | □Labeling | □Technology | □Performance |
| | □Engineering | □Materials | □Others | ☑ None (See rationale statement) |
| HITACHI<br>Rationale<br>Statement | Modified specification doesn't constitute a new intended use. There are no significant changes in technological<br>characteristics. For safety, gradient system and RF system is controlled according to same regulation as OASIS<br>(K192851). So, safety and effectively of the device are same as OASIS (K192851). | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Hitachi. The word "HITACHI" is written in large, bold, black letters. Below the company name is the slogan "Inspire the Next" in a smaller, black font.
The technological characteristics in regards to coils of the OASIS MRI system and the predicate are listed in the table below:
| | ITEM | OASIS (K192851) | OASIS | DIFFERENCE |
|----------|----------------|-----------------|-----------------------|-------------|
| RF Coils | Transmit Coil | T/R Body | T/R Body | No |
| | Receiver Coils | Rapid Head | Head Coil | See Table 2 |
| | | Rapid Body | - | See Table 2 |
| | | Rapid Knee | Extremity Coil | See Table 2 |
| | | Rapid C-spine | - | See Table 2 |
| | | Rapid Wrist | Wrist Coil | See Table 2 |
| | | Rapid Shoulder | Shoulder Coil | See Table 2 |
| | | Large Joint | Multipurpose Coil | See Table 2 |
| | | Micro coil (S) | Micro Coil | See Table 2 |
| | | Foot/Ankle coil | Foot/Ankle Coil | See Table 2 |
| | | | Breast Coil | See Table 2 |
| | | | Large Flex Coil | See Table 2 |
| | | | Extra Large Flex Coil | See Table 2 |
| | | | WIT Spine Coil | See Table 2 |
The coil differences from the OASIS MRI system to the predicate device are analyzed in the table below:
| FDA Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does<br>not constitute a new intended use; and any differences in technological characteristics are accompanied by<br>information that demonstrates the device is as safe and effective as the predicate and do not raise different questions<br>of safety and effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety<br>and effectiveness (see section 513(i)(1)(A) of the FD&C Act and<br>21 CFR 807.87(f)). | | | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-------------|---------------------------------|
| Device Modification<br>Summary | ● WIT Spine Coil is added as new 16 channel receiver coil.<br>● RAPID Body Coil and RAPID C-spine Coil are deleted.<br>● Head Coil, Extremity Coil, Wrist Coil, Shoulder Coil, Multipurpose Coil, Micro Coil and Foot/Ankle Coil are<br>same coils as in the lineup of OASIS (K192851), but its name and cable connector is changed to match the<br>interface of the new system. The performance and technological characteristics of the coils are the same as<br>OASIS (K192851). The methodology to check the signal-to-noise performance is changed from NEMA MS 1<br>method 1 to NEMA MS 1 method 4.<br>● Breast Coil, Large Flex Coil and Extra Large Flex Coil are same coils as approved as K073310 and<br>K080062, but its name and cable connector is changed to match the interface of the new system. The<br>performance and technological characteristics of the coils are the same as K073310 and K080062. The<br>methodology to check the signal-to-noise performance is changed from NEMA MS 1 method 1 to NEMA MS<br>1 method 4. | | | |
| Significant Changes | ☐ Manufacturing Process | ☐ Labeling | ☐Technology | ☐Performance |
| | ☐Engineering | ☐Materials | ☐Others | ☑None (See rationale statement) |
| HITACHI Rationale<br>Statement | Additional or modified coils did not constitute a new intended use. There are no significant changes in technological<br>characteristics. During transmitter coil operation, RF Coils are de-resonated by same scheme as OASIS (K192851).<br>WIT Spine Coil is a new receiver coil has the capability of sixteen (16) channels. This is an expansion of maximum<br>number of receiver channel compared to OASIS (K192851) | | | |
#### Table 2 Coil Comparison Analysis
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Hitachi. The word "HITACHI" is written in large, bold, black letters. Below the company name is the slogan "Inspire the Next" in a smaller, black font. The letter "t" in the word "Next" has a red accent mark.
The technological characteristics in regards to changes in functionality of the OASIS MRI System as compared to the predicate are listed in the table below:
| ITEM | DIFFERENCES | ANALYSIS |
|-----------------------|---------------------------------------------------------------|-------------|
| Operating System | Going from Windows 7 and 10 loT to Windows 10 loT. | See Table 3 |
| CPU Platform | Going from Xeon 3.5 GHz to Xeon 3.8 GHz. | See Table 3 |
| Application Software | Going from V6.0F to V7.0D. | See Table 3 |
| Scan Tasks | None. | No |
| 2D Processing Tasks | None. | No |
| 3D Processing Tasks | None. | No |
| Analysis Tasks | None. | No |
| Maintenance Tasks | None. | No |
| Viewport Tools | None. | No |
| Film, Archive Tools | None. | No |
| Network Tools | None. | No |
| Protocol Enhancements | Maximum available Presaturation pulse number is changed to 6. | See Table 3 |
| Pulse Sequences | 2D IR, 2D TRSG and 3D TRSG sequences are deleted. | See Table 3 |
The functionality differences from the OASIS MRI System to the predicate device are analyzed in the table below. Features have been added since the predicate device through the Memo to File process.
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see<br>section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Modification<br>Summary | • Windows 7 in Operating System is deleted from product specification. Only Windows 10 IoT is used.<br>• CPU performance is improved from Xeon 3.5 GHz to Xeon 3.8 GHz.<br>• Application software is changed to V7.0D.<br>• Maximum available Presaturation pulse number is changed from eight (8) to six (6).<br>• 2D IR, 2D TRSG and 3D TRSG sequences are deleted. | | | |
| Significant<br>Changes | □ Manufacturing Process | □ Labeling | □Technology | □Performance |
| HITACHI<br>Rationale<br>Statement | □Engineering | □Materials | □Others | ☑None (See rationale statement)<br>Modified functions do not constitute a new intended use. There are no significant changes in technological characteristics.<br>For safety, pulse sequences are controlled according to same safety limits as OASIS (K192851). So safety and effectivity<br>of the device are equivalent to the OASIS (192851). |
### Table 3 Functionality Comparison Analysis
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Hitachi. The word "HITACHI" is in large, bold, black letters. Below it, in a smaller font, are the words "Inspire the Next" in black, with a small red accent mark over the "t" in "Next."
## Substantial Equivalence
A summary decision was based on analysis of results in the table below:
| Table 4 Rationale Analysis: OASIS MRI vs. Predicate | |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | Overall Rationale Analysis |
| Hardware | Modified specification doesn't constitute a new intended use. There are no significant changes in technological<br>characteristics. For safety, gradient system and RF system is controlled according to same regulation as OASIS<br>(K192851). So, safety and effectively of the device are same as OASIS (K192851). |
| Coils | Additional or modified coils did not constitute a new intended use. There are no significant changes in technological<br>characteristics. During transmitter coil operation, RF Coils are de-resonated by same scheme as OASIS (K192851). WIT<br>Spine Coil is a new receiver coil has the capability of sixteen (16) channels. This is an expansion of maximum number of<br>receiver channel compared to OASIS (K192851). |
| Functionality | Modified functions do not constitute a new intended use. There are no significant changes in technological characteristics.<br>For safety, pulse sequences are controlled according to same safety limits as OASIS (K192851). So safety and effectivity<br>of the device are equivalent to the OASIS (192851). |
Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed OASIS MRI is considered substantially equivalent to the currently marketed predicate device in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
## Summary of Non-Clinical Testing
The OASIS MRI System was subjected to the following laboratory testing.
- NEMA MS-1-2008 (R2014), Determination of Signal-to-Noise Ratio (SNR) In Diagnostic Magnetic . Resonance Imaging (All receiver coils)
- NEMA MS 3-2008 (R2014), Determination of Image Uniformity in Diagnostic Magnetic . Resonance Images (All receiver coils)
- NEMA MS 8-2016, Characterization of the Specific Absorption Rate (SAR) for Magnetic . Resonance Imaging Systems
- IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 ●
- . IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-33 Ed. 3.2 b:2015. Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis
- IEC 62304 Ed.1.1:2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes
The revisions to the OASIS MRI System and standards will have no effect on the standards tests, which were conducted on the OASIS MRI System (K192851) and included in the original submission.
- . NEMA MS 2-2008 (R2014), Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
- NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance ● Imaging Devices
- NEMA MS 5-2018, Determination of Slice Thickness in Diagnostic Resonance Imaging ●
In addition, tests were conducted on the new WIT Spine Coil, which includes 16 channels.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Hitachi. The word "HITACHI" is written in large, bold, black letters. Below the company name is the slogan "Inspire the Next" in a smaller, black font.
## Summary of Clinical Testing
Clinical images were collected and analyzed, to ensure that images from the new feature meet user needs. As a result of the analysis:
| Testing Type | Rationale Analysis |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing -<br>Clinical | Clinical image examples are provided for each applicable new feature, which is the WIT Spine Coil, and that we judged to be sufficient to evaluate clinical usability. |
## Conclusions
It is the opinion of Hitachi, the OASIS MRI system is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the OASIS MRI System (K192851).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.