uMR570

K201540 · Shanghai United Imaging Healthcare Co., Ltd. · LNH · Jul 6, 2020 · Radiology

Device Facts

Record IDK201540
Device NameuMR570
ApplicantShanghai United Imaging Healthcare Co., Ltd.
Product CodeLNH · Radiology
Decision DateJul 6, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.

Device Story

uMR 570 is a 1.5T superconducting MRI system with 70cm bore. Inputs include RF signals captured via various anatomical coils (head, body, cardiac, etc.) and vital signal gating (ECG, respiratory). System transforms inputs into cross-sectional and spectroscopic images. New feature: gre_senc_spiral pulse sequence for cardiac Strain ENCoding (SENC) imaging. Images stored in DICOM format; processed by 3rd party software for strain analysis. Used in clinical settings by trained physicians to assist diagnosis. Benefits include non-invasive visualization of internal anatomy and cardiac strain assessment.

Clinical Evidence

No clinical testing was conducted. Substantial equivalence is supported by bench testing, specifically a performance evaluation report for the SENC pulse sequence.

Technological Characteristics

1.5T superconducting magnet; 70cm bore; 45mT/m max gradient amplitude; 200T/m/s max slew rate; 1 transmit channel; up to 48 receive channels; 20kW amplifier peak power per channel. Conforms to NEMA and DICOM standards. Biocompatibility per ISO 10993-5 and ISO 10993-10. Electrical safety per ES60601-1; EMC per IEC60601-1-2; SAR/dB/dt per IEC 60601-2-33.

Indications for Use

Indicated for use as an MRDD to produce cross-sectional and spectroscopic images of the head, body, and extremities to display internal anatomical structure and/or function. Used by trained physicians to assist in diagnosis. Contrast agents may be used.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". Shanghai United Imaging Healthcare Co., Ltd. % Xin GAO Regulatory Affairs Specialist NO. 2258 Chengbei Road Shanghai, Shanghai 201807 CHINA July 6, 2020 ## Re: K201540 Trade/Device Name: uMR570 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: June 5, 2020 Received: June 9, 2020 Dear Xin GAO: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {2}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201540 Device Name uMR 570 Indications for Use (Describe) The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of the scan. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> </div> | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other and are in a bold, sans-serif font. To the right of the words is a stylized "U" shape that is dark gray. The logo is simple and modern. # 510 (k) SUMMARY ### 1. Date of Prepared June 5, 2020 ### 2. Sponsor Identification Shanghai United Imaging Healthcare Co., Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com ### 3. Identification of Proposed Device(s) Trade Name: uMR 570 Common Name: Magnetic Resonance Diagnostic Device Model: uMR 570 Product Code: LNH Regulation Number: 892.1000 Device Class: II # 4. Identification of Predicate Devices(s) Predicate Device #1 510(k) Number: K200024 Device Name: uMR 570 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH Reference Device #1 510(k) Number: K183063 Device Name: Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR {5}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized letter "U" that is also in bold. The color of the logo is a dark teal. ### Systems Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH # 5. Device Description The uMR 570 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 570 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards. The predicate device is uMR 570 (K200024) and the modification to the predicate device in this submission is the addition of a new pulse sequence, gre senc spiral which also exists in reference device of Philips Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K183063) . The modification performed on the predicate uMR 570 (K200024) in this submission is due to the addition of a new pulse sequence, gre_senc_spiral. The modification, which does not affect the intended use nor alters the fundamental scientific technology of the device, is as the following: > Introduce gre senc spiral as a new pulse sequence # 6. Intended Use The uMR 570 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan. # 7. Technological Characteristic {6}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in bold. The color of the logo is a dark teal. The differences from the predicate device are discussed in the comparison table in this submission is added below. The new sequence, gre_senc_spiral, is intended for acquisition of SENC(Strain ENCoding) cardiac images. The images are stored in DICOM format and processed by 3rd party software for strain analysis and report. | ITEM | Proposed Device<br>uMR 570 | Predicate Device<br>uMR 570 (K200024) | Remark | | | | | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------------------------------------|---------------------------------------|------|--------| | Product Code | LNH | LNH | Same | | | | | | Regulation No. | 21 CFR 892.1000 | 21 CFR 892.1000 | Same | | | | | | Class | II | II | Same | | | | | | Indications For<br>Use | The uMR 570 system is<br>indicated for use as a magnetic<br>resonance diagnostic device<br>(MRDD) that produces sagittal,<br>transverse, coronal, and oblique<br>cross sectional images, and<br>spectroscopic images, and that<br>display internal anatomical<br>structure and/or function of the<br>head, body and extremities.<br>These images and the physical<br>parameters derived from the<br>images when interpreted by a<br>trained physician yield<br>information that may assist the<br>diagnosis. Contrast agents may<br>be used depending on the<br>region of interest of the scan. | The uMR 570 system is indicated<br>for use as a magnetic resonance<br>diagnostic device (MRDD) that<br>produces sagittal, transverse,<br>coronal, and oblique cross<br>sectional images, and<br>spectroscopic images, and that<br>display internal anatomical<br>structure and/or function of the<br>head, body and extremities.<br>These images and the physical<br>parameters derived from the<br>images when interpreted by a<br>trained physician yield<br>information that may assist the<br>diagnosis. Contrast agents may<br>be used depending on the region<br>of interest of the scan. | Same | | | | | | Field Strength | 1.5 Tesla | 1.5 Tesla | Same | | | | | | Type of Magnet | ITEM | Superconducting | Proposed Device<br>uMR 570 | Superconducting | Predicate Device<br>uMR 570 (K200024) | Same | Remark | | Patient-<br>accessible bore<br>dimensions | | width 700mm,<br>height 530mm,<br>length 1500mm | 0.240ppm @ 20cm DSV | width 700mm,<br>height 530mm,<br>length 1500mm | 0.240ppm @ 20cm DSV | Same | | | Type of<br>Shielding | | Actively shielded, OIS<br>technology | 0.040ppm @ 10cm DSV | Actively shielded, OIS<br>technology | 0.040ppm @ 10cm DSV | Same | | | Magnet<br>Homogeneity | Gradient system | 1.4ppm @ 50cm DSV<br>0.9ppm @ 45cm DSV<br>0.72ppm @ 40cm DSV<br>0.420ppm @ 30cm DSV | | 1.4ppm @ 50cm DSV<br>0.9ppm @ 45cm DSV<br>0.72ppm @ 40cm DSV<br>0.420ppm @ 30cm DSV | | Same | | | Max gradient<br>amplitude | 45mT/m | 45mT/m | Same | | | | | | Max slew rate | 200T/m/s | 200T/m/s | Same | | | | | | Shielding | active | active | Same | | | | | | Cooling | water | water | Same | | | | | | RF system | | | | | | | | | Resonant<br>frequencies | 63.87 MHz | 63.87 MHz | Same | | | | | | Number of<br>transmit<br>channels | 1 | 1 | Same | | | | | | Number of<br>receive channels | Up to 48 | Up to 48 | Same | | | | | | Amplifier peak<br>power per<br>channel | 20 kW | 20 kW | Same | | | | | | RF Coils | | | | | | | | | Head & Neck<br>Coil -16 | Yes | Yes | Same | | | | | | Body Array Coil<br>- 6 | Yes | Yes | Same | | | | | | Body Array Coil<br>- 12 | Yes | Yes | Same | | | | | | Breast Coil - 10 | Yes | Yes | Same | | | | | | Flex Coil Large<br>- 4 | Yes | Yes | Same | | | | | | Flex Coil Small<br>- 4 | Yes | Yes | Same | | | | | | Knee Coil - 12 | Yes | Yes | Same | | | | | | Lower Extremity<br>Coil - 24 | Yes | Yes | Same | | | | | | Shoulder Coil -<br>12 | Yes | Yes | Same | | | | | | Small Loop Coil | Yes | Yes | Same | | | | | | Spine Coil - 24 | Yes | Yes | Same | | | | | | Wrist Coil - 12 | Yes | Yes | Same | | | | | | Cardiac Coil -<br>24 | Yes | Yes | Same | | | | | ## Table 1 Comparison of Hardware configuration {7}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is made up of two vertical bars with a horizontal bar connecting them at the top. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted. | ITEM | Proposed Device<br>uMR 570 | Predicate Device<br>uMR 570 (K200024) | Remark | |----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Foot & Ankle<br>Coil - 24 | Yes | Yes | Same | | Temporomandibular Joint Coil -<br>4 | Yes | Yes | Same | | Head Coil - 24 | Yes | Yes | Same | | Head Coil - 12 | Yes | Yes | Same | | Carotid Coil - 8 | Yes | Yes | Same | | Patient table | | | | | Dimensions | width 640mm, height 880mm,<br>length 2620mm | width 640mm, height 880mm,<br>length 2620mm | Same | | Maximum<br>supported<br>patient weight | 250 kg | 250 kg | Same | | Accessories | | | | | Vital Signal<br>Gating | ECG, Peripheral Pulse Gating,<br>Respiratory Gating | ECG, Peripheral Pulse Gating,<br>Respiratory Gating | Same | | Safety | | | | | Electrical Safety | Comply with ES60601-1 | Comply with ES60601-1 | Same | | EMC | Comply with IEC60601-1-2 | Comply with IEC60601-1-2 | Same | | Max SAR for<br>Transmit Coil | Comply with IEC 60601-2-33 | Comply with IEC 60601-2-33 | Same | | Max dB/dt | Comply with IEC 60601-2-33 | Comply with IEC 60601-2-33 | Same | | Biocompatibility | Patient Contact Materials were<br>tested and demonstrated no<br>cytotoxicity (ISO 10993-5), no<br>evidence for irritation and<br>sensitization (ISO 10993-10) | Patient Contact Materials were<br>tested and demonstrated no<br>cytotoxicity (ISO 10993-5), no<br>evidence for irritation and<br>sensitization (ISO 10993-10) | Same | Table 2 provides the new application software features of the proposed device in comparison to the predicate device. Table 2 Comparison of the new Application Software Features | ITEM | Proposed Device<br>uMR 570 | Reference Device<br>Ingenia and Ingenia CX, Ingenia<br>Elition, and Ingenia Ambition<br>MR Systems (K183063) | Remark | |------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------|--------| | Imaging Features | | | | | Strain Encoding (SENC) | Yes | Yes | Same | # 8. Non-Clinical Tests The following testing was conducted on the proposed devices: - Performance Evaluation Report SENC A {9}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold, sans-serif font. The color of the logo is a dark teal. The test results demonstrated that the device performs as expected and thus, it is substantial equivalent to the predicate devices to which it has been compared. # 9. Clinical Tests No clinical testing was conducted on the proposed devices. # 10. Conclusion In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we conclude that uMR 570 Magnetic Resonance Diagnostic Device is substantially equivalent to the predicate device. It does not introduce new indications for use, and has the same technological characteristics and does not introduce new potential hazards or safety risks.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...