The Innovision Audio system is intended to provide entertainment and facilitate communication from the operator to the patient in the MRI scanner environment. The system is intended for "MR Conditional" use during supine MRI examinations at 1.5T and 3.0T.
Device Story
Innovision Audio is a wireless system providing entertainment and communication between MRI operators and patients. Components include a wireless patient pillow, wireless bridges (data relays), and a Console Unit. Used in MRI scanner environments by healthcare professionals. The system transmits audio to the patient via the pillow, replacing traditional headphones. It is MR Conditional at 1.5T and 3.0T. Benefits include patient comfort and improved communication during imaging. The device is designed to be minimally disruptive to MRI workflows and does not interfere with MR image quality (SNR, uniformity, geometric distortion).
Clinical Evidence
Bench testing only. Testing included MR safety (RF/gradient heating <41°C), MR immunity, and NEMA MS 1, MS 2, and MS 3 standards to assess impact on image quality (SNR, uniformity, geometric distortion). Usability testing confirmed minimal disruption to MRI workflow. Compliance with IEC 60601-1, IEC 60601-1-2, and ISO 10993-1 was demonstrated.
Technological Characteristics
Wireless multi-component system (pillow, bridges, console). MR Conditional at 1.5T and 3.0T. Noise reduction rating >32dB. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), and NEMA MS 1-3 (MR image quality).
Indications for Use
Indicated for patients undergoing supine MRI examinations at 1.5T and 3.0T requiring entertainment or communication with the operator.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of a human figure, while the FDA name on the right is written in blue and includes the words "U.S. Food & Drug Administration."
April 7, 2020
Innovere Medical Inc. % Ms. Lynsie Thomason Regulatory & Operations Lead 6-250 Shields Court Markham, Ontario L3R 9W7 CANADA
## Re: K193218
Trade/Device Name: Innovision Audio Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: March 6, 2020 Received: March 9, 2020
Dear Ms. Thomason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193218
#### Device Name
Innovision Audio
#### Indications for Use (Describe)
The Innovision Audio system is intended to provide entertainment and facilitate communication from the MRS scanner environment. The system is intended for "MR Conditional" use during supine MRI examinations at 1.5T and 3.0T.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."
{3}------------------------------------------------
## SECTION 5 : 510(k) Summary
### 5.1 Applicant/Submitter
| Company Name | Innovere Medical Inc. |
|-----------------------------------|-------------------------------------|
| Establishment Registration Number | |
| Phone Number | 1-888-262-0408 |
| Company Street Address | 6-250 Shields Court |
| Fax Number | |
| City | Markham |
| State | ON |
| Country | Canada |
| Zip Code | L3R 9W7 |
| 5.2 Contact Person | |
| Full Name | Lynsie Thomason |
| Job Title | Regulatory & Operations Lead |
| Phone | 1-888-262-0408 (x104) |
| Email | lynsie.thomason@innoveremedical.com |
#### 5.3 Date of Preparation
| Date of Preparation | : 12/13/2019 |
|---------------------|--------------|
|---------------------|--------------|
#### 5.4 Device Information
#### Table - 5.1 Device Information
| Trade Name | Innovision Audio |
|----------------------|---------------------------------------|
| Common or Usual Name | Magnetic resonance diagnostic device. |
| Classification Name | 21 CFR 892.1000 |
| Regulatory Class | 2 |
| Product Code | LNH |
#### 5.5 Device Description
Innovision Audio is a wireless device intended to provide entertainment and facilitate communication from the operator to the patient in the MRI scanner environment. The Innovision Audio system is intended to be used by healthcare professionals.
Innovision Audio is a multi-component system comprised of a wireless patient pillow, wireless bridges (data relays), and a Console Unit (entertainment control centre).
#### 5.6 Predicate Device(s)
Table - 5.2 Predicate Device(s)
{4}------------------------------------------------
| Predicate Type | 510(k) Number | Device Name | Manufacturer |
|----------------|---------------|-----------------------|--------------|
| Primary Device | K173409 | Wireless Audio System | NeoCoil, LLC |
### 5.7 Comparison of Technological Characteristics with Predicate
Based on the evidence provided, the Innovision Audio System is substantially equivalent to the legally marketed device NeoCoil Wireless Audio System (K173409, cleared on 02/16/2018).
Use of the devices with the MRI scanner is similar.
The Innovision Audio system differs from the predicate in the following ways:
- · Innovision Audio has a higher Noise Reduction Rating of >32dB
- · Innovision Audio delivers audio to the patient through a pillow instead of traditional headphones/earbuds
- · Innovision Audio includes an entertainment audio source with the device
Bench testing demonstrates that the use of the Innovision Audio system is safe and effective and does not affect the safety or effectiveness of the MRI system when used as intended.
#### 5.8 Indications for Use
The Innovision Audio system is intended to provide entertainment and facilitate communication from the operator to the patient in the MRI scanner environment. The system is intended for "MR Conditional" use during supine MRI examinations at 1.5T and 3.0T.
#### 5.9 Testing
The testing summarized in Table - 5.3 Performance Testing - Bench and Table - 5.4 Published Standards Testing have been submitted, referenced or relied on to demonstrate the safety and effectiveness of Innovision Audio. The device performs as intended.
| Test | Pass/Fail Criteria | Results |
|-------------------------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| MR safety | Pre-defined performance standards | PASS<br>RF, gradient and combined fields<br>heating of applied parts is not greater<br>than 41°C. |
| NEMA MS 1<br>NEMA MS 2<br>NEMA MS 3 | Pre-defined performance standards | PASS<br>Innovision Audio has no appreciable<br>effect on MR image quality through<br>assessment of image SNR, signal<br>uniformity, geometric distortion. |
| MR Immunity | Pre-defined performance standards | PASS<br>Innovision Audio is immune to the<br>fields generated during MR imaging. |
| Usability of device | Pre-defined performance standards | PASS<br>Innovision Audio is usable and<br>minimally disruptive to the MRI<br>workflow. |
#### Table - 5.3 Performance Testing - Bench
#### Table - 5.4 Published Standards Testing
Standard
{5}------------------------------------------------
| IEC 60601-1 | Electrical safety |
|---------------|--------------------------------------------------|
| IEC 60601-1-2 | Electromagnetic compatiblity |
| ISO 10993-1 | Biocompatiblity |
| NEMA MS 1 | Signal-to-Noise Ratio (SNR) in MR images |
| NEMA MS 2 | 2-Dimensional geometric distortions in MR images |
| NEMA MS 3 | Image uniformity of MR images |
## 5.10 Conclusion
This submission demonstrates that for the specified indications for use the Innovision Audio system is as safe and effective as the predicate device, NeoCoil Wireless Audio System, K173406, as cleared on 02/16/2018.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.