K162183 · Toshibamedical Systems Corporation · LNH · Nov 25, 2016 · Radiology
Device Facts
Record ID
K162183
Device Name
Vantage Galan 3T, MRT-3020, V4.0
Applicant
Toshibamedical Systems Corporation
Product Code
LNH · Radiology
Decision Date
Nov 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density) - Spin-lattice relaxation time (T1) - Spin-spin relaxation time (T2) - Flow dynamics - Chemical Shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Galan 3T is a 3 Tesla MRI system utilizing a 1.6m short, 6.4-ton magnet and Pianissimo scan noise reduction technology. It captures proton density, T1, T2, flow dynamics, and chemical shift signals to generate cross-sectional images. Operated by trained clinicians in clinical settings, the system processes inputs via pulse sequences and reconstruction algorithms to produce diagnostic images for physician interpretation. Modifications include increased RF duty cycle (up to 100%), PURERF Rx/Tx, and new sequences (mUTE, mUTE 4D-MRA, 2D-RMC, MP2RAGE, mASTAR, FASE DWI). Software enhancements include EasyTech Cardiac (SUREVOI, CardioLine+) and improved SUREECG R-wave detection. Output assists physicians in diagnosing anatomical structures and pathologies. Benefits include reduced acoustic noise, improved motion correction, and advanced contrast/mapping capabilities.
Clinical Evidence
Bench testing, phantom imaging, and volunteer clinical imaging were performed. No clinical patient trials were required. Testing confirmed that new software functionalities and hardware modifications meet performance requirements and safety standards (IEC 60601-2-33).
Technological Characteristics
3T MRI system; 1.6m short magnet; 50x50x45 cm FOV. Features Pianissimo noise reduction, PURERF Rx/Tx, and 100% RF duty cycle. Complies with IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-33, 60825-1, 62304, 62366, and NEMA MS-1 through MS-5. Software includes sequence enhancements for fat suppression, motion correction, and cardiac/ECG-gated imaging.
Indications for Use
Indicated for diagnostic imaging of head or body anatomic structures in patients requiring cross-sectional transaxial, coronal, sagittal, or oblique MRI, including non-contrast MRA. Contraindicated for use with contrast agents outside of approved drug indications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2016
Toshiba Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K162183
Trade/Device Name: Vantage Galan 3T, MRT-3020, V4.0 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 14, 2016 Received: November 15, 2016
Dear Ms. Reyes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
For
Enclosure
{2}------------------------------------------------
Toshiba Medical Systems Corporation
510(k) Premarket Notification
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K162183
Device Name Vantage Galan 3T, MRT-3020, V4.0
Indications for Use (Describe)
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- · Proton density (PD) (also called hydrogen density)
- · Spin-lattice relaxation time (T1)
- · Spin-spin relaxation time (T2)
- · Flow dynamics
- · Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <b> <svg height="15" width="15"> <path d="M2 2 L13 13 M2 13 L13 2" stroke="black" stroke-width="2"></path> </svg> </b> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="15" width="15"> <rect fill="white" height="15" stroke="black" stroke-width="1" width="15"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the words "TOSHIBA MEDICAL" stacked on top of each other. The word "TOSHIBA" is on the top line, and the word "MEDICAL" is on the bottom line. Both words are in a bold, sans-serif font and are colored red. The background of the image is white.
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92
#### CLASSIFICATION AND DEVICE NAME: 1.
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Galan 3T, MRT-3020, V4.0 |
| Model Number: | MRT-3020 |
#### SUBMITTER'S NAME: 2.
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
#### OFFICIAL CORRESPONDENT: 3.
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
#### 4. CONTACT PERSON, U.S. AGENT AND ADDRESS:
#### Contact Person:
Janine Faith Reyes Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone No .: (714) 669-7853 E-mail: JFReyes@tams.com
#### 5. MANUFACTURING SITE:
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
### 6. ESTABLISHMENT REGISTRATION: 9614698
- 7. Date Prepared: August 2, 2016
- 8. DEVICE NAME: Vantage Galan 3T, MRT-3020, V4.0
**TOSHIBA AMERICA MEDICAL SYSTEMS, INC.**
2441 Michelle Drive, Tustin, CA 92780 / 714.730.5000 / medical.toshiba.com
### U.S. Agent:
Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone No.: (714) 669-7808 E-mail: PBiggins@tams.com
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is written in large, bold, red letters. Below it, the word "MEDICAL" is written in the same font and color, but slightly smaller. The logo is simple and clean, with a focus on the company name.
## 9. TRADE NAME(S):
Vantage Galan 3T, MRT-3020, V4.0
#### 10. CLASSIFICATION NAME:
Magnetic Resonance Diagnostic Device (MRDD)
#### 11. CLASSIFICATION PANEL:
Radiology
#### 12. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
#### 13. PRODUCT CODE:
90-гин
#### 14. PREDICATE DEVICE:
Primary Predicate Device (system): Vantage Titan 3T, MRT-3010, V3.5 Reference Predicate Device (added software functionalities): Vantage Titan 1.5T, MRT-1510, V3.6
| | Subject | Primary Predicate | Reference Predicate |
|----------------|-------------------------------------|-------------------------------------|---------------------------------------|
| System | Vantage Galan 3T,<br>MRT-3020, V4.0 | Vantage Titan 3T,<br>MRT-3010, V3.5 | Vantage Titan 1.5T,<br>MRT-1510, V3.6 |
| Marketed By | Toshiba America<br>Medical Systems | Toshiba America<br>Medical Systems | Toshiba America<br>Medical Systems |
| 510(k) Number | This Submission | K152371 | K160632 |
| Clearance Date | | October 23, 2015 | 510(k) Pending |
#### 15. REASON FOR SUBMISSION
Modification of a cleared device.
#### 16. DEVICE DESCRIPTION
The Vantage Galan 3T (Model MRT-3020) is a 3 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Galan 3T uses 1.6m short and 6.4 tons light weight magnet. It includes the Toshiba Pianissimo™ technology (scan noise reduction technology). The design of the gradient coil and the WB coil of the Vantage Galan 3T provides the maximum field of view of 50 x 50 x 45 cm. The 3T Vantage Galan MRI System is comparable to the current 3T Vantage Titan MRI System (K152371), cleared October 23rd, 2015 with the following modifications.
#### 17. SUMMARY OF HARDWARE CHANGES
This submission is to report the following hardware changes have been made:
- RF system: ●
- Maximum RF duty cycle changed from 50% to 100% o
- PURERF Rx/Tx O
AMERICA MEDICAL SYSTEMS.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is written in large, bold, red letters. Below it, the word "MEDICAL" is also written in large, bold, red letters, but slightly smaller than "TOSHIBA". The logo is simple and clean, with a focus on the brand name.
#### 18. SUMMARY OF SOFWARE CHANGES
This submission is to report the following software functionalities have been added:
- Sequence Enhancement:
- FFE3D: o
●
- SPAIR - pre-pulse for fat suppression
- T2 Prep - pre-pulse for T2 weighted contrast
- mUTE - Ulta short TE T1 weighted sequence with reduced acoustic noise
- mUTE 4D-MRA Ultra short TE time resolved MR angiography with reduced ■ acoustic noise
- 2D-RMC - Two dimensional real-time motion correction
- O MP2RAGE - Images with multiple TI (Inversion time) to create T1 map
- FSE2D:mEcho Images with multiple TE (Echo time) to create T2 map O
- FFE2D:mEcho Images with multiple TE (Echo time) to create T2* map O
- mASTAR Multiple phase ASL (Arterial Spin Labeling) O
- FASE DWI Spin Echo based Diffusion Weighted Imaging with reduced acoustic O noise
- Enhancement of SPEEDER post-processing
- EasyTech Cardiac - SUREVOI and CardioLine+
- lmprovement of SUREECG R-wave detection ●
#### 19. SAFETY PARAMETERS
| Item | Subject Device: Vantage Galan<br>3T, MRT-3020, V4.0 | Predicate Device: Vantage<br>Titan 3T, MRT-3010, V3.5<br>(K152371) | Notes |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------|
| Static field strength | 3T | 3T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter<br>display | SAR dB/dt | SAR dB/dt | Same |
| ii. Operating mode<br>access requirements | Allows screen access to 1st level<br>operating mode | Allows screen access to 1st level<br>operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33(2010)) | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33(2010)) | Same |
| Maximum dB/dt | 1st operating mode specified in<br>IEC 60601-2-33 (2010) | 1st operating mode specified in<br>IEC 60601-2-33 (2010) | Same |
| Potential emergency<br>condition and means<br>provided for shutdown | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Same |
TOSHIBA AMERICA MEDICAL SYSTEMS, INC. 441 Michelle Drive, Tustin, CA 92780 / 714.730.5000 / medical.tosh
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is written in a bold, red font on the top line. Below it, the word "MEDICAL" is written in the same bold, red font. The logo is simple and clean, with a focus on the company name.
#### 20. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K152371.
#### 21. INDICATIONS FOR USE
Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift ●
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
No changes to the previously cleared indication, K152371.
#### 22. SUMMARY OF DESIGN CONTROL ACTIVITIES
Hazard analysis has been performed and documentation is included in this submission. PS Risk List for new software functionalities and pulse sequences are included in this submission. The test methods used are the same as those submitted in the previously cleared submissions (K152371). A declaration of conformity with design controls is included in this submission.
#### 23. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):
HIBA AMERICA MEDICAL SYSTEMS, INC · Drive, Tustin, CA 92780 / 714.730.5000
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is in large, bold, red letters. Below it, the word "MEDICAL" is also in large, bold, red letters. The logo is simple and eye-catching, with the red color conveying a sense of urgency and importance.
### LIST OF APPLICABLE STANDARDS
- IEC60601-1:2005 ●
- IEC60601-1-2:2007
- IEC60601-1-8:2006,Amd.1:2012 ●
- IEC60601-2-33:2010 ●
- IEC60825-1: 2007 ●
- IEC62304:2006 ●
- o IEC62366:2007+Amd.1:2014
- NEMA MS-1:2008 ●
- NEMA MS-2:2008 ●
- NEMA MS-3:2008 ●
- NEMA MS-4:2010 ●
- NEMA MS-5:2010 ●
#### 24. TESTING
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
#### 25. SUBSTANTIAL EQUIVALENCE
Toshiba Medical Systems Corporation believes that the Vantage Galan 3T, MRT-3020, V4.0 Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate device referenced in this submission. Toshiba Medical Systems Corporation believes that the changes incorporated into the Vantage Galan 3T, MRT-3020, V4.0 are substantially equivalent to the previously cleared predicate device.
#### 26. CONCLUSION
The modifications incorporated into the Vantage Galan 3T, MRT-3020, V4.0 do not change the indications for use or the intended use of the device. Based upon bench testing, phantom imaging, volunteer clinical imaging, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.