← Product Code [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH) · K143606

# MultiBand SENSE (K143606)

_Philips Medical Systems Nethrlands BV · LNH · May 8, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K143606

## Device Facts

- **Applicant:** Philips Medical Systems Nethrlands BV
- **Product Code:** [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH.md)
- **Decision Date:** May 8, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

MultiBand SENSE is a software option intended for use on Achieva and Ingenia 1.5T & 3.OT MR Systems. It's indicated for use in magnetic resonance imaging of the brain for BOLD fMRI. MultiBand SENSE consists of an acquisition and reconstruction technique allowing simultaneous excitation of multiple volumes to accelerate imaging acquisition times or increasing coverage or resolution without increasing scan time.

## Device Story

MultiBand SENSE is a software option for Philips 1.5T/3.0T Achieva and Ingenia MR systems; enables simultaneous excitation/acquisition of multiple volumes/slices. Input: MR signals acquired via pulse sequences using multi-band RF pulses and blip-gradients. Transformation: SENSE algorithm unfolding; linear phase shift introduced in k-space via blip-gradients; coil sensitivity data translated during reconstruction to correct for phase shift. Output: MR images with accelerated acquisition times or increased coverage/resolution. Used in clinical radiology settings; operated by MR technicians/radiologists. Healthcare providers view/archive images via standard clinical workflows. Benefits: reduced scan times or increased imaging resolution/coverage for BOLD fMRI.

## Clinical Evidence

Bench testing and clinical validation performed on Achieva and Ingenia 1.5T/3.0T systems. Bench testing verified loading of examcards, parameter accuracy, and image generation with SNR deviation <10%. Clinical validation confirmed that MultiBand SENSE enabled shorter scan times for fMRI compared to standard SENSE, or allowed increased slice counts within the same scan time. No new hazards identified.

## Technological Characteristics

Software-only feature for existing MR systems. Utilizes modulated RF pulses and blip-gradients for signal acquisition. Reconstruction uses SENSE unfolding with new parameterization and coil sensitivity shifting. Runs on host computer (HP Z420, 3.5 GHz, 64 GB RAM, Windows 7 64-bit). No hardware changes required. Not networked to external systems.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SENSE ([K110151](/device/K110151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other, with flowing lines connecting them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2015

Philips Medical Systems, Nederland B.V. % Ms. Susan Quick Regulatory Affairs Specialist Philips Medical Systems (Cleveland), Inc. 595 Miner Road CLEVELAND OH 44143

Re: K143606

Trade/Device Name: Multiband Sense Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH Dated: April 17, 2015 Received: April 20, 2015

Dear Ms. Quick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara for

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| PSC Publishin<br>Services (301 ) 443-6741)<br>नज                                               | ။ ၂၀၂ ခုနှစ်မြ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | (ANS) 1888 AD3 MAO3                                           |
|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| 11                                                                                             | toodsa redto your and elamitee nebud sint guibregen sinematii to noitemorii to noitent pre<br>ə qaran bərən məbədə qalan mənələr çənin ərazi varian qalınır. Bu və və və qalında və və və və və və və və və və və və və və və və və və v<br>:of ,nebud sint prisuber 10 ຂາດ ນ້າ ຂອງ ຫຼຸບ ຂ<br>əd bulunur. Əsnoqsən rəq sunda çələnirə əsi notibatinə sinfi növünəlir. İstinadlar   fi<br>to woll and production in periupes ton si nosted a business to toubuco you year your and<br>. WOJER 229800A JIANS 37718 ASS 3HT OT MAO3 GITTY MAO AU0Y GN3S TON OG*<br>See to to to A noticuben your and to stremerings of your seilgge noilibes sin T<br>้างอิง ที่ 2000 biles ประเทศที่ 19 รอบคน แบบสถานี<br>Designes for Health and Human Services<br>the started in Art Art (Provised State) States<br>Office of Chief Information Officer<br>Food and Drug Administration<br>vop.syy.repl@yglys?SAAA |                                                               |
|                                                                                                | MINDO STARATE PARATE PARATE PARRETE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                               |
| (2 ກາຣປຸດັກ 108 ກຸລັງ 12) ອະປີ 1970 ໄດ້ ປະຈຸບັນດວົງອາໄປ-1990 ປີ                                | (0) heqdus 108 Я72 12 Fra 81 Craped D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Type of Use (Select one or on ont, as applicable)             |
|                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | .emit mass guisseroni inodition noitulosen no sgrings son tri |
|                                                                                                | າດ ອອການ ແດງໃນເມືອງທຸ່ມອາລາວໄທວາ ວ່າ ເປນວັນທີ່ 10 ກຸດທີ່ສຳວັນ ຂອງທີ່ເກີດ ອັກສາຫະກຳການປະກວດນາງງາມ<br>ນາເຣນນີ້ໄຊ້ເຖລາ ແລະ ໃນ ຂ່າວປະເທດ ປີ ເປັນ ປີ ເດີຍ ປີ ເດີຍນີ ອກ ການສະຫຼຸບ ອາຫາດຂະໜາວທີ່ອາງສາຕາ ແລະມາການ ການ ການປະກວ<br>ini a fil . smatze AM T0 . E & T2.1 sincent bass svoids A no seu rot bebreini noideo e zi ZNGC bara iti                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                               |
|                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | (addrosed) est Use (Describe                                  |
|                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | עוואשער אין אין אין אין אין איינער פ<br>Device Name           |
|                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 500(k) Number (if known)<br>ג 1436666                         |
| 02/0-0180 .0M 8MO :bevonqqA mio	T<br>see PRA Statement below<br>and the Date: January 31, 2017 | 2301VЯЭг ИАМИН ОНА НТЈАЭН ЗО ТИЭМТЯАЧЭО<br>Indications for Use<br>FoireinimbA ഉมาQ bus bood                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                               |
|                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                               |

{3}------------------------------------------------

# PHILIF

# Philips Medical Systems Nederland B.V.

# 510(k) Summary

# MultiBand SENSE

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

- I. General Information 21 CFR 807.92 (a)(1), (2)

| Company Name:         | Philips Medical Systems Nederland B.V.                                                                                              |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Address:              | Veenpluis 4-6<br>5684 PC  Best<br>The Netherlands                                                                                   |
| Registration No .:    | 3003768277                                                                                                                          |
| Contact Person:       | Susan Quick<br>595 Miner Rd<br>Cleveland, Oh 44143<br>Tel: (440)-483-2291<br>Fax: (440)-483-4799<br>E-mail: susan.quick@philips.com |
| Prepared (date):      | 2014 December 12                                                                                                                    |
| Trade Name of Device: | MultiBand SENSE                                                                                                                     |
| Classification:       | Class II                                                                                                                            |
| Regulatory Section:   | Magnetic Resonance Diagnostic Device.<br>892.1000                                                                                   |
| Product Code:         | 90LNH                                                                                                                               |

{4}------------------------------------------------

510(k) Summary MultiBand SENSE

# 21 CFR 807.92(a)(3):Legally marketed predicate device to which substantial equivalence is claimed:

| 1. Primary Predicate Device: SENSE |                                      |
|------------------------------------|--------------------------------------|
| Manufacturer:                      | Philips Medical Systems Nederland BV |
| Predicate Device k#:               | K110151                              |

# 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification:

#### Summary of functions of the device and its major components

The MultiBand SENSE technique enables simultaneous excitation and acquisition of multiple volumes or slices for the purpose of speeding up acquisition times or increasing coverage or resolution at constant scan time. The simultaneous excitation is done using a multi-band radiofrequency pulse. The unfolding of the simultaneously acquired volumes is done using the SENSE algorithm. The unfolding process (solving the linear equation of the SENSE algorithm) is improved by introducing a linear phase over k-space in the volume direction resulting in a spatial shift of the aliased pixels. The phase shift is applied by additional blip-gradients in the slice direction or switching between different RF pulses, and compensated for in reconstruction by a translation of the coil sensitivity data before the SENSE unfolding. The feature consists of:

- Modulated RF pulses exciting 2 or more slices
- Blip-gradients to introduce a phase shift for improved unfolding ●
- New parameterization of the SENSE calculations ●
- Shifting coil sensitivities in reconstruction to correct for linear phase shift. ●

MultiBand SENSE is supported on the following systems:

- . 3.0T Ingenia
- 3.0T Achieva ●
- 1.5T Ingenia ●
- . 1.5T Achieva

The functionality is supported on all available gradient performance levels. Optimized protocols will be provided for the different performance points. MultiBand SENSE is supported on the centralized data acquisition systems of the Achieva systems as well as the digitally networked data acquisition system of the Ingenia systems. The data acquisition system is fully transparent to the MultiBand SENSE pulse sequences and reconstructions.

{5}------------------------------------------------

## 21 CFR 807.92(a)(5): Intended Use

MultiBand SENSE is a software option intended for use on Achieva and Ingenia 1.5T & 3.OT MR Systems. It's indicated for use in magnetic resonance imaging of the brain for BOLD fMRI. MultiBand SENSE consists of an acquisition and reconstruction technique allowing simultaneous excitation of multiple volumes to accelerate imaging acquisition times or increasing coverage or resolution without increasing scan time.

#### 21 CFR 807.92(a)(6): Technological Characteristics:

The main functional units in the software are:

- Methods (acquisition of MR signals by means of MR pulse sequences) -
- Reconstruction (transforming the MR signals to images) -
- -Patient Administration (storing of the images in the database and providing access)
- Viewing (display of images) -

The technical impact of the feature MultiBand SENSE comprises:

- -Methods: Introduction of a modulated RF pulse. Apply blipped gradients. Provide new parameterization for SENSE reconstruction.
- Reconstruction: Read new parameterization of SENSE calculations. Shift coil sensitivities before SENSE calculations.

No off-the-shelf software is used for the feature MultiBand SENSE. The off-the-shelf software used in the basic MR system is cleared. MultiBand SENSE is not designed to be connected to an external network.

MultiBand SENSE does not require any change of the hardware platform. The extension introduced by Multiband SENSE, are in methods pulse sequence code, and in reconstruction only for a new parameterization of a cleared SENSE unfolding calculation. Those run on the host computer characteristics :

- Manufacturer: HP; Model: Z420; Processor clock: 3.5 GHz; RAM: 64 GB RAM; -Processors: six core with hyper threading
- Operating system: Windows 7, 64 bits -

The only new element for the operator of the Multiband SENSE feature in this clinical routine workflow is:

- Protocol selection: The operator selects an ExamCard with Multiband SENSE protocols -
- Planscan phase: Optionally the operator may want to change the predefined Multiband acceleration factor.

All other steps are not changed. The generated image types can be viewed, post-processed, printed and archived as any other image type.

{6}------------------------------------------------

### 21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:

MultiBand SENSE has been verified to function with the Achieva and Ingenia 1.5T and 3.0T MR systems.

The verification testing showed the MultiBand SENSE examcards could be loaded, the correct parameters were listed for each scan. MultiBand SENSE functioned properly together, the average SNR deviation was <10% for every slice, all scans ran properly and images were provided.

The conclusion from this report is: All the tests performed for MultiBand SENSE were successful. Workflow was smooth and no problems occurred. No defects were reported.

## 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:

Clinical user needs are tested as part of validation. The validation testing showed that for fMRI scans with MultiBand SENSE enabled the scan time was shorter when compared to fMRI with MultiBand disabled for the same number of slices. Also the user could increase the number of slices when using MultiBand SENSE under the scan time recorded without MultiBand SENSE.

The conclusion from testing the device is:

The clinical validation of MultiBand SENSE has completed successfully. All clinical user needs have passed for MultiBand SENSE on the Achieva and Ingenia 1.5T and 3T systems. No new hazards were identified.

## 21 CFR 807.92(b)(3): The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section:

The nonclinical and clinical tests have demonstrated that the device is safe and works according to its intended use.

MultiBand SENSE software does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers MultiBand SENSE Software to be substantially equivalent to the above mentioned predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K143606](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K143606)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
