VANTAGE TITAN WITH HELIOS GRADIENT MODEL MRT-1504/US
K122613 · Toshiba America Medical Systems, In.C · LNH · Dec 13, 2012 · Radiology
Device Facts
Record ID
K122613
Device Name
VANTAGE TITAN WITH HELIOS GRADIENT MODEL MRT-1504/US
Applicant
Toshiba America Medical Systems, In.C
Product Code
LNH · Radiology
Decision Date
Dec 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MRI system is indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: Proton density (PD) (also called hydrogen density), Spin-lattice relaxation time (T1), Spin-spin relaxation time (T2), Flow dynamics, Chemical shift. Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Titan (MRT-1504/U5) is a 1.5T MRI system; utilizes proton density, T1/T2 relaxation, flow dynamics, and chemical shift to generate diagnostic images. System incorporates Helios gradient hardware (amplifier and coil) for modified gradient performance. Operated by trained clinicians in clinical settings to produce cross-sectional images of head/body. Output interpreted by physicians to aid diagnosis. System supports non-contrast MRA. Improvements include 3D Advanced Fourier Imaging (AFI), updated database software (Altibase), and new specialized coils (cardiac, head, spine, flexible SPEEDER). Safety features include SAR and dB/dt monitoring, emergency ramp-down unit, and compliance with IEC 60601-2-33. Device provides anatomical visualization to assist clinical decision-making.
Clinical Evidence
Bench testing only. Image quality metrics were evaluated using phantoms. System performance was verified in accordance with applicable IEC and NEMA standards for MRI safety and imaging performance. No clinical data provided.
Technological Characteristics
1.5T MRI system. Components: Helios gradient amplifier and coil. Connectivity: DICOM (NEMA PS 3.1-18). Standards: IEC 60601-1, IEC 60601-2-33, NEMA MS-1 through MS-5. Software: Altibase database. Features: 3D Advanced Fourier Imaging (AFI), wireless gating, and specialized coils (cardiac, head, spine, flexible SPEEDER).
Indications for Use
Indicated for diagnostic imaging of head or body anatomic structures in patients requiring cross-sectional transaxial, coronal, sagittal, or oblique MRI or non-contrast MRA. Contraindicated for use with contrast agents outside of approved drug indications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# ATTACHMENT G
Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) Vantage Titan, v2.1, MRT-1504/U5
# 510(k) SUMMARY AND EFFECTIVENESS
# DEC 1 3:2012
Page 1 of 5
# 1. SUBMITTER'S NAME: Toshiba America Medical Systems, Inc.
- 2. ESTABLISHMENT REGISTRATION: 2020563
#### CONTACT PERSON AND U.S AGENT INFORMATION:
Charlemagne Chua Contact Person: (714) 730-5000
U.S. Agent Name:
Paul Biggins (714) 730-5000
2441 Michelle Drive Tustin, Ca. 92780
Establishment Name and Address:
## MANUFACTURING SITE
Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550
Toshiba America Medical Systems, Inc.
Japan
## 5. DATE OF SUBMISSION:
- 6. DEVICE NAME:
Generic Name: Model Name: Trade/ Proprietary Name: Magnetic Resonance Diagnostic Device MRT-1504/U5 Vantage Titan with Helios gradient
August 24, 2012 (Revised 11-5-2012)
## 7. CLASSIFICATION AND CLASS OF DEVICE
90-LNH, Class II per 21 CFR 892.1000
K122613-S1: Response to Al (10-9-12) Page 33 of 65
{1}------------------------------------------------
#### PREDICATE DEVICE(S):
- K120638: Vantage Titan, MRT-1504/U4
- K112003: Vantage Titan HSR, MRT-1504/A5
#### REASON FOR SUMBISSION 9.
Modification of a cleared device
## 10. Submission Type Traditional 510(k)
#### 11. DEVICE DESCRIPTION
The Vantage Titan with Helios gradient (Model MRT-1504/U5) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Titan with Helios gradient uses the same magnet as the Vantage Titan (K120638). The gradient performance was modified using. the same gradient amplifier and gradient coil as Vantage Titan HSR (K112003).
# 12. INTEDED USE
The MRI system is indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body : In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
Proton density (PD) (also called hydrogen density),
- Spin-lattice relaxation time (T1),
Spin-spin relaxation time (T2);
- Flow dynamics,
Chemical shift.
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
No changes from the previous submission, K120638
{2}------------------------------------------------
Page 3 of 5
#### 13. SUMMARY OF MAJOR HARDWARE CHANGES
- Gradient Amplifier
- Gradient Coil : b.
-
#### 14. SUMMARY OF MAJOR SOFTWARE CHANGES
- a. There are no software changes.
### 15. SUMMARY OF IMPROVEMENTS
- MRS Scan a.
- b. Altibase (Database SW), version upgrade
... ...
- c. . Scan parameter window -----
- ರ. Reconstruction speed
- e. Locator window operation
- Clinical application operability f.
- Autoview GUI g.
- Automatic map-scan h.
- Implementation of 3D Advanced Fourier Imaging (AFI) i.
- New coils (cardiac, Head, Spine and Flexible SPEEDER) i:
: : :
### <Next page>
K122613-S1: Response to AI (10-9-12) Page 35 of 65
{3}------------------------------------------------
Toshiba America Medical Systems, Inc.
Pre-Market Notification 510(k) Vantage Titan, v2.1, MRT-1504/U5
#### 16. SAFETY PARAMETERS
| | Vantage Titan with<br>Helios gradient<br>(Subject device) | Vantage Titan<br>(K120638) | Vantage Titan HSR<br>(K112003) |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| a. Static field<br>strength: | 1.5T | 1.5T | 1.5T |
| b. Peak and A-<br>weighted acoustic<br>noise: | 112.0 dB (A-weighed)<br>122.9 dB(peak) | 106.2 dB (A-weighted)<br>115.4 dB (peak) | 113.0 dB (A-weighted)<br>121.6 dB (peak) |
| c. Operational<br>modes: | 1st operating mode | 1st operating mode | 1st operating mode |
| i. Safety<br>parameter<br>display: | SAR, dB/dt | SAR, dB/dt | SAR, dB/dt |
| ii. Operating<br>mode access<br>requirements: | Allows screen access to<br>1st level operating mode | Allows screen access to<br>1st level operating mode | Allows screen access to<br>1st level operating mode |
| d. Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified<br>in IEC 60601-2-33 (2002)) | 4W/kg for whole body (1st<br>operating mode specified<br>in IEC 60601-2-33 (2002)) | 4W/kg for whole body (1st<br>operating mode specified<br>in IEC 60601-2-33 (2002)) |
| e. Maximum dB/dt | <1st operating mode<br>specified in IEC 60601-2-<br>33 (2002) | <1st operating mode<br>specified in IEC 60601-2-<br>33 (2002) | <1st operating mode<br>specified in IEC 60601-2-<br>33 (2002) |
| f. Potential<br>emergency<br>conditions and<br>means provided for<br>shutdown: | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard by<br>ferromagnetic objects | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard by<br>ferromagnetic objects | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard by<br>ferromagnetic objects |
| g. Biocompatibility<br>of materials | Confirmed for electrodes<br>and accessories for<br>wireless gating | Confirmed for electrodes<br>and accessories for<br>wireless gating | Confirmed for electrodes<br>and accessories for<br>wireless gating |
# 16. IMAGING PERFORMACE PARAMETERS
No change from the previous predicate submission (K120638).
{4}------------------------------------------------
#### 17. DESIGN CHANGE
The Vantage Titan with Helios gradient MRI System is comparable to the existing 1.5T Vantage Titan MRI System (K120638), with the following modifications.
- Maximum gradient slew rate have been changed. a.
- Maximum gradient strength has been changed. b.
#### 18. SUMMARY OF DESIGN CONTROL ACTIVITIES
PS Risk List for software and hardware changes have been is included in this submission. The test methods used are the same as those submitted in the previously cleared submissions (K120638). A declaration of conformity with design controls is included in this submission.
#### 19. TRUTHFUL AND ACCURACY CERTIFICATION
A certification of the truthfulness and accuracy of the Vantage Titan with Helios gradient described in this submission is provided in this submission.
#### 20. SUBSTANTIAL EQUIVALENCE
Toshiba Medical Systems Corporation believes that the Vantage Titan with Helios gradient (model MRT-1504/U5). Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate devices referenced in this submission.
Testing was done in accordance with applicable recognized consensus standards as listed below.
#### List of Applicable Standards
- IEC60601-1:1988, Amd.1:1991, Amd.2:1995
- IEC60601-1-1:2000
- IEC60601-1-2:2001, Amd.1:2004 ........
- IEC60601-1-4:1996, Amd.1:1999
- IEC60601-1-6:2006
- IEC60601-1-8:2003,Amd.1:2006
- IEC60601-2-33:2002, Amd.1:2005, Amd.2:2007
- IEC60825-1: 2007
- IEC62304:2006
- IEC62366:2007
- NEMA MS-1:2008
- NEMA MS-2:2003
- NEMA MS-3:2008
- NEMA MS-4:2006
- NEMA MS-5:2003
- NEMA PS 3.1-18 (2008)
#### 21. TESTING
Image Quality metrics utilizing phantoms are provided in this submission. Additionally, testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices.
#### 22. CONCLUSION :
The modifications incorporated into Vantage Titan with Helios Gradient (MRT-1504/U5), SW V2.20, do not affect the indications for use or the intended use of the device. Safety and effectiveness have been verified via risk management and application of design controls to the modifications,
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 13, 2012
Toshiba Medical Systems Corporation C/O Mr. Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K122613
Trade/Device Name: Vantage Titan, MRT-1504/U5 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 5, 2012 Received: November 6, 2012
Dear Mr. Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rollowed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the encreate) to togen. Jo togen. Januaries date of the Medical Device Amendments, or to conninered prior to May 20, sire) in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic rice (110) that the novice, subject to the general controls provisions of the Act. The I ou may, dierelore, manol of the Act include requirements for annual registration, listing of general obliations provision practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Dr o levice complies with other requirements of the Act that I Drimas intatutes and regulations administered by other Federal agencies. You must or any I outlar statutes and registements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CI K Fart 607); mooning (DI CFR 803); good manufacturing practice requirements as set de rice-relation daverse overles) (1 CFR Part 820); and if applicable, the electronic forth in the quality of evelsions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Janine M. Morris -S
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
Vantage Titan, MRT-1504/U5
Indications for Use:
The MRI system is indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density),
K122613
- Spin-lattice relaxation time (T1),
- Spin-spin relaxation time (T2),
- Flow dynamics,
- Chemical shift.
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Over-The-Counter Use AND/OR Prescription Use × (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Janine M. Morris -S 2012.12.13 13:47:51 -05'00'
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
K122613 510(k) Number
Page 1 of _1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.