K121545 · Medtec Inc. Dba Civco Medical Solutions · LNH · Jun 8, 2012 · Radiology
Device Facts
Record ID
K121545
Device Name
BELLYBOARD, SBRT ACCESSORIES
Applicant
Medtec Inc. Dba Civco Medical Solutions
Product Code
LNH · Radiology
Decision Date
Jun 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
CIVCO Patient Positioning Devices are used to aid in the support and positioning of patients during MR, radiological, and other procedures.
Device Story
Patient positioning and immobilization devices for radiotherapy procedures; includes Bellyboard and SBRT accessories (transfer systems, plates, straps). Devices support patient during MR and radiological imaging. Constructed from non-magnetic, plastic, and fiberglass materials. Used in clinical settings by healthcare professionals. Provides stable patient positioning to facilitate accurate imaging and treatment delivery. MR conditional labeling applied due to observed image artifacts in specific SBRT components.
Clinical Evidence
Bench testing only. MR compatibility testing performed per ASTM F2182-9, F2119-07, F2052-06, and F2213-06. Results demonstrated safety for RF heating, magnetic induced torque, and displacement force. Biocompatibility testing completed for patient-contacting materials. Image artifact observed in SBRT Patient Transfer System addressed via MR conditional labeling.
Technological Characteristics
Non-implanted, non-magnetic plastic and fiberglass construction. MR conditional. ASTM standards used: F2182-9 (RF heating), F2119-07 (image artifact), F2052-06 (displacement force), F2213-06 (torque). Sold non-sterile; reusable. No electronic or software components.
Indications for Use
Indicated for support and positioning of patients during MR, radiological, and other procedures.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the CIVCO Medical Solutions logo, which includes a stylized globe graphic and the company name in bold letters. To the right of the logo, there is a handwritten number "k121545" and the date "JUN - 8 2012" printed below it. The date indicates that the image or document is from June 8, 2012.
WWW.CIVCO.COM
## Section 5.0 - 510(k) Summary
## CIVCO MR & Radiological Patient Positioning Devices
### A. General Information
| Submitter's Name: | MEDTEC, Inc. d/b/a CIVCO Medical Solutions |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1401 8th Street SE, Orange City, Iowa 51041 |
| Telephone No.: | 319-248-6628 |
| Fax No.: | 877-218-0324 |
| Contact Person: | Amanda Stahle, Regulatory Affairs Specialist |
| Date Summary Prepared: | March 16, 2012 |
| Trade Name: | Bellyboard, SBRT Accessories |
| Common Name: | MR & Radiological Patient Positioning Devices |
| Classification Names: | System, Nuclear Magnetic Resonance Imaging<br>(21 CFR 892.1000, Product Code LNH)<br>Accelerator, Linear, Medical<br>(21 CFR 892.5050. Product Code IYE) |
### B. Predicate Devices
MEDTEC, Inc., doing business as CIVCO Medical Solutions, claims the proposed devices to be substantially equivalent to the following devices devices previously cleared by the FDA in the following 510(k)s:
- Bellyboard
- O K060737 Sinmed Immobilization Systems (Head and Shoulder, Lung and Thorax, Pelvis and Lower Extremities, and Posicast) and Sinmed Repovac Cushions
#### SBRT Accessories ●
- o K973842 MEDTEC, Inc. Carbon Fiber Conformal Couch Top
The predicate devices have been used for many years in radiological and other medical procedures. The purpose of this 510(k) is to have these products cleared for use in the MR environment.
The predicate devices have the same intended use with the exception of use in the MR environment. The design and materials have been changed where necessary to use these products in the MR environment. CIVCO has conducted testing to demonstrate that the changes in design and materials do not affect the safety and effectiveness of the device.
| Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | ⓞ 319.656.4447 | ⓞ 319.656.4451 |
|------------------------|------------------------|-----------------------|-----------------|---------------------|---------------------|
| Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | ⓞ +31(0) 182.394495 | ⓞ +31(0) 182.395014 |
| Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | ⓞ 712.737.8688 | ⓞ 712.737.8654 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the text "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font. The overall design is clean and modern.
### WWW.CIVCO.COM
### C. Device Description
The MR & Radiological Patient Positioning devices are used to aid in patient positioning and immobilization during radiotherapy procedures. The following models of the Patient Positioning Devices are included in this submission:
- . Bellyboard
126000 Bellyboard, which includes the following: 301101 Bellyboard Cushion
#### o SBRT Accessories
MTSBRT011 SBRT Patient Transfer System, which includes the following: MTSBRT012 Gurney Plate Assembly
MTSBRT032 Transfer System Trans Slot MTSBRT033 Transfer System Longitudinal Slot MTSBRT014 Ski Assembly 20-CFHN-017 Type-S Strap
Devices are sold non-sterile and may be reused for multiple patient or single patient use if set-up for a single patient. The proposed devices are non-implanted devices i that are large in size and manufactured of non-magnetic and plastic materials.
### D. Intended Use/Indications for Use
CIVCO Patient Positioning Devices are used to aid in the support and positioning of patients during MR, radiological, and other procedures.
### E. Technological Characteristics
Technological characteristics which have changed between the proposed and predicate devices include changes in the design and materials which were tested for use in a MR environment to ensure these differences do not affect the safety and effectiveness of the device. These changes include the use of fiberglass instead of carbon fiber, and the addition of a third hole to secure the device in the MR environment, preventing non-MR compatible devices from being used in the MR environment.
### F. Non-Clinical Testing
MR compatibility test methodology generally followed ASTM Standards F2182-9, F2119-07, F2052-06 and F2213-06, but modifications were made to accommodate the large size of the proposed devices and to accommodate their external use (not implanted). The devices passed the acceptance criteria for RF heating, magnetic
| Corporate Headquarters | | 102 First Street South | | Kalona, IA 52247 | | USA | | ⓮ 319.656.4447 | | ⓕ 319.656.4451 |
|------------------------|--|------------------------|--|-----------------------|--|-----------------|--|---------------------|--|---------------------|
| Europe Office | | Pasteurstraat 6 | | 2811 DX Reeuwijk | | The Netherlands | | ⓮ +31(0) 182.394495 | | ⓕ +31(0) 182.395014 |
| Orange City Office | | 1401 8th Street SE | | Orange City, IA 51041 | | USA | | ⓮ 712.737.8688 | | ⓕ 712.737.8654 |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the company name in bold, sans-serif font. The words "MEDICAL SOLUTIONS" are printed in a smaller font size below the company name.
### WWW.CIVCO.COM
induced torque, and magnetically induced displacement force and demonstrate that the device is safe for use in the MR environment. Image artifact was observed in a specific area of the SBRT Patient Transfer System, but the device has been labeled MR conditional to account for this artifact. Biocompatibility testing was also completed for patient contacting materials.
### G. Conclusion
ﺎ ﻓﻲ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
This premarket submission for the CIVCO MR & Radiological Patient Positioning Devices has demonstrated Substantial Equivalence as defined and understood in the Federal Food, Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. Based on comparison against current product offering and predicate devices, CIVCO MR & Radiological Patient Positioning Devices are safe and effective for their intended and indicated use.
| Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | 319.656.4447 | 319.656.4451 |
|------------------------|------------------------|-----------------------|-----------------|-------------------|-------------------|
| Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | +31(0) 182.394495 | +31(0) 182.395014 |
| Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | 712.737.8688 | 712.737.8654 |
1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health and Human Services. The logo consists of a blue seal on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The seal features a stylized human figure with outstretched arms.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MEDTEC Inc. dba CIVCO Medical Solutions % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
JUL 9 2012
Re: K121545
Trade/Device Name: MR & Radiological Patient Positioning Devices Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH and IYE Dated: May 24, 2012 Received: May 25, 2012
Dear Mr. Job:
This letter corrects our substantially equivalent letter of June 8, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): _TBD
MR & Radiological Patient Positioning Devices Device Name:
Indications for Use:
CIVCO Patient Positioning Devices are used to aid in the support and positioning of patients during MR, radiological, and other procedures.
Over-The-Counter Use Prescription Use_x Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lalut D. Oth
ivision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K121545.
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.