← Product Code [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH) · K120638

# VANTAGE TITAN (K120638)

_Toshibamedical Systems Corporation · LNH · Jun 1, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K120638

## Device Facts

- **Applicant:** Toshibamedical Systems Corporation
- **Product Code:** [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH.md)
- **Decision Date:** Jun 1, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MRI system is indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density), - Spin-lattice relaxation time (T1) - Spin-spin relaxation time (T2) - Flow dynamics - Chemical shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

## Device Story

Vantage Titan (MRT-1504/U4) is a 1.5T MRI system; utilizes nuclear magnetic resonance of protons to generate cross-sectional images (transaxial, coronal, sagittal, oblique). Input: proton density, T1/T2 relaxation times, flow dynamics, chemical shift. Output: diagnostic images for physician interpretation. Modifications from predicate include redefined gradient strength, updated magnet cover/cabinet configuration, and two new software applications: Diffusion Tensor Tractography (DTT) and PaceMaker (FBI with Auto ECG). Operated by trained clinicians in clinical settings. DTT provides tractography visualization; PaceMaker (FBI with Auto ECG) automates ECG-gated imaging. System aids physicians in clinical diagnosis by visualizing anatomic structures and flow dynamics.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, design control activities, and risk analysis consistent with previously cleared submissions (K080038, K112003).

## Technological Characteristics

1.5T MRI system; superconducting magnet. Hardware: redefined gradient coils, modified cabinet/covers. Software: DTT and PaceMaker applications. Safety: SAR/dB/dt monitoring per IEC 60601-2-33. Connectivity: DICOM (NEMA PS 3.1-18). Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, NEMA MS series.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Vantage Titan ([K080038](/device/K080038.md))
- Vantage Titan HSR ([K112003](/device/K112003.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN - 1 2012

K12D 638

Toshiba America Medical Systems, Inc Pre-Market Notification 510(k) Vantage Titan, v2.1, MRT-1504/U4

#### 510(k) SUMMARY AND EFFECTIVENESS

#### . 1. DEVICE NAME:

Generic Name: Model Name: Trade/ Proprietary Name: Magnetic Resonance Diagnostic Device MRT-1504/U4 Vantage Titan

#### 2. ESTABLISHMENT REGISTRATION: 2020563

#### 3. U.S AGENT INFORMATION:

U.S. Agent Name:

Paul Biggins (714) 730-5000

Establishment Name and Address:

Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, Ca. 92780

4. MANUFACTURING SITE:

Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550 Japan

5. DATE OF SUBMISSION:

February 29, 2012

#### 6. DEVICE DESCRIPTION

The new Vantage Titan (Model MRT-1504/U4) MRI System is comparable to the current 1.5T Vantage Titan MRI System. The Vantage Titan uses the same magnet as the current Vantage Titan (K080038). However, the following modifications have been added: a redefinition of gradient strength, magnet cover and cabinet's configuration changes. In addition, two applications software: Diffusion Tensor Tractography (DTT), and PaceMaker.

#### 7. SUMMARY OF MAJOR HARDWARE CHANGES

- a. Gradient Strength (redefinition)
- b. Magnet Cover
- Changes to the cabinet configuration C.

{1}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

# 8. SUMMARY OF MAJOR SOFTWARE CHANGES

a. Addition of 2 applications software

- i. Diffusion Tensor Tractography (DTT)
ii. PaceMaker (FBI with Auto ECG)

### 9. SAFETY PARAMETERS

|                                                                             | New Vantage Titan<br>(Subject device)                                                        | Vantage Titan<br>(Predicate, K080038)                                                        | Vantage Titan HSR<br>(Predicate, K112003)                                                    |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| a. Static field<br>strength:                                                | 1.5T                                                                                         | 1.5 T                                                                                        | 1.5T                                                                                         |
| b. Peak and A-<br>weighted acoustic<br>noise:                               | 106.2 dB (A-weighed)<br>115.4 dB(peak)                                                       | 105.7dB (A-weighted)<br>115.7dB (peak)                                                       | 113.0 dB (A-weighted)<br>121.6 dB (peak)                                                     |
| c. Operational modes:                                                       | 1st operating mode                                                                           | 1st operating mode                                                                           | 1st operating mode                                                                           |
| i. Safety parameter<br>display:                                             | SAR, dB/dt                                                                                   | SAR, dB/dt                                                                                   | SAR, dB/dt                                                                                   |
| ii. Operating mode<br>access<br>requirements:                               | Allows screen access to<br>1st level operating mode                                          | Allows screen access to 1st<br>level operating mode                                          | Allows screen access to<br>1st level operating mode                                          |
| d. Maximum SAR                                                              | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33 (2002))           | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33 (2002))           | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33 (2002))           |
| e. Maximum dB/dt                                                            | <1st operating mode<br>specified in IEC 60601-2-33<br>(2002)                                 | <1st operating mode<br>specified in IEC 60601-2-33<br>(2002)                                 | <1st operating mode<br>specified in IEC 60601-2-33<br>(2002)                                 |
| f. Potential emergency<br>conditions and means<br>provided for<br>shutdown: | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard by<br>ferromagnetic objects | Shut down by Emergency<br>Ramp Down Unit for collision<br>hazard by ferromagnetic<br>objects | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard by<br>ferromagnetic objects |
| Biocompatibility of<br>materials                                            | Confirmed for electrodes<br>and accessories for<br>wireless gating                           | Not applicable                                                                               | Confirmed for electrodes<br>and accessories for wireless<br>gating                           |

{2}------------------------------------------------

## 10. IMAGING PERFORMACE PARAMETERS

No change from the previous predicate submission (K080038).

#### 11. INTEDED USE

The MRI system is indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nucle) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

- Proton density (PD) (also called hydrogen density),

- Spin-lattice relaxation time (T1)

- Spin-spin relaxation time (T2)

- Flow dynamics

- Chemical shift

Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

No changes to the previously cleared indication (K080038).

#### 12. DESIGN CHANGE

The Vantage Titan MRI System is comparable to the existing 1.5T Vantage Titian MRI System (K080038) and 1.5T Vantage Titan HSR MRI System (K112003) with the following modifications.

- Definition of gradient strength has been changed. a.
- ﻟﻘ Magnet covers have been changed.
- C. Cabinet configurations have been changed.
- Two application software have been added d.
	- 1. Diffusion Tensor Tractography (DTT)
	- 2. PaceMaker (FBI with Auto ECG)

{3}------------------------------------------------

#### 13. SUMMARY OF DESIGN CONTROL ACTIVITIES

PS Risk List for software and hardware changes have been is included in this submission. The test methods used are the same as those submitted in the previously cleared submissions (K080038 and K112003). A declaration of conformity with design controls is included in this submission.

## 14. TRUTHFUL AND ACCURACY CERTIFICATION

A certification of the truthfulness and accuracy of the Vantage Titan described in this submission is provided in this submission.

#### 15. SUBSTANTIAL EQUIVALENCE

Toshiba Medical Systems Corporation believes that the Vantage Titan (model MRT-1504/U4). Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate devices referenced in this submission.

Testing was done in accordance with applicable recognized consensus standards as listed below.

#### List of Applicable Standards

- · IEC60601-1:1988, Amd.1:1991, Amd.2:1995
- IEC60601-1-1:2000
- IEC60601-1-2:2001, Amd.1:2004
- IEC60601-1-4:1996, Amd.1:1999
- IEC60601-1-6:2006
- · IEC60601-1-8:2003.Amd.1:2006
- IEC60601-2-33:2002, Amd.1:2005, Amd.2:2007
- IEC60825-1: 2007
- · IEC62304:2006
- · IEC62366:2007
- · NEMA MS-1:2008
- · NEMA MS-2:2003
- · NEMA MS-3:2008
- · NEMA MS-4:2006
- · NEMA MS-5.2003
- · NEMA PS 3.1-18 (2008)

Page 4 of 4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Toshiba Medical Systems Corporation % Mr. Paul Biggins Director Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K120638

Trade/Device Name: MRT-1504/U4, Vantage Titan Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: April 25, 2012 Received: April 26, 2012

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

JUN - 1 2012

{5}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket whitication. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K120638

Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) Vantage Titan 1.5T, v2.1, MRT-1504/U4

# Indications for Use

510(k) Number (if known):

Device Name:

MRT-1504/U4. Vantage Titan

#### Indications for Use:

The MRI system is indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. In addition, this system supports non-contrast MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical shift

Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Andrew D. Kim

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

**510(k) Number** K120638

Page 1 of 1

Indication for Use Page 1 of 1

510(k): Vantage Titan, MRT-1504/U4, v2.1 Page 15 of 2798

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K120638](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K120638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
