← Product Code [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH) · K102489

# VANTAGE TITAN 3T (K102489)

_Toshiba Medical Systems Corporation, Japan · LNH · Aug 2, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K102489

## Device Facts

- **Applicant:** Toshiba Medical Systems Corporation, Japan
- **Product Code:** [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH.md)
- **Decision Date:** Aug 2, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: · Proton density (PD) (also called hydrogen density) - · Spin-lattice relaxation time (T1) - · Spin-spin relaxation time (T2) - · Flow dynamics - · Chemical Shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

## Device Story

Vantage Titan 3T is a 3 Tesla MRI system; utilizes 1.6m short magnet; incorporates Pianissimo scan noise reduction technology. Inputs: nuclear magnetic resonance signals from protons in body tissues/fluids. System transforms signals into cross-sectional images (transaxial, coronal, sagittal, oblique) representing proton density, T1/T2 relaxation times, flow dynamics, and chemical shift. Operated by trained clinicians in clinical settings. Output: diagnostic images for physician interpretation to aid clinical decision-making. Features include optional wireless gating, 16-channel (standard) or 32-channel (optional) receiver, and 2-channel RF amplifier. Benefits include high-field diagnostic imaging capabilities for anatomical assessment.

## Clinical Evidence

Bench testing performed per IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-33 (2002) + Amd.1/2, NEMA MS1-MS5, and ACR/NEMA DICOM 3.0. Volunteer studies conducted to verify imaging performance. No clinical diagnostic trials reported.

## Technological Characteristics

3T MRI system; 1.6m short magnet; 50x50x45 cm field of view. RF frequency 123MHz; 2-channel RF amplifier; 16/32-channel receiver. Pianissimo noise reduction technology. Optional wireless gating. Compliance with IEC 60601-2-33. Software-controlled SAR, dB/dt, and distortion correction. Biocompatibility confirmed for wireless gating accessories.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- EXCELART Vantage Titan ([K080038](/device/K080038.md))

## Reference Devices

- SIEMENS Verio ([K072237](/device/K072237.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Toshiba America Medical Systems, Inc. AUG – 2 2011 Pre-Market Notification 510(k) MRT-3010/A5 Vantage Titan 3T

Attachment 1

## 510(k) SUMMARY

Magnetic Resonance Diagnostic Device Accessory 1. DEVICE NAME: MRT-3010/A5 Model Name: Trade/Proprietary Name: Vantage Titan 3T

#### 2. ESTABLISHMENT REGISTRATION:

- U.S. Agent Name and Address: Toshiba America Medical Systems, Inc. 3. 2441 Michelle Drive Tustin, Ca 92780

| Contact Person:     | Paul Biggins<br>(714) 730 - 5000 |  |
|---------------------|----------------------------------|--|
| Manufacturing Site: | TOSHIBA CORPORATION              |  |

- 4. TOSHIBA CORPORATION lanufacturing Site: MEDICAL SYSTEMS COMPANY 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
- ડ. August 30, 2010 DATE OF SUBMISSION: .

#### 6. DEVICE DESCRIPTION

The Vantage Titan 3T (Model MRT-3010/A5) is a 3 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Titan 3T uses 1.6m short and 6.4 tons light weight magnet. It includes the Toshiba Pianissimo™ technology (scan noise reduction technology). The design of the gradient coil and the WB coil of the Vantage Titan 3T provides the maximum field of view of 50 x 50 x 45 cm.

The Vantage Titan 3T MRI System is comparable to the current 1.5T EXCELART Vantage Titan MRI System (K080038), cleared January 22, 2008 with the following modifications.

- . Field strength of magnet has been changed to 3T.
- . RF amplifier, transmitter and receiver (RF cabinet) are modified to adopt the change in resonance frequency (63MHz -> 123MHz).
- . Number of RF amplifier channels is increased from 1 to 2.
- Number of receiver channels is 16ch for standard (same as Titan 1.5T), 32ch for optional. .
- Software platform has been changed. .

NOTE: The Filed Strength Vantage Titan 3T MRJ System is equivalent to that of the SIEMENS Verio, which is cleared on October 10, 2007 (K072237).

{1}------------------------------------------------

Attachment 1

## 6.1. SUMMARY OF MAJOR HARDWARE CHANGES

- Field strength of magnet has been changed to 3T. a.
- RF amplifier, transmitter and receiver (RF cabinet) are modified to adopt the change in b. resonance frequency (63MHz -> 123MHz).
- Number of RF amplifier channels is increased from 1 to 2. C.
- Number of receiver channels is 16ch for standard (same as Titan 1.5T), 32ch for d. optional.
- Cooling fan box is separated from filter cabinet. e.
- Transformer cabinet has been modified to cover the system power requirement. f.
- Wireless gating is added (optional). g.

#### SUMMARY OF MAJOR SOFTWARE CHANGES 6.2.

- New RF amplifier cabinet control. a.
- b. New gradient power supply control.
- New Magnet supervisory control. c.
- d. Modified the data base for SAR control for new WB coil.
- Modified the data base for distortion correction and dB/dt calculation for new gradient e. coil.
- f. New software platform.

{2}------------------------------------------------

Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) MRT-3010/A5 Vantage Titan 3T

K102484
p3/4

### 7. SAFETY PARAMETERS

Attachment l

|                                                                          | EXCELART Vantage Titan<br>(K080038)                                                       | New Vantage Titan 3T                                                        |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| a. Static field strength:                                                | 1.5 T                                                                                     | 3T                                                                          |
| b. Peak and A-weighted<br>acoustic noise:                                | 105.7dB (A-weighted)<br>115.7dB (peak)                                                    | 111.8 dB (A-weighted)<br>121.3 dB (peak)                                    |
| c. Operational modes:<br>i. Safety parameter<br>display:                 | 1st operating mode<br>SAR, dB/dt                                                          | Same                                                                        |
| ii. Operating mode access<br>requirements:                               | Allows access to 1st level<br>operating mode                                              | Same                                                                        |
| d. Maximum SAR                                                           | 4 W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33 (2002))       | Same                                                                        |
| e. Maximum dB/dt                                                         | <1st operating mode specified in<br>IEC 60601-2-33 (2002)                                 | Same                                                                        |
| and Gradient coil<br>dimensions:                                         | 760 x 893 x 1405<br>(inner diameter x outer diameter x<br>length, unit: mm)               | 749 x 893 x 1405<br>(inner diameter x outer diameter<br>x length, unit: mm) |
| f. Potential emergency<br>conditions and means<br>provided for shutdown: | Shut down by Emergency Ramp<br>Down Unit for collision hazard by<br>ferromagnetic objects | Same                                                                        |
| g. Biocompatibility of<br>materials:                                     | Not applicable                                                                            | Confirmed for electrodes and<br>accessories for wireless gating             |

## 8. IMAGING PERFORMANCE PARAMETERS

No changes from the previous submission, K080038.

{3}------------------------------------------------

## Attachment 1

#### 9. INTENDED USE

Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

· Proton density (PD) (also called hydrogen density)

- · Spin-lattice relaxation time (T1)
- · Spin-spin relaxation time (T2)
- · Flow dynamics
- · Chemical Shift

Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

#### 10. EQUIVALENCY INFORMATION

Toshiba Medical Systems Corporation believes that the Vantage Titan 3T (model MRT-3010/A5) Magnetic Resonance Imaging (MRI) System is substantially equivalent to the current Vantage Titan (model MRT-1504/S3) cleared on January 22, 2008 (K080038) except for magnetic field strength (3T instead of 1.5T). Field strength of this system is equivalent to that of SIEMENS Verio, which is cleared on October 10, 2007 (K072237).

Testing was done in accordance with applicable recognized consensus standards as listed below. Additionally, volunteer studies were conducted to verify imaging performance.

IEC 60601-1-1 IEC 60601-1-2 IEC 60601-2-33(2002) + Amd.1 (2005) + Amd.2 (2007) NEMA MS1, MS2, MS3, MS4, MS5 ACR/NEMA DICOM 3.0

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Toshiba Medical Systems Corporation, Japan % Mr. Paul Biggins Director Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

AUG - 2 2011

Re: K102489

Trade/Device Name: Vantage Titan 3T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH and MOS Dated: April 8, 2011 Received: April 18, 2011

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of

{5}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# K102489

# Indications for Use

510(k) Number (if known):

Device Name: Vantage Titan 3T

Indications for Use:

Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1) ·
- . Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift

Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Prescription Use (Part 21 CFR 801Subpart D) 801Subpart C)

OR

Over-The-Counter Use (Part 21 CFR

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S Patel

---

(Division Sign Off)

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K102-189

Company Confidential AUG 2010

510(k): Vantage Titan 3T Page 37 of 3857

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K102489](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K102489)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
