3TP SOFTWARE OPTION

K031350 · 3tp, LLC · LNH · Jun 23, 2003 · Radiology

Device Facts

Record IDK031350
Device Name3TP SOFTWARE OPTION
Applicant3tp, LLC
Product CodeLNH · Radiology
Decision DateJun 23, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The 3TP Software Option is intended to be used as a post processing software package designed to provide a reliable means for visualizing the presence and pattern of contrast induced enhancement on MR datasets. 3TP supports the evaluation of dynamic MR data gathered during the injection of a bolus of contrast media. The resulting time course information can be displayed in a variety of formats, including a parametric image overlaid onto source MR images. In the hands of a trained physician the information provided by the 3TP Software Option could vield information that may assist in the interpretation of dynamic contrast enhanced studies.

Device Story

3TP Software Option is a post-processing software module for dynamic MR imaging. Input consists of multi-sliced MR datasets acquired over time following a bolus contrast media injection. The software processes these datasets to analyze signal intensity changes over time, generating a composite '3TP Image' (parametric image) overlaid on source MR images. Used in clinical settings by trained physicians to assist in diagnostic interpretation of dynamic contrast-enhanced studies. The output provides visual patterns of contrast enhancement, aiding clinical decision-making regarding tissue characterization.

Clinical Evidence

Bench testing only. The device undergoes integration testing and verification prior to beta validation. No clinical study data provided.

Technological Characteristics

Post-processing software module for MR imaging. Analyzes dynamic signal intensity changes from multi-slice MR datasets. Outputs parametric images overlaid on source data. Operates as a software-only package.

Indications for Use

Indicated for use by trained physicians to assist in the interpretation of dynamic contrast-enhanced MR studies by visualizing contrast enhancement patterns in MR datasets.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # ATTACHMENT 7 JUN 2 3 2003 # Summary of Information Respecting Safety And Effectiveness This 510(K) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h). | Contact: | Vincent J. McGill | |----------|----------------------------------------------------------| | Phone: | (212) 779-9910 | | Fax: | (212) 779-9928 | | Product: | 3TP Software Option<br>Imaging Processing Software for N | ### MR Devices Manufactured by: 3TP LLC Distributed by: 3TP LLC #### 1) Indications for Use The first to the state the country of the county of the county of the county of the county of the county of the county of the county of the county of the county of the county Callery Controller Comments of Children Comments of Children LA LA BERESTER DE MALES A LEI BERE LE FERELARD The 3TP Software Option is intended to be used as a post processing software package designed to provide a reliable means for visualizing the presence and pattern of contrast induced enhancement on MR datasets. 3TP supports the evaluation of dynamic MR data gathered during the injection of a bolus of contrast media. The resulting time course information can be displayed in a variety of formats, including a parametric image overlaid onto source MR images. In the hands of a trained physician the information provided by the 3TP Software Option could vield information that may assist in the interpretation of dynamic contrast enhanced studies. #### Device Description 2) The 3TP Software Option is a post processing software module. The 3TP Software Option supports the evaluation of dynamic MR data gathered after the injection of a bolus of a contrast agent. Multi-sliced MR datasets with specified time intervals are used for input. The purpose is to provide a new composite "3TP Image" that provides a trained physician with information useful for diagnostic purposes. ### Marketing History 3) The software has not yet been marketed. #### Predicate Devices 4) GE Advantage Windows With Functool Option (K960265) GE Medical Systems 300 N. Grandview Blvd. Waukesha, WI 53186 GE Prostate Spectroscopy and Imaging Exam (PROSE) (K011604) GE Medical Systems 300 N. Grandview Blvd. Waukesha, WI 53186 Philips EasyVision (Quantitative Analysis Option) (K971965) Philips Medical Systems 710 Bridgeport Ave Shelton, Ct 06484 {2}------------------------------------------------ Similar to each of the named predicate devices the 3TP Software Option provides a postprocessing means for analyzing changes in signal intensity of a contrast agent as reflected in MR images. The use of the 3TP Software Option does not result in any additional hazards when compared to the post-processing software packages (Functyool, PROSE and Philips Easy Vision) currently being marketed by GE Medical Systems and Philips Medical Systems. The 3TP Software Option does not include any new indications for use nor does the use of this device result in any new potential hazards. #### 5. Performance Testing The 3TP Software Option will successfully complete integration testing and verification prior to beta validation and the software beta testing will be successfully completed validating the 3TP Software Option prior to market release. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 3 2003 3TP LLC % Mr. Vincent J. McGill Eaton & Van Winkle, LLP 3 Park Avenue NEW YORK NY 10016 Re: K031350 Trade/Device Name: 3TP Software Option Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: April 28, 2003 Received: April 29, 2003 Dear Mr. McGill: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {5}------------------------------------------------ 1350 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: Indications for Use: The 3TP Software Option is intended to be used as a post processing software package designed to provide a reliable means for visualizing the presence and pattern of contrast induced enhancement on MR datasets. 3TP supports the evaluation of dynamic MR data gathered during the injection of a bolus of contrast media. The resulting time course information can be displayed in a variety of formats, including a parametric image overlaid onto source MR images. In the hands of a trained physician the information provided by the 3TP Software Option could yield information that may assist in the interpretation of dynamic contrast enhanced studies. ### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 C.F.R. 801.109) ાર Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) David h. Symm Division Sign-Off ivision of Reproductive, Abdomina nd Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...