← Product Code [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH) · K012875

# SIGNA OPENSPEED MAGNETIC RESONANCE SYSTEM (K012875)

_GE Medical Systems · LNH · Nov 20, 2001 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K012875

## Device Facts

- **Applicant:** GE Medical Systems
- **Product Code:** [LNH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH.md)
- **Decision Date:** Nov 20, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The Signa OpenSpeed Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa OpenSpeed Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa OpenSpeed Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. Due to the 'open' design of the system, the Signa OpenSpeed may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles.

## Device Story

Signa OpenSpeed is an open-style, 0.7T superconducting MRI system. It acquires 2D single/multi-slice and 3D volume images by measuring proton density, T1, T2, and flow. Input signals are processed via 2D and 3D Fourier transformation techniques to produce anatomical images. Operated by trained clinicians in clinical settings, the system supports high-resolution imaging and shorter scan times. The open architecture allows for use during interventional procedures when paired with MR-compatible accessories like in-room displays and biopsy needles. Physicians interpret the resulting images to assist in clinical diagnosis. The device offers diffusion imaging as an optional modality.

## Clinical Evidence

Bench testing only. The device was evaluated against NEMA and IEC 60601-2-33 international medical equipment safety standards for magnetic resonance equipment. No clinical data provided.

## Technological Characteristics

0.7T superconducting open-style magnet. Gradient strength and slew rate (SR40) improved over predicate. Imaging planes: sagittal, coronal, axial, oblique, double oblique. Reconstruction: 2D and 3D Fourier transformation. Supports diffusion imaging. Compatible with previously cleared coils and accessories.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Signa HFO/i Magnetic Resonance System ([K992746](/device/K992746.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a black and white logo. The logo is circular and contains two intertwined letters that appear to be "G" and "E". The logo is likely for General Electric (GE).

NOV 2 0 2001

Page 1 of 2

General Electric Company P.O. Box 414, Milwaukee, WI 53201

# 510(K) SUMMARY

Ko12875

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).

| Submitter:      | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|---------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs   |
| Telephone:      | 262-544-3894                                            |
| Fax:            | 262-544-3863                                            |
| Date Prepared:  | August 23, 2001                                         |

#### Device Name:

GE Signa OpenSpeed Magnetic Resonance System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH

#### Marketed Device:

The Signa OpenSpeed Magnetic Resonance System is substantially equivalent to the currently marketed The organ of Magnetic Resonance System (K992746) with the main differences being an increase in gradient strength from 15mT/m, and an increase in slew rate from SR25 to SR40. In addition, Diffusion imaging is now offered as an imaging option.

#### Device Description:

The Signa OpenSpeed Magnetic Resonance System is a modification to the Signa HFO/i Magnetic Resonance System (K992746) which utilizes a superconducting magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The Signa OpenSpeed Magnetic Resonance System is an open style magnet operating at 0.7T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high resolution anatomical applications and shorter scan times. Previously cleared software options, coils, and other accessories may be used with the Signa OpenSpeed Magnetic Resonance System.

#### Indications for Use:

The Signa OpenSpeed Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa OpenSpeed Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa OpenSpeed Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a circular logo with a stylized design. The logo appears to be a vintage or retro design, possibly representing a company or brand. The design consists of intertwined or overlapping elements within the circle, creating a visually complex and abstract pattern. The logo is black and white, with the design elements appearing in black against a white background.

Page 2 of 2

General Electric Company P.O. Box 414, Milwaukee, WI 53201

physician, these images provide information that can be useful in determining a diagnosis.

Due to the 'open' design of the system, the Signa OpenSpeed may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles.

#### Comparison with Predicate Device:

The Signa OpenSpeed Magnetic Resonance System is a modification of the Signa HFP/i MR system (K992746) with the main differences being an increase in gradient strength from 15mT/m, (1892140) with the main differences oong an more and and the maging is now offered as an imaging option.

#### Summary of Studies:

The Signa OpenSpeed Magnetic Resonance System was evaluated to the appropriate NEMA The Signa Openspece Magnetic Robertanoon J. International Medical Equipment Safety standard and performance buildiana as as "quirements for Safety of Magnetic Resonance Equipment for Medical IEC 001-2-53 I articular Nequilions for SReey of MSS. Somparable to the currently marketed Signa HFO/i Magnetic Resonance System.

#### Conclusion:

It is the opinion of GE that the Signa OpenSpeed Magnetic Resonance System is substantially equivalent it is the Signa HFO/i Magnetic Resonance System. The Signa OpenSpeed Magnetic Resonance System to the bight III Ori Haghetions for use, nor does use of this device result in any new potential hazards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### NOV 2 0 2001

Larry Kroger, Ph.D. Regulatory Programs Manager GE Medical Systems General Electric Company P.O. Box 414 MILWAUKEE WI 53201

Re: K012875 Trade/Device Name: Signa OpenSpeed Magnetic

Resonance System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: August 23, 2001 Received: August 27, 2001

Dear Dr. Kroger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouthoution. The I Drivinention for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you done office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Actively reliable at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on of reference to premainted from the obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Snugden

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{4}------------------------------------------------

## NOV 2 0 2001

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Signa OpenSpeed Magnetic Resonance System

#### Indications For Use:

The Signa OpenSpeed Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa OpenSpeed Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa OpenSpeed Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Due to the 'open' design of the Signa OpenSpeed may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use (Per 21 CFR 801.109) | <div> <span style="text-decoration: overline;"></span> <span style="vertical-align: middle;">✓</span> </div> |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------|
|---------------------------------------|--------------------------------------------------------------------------------------------------------------|

OR Over-The-Counter Use 
  

Nancy brigton
  

Division Sign-Off
  

Division of Reproductive, Abdominal,
  

Radiological Devices
  

510(K) Number K012875

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K012875](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LNH/K012875)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
