← Product Code [LMD](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LMD) · K980220

# IISYS PACS SYSTEM (K980220)

_Sterling Diagnostic Imaging, Inc. · LMD · Feb 13, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LMD/K980220

## Device Facts

- **Applicant:** Sterling Diagnostic Imaging, Inc.
- **Product Code:** [LMD](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LMD.md)
- **Decision Date:** Feb 13, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2020
- **Device Class:** Class 1
- **Review Panel:** Radiology

## Indications for Use

The iiSYS PACS System is a complete PACS System for the Transmission, Display, Archive, and printing of patient images and demographic information. The system is indicated for the assembly, organization, sharing, and display of patient images and demographic information for diagnostic and referral purposes. Application areas include radiologist central reading rooms or any location where a medical professional would require or desire access to patient image and demographic information.

## Device Story

iiSYS PACS System integrates previously cleared diagnostic imaging components to facilitate transmission, archiving, display, and printing of patient images and demographic data. Inputs include digital information, video signals, and hard copy prints. System components include viewing/reading workstations, teleradiology devices for network transmission, archive storage, LINX network system for secondary capture, and printers. Operated by trained medical professionals in clinical environments (e.g., reading rooms, remote facilities). Device does not alter original image data; output is reviewed by clinicians to support diagnostic and referral decision-making. Integration is verified via a comprehensive testing protocol to ensure connectivity and efficiency across components.

## Clinical Evidence

No clinical data. Bench testing only, specifically focusing on integration testing protocols to ensure connectivity and compatibility between previously cleared PACS components.

## Technological Characteristics

System comprises viewing workstations, teleradiology digitizers, archive storage, LINX network system, and printers. Connectivity via local/wide area networks. No patient-contacting materials. Software-based image management and transmission. Integration verified via proprietary test protocol.

## Regulatory Identification

A medical image communications device provides electronic transfer of medical image data between medical devices. It may include a physical communications medium, modems, and interfaces. It may provide simple image review software functionality for medical image processing and manipulation, such as grayscale window and level, zoom and pan, user delineated geometric measurements, compression, or user added image annotations. The device does not perform advanced image processing or complex quantitative functions. This does not include electronic transfer of medical image software functions.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K980220 Page 1 of 2

# Summary of Safety and Effectiveness

| CLASSIFICATION NAME:    | The proposed classification is 21CFR892.2050 (FR 63774, Dec 2, 1997) This device is regulated as Class II Tier I. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------|
| COMMON/USUAL NAME:      | PACS System                                                                                                       |
| TRADE/PROPRIETARY NAME: | iiSYS PACS System                                                                                                 |
| ESTABLISHMENT No.       | 1043882                                                                                                           |

#### PREDICATE DEVICE

FEB 1 3 1998

There are no significant differences between the System and its assembled component devices other than the guaranteed connectivity provided by the integration testing. The integration of the devices into the Sterling Diagnostic Imaging iiSYS PACS System is accomplished via comprehensive testing protocol.

There is no change in intended use from the component devices. There is no change in indications for use of the component devices. The intended use and indications for use of the system is the compiled uses of the integrated components.

#### REASON FOR 510(k):

The submission is being made to identify Sterling Diagnostic Imaging, Inc. as offering a complete PACS System which may become a separate device category under proposed regulations. All integral parts of the system have been previously reviewed and cleared by the Agency.

## SYSTEM DESCRIPTION:

The iiSYS PACS System is a full featured PACS System capable of transmission, archive, display, and print of patient image and demographic information. Its purpose is to facilitate these operations utilizing shared data to promote the availability of information at remote facilities and at locations other than that at which it was acquired. Data may be received as digital information, video signals, or hard copy prints and may be reviewed via monitor or printed hardcopy. The iiSYS PACS system consists of the following major components:

- Series of Viewing and Reading Workstations ●
- Teleradiology devices for digitizing and transmission of images over wide area or . local area networks for remote or at home review
- Archive for short or long term storage .
- LINX Network system for secondary capture and transmission of images to other . devices (workstation, telerad, printers, etc.)
- . Printers

The major elements of the System are previously cleared devices as identified in the following submission. Sterling is not altering the specifications, claims, indications, or intended use of any component. The Sterling iiSYS PACS System is a compilation of components that has been professionally tested to insure their connectivity and efficiency when utilized together. This testing is optimized to give the user assurance that the devices will provide a maximum value-added to the clinical environment without impacting safety or efficacy of the individual components. Connectivity and compatibility are evaluated and ensured via the Integration Test Protocol.

{1}------------------------------------------------

#### INTENDED USE

The iiSYS PACS System is a complete PACS System for the Transmission, Display, Archive, and printing of patient images and demographic information. The system is indicated for the assembly, organization, sharing, and display of patient images and demographic information for diagnostic and referral purposes.

Application areas include radiologist central reading rooms or any location where a medical professional would require or desire access to patient image and demographic information.

## SAFETY INFORMATION:

The System has no patient contacting materials and is utilized only by trained professionals. The output of the device is evaluated by additional trained professionals allowing sufficient review to afford identification and intervention in the event of a malfunction. The device does not impact the quality or status of the original acquired image data.

Selection of the specific components of the system has been made to allow the system to remain cost effective and consistent with current technology through the substitution of components as technology develops. By retaining criteria for the substitution of components, any concerns about safety or efficacy and substantial equivalence can be satisfactorily met by a determination that the component substitution is not a significant change in this is consistent with existing Agency guidance.

## CONCLUSION [21 CFR: 807.92(b)(3)]

The subject device has no patient contact, nor does it control, monitor, or effect any devices directly connected to or effecting such a patient contacting device. The images generated by the subject device is observed by medical personnel, offering ample opportunity for competent human intervention in the event of a failure.

Sterling believes sufficient information is included to reach a determination of substantial equivalence. We conclude that the subject device is as safe and effective as the component devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The words are stacked on a single line and are easily readable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 3 1998

Timothy W. Capehart Manager of Regulatory Affairs and Compliance Sterling Diagnostic Imaging, Inc. 10 South Academy Street Mailstop 102 P.O. Box 19048 Greenville, SC 29602-9048

Re:

K980220 iiSYS PACS System Dated: January 22, 1998 Received: January 22, 1998 Regulatory class: Unclassified Procode: 90 LMD

Dear Mr. Capehart:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.btml".

Sincerely yours.

William Yu, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

510(k) Number (if known) :

K980220

Device Name :

iiSYS PACS System

Indications for Use:

The iiSYS PACS System is a complete PACS System for thhe Transmission, Display, Archive, and printing of patient images and demographic information. The system is indicated for the assembly, organization, sharing, and display of patient images and demographic information for diagnostic and referral purposes.

Application areas include radiologist central reading rooms or any location where a medical professional would require or desire access to patient image and demographic information.

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Varial A. Seym

---

(Division Sign-Off)

(Division of Reproductive, Abdominal, E and Radiological Dev 510(k) Number

Prescription Use
(Per 21 CFR 801.109) ✓

Over the Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LMD/K980220](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LMD/K980220)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
