K974102 · Mitra Imaging, Inc. · LMD · Jan 20, 1998 · Radiology
Device Facts
Record ID
K974102
Device Name
MILTRA IMAGING INC., IMAGE VAULT
Applicant
Mitra Imaging, Inc.
Product Code
LMD · Radiology
Decision Date
Jan 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2020
Device Class
Class 1
Attributes
Software as a Medical Device
Indications for Use
Vault is intended to act as a long term storage device (archive) for medical images. Vault is a networked device that acts as a DICOM Storage Service Class Provider (SCP): - Images can be sent to it via the DICOM protocol, from modalities or other devices that are sources of medical images. - Images can subsequently be queried and retrieved via the DICOM protocol, to imaging workstations and other processing devices. - Internally, Vault is optimized to provide secure storage for a very large number of images, and to provide rapid retrieval on demand. To that end, Vault includes high capacity storage media, and has implemented hierarchical storage management software to manage cache storage effectively.
Device Story
Image Vault is a PC-based, DICOM-compliant PACS software solution for medical image archiving. It functions as a DICOM Storage Service Class Provider (SCP), receiving images from modalities via TCP/IP, storing them on magnetic tape or hard disk, and enabling retrieval to workstations. It supports JPEG-compressed and non-compressed DICOM images. The system includes a display utility for patient and study identification, though it is not intended for diagnostic use. Operated by hospital staff or administrators, it manages large-scale image data using hierarchical storage management. It provides no image processing or quality enhancement; its primary benefit is secure, rapid, long-term storage and retrieval of clinical imaging data.
Clinical Evidence
Bench testing only. The software was validated through internal procedures and testing by programmers and potential customers. No clinical data was required as the software is used for storage and control purposes and does not affect image quality.
Technological Characteristics
PC-based software; DICOM-compliant; TCP/IP communication; supports magnetic tape and hard disk storage; CD-R media support; 1024x768 display resolution. No image processing algorithms. Software is used for control and storage management only.
Indications for Use
Indicated for long-term storage and retrieval of medical images in a hospital department setting. Users include service personnel, system administrators, and clinical staff (e.g., chief technologist).
Regulatory Classification
Identification
A medical image communications device provides electronic transfer of medical image data between medical devices. It may include a physical communications medium, modems, and interfaces. It may provide simple image review software functionality for medical image processing and manipulation, such as grayscale window and level, zoom and pan, user delineated geometric measurements, compression, or user added image annotations. The device does not perform advanced image processing or complex quantitative functions. This does not include electronic transfer of medical image software functions.
Predicate Devices
Kodak Cardiology Digital Archive & Review System (Cardiac Archive Station, CAS 6000) (K960043)
Submission Summary (Full Text)
{0}------------------------------------------------
K974102
P 1/4
JAN < U 1998
# 510(k) SUMMARY of Safety and Effectiveness
- 1.1 Company Identification
Mitra Imaging, Inc. 115 Randall Drive Waterloo, Ontario, Canada N2V IC5 Tel: (519) 746-2900 Fax: (519) 746-3745
- 1.2 Official Correspondent
Gary J. Allsebrook Regulatory Management Services 16303 Panoramic Way San Leandro, CA, USA, 94578-1116
- 1.3 Date of Submission
October 31, 1997
- 1.4 Device Name
CCCCCCCCCGGGG
Classification Name: Common/Usual Name: PACS Image Archive System/ Teleradiology System Image Vault
Proprietary Name:
- Substantial Equivalence 1.5
The Image Vault software is substantially equivalent to the Kodak Cardiology Digital Archive & Review System (Cardiac Archive Station, CAS 6000), K960043.
{1}------------------------------------------------
K974102 2/4
#### 1.6 Device Description and Intended Use
Image Vault is a software package which may be marketed as a software only solution, as well as in conjunction with standard PC hardware. Image Vault is a PC-based, DICOMcompliant PACS device that is able to send & receive, display, store, and archive DICOM non-compressed and JPEG-compressed images. Images can be archived using standard magnetic tapes. It also has the ability to read/write a collection of images to/from CD-R media in a method defined by DICOM.
Images can be viewed using a simple display utility, for the purposes of properly identifying patients and their corresponding studies, and is not intended to be used for diagnosis.
Image Vault uses standard "off-the-shelf" PC hardware and communicates using the standard TCP/IP stack. The network hardware used to support the TCP/IP stack is superfluous to Image Vault.
#### 1.7 Software Development
Mitra certifies that the Image Vault software is designed, developed, tested and validated according to written procedures. These procedures identify individuals within the organization responsible for developing and approving product specifications, coding and testing, validation testing and field maintenance. The software used in this product is used for cardiology and storage only and does not affect image quality.
{2}------------------------------------------------
K974102
### Substantial Equivalence Comparison Chart 1.8
Substantial Equivalence Comparison Chart (Image Vault)
| Specification | Image Vault | Kodak Cardiac Archive Station CAS 6000 |
|---------------------------------|------------------------------------|----------------------------------------|
| Graphical User Interface | Yes | Yes |
| Patient Demographics | Yes | Yes |
| Display Resolution | 1024x768 | 1024x768 |
| Communications | TCP/IP | TCP/IP |
| Image Review | Flip/Rotate/Pan/Zoom/Sequential | Flip/Rotate/Pan/Zoom/Sequential |
| CD Writers Supported | Yamaha CDR400<br>Yamaha CDR100 | Kodak PCD Writer 600 |
| CD Labeling | Yes | Yes |
| Temporary Image Storage | Yes | Yes |
| Image Archiving (Hard Disk) | Yes | Yes |
| Image Archiving (Magnetic Tape) | Yes | No |
| Magnetic Tape Drives Supported: | ADIC Scalar 218<br>ADIC Scalar 448 | No |
| Platform | PC | PC |
| | Yes | Yes |
#### Safety and Effectiveness 1.9
In accordance with the provisions of the Safe Medical Device Act of
the same with the providing a summary of safety and In accordance with the provisions of the Summary of safety and 1990, Mitra Imaging, Inc. Is providing a Sammary
effectiveness information regarding the Image Vault software.
# General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and indication for use.
The hardware components specified (but not supplied) are all "off the shelf" computer components.
# Validation and Effectiveness:
Extensive testing of the software package has been performed by
entify of the software commerce and by potential customers. Extensive testing of the soltware passes and by potential customers.
programmers, by non-programmers, and by potential customers. programmers, by non-programmers, and by p on any posses and has no bearing on Software is only used for control purposso and has and master in this software.
image quality. There is no image processing used with this software.
Images that are compressed are properly identified in the images Images that are compressed are propenty tierhinet in the DCOM standard.
information as being compressed as specified by the entire information as being compressed as spoolibus by and the entire life of the image.
{3}------------------------------------------------
K974102
## Substantial Equivalence:
The Image Vault software is a software package used for archiving cardiac images and for writing to CD-ROM media. The intended use and technological characteristics of the system are similar to the Kodak Cardiac Archive Station CAS 6000. Any differences between the Image Vault software and the equivalent device have no significant influence on safety or effectiveness.
Its is our conclusion that there is no software component that we know of in the Image Vault software whose failure or latent design flaw would be expected to result in death or injury to a patient.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of a bird.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Image /page/4/Picture/4 description: The image shows the text "JAN 20 1998" in a bold, sans-serif font. The text is arranged horizontally, with the month abbreviation "JAN" on the left, followed by the day "20", and then the year "1998". The text is black against a white background.
Mitra Imaging, Inc. c/o Gary J. Állesbrook Regulatory Management Services 16303 Panoramic Way San Leandro, CA 94578-1116
Re:
K974102 Image Vault Dated: October 31, 1997 Received: October 31, 1997 Regulatory class: Unclassified Procode: 90 LMD
Dear Mr. Allesbrook:
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitry diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html":.............................................................................................
Sincerely yours.
W.Liau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
K974102
510(k) Number (if known):
## Device Name: Image Vault
Indications for Use:
Vault is intended to act as a long term storage device (archive) for medical images.
Vault is a networked device that acts as a DICOM Storage Service Class Provider (SCP):
- Images can be sent to it via the DICOM protocol, from modalities or other devices that are sources of medical images.
- [ Images can subsequently be queried and retrieved via the DICOM protocol, to imaging workstations and other processing devices.
- O Internally, Vault is optimized to provide secure storage for a very large number of images, and to provide rapid retrieval on demand. To that end, Vault includes high capacity storage media, and has implemented hierarchical storage management software to manage cache storage effectively.
Users:
The Vault will be located in a hospital department. The expected users are as follows:
Service personnel for installation, configuration and support.
- [ System administrator within the department, who in many cases will be chief technologist or other designated clinical staff.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|----------------------|--|
| (Per 21 CFR 901.109) | |
Over-the-Counter Use
(Optional Format 1-2-96)
Gerrit C. Seymon
OR
510(k) Number
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.