KODAK DIGITAL SCIENCE 9000D MEDICAL LASER PRINTER (9000D)
K972847 · Eastman Kodak Company · LMC · Oct 7, 1997 · Radiology
Device Facts
Record ID
K972847
Device Name
KODAK DIGITAL SCIENCE 9000D MEDICAL LASER PRINTER (9000D)
Applicant
Eastman Kodak Company
Product Code
LMC · Radiology
Decision Date
Oct 7, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
The KODAK Digital Science 9000D Medical Laser Printer is a laser printer designed to produce hard-copy films of digitized medical images. It is a newly designed printer to be added to Kodak's family of medical laser printer products that is intended for the centralized network printing solution for high volume sites for maximum productivity. This printer will be environmentally friendly by eliminating wet chemistry, site concerns, costs and will reduce chemical disposal problems. This is accomplished by using a totally proprietary imaging process. Images printed on film from the laser printer will be used for primary diagnosis, similar to other laser printers and CRT mult-format cameras. Evaluation of the hard copy output by trained health care professionals provides adequate opportunity for competent human intervention. Laser printers are not represented to be of use in supporting or sustaining human life, nor do they present a potential for unreasonable risk of illness or injury.
Device Story
Secondary imaging device; receives digital, video, or network image data from medical modalities; transforms inputs into hard-copy film output via laser printing system; utilizes dry thermal processor instead of wet chemistry; operated by clinicians/technicians via keypad or autofilming link at system console; used in clinical environments for centralized network printing; provides radiologists with hard-copy images for primary diagnosis; benefits include increased productivity, elimination of wet chemistry, and reduced chemical disposal requirements.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through compliance with safety and electromagnetic interference standards (EN 60601-1, EN 60825, UL 2601-1, IEC 801 series).
Technological Characteristics
Red diode laser; polygon deflection; 4096x5120 image matrix; 12-bit depth; dry thermal processor; 79-micron pixel pitch; 50mw laser power; supports multiple network/digital/video inputs; seismic kit available; compliant with EN 60601-1, EN 60825, UL 2601-1.
Indications for Use
Indicated for production of hard-copy films of digitized medical images for primary diagnosis by trained healthcare professionals in high-volume clinical settings.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Kodak Ektascan 2180 Laser Printer (KELP 2180)
Reference Devices
DV 8700
Submission Summary (Full Text)
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### INTRODUCTION I.
The Kodak Digital Science 9000D Medical Laser Printer (MLP 9000D) is a secondary imaging device that receives data from various modalities that can be stored and then printed on film for evaluation by a radiologist. The MLP 9000D is an integrated system which is based on the overall system concept of the Kodak Ektascan 2180 Laser Printer (KELP 2180), Common Protocol (CP) interfaces and the Kodak X-Omat 180LP/LPS Processor system. The MLP 9000D's film transport and exposure mechanisms will be a modification of the current KELP 2180 design. The Kodak X-Omat 180LP/LPS will be replaced by a dry thermal processor that will be used to develop a new photothermographic film.
The MLP 9000D will consist of many components which are similar to the KELP 2180. Examples are various input sections that are either video, digital, digital video, or network interfaces that interact between the host and the MLP 9000D to perform its tasks. The operator interfaces with the system by either using a keypad to store and print images or with an autofilming link to perform the same storing and printing functions at a system console.
### II. GENERAL INFORMATION
Name of Manufacturer: Eastman Kodak Company Health Imaging Division 901 Elmgrove Road Rochester, New York 14653-5513
Establishment Registration Number: 1317307
Device Name: KODAK Digital Science 9000D Medical Laser Printer
Device Classification: Class II Device
UCT - 7 1997
{1}------------------------------------------------
#### III. THE DEVICE
The general hardware configuration of the KODAK Digital Science 9000D Medical Laser Printer contains the following major components / subsystems:
Modality Interfaces Laser Printing System Processor with optional Film Sorter
#### OPTIONAL ACCESSORIES:
Image Buffer Controller / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
High Capacity Image Buffer Disk / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
Memory Board / 6 MP / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
Seismic Kit / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
Standard Image Buffer / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
Basic Image Buffer / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
Enhanced Image Buffer / for KODAK DIGITAL SCIENCE 9000D Medical.Laser Printer and KODAK EKTASCAN 2180 Laser Printer
High Capacity Enhanced Image Buffer / for KODAK DIGITAL SCIENCE 9000D Medical Laser Printer and KODAK EKTASCAN 2180 Laser Printer
#### IV. INDICATION FOR USE
Images printed on film from the Laser Printer will be used for primary diagnoses, similar to other laser printers and CRT multi-format cameras. Evaluation of the hard copy output by trained health care professionals provides adequate opportunity for competent human intervention. Laser printers are not represented to be of use in supporting or sustaining human life, nor do they present a potential for unreasonable risk of illness or injury.
2
{2}------------------------------------------------
## V. Comparison of features
| IMAGING FEATURES | KELP 2180 | 9000 D | DV 8700 |
|-----------------------|------------------------------------|-----------------------------------|-----------|
| Laser Type | Red Diode | Red Diode | IR-Diode |
| Deflection | Polygon | Polygon | Galvo |
| xxxxScan | Scan Roller | Scan Roller | Galvo |
| Image Matrix | 4096x5120 | 4096x5120 | 4096x5120 |
| Bits/Pixel | 12/12 | 12/12 | 12/12 |
| Trim Option | Yes | Yes | Yes |
| Pixel Pitch (Microns) | 79 | 79 | 80 |
| Laser Power | 5mw | 50mw | |
| TONE SCALE FEATURES | | | |
| Contrast | -5 to +5 | -5 to +5 | |
| Density Max | 3.2 | 3.2 | |
| Curve Shape | 6 | 6 | |
| Interpolation | Cubic/Rep/<br>Sharp Cubic/Bilinear | Cubic/Rep<br>Sharp Cubic/Bilinear | Cubic/Rep |
| PHYSICAL FEATURES | | | |
| Footprint (in.) | 31 x 52 | 31 x 52 | 32 x 26 |
| Kodak Processor | 180LP Wet | Thermal | Thermal |
| Roomlight Film Supply | Yes | Yes | Yes |
| Film Sizes | 4 | 4 | 1 |
| PRODUCTIVITY | | | |
| Lockout Time (sec) | 0 | 0 | 0 |
| Cycle Time (sec) | 20 | 30 | 30 |
| Print Time | 12 | 26 | |
| Memory Capacity (MB) | 4G | 4G | 128MB |
| Autoprint | Yes | Yes | Yes |
| Random Erase | Yes | Yes | Yes |
| MULTIPLE INPUTS | KELP 2180 | 9000 D | DV 8700 |
| Max Inputs | 8 | 8 | 8 |
| Volatile Memory | No | No | Yes |
| Keypad/Autofilm | Yes | Yes | Yes |
3
{3}------------------------------------------------
#### VI. REGULATORY COMPLIANCE
The Kodak Digital Science 9000D Medical Laser Printer (MLP 9000D) is being designed and tested in conformance with the following standards.
Product Safety
| EN 60601-1 | Medical Electrical Equipment, General Requirements |
|---------------------|----------------------------------------------------------------------------------------------|
| EN 60825 | Safety of Laser Products, Part 1: Equipment Classification,<br>Requirement, and User's Guide |
| CSA C22.2 No. 601.1 | Medical Electrical Equipment, Part 1: General Requirements |
| UL 2601-1 | Medical Electrical Equipment, General Requirements |
| Laser safety: | Title 21CFR CHAPTER 1, /subchapter J, Part 1040.10 |
#### Electromagnetic Interference
EN 55011 Class B/1992, EN 50082-1/1992, EN 60950/1993 following the provisions of the applicable directives 89/336/EEC and amendments 72/23/EEC and amendments
EN 61000-3-2 Power Line Harmonics
EN 61000-3-3 Flicker
Electromagnetic Noise Susceptibility
EN 50082-1:92, Generic Immunity as tested per the following:
(a) IEC 801-2:91; ESD, 4 KV direct, 8 KV air.
(b) IEC 801-3:84; RF Immunity Severity Level 2.
- (c) IEC 801-4:88; EFT Severity Level 2
#### Seismic
Requirements of the State of California, Office of Statewide Health Planning and Development for "Pre-Approval" in accordance with California Administrative Code, Title 24, Part 2 and Title 22, Division 7, Chapter 7.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the edge. In the center of the seal is a symbol consisting of three stylized human figures. The figures are arranged in a vertical stack, with each figure represented by a simple, curved line. The overall design is clean and minimalist.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 7 1997
Robert L. Hiller Director, HI Regulatory Compliance Eastman Kodak Company Health Imaging Division 901 Elmgrove Road Rochester, NY 14653
K972847 Re:
Kodak Digital Science 9000D Medical Laser Printer Dated: July 30, 1997 Received: August 1, 1997 Regulatory class: Unclassified Procode: 90 LMC
Dear Mr. Hiller:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(K) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
hDliau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _ K97284
Device Name: KODAK DIGITAL SCIENCE 9000D MEDICAL LASER PRINTER
INDICATION FOR USE:
The KODAK Digital Science 9000D Medical Laser Printer is a laser printer designed to produce hard-copy films of digitized medical images. It is a newly designed printer to be added to Kodak's family of medical laser printer products that is intended for the centralized network printing solution for high volume sites for maximum productivity. This printer will be environmentally friendly by eliminating wet chemistry, site concerns, costs and will reduce chemical disposal problems. This is accomplished by using a totally proprietary imaging process.
Images printed on film from the laser printer will be used for primary diagnosis, similar to other laser printers and CRT mult-format cameras. Evaluation of the hard copy output by trained health care professionals provides adequate opportunity for competent human intervention. Laser printers are not represented to be of use in supporting or sustaining human life, nor do they present a potential for unreasonable risk of illness or injury.
Sincerely yours,
Robert L. Hiller
Robert L. Hiller
7/30/97
Robert L. Hiller Director, Regulatory Affairs KODAK Health Imaging
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CRF 801.109)
OR Over-the-
(Division Sign-Off)
Over-the-Counter Use
(Optional Format 1-2-96)
Division of Reproductive, Abdominal, EN and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.