K090469 · Carestream Health, Inc. · LMC · May 22, 2009 · Radiology
Device Facts
Record ID
K090469
Device Name
CARESTREAM DRYVIEW 5850 LASER IMAGER
Applicant
Carestream Health, Inc.
Product Code
LMC · Radiology
Decision Date
May 22, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
The CARESTREAM DRYVIEW 5850 Laser Imager is intended to provide highresolution hard copy images from digital imaging source output signals. The device is intended for use with KODAK DRYVIEW media including DVM (DryView mammography) films. The imager will interface with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography), MRI (Magnetic Resonance Imaging) and FFDM (Full Field Digital mammography). Image resizing is used to preserve true geometric size images. The images are to be used for medical diagnosis and referral to physicians and their patients.
Device Story
Laser imager producing high-resolution hard copy medical images from digital modality inputs (CR, DR, CT, MRI, FFDM). Device receives digital signals; merges image data with KODAK DRYVIEW media at laser imaging station; scans film; transports through integrated processor. Features Automated Image Quality Control (AIQC) to match printing power with film characteristics. Used in clinical settings; operated by medical personnel. Output provides physical film for diagnostic review and referral. No patient contact; device does not monitor or affect patients directly. Human intervention possible in case of malfunction.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness demonstrated through adherence to voluntary standards and performance testing of the AIQC system and test pattern generator to ensure consistency between input signals and output density.
Technological Characteristics
Laser printer; digital input interface. Materials: KODAK DRYVIEW media. Standards: DICOM, SMPTE, UL 60950, IEC 60601-1-1, IEC 60825-1, ISO 12207, ISO 14971. Features AIQC (Automated Image Quality Control) and test pattern generator. Connectivity: Modality-interfaced.
Indications for Use
Indicated for producing high-resolution hard copy medical images from digital modalities (CR, DR, CT, MRI, FFDM) for diagnostic use and physician/patient referral. No specific age or gender restrictions; intended for use with KODAK DRYVIEW media.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Kodak DRYVIEW 8900 Laser Imager with Mammography Accessory (K033821)
Submission Summary (Full Text)
{0}------------------------------------------------
K090469
Page 1 of 2
# 510(k) Summary
May 20, 2009
MAY 22 2009
CARESTREAM Health, Inc. 150 Verona Street Rochester NY 14608
Contact: Gayle Carroll 1 Imation Wav Oakdale, MN 55128 Phone: 651-393-1501 FAX: 651-393-1160
### Device
| Trade name: | CARESTREAM DRYVIEW 5850 Laser Imager |
|----------------------|-------------------------------------------------|
| Common name: | Laser Printer |
| Classification name: | Medical Image Hardcopy Device (21 CFR 892.2040) |
Predicate device
Kodak DRYVIEW 8900 Laser Imager with Mammography Accessory (K033821)
### Description and Intended Use of Device
The CARESTREAM DRYVIEW 5850 Laser Imager is intended to provide highresolution hard copy images from digital imaging source output signals. The device is intended for use with KODAK DRYVIEW media including DVM (DryView mammography) films. The imager will interface with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography), MRI (Magnetic Resonance Imaging) and FFDM (Full Field Digital mammography). Image resizing is used to preserve true geometric size images. The images are to be used for medical diagnosis and referral to physicians and their patients.
#### Technological Characteristics
The subject device and predicate devices use the same technical design base. The printers receive image data from the modality. User control is performed directly by the modality or through the host control. KODAK DRYVIEW imaging media is removed from a daylight cartridge and transported to the laser imaging station. Image data and media merge at the laser station and the film is scanned. The exposed media is transported through the integrated processor and exits the printer.
Software is used to control the image management and machine functions. AIQC (Automated Image Quality Control) matches printing power with film characteristics to provide consistently high image quality.
Performance Data
{1}------------------------------------------------
K090469
Page 2 of 2
Safety and effectiveness are assured via meeting voluntary standards, including: DICOM, SMPTE, UL 60950, IEC 60601-1-1, IEC 60825-1, ISO 12207 and ISO 14971.
#### Conclusion
The subject device, like the predicate, has no patient contact. The device does not control, monitor or otherwise affect any devices directly connected to or affecting the patient. Medical personnel review images displayed by the subject device and its predicate. This offers ample opportunity for competent human intervention in case of a malfunction or other failure.
The subject CARESTREAM DRYVIEW 5850 Laser Imager and predicate device KODAK DRYVIEW 8900 with Mammography Accessory have both been designed to the equivalent safety standards. As with the predicate device, a test pattern generator and automatic image quality control (AIQC) system are incorporated to assure consistency between input signals and output density. Both are high resolution printers with incorporated test patterns to assist in MQSA quality assurance testing.
Carestream Health therefore concludes that the CARESTREAM DRYVIEW 5850 Laser Imager is as safe and effective as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
MAY 22 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Gayle Carroll Regulatory Affairs Manager Carestream Health, Inc. 1 Imation Way, 304-4B-28 OAKDALE MN 55128-3414
Re: K090469
Trade/Device Name: Carestream DRYVIEW 5850 Laser Imager · Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: LMC Dated: February 20, 2009 Received: February 23, 2009
#### Dear Ms. Carroll:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
(Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 876.xxx (240) 276-0115 21 CFR 884.xxx (Obstetrics/Gynecology) (240) 276-0120 21 CFR 892.xxx (Radiology) (240) 276-0100 Other
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use:
510(K) Number (if known): __
K090469
Device Name: Carestream DRYVIEW 5850 Laser Imager
Indications for Use:
The CARESTREAM DRYVIEW 5850 Laser Imager is intended to provide highresolution hard copy images from digital imaging source output signals. The device is intended for use with KODAK DRYVIEW media including DVM (DryView mammography) films. The imager with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography), MRI (Magnetic Resonance Imaging) and FFDM (Full Field Digital mammography). Image resizing is used to preserve true geometric size images. The images are to be used for medical diagnosis and referral to physicians and their patients.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hulu Palmer
Page 1 of 1
(Division Sign-Off) Division of Reproductive. Abdominal and Radiological Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.