K011617 · Ferrania S.P.A. Stabilimento DI Ferrania · LMC · Jun 8, 2001 · Radiology
Device Facts
Record ID
K011617
Device Name
LIFEJET PRINTER 400
Applicant
Ferrania S.P.A. Stabilimento DI Ferrania
Product Code
LMC · Radiology
Decision Date
Jun 8, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The LifeJet™ Printer 400 is an ink jet printer designed to provide referral high quality color and/or gray scale hard copies of medical images generated by medical equipment. The LifeJet™ Printer 400 has been developed to be used together with LifeJet™ Medical Paper and LifeJet™ Medical Ink Jet Cartridges to achieve the best quality medical images.
Device Story
LifeJet™ Printer 400 is an inkjet-based medical image hardcopy device. It receives digital medical image data from medical equipment and produces color or gray-scale prints on coated paper. The system includes the printer, LifeJet™ Medical Paper, and LifeJet™ Medical Ink Jet Cartridges. Operated in clinical environments, the device driver allows users to format print layouts (one to eight images per page, portrait or landscape) during ongoing diagnostic examinations. The output provides clinicians with physical documentation for referral and review, facilitating medical workflow.
Clinical Evidence
Bench testing only. Performance testing compared the LifeJet™ Printer 400 to the predicate device by measuring printing time for a 203x203 mm SMPTE pattern (512x512 pixels), maximum optical density, and image mottling/bleeding at 100% density. Results demonstrated the device performs as well as or better than the predicate.
Technological Characteristics
Inkjet printing technology; 1200x1200 dpi resolution; 256 x CMYK color/gray scale capability. Interfaces via IEEE 1284 parallel port. Operates on Windows NT 4.0. Includes 2 MB RAM. Complies with UL, CSA, FCC, CE Mark, IEC 60601-1, and Energy Star standards. Uses proprietary coated paper and ink cartridges.
Indications for Use
Indicated for the production of high-quality color and/or gray-scale hard copies of medical images generated by medical equipment for referral purposes. Intended for prescription use.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
KODAK DIGITAL SCIENCE 1200 Distributed Medical Imager (K983905)
Submission Summary (Full Text)
{0}------------------------------------------------
JUN - 8 2001 Section 10 - 510(k) Summary of Safey and Effectiveness as required by 21 CFR 807.92(c) Submitted by Ferrania S.p.A. 10.1 Viale della Libertà 57 17014 Ferrania Savona (Italy) Ing. Mannella Paolo 10.2 Contact person Regulatory Manager +39 0195224055 (phone) +39 0195224546 (fax). E-mail: pmannella@ferraniait.com March 30, 2001 10.3 Date Summary Prepared: Common Name: Medical Ink-jet printer 10.4 Device name: Trade name: LifeJet™ Printer 400 Classification Name: Medical image hardcopy device (per 21 CFR 892.2040) KODAK DIGITAL SCIENCE 1200 Distributed 10.5 Predicate device: Medical Imager (510(k) number: K983905) Eastman Kodak Company Health Imaging Division 343 State Street Rochester, NY 14650
#### 10.6 Description of device
The LifeJet™ Printer 400 is an inkjet printer designed to produce high quality color and gray-scale referral medical images. Together with specifically designed medical paper and inks (LifeJet™ Medical Paper and LifeJet™ Medical Ink Jet Cartridges) the LifeJet™ Printer 400 forms the LifeJet™ 400 System, which is an innovative, convenient and work-flow improving medical printing platform, based on the state of the art inkjet technology, capable of delivering superior and reliable printed images for any medical documentation requirement.
#### 10.7 Statement of intended use
The LifeJet™ Printer 400 is an ink jet printer designed to provide referral high quality color and/or gray scale hard copies of medical images generated by medical equipment.
#### 10.8 Comparison with predicate device
The purpose and functionality of the Ferrania LifeJet™ 400 Printer are substantially equivalent to the KODAK DIGITAL SCIENCE 1200 Distributed
KOII617
{1}------------------------------------------------
## 510(k) PREMARKET NOTIFICATION FOR FERRANIA S.p.A. LifeJet™ Printer 400
## Section 10 - 510(k) Summary of Safey and Effectiveness as required by 21 CFR 807.92(c)
The basis for the equivalence is that both of them are computer peripheral printing devices which print monochrome or color images on paper, using inkjet technology.
The only difference is that the LifeJet™ Printer 400 driver allows the user to format the print layout by printing one to eight images per page, in strips of two, either in portrait or landscape mode, as soon as they are available and while the diagnostic examination goes on.
The technological equivalence of the two printers and the equivalence in terms of their features is summarized in the table below.
| Feature/<br>Specification | LifeJet™ Printer 400 | Kodak Digital Science 1200<br>Distributed Medical Imager |
|---------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Printing Technology | Ink-jet | Ink-jet |
| Operating System | Windows NT 4.0 | Windows 95/98, NT 4.0 |
| Printing Resolution | 1200x1200 dpi | 1200x1200 dpi |
| Tones | Color and gray scale | Color and gray scale |
| Number of Colors | 256 x CMYK | 256 x CMYK |
| Standards | UL, CSA, FCC, CE Mark, IEC 60601-<br>1, Energy Star | UL, CSA, FCC, EN50082, CISPR 22,<br>VCCI, CE Mark, C-Tick, Energy Star |
| Interfaces | Parallel: IEEE 1284 | Parallel: IEEE 1284 |
| Cartridges | Black LifeJet™ Medical Ink Jet<br>Cartridge,<br>Color LifeJet™ Medical Ink Jet<br>Cartridges | DMI Ink cartridge 1<br>DMI Ink Cartridge 2, or<br>DMI Black Cartridge |
| Printing Media | Coated paper | Coated paper<br>Coated film (blue) |
| Sheet sizes | A4, A5, A6, US letter, 4"x6",<br>personalized | A4, A5, A6, B5, Executive, Legal, US<br>letter, 4"x6", 8"x10" |
| Supply Tray<br>Capacity | 50 sheets | 45 sheets |
| RAM | 2 MB | Not available |
## Table of comparison of features of Predicate Device
### 10.9 Performance testing
The tests carried out to measure the printing time and image quality for the LifeJet™ Printer 400 in comparison with the predicate device confirmed their substantial equivalence also in terms of performances.
Objective test procedures were established to measure:
- the time to print one 203x203 mm image of a SMPTE pattern (512x512 pixel), .
- . the printed maximum optical density,
30-03-2001
{2}------------------------------------------------
## 510(k) PREMARKET NOTIFICATION FOR FERRANIA S.p.A. LifeJet™ Printer 400
# Section 10 - 510(k) Summary of Safey and Effectiveness as required by 21 CFR 807.92(c)
- the image mottling and bleeding @ 100% density. ●
### Conclusion 10.6
Based on the analysis of the comparisons made between the LifeJet™ Printer 400 Dason on the untaryse or are on the results of performance test results, Ferrania and the producted that the LifeJet™ Printer 400 is safe, effective and performs as well as or better than the predicate device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure with three flowing lines above it, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 8 2001
Ferrania S.P.A. % Ms. Chantel Carson Underwriters Laboratories Inc. 333 Pfingsten Road NORTHBROOK IL 60062
Re: K011617 LifeJet™ Printer 400 Dated: May 15, 2001 Received: May 25, 2001 Regulatory Class: II 21 CFR 892.2040/Procode: 90 LMC
### Dear Ms. Carson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Nancy Crogdon
Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure (s)
{4}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known)
# Device Name: LifeJet™ Printer 400
## Indications for Use:
The LifeJet™ Printer 400 is an ink jet printer designed to provide referral high quality color and/or gray scale hard copies of medical images generated by medical equipment.
The LifeJet™ Printer 400 has been developed to be used together with LifeJet™ Medical Paper and LifeJet™ Medical Ink Jet Cartridges to achieve the best quality medical images.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use √
OR
Over the Counter Use
David A. Stegman
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.