FILM SCANNER 300
Device Facts
| Record ID | K960839 |
|---|---|
| Device Name | FILM SCANNER 300 |
| Applicant | Cannon U.S.A., Inc. |
| Product Code | LMA · Radiology |
| Decision Date | Dec 18, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2030 |
| Device Class | Class 2 |
Indications for Use
Both devices are designed to digitize medical images from x-ray film.
Device Story
Desktop x-ray film digitizing system; utilizes CCD linear sensor; features automatic film feeder with 100-sheet capacity; high-speed scanning; supports DICOM and FTP protocols over TCP/IP; operated by clinical staff; converts analog film images to digital format for display and diagnostic review; facilitates digital storage and transmission of medical images.
Clinical Evidence
No clinical data; bench testing via software evaluation performed to demonstrate substantial equivalence.
Technological Characteristics
Desktop form factor; CCD linear sensor; automatic film feeder; connectivity via TCP/IP (DICOM, FTP protocols).
Indications for Use
Indicated for digitizing medical images from x-ray film for clinical use.
Regulatory Classification
Identification
A medical image digitizer is a device intended to convert an analog medical image into a digital format. Examples include systems employing video frame grabbers, and scanners which use lasers or charge-coupled devices.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std.). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Vidar VXR-12 (number not available)