K982563 · Ramsoft, Inc. · LLZ · Oct 6, 1998 · Radiology
Device Facts
Record ID
K982563
Device Name
ULTRAPRO MODEL 98
Applicant
Ramsoft, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed. Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.
Device Story
Ultrapro is a PACS workstation for medical imaging practices. It acquires images from standard medical video sources or digital inputs (DICOM, JPEG, TIFF) via video frame grabber or scanner. The system stores images on PC hardware; supports lossless RLE and lossy JPEG/Motion JPEG compression (ratios up to 60:1). Radiologists use the workstation to review images, adjust display parameters (contrast, brightness, zoom), perform annotations, and make distance measurements. Reports are created within the system. Data is communicated to other workstations or third-party DICOM 3.0 components via network/modem and archived to optical media. The system is intended for use by medical professionals in clinical settings to facilitate diagnostic review and clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence determination is not based on an assessment of performance tests.
Technological Characteristics
PC-based PACS workstation. Inputs: video frame grabber, TWAIN scanner, DICOM/PC formats. Outputs: display, network communication, optical media archive. Features: NTSC/PAL video support, RLE/JPEG/Motion JPEG compression, DICOM 3.0 compliance. Software-based image processing for contrast, brightness, zoom, annotation, and measurement. Standalone or networked configuration.
Indications for Use
Indicated for use by medical professionals to acquire, display, communicate, and archive diagnostic images from FDA-listed imaging equipment, including ultrasound, CT, MRI, x-ray, and video cassette recorders.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
ALI UltraPACS (K925965/A) with ALI comPACS data compression module (K963610)
CEMAX-ICON Archive Manager 2.0 and Autorad modules (K955092)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo on the left and a date on the right. The logo is circular and contains the text "RamSoft". The date is "OCT 6 1998". The text is in a bold, sans-serif font.
K982 56.3
# 510(k) Summary for Ultrapro 98 1
# 1.1 General Information
# 1.1.1 Name and Address of Manufacturer
RamSoft Inc. 37 Bankview Circle Toronto. ON M9W 6S6, CANADA Tel: (416) 674-1347 Fax: (416) 674-7147 Email: ram@ramsoft.on.ca Contact Person: Naganatha S Ramanathan
# 1.1.2 Establishment Registration Number
9006410
1.1.3 Common and Proprietary names of Device
Trade or Proprietary Name: Ultrapro™ Common Name: Picture Archiving and Communication System
### 1.1.4 Class
Systems, digital image communications - Class | 892.2020 LMD
### 12 Substantially Equivalent Devices
ALI UltraPACS (K925965/A) with ALI comPACS data compression module (K963610) Manufactured by: ALI Technologies, Inc
CEMAX-ICON Archive Manager 2.0 and Autorad modules (K955092) Manufactured by: CEMAX-ICON Inc.
### 1.3 Device Description
Ultrapro is a tool for a medical imaging practice. It can acquire medical images from any standard medical video source. Digital data in the DICOM format or other popular PC formats such as JPEG, TIFF, etc. can also be acquired.
Images may be acquired in uncompressed or compressed format (lossless or lossy). The lossy compression ratio is selected by viewing the original image and the lossy compressed image side by side and increasing the ratio while retaining clinically significant characteristics. Images that have been lossy compressed have the text "LOSSY xx:1" where xx is the compression ratio placed in the visible, non-diagnostic area of the image.
Related patient information can also be entered or scanned in with a TWAIN compliant scanner. Once acquired, Ultrapro serves as a digital storage for these images. Images are stored on the computer. Images may be communicated to other Ultrapro workstations or to third-party DICOM 3.0 compliant PACS components. The radiologist may review the images with Ultrapro and create reports. The display of images (contrast, brightness, zoom, etc.) may be adjusted at any time. Annotations may be drawn on images. Distance measurements may be made on images. These image parameters may be saved along with the image while still preserving the original image. This information can be communicated to other workstations through a network or modem. Finally, all of the information may be archived onto optical media for permanent storage.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called "RalliSoft". The logo features the company name in a stylized font inside of an oval shape. The oval is positioned in the center of a globe graphic. The globe is depicted with curved lines to represent the continents and oceans.
# 1.4 Intended Use Statement
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use with FDA approved imaging equipment that provides a video output or a digital output.
| | ALI UltraPACS with ALI comPACS | RamSoft Ultrapro |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Acquisition | Video frame grabber or DICOM<br>input | Video frame grabber, FDA<br>approved film scanner or DICOM<br>input |
| Compatible Video Signals | NTSC (up to 30 frames/s,<br>640x480, 256 grayscale, 16.7<br>million colors)<br>PAL (up to 25 frames/s, 768x576.<br>256 grayscale, 16.7 million colors) | NTSC (up to 30 frames/s,<br>640x480, 256 grayscale, 16.7<br>million colors)<br>PAL (up to 25 frames/s, 768x576,<br>256 grayscale, 16.7 million colors) |
| Cine Capability (motion capture) | Yes | Yes |
| Image Compression | Lossless RLE and Lossy JPEG | Lossless RLE and Lossy JPEG |
| Lossy Compression Ratio<br>Range | 4:1 to 30:1 | 4:1 to 60:1 |
| Uses JPEG lossy compression<br>in viewing, transmitting and<br>archiving still images | Yes | Yes |
| Uses Motion JPEG lossy<br>compression for viewing,<br>transmitting and archiving cine<br>images | Yes | Yes |
| DICOM 3.0 compliant | Yes | Yes |
| Imaging Modalities | Multi-modality | Multi-modality |
| Image Display Capability | Yes | Yes |
| Remote and Local Area Network<br>Communication | Yes | Yes |
| Image Annotation Capability | Yes | Yes |
| Archives Image on Optical<br>Media | Yes | Yes |
# 1.5 Companson of Technological Characteristics
# 1.6 Conclusion
In conclusion, Ultrapro is substantially equivalentitititi to the ALI product. The determination of substantial equivalence is not based on an assessment of performance tests.
ﺎ
55555 : 35453 :
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with its wings spread, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 6 1998
Naganatha S. Ramanathan, Ph.D. President Ramsoft, Inc. 37 Bankview Circle Toronto, Ontario Canada M9W 6S6
Re:
K982563 "Ultrapro 98, Digital Image Communications System Dated: July 22, 1998 Received: July 23, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Dr. Ramanathan:
We have reviewed your Section 5100% notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments on to devices that have been reclassified in accordance with the Feceral Food Drues and Cosmetic Act (Act). You may, therefore, narket the device, subject to the general controls of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical-Devices: Generaliation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition; FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odch/dsmamain.html"
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and -Radiological Health
Enclosure
{4}------------------------------------------------
Page of
t
| 510(k) Number (if known): | K982563 |
|---------------------------|-------------|
| Device Name: | ULTRAPRO 98 |
Indicatic is For Use:
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed.
Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.
(PL) ASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Gimel li. Syson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number
OR
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.