← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K242314

# Minimally Invasive Prostate Surgery Navigation System(Model:AmaKris SR1-A-2) (K242314)

_Augment Intelligent Medical System (China) Co., Ltd. · LLZ · Sep 11, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K242314

## Device Facts

- **Applicant:** Augment Intelligent Medical System (China) Co., Ltd.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Sep 11, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Intended Use

Minimally Invasive Prostate Surgery Navigation System(Model:AmaKris SR1-A-2) is intended to guide trained urologists in the planning and positioning of insertion tools during the performance of trans-perineal prostate biopsies and interventional procedures under trans-rectal biplane ultrasound guidance. The procedures include clinical diagnosis and treatment of prostate conditions: biopsy, single or multi-needle ablation, and particle implantation. The device provides real-time 2D ultrasound scanning of the prostate to construct a 3D model and fuses it with the MRI 3D model, which can assist the surgeon in planning the puncture path avoiding the urethra and pubic bones. The device also provides a 3D simulation showing the current and planned puncture path of insertion tools such as biopsy needle and therapeutic needle, taking into account patient movement.

## Device Story

System integrates computer, imaging, navigation, and robotic technology to guide trans-perineal prostate procedures. Inputs: multi-parameter MRI images and real-time 2D ultrasound from compatible third-party systems (SmartUs EXT-1M/3M). Processing: constructs 3D models of prostate, tumor, urethra, and pubis; performs MRI-TRUS fusion; calculates puncture paths. Output: 3D visualization of anatomy, planned puncture paths, and real-time needle tracking on a console display. Used in clinical settings by urologists. Robotic arm provides mechanical guidance for manual needle insertion. Benefits: assists in precise targeting while avoiding critical structures like urethra and pubic bones; accounts for patient movement during procedures.

## Clinical Evidence

No clinical data. Evidence based on non-clinical bench testing: safety (IEC 60601-1), EMC (IEC 60601-1-2), puncture precision (mean deviation 0.397 mm), phantom simulation accuracy (mean 0.435 mm), and model fusion accuracy (mean 1.4270 mm) using ICP algorithm.

## Technological Characteristics

System comprises an Intelligent Surgical Console and a navigation robotic arm. Connectivity: USB interface to third-party ultrasound beamformer. Software: Windows-based, supports DICOM import/export, 3D rendering, multi-planar reconstruction, and segmentation. Modeling algorithm uses Iterative Closest Point (ICP) for fusion. Sterilization: compatible with sterile probe sleeves and protective covers.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Minimally Invasive Prostate Surgery Navigation System (Model: AmaKris SR1-A-1) ([K221499](/device/K221499.md))

## Reference Devices

- SmartUs EXT-1M/3M ([K163121](/device/K163121.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

September 11, 2025

Augment Intelligent Medical System (China) Co., Ltd.
Yongwei Wang
COO
1801-1812, Floor 18, Block B, Kechuang No.1 Building, No.320
Pubin Road, Jiangpu Sub-District, Pukou District
Nanjing, Jiangsu 211808
China

Re: K242314
Trade/Device Name: Minimally Invasive Prostate Surgery Navigation System(Model:AmaKris SR1-A-2)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: July 29, 2025
Received: July 29, 2025

Dear Yongwei Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K242314 - Yongwei Wang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K242314 - Yongwei Wang

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Jessica Lamb, Ph.D.

Assistant Director

Imaging Software Team

DHT8B: Division of Radiological

Imaging and Radiation Therapy Devices

OHT8: Office of Radiological Health

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K242314  |   |
|  Device Name Minimally Invasive Prostate Surgery Navigation System(Model:AmaKris SR1-A-2)  |   |
|  Indications for Use (Describe) Minimally Invasive Prostate Surgery Navigation System(Model:AmaKris SR1-A-2) is intended to guide trained urologists in the planning and positioning of insertion tools during the performance of trans-perineal prostate biopsies and interventional procedures under trans-rectal biplane ultrasound guidance. The procedures include clinical diagnosis and treatment of prostate conditions: biopsy, single or multi-needle ablation, and particle implantation. The device provides real-time 2D ultrasound scanning of the prostate to construct a 3D model and fuses it with the MRI 3D model, which can assist the surgeon in planning the puncture path avoiding the urethra and pubic bones. The device also provides a 3D simulation showing the current and planned puncture path of insertion tools such as biopsy needle and therapeutic needle, taking into account patient movement.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

中交医疗

ZHONGYI MEDICAL

# 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: September.11,2025

# Submitter's Information

The submitter of this pre-market notification is:

|  Name: | Augment Intelligent Medical System (China) Co., Ltd.  |
| --- | --- |
|  Address: | 1801-1812, Floor 18, Block B, Kechuang No.1 Building, No.320 Pubin Road, Jiangpu Sub-District, Pukou District, Nanjing, Jiangsu 211808 China  |
|  Contact person: | Yongwei Wang  |
|  Title: | COO  |
|  E-mail: | wangyw@aimsrobots.com  |
|  Tel: | 86-13705163717  |

Device Identification

|  510(K) number: | K242314  |
| --- | --- |
|  Trade/Device Name: | Minimally Invasive Prostate Surgery Navigation System(Model: AmaKris SR1-A-2)  |
|  Common name: | System, Image Processing, Radiological  |
|  Regulation Number: | 892.2050  |
|  Regulation Name: | Medical image management and processing system  |
|  Regulation Class: | Class 2  |
|  Panel: | Radiology  |
|  Product Code: | LLZ  |

Predicate Device

|  510(K) number: | K221499  |
| --- | --- |
|  Trade/Device Name: | Minimally Invasive Prostate Surgery Navigation System(Model: AmaKris SR1-A-1)  |
|  Common name: | System, Image Processing, Radiological  |
|  Regulation Number: | 892.2050  |
|  Regulation Name: | Medical image management and processing system  |
|  Regulation Class: | Class 2  |
|  Panel: | Radiology  |
|  Product Code: | LLZ  |

510(k) Summary

{5}

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

中交医疗

ZHONGYI MEDICAL

# 4. Indication for Use

Minimally Invasive Prostate Surgery Navigation System(Model: AmaKris SR1-A-2) is intended to guide trained urologists in the planning and positioning of insertion tools during the performance of trans-perineal prostate biopsies and interventional procedures under trans-rectal biplane ultrasound guidance. The procedures include clinical diagnosis and treatment of prostate conditions: biopsy, single or multi-needle ablation, and particle implantation.

The device provides real-time 2D ultrasound scanning of the prostate to construct a 3D model and fuses it with the MRI 3D model, which can assist the surgeon in planning the puncture path avoiding the urethra and pubic bones. The device also provides a 3D simulation showing the current and planned puncture path of insertion tools such as biopsy needle and therapeutic needle, taking into account patient movement.

# 5. Device Description

Minimally Invasive Prostate Surgery Navigation System (Model: AmaKris SR1-A-2) is a clinical device that integrates computer technology, medical imaging technology, navigation technology and robot technology. It can be used to guide surgeons in the planning and positioning of insertion tools (e.g., biopsy needles, therapeutic needles, etc.) during trans-perineal prostate diagnostic and interventional procedures with the guidance of MRI and ultrasound (MRI-TRUS) fusion.

The device allows the importation of multi-parameter MRI images to construct 3D models of the prostate, tumor, urethra, and pubis. It can also construct 3D models based on real-time images from commercial ultrasound system, and provide the fusion between ultrasound and MRI 3D models as the guidance and visualization references for prostate biopsy and interventional procedures. The insertion of biopsy needles or therapeutic needles is performed manually by the surgeon. Other software functions include patient data management, multi-planar reconstruction and segmentation, image measurement, surgical planning, and report generation.

AmaKris SR1-A-2 is composed of two modules: Navigation robotic arm and Intelligent Surgical Console.

The device is used in conjunction with a commercial urological ultrasound device supplied by the hospital. The third-party ultrasound system is composed of a beamformer and a bi-plane transducer (probe) with a linear array and a convex array. The beamformer is connected to AmaKris SR1-A-2's IPC by a USB cable and is powered by a 12V power supply cable. AmaKris SR1-A-2 receives the image signal through the driver software provided by the ultrasound manufacturer and displays it on the screen.

# 6. Modifications between AmaKris SR1-A-2 and AmaKris SR1-A-1

a. Add an IFU claim to state a new function for planning and navigation of interventional procedures, such as soft tissue ablations.

510(k) Summary

{6}

中交医疗
ZHONGYI MEDICAL

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

b. Software update: add ablation plan and puncture function.
c. Navigation robotic arm's construction was upgraded.
d. Navigation robotic arm parameters were modified.
e. Add a Robotic arm bracket, which can lower the working height of the navigation robotic arm and expand the operating space.
f. Modeling algorithm was upgraded.
g. AmaKris SR1-A-2 needs to be used in conjunction with recommended models of ultrasound equipment, SmartUs EXT-1M/3M(K163121).
f. Other minor changes, such as IPC configuration upgrade, software UI upgrade, These changes do not significantly affect the use of the device, nor do they raise new or additional safety risks.

# 7. Compared to Predicate Device

Compared to the predicate devices, the subject device has the same intended use, similar product design, similar performance, same safety as the predicate device, the summarized comparison information is listed in the following table

|  SE Comparisons | Subject Device Minimally Invasive Prostate Surgery Navigation System (Model: AmaKris SR1-A-2) | Predicate Device Minimally Invasive Prostate Surgery Navigation System (Model: AmaKris SR1-A-1) K221499 | Similarities/ Differences  |
| --- | --- | --- | --- |
|  Indication for Use | Minimally Invasive Prostate Surgery Navigation System is intended to guide trained urologists in the planning and positioning of insertion tools during the performance of trans-perineal prostate biopsies and interventional procedures under trans-rectal biplane ultrasound guidance. Types of procedures include, but are not limited to, clinical diagnosis and treatment of prostate conditions such as: biopsy, single or multi-needle ablation, and particle implantation, etc. The device provides real-time 2D ultrasound scanning of the prostate to construct a 3D model and fuses it with the MRI 3D model, which can assist the surgeon in planning the puncture path avoiding the urethra and pubic bones. The | Minimally Invasive Prostate Surgery Navigation System (Model: AmaKris SR1-A-1) is intended for use by the trained physician or urologist to perform the computer-assisted prostate surgical procedures through transperineal skin under real- time transrectal ultrasound guidance. It provides the capability to register and fuse with MRI medical images in DICOM format. It provides real- time 3D visualization and localization for prostate, biopsy needle, and probe. It also provides the ability to display an image coordinates and guidewire that means the current and the projected future path of the biopsy needle. Other software features include patient data management, prostate and tumor modeling,3D image registration. | The subject device claims a new function for planning and navigation of interventional procedures. However,its work principle and operation process are the same.  |

510(k) Summary

{7}

9

中奖医疗

ZHONGYI MEDICAL

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

|  SE Comparisons | Subject Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-2) | Predicate Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-1)
K221499 | Similarities/
Differences  |
| --- | --- | --- | --- |
|   | device also provides a 3D simulation showing the current and planned puncture path of insertion tools such as biopsy needle and therapeutic needle, taking into account patient movement. | Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-1) is intended for treatment planning and guidance for clinical, interventional and/or diagnostic procedures. |   |
|  Product code | LLZ | LLZ | Same  |
|  Class | II | II | Same  |
|  Target anatomy | Prostate | Prostate | Same  |
|  Anatomy access | Transperineal | Transperineal | Same  |
|  Software  |   |   |   |
|  Windows OS | Yes | Yes | Same  |
|  Medical Imaging Software | Yes | Yes | Same  |
|  Image display  |   |   |   |
|  General Image 2D/3D Review | Yes | Yes | Same  |
|  3D Rendering View | Yes | Yes | Same  |
|  Live 2D Ultrasound | Yes | Yes | Same  |
|  Image Process  |   |   |   |
|  Gland Segmentation | Yes | Yes | Same  |
|  Image Registration | Yes | Yes | Same  |
|  Rigid Registration | Yes | Yes | Same  |
|  Elastic Registration | Yes | Yes | Same  |
|  Multi-Planar Reformation | Yes | Yes | Same  |
|  Connectivity  |   |   |   |
|  DICOM Import/Export | Yes | Yes | Same  |
|  Ultrasound | Yes | Yes | Same  |

510(k) Summary
Page 4 of Page 7

{8}

9

中奖医疗

ZHONGYI MEDICAL

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

|  SE Comparisons | Subject Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-2) | Predicate Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-1)
K221499 | Similarities/
Differences  |
| --- | --- | --- | --- |
|  Video |  |  |   |
|  Connect to ultrasound device | be used in conjunction with recommended models of ultrasound equipment, SmartUs EXT-1M/3M (K163121). | be used in conjunction with a third-party ultrasound machine and endorectal probe that supports type-B ultrasound | Different.
Only specified device can be used with the subject device.  |
|  Review Tools  |   |   |   |
|  Standard Image Viewing Tools | Yes | Yes | Same  |
|  Measurement Tools | Yes | Yes | Same  |
|  Annotation Tools | Yes | Yes | Same  |
|  Segmentation Tools | Yes | Yes | Same  |
|  Reporting Tools | Yes | Yes | Same  |
|  Video Capture | Yes | Yes | Same  |
|  Image Overlays | Yes | Yes | Same  |
|  Planning & Navigation  |   |   |   |
|  Import Prior Plan | Yes | Yes | Same  |
|  Import/Add Targets | Yes | Yes | Same  |
|  Plan/Mark Locations | Yes | Yes | Same  |
|  Navigation Type | Mechanical | Mechanical | Same  |
|  Hardware  |   |   |   |
|  Navigation robotic arm | comprises guider,control arm and support arm | comprises guider,manipulator, moving arm and rotating arm. | Different.
The construction was modified.  |

510(k) Summary
Page 5 of Page 7

{9}

9

中交医疗

ZHONGYI MEDICAL

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

|  SE Comparisons | Subject Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-2) | Predicate Device
Minimally Invasive Prostate Surgery Navigation System
(Model: AmaKris SR1-A-1)
K221499 | Similarities/
Differences  |
| --- | --- | --- | --- |
|  Robotic arm bracket | With | Without | Different.
It can lower the working height of the navigation robotic arm and expand the operating space.  |
|  Accessory & Disposable | Medical ultrasonic coupling agent,
biopsy needle/gun,
coaxial guiding needle,
ablation needle,
particle implantation gun/needle,
interventional needle,
probe cover,
rubber band,
adhesive tape,
sterile probe sleeve and device protective cover | Medical ultrasonic coupling agent,
biopsy needle/gun,
needle sheath,
anal dilator,
condom,
rubber band,
duct tape,
disposable plastic probe cover and protective cover. | Different.  |

The new device and predicate devices are substantially equivalent in the areas of technological characteristics such as basic design, features, energy source, method of operation, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.

# 8. Performance Data

## Clinical test:

Clinical testing is not required.

## Safety and Effectiveness

The AmaKris SR1-A-2 labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via the product (software and hardware) development process, verification and validation testing.

## Non-clinical data

The following testings were conducted on the subject device AmaKris SR1-A-2.

- Safety test were conducted according to IEC 60601-1:2005+AMD1:2012+AMD2:2020.

510(k) Summary

{10}

9

中奖医疗

ZHONGYI MEDICAL

Traditional 510(k) Submission of Minimally Invasive Prostate Surgery Navigation System

- EMC testings were conducted according to IEC 60601-1-2:2014+A1:2020.
- Verification of Accuracy and Precision:
Method: 25 Points Test
Acceptance criteria: Mean Deviation ≤ 2.1 mm, Standard Deviation ≤ 1.3 mm
Results: the total mean of the target point puncture precision was 0.397 mm, and the mean variance was 0.2043.
- Simulation Testing for Verifying Accuracy and Precision:
Method: a. Egg Phantom Test b. Metal Needle Phantom Test
Acceptance criteria: Mean Deviation ≤ 2.1 mm, Standard Deviation ≤ 1.3 mm
Results: The test results showed the device system is able to navigate puncture needles to targets within a defined boundary. The average navigation accuracy is 0.435 mm with a mean standard deviation of 0.1903, which means the navigation accuracy is within 1.0 mm radius
- Model Fusion Accuracy and Precision verification
Method: used ICP (Iterative Closest Point) algorithm for model fusion.
Acceptance criteria: Mean Deviation ≤ 2.4 mm, Standard Deviation ≤ 1.2 mm
Results: According to the analysis of 4 cases, the mean value of the distance between A and B is 1.4270 mm with a mean standard deviation 0.86617. Furthermore, the difference of target direction does not have statistically significant effect on model fusion accuracy.
- Software Verification and Validation: To ensure all software updates meets the specifications and the intended purpose.

# 9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicated device (K221499).

510(k) Summary

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K242314](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K242314)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
