K241495 · Jazz Imaging · LLZ · Nov 21, 2024 · Radiology
Device Facts
Record ID
K241495
Device Name
Jazz Classic
Applicant
Jazz Imaging
Product Code
LLZ · Radiology
Decision Date
Nov 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Intended Use
Jazz Classic is a clinical software application that captures images and data from Jazz intraoral xray sensors and various imaging sources (e.g., radiographic devices, digital still/video capture devices, and generic image devices such as scanners). Jazz Classic enables the storage of images, clinical notes, and clinical exam data. Jazz Classic is intended to be used for general populations which includes use for pediatric populations. It is intended to acquire, display, edit (e.g., resize, enhance), review, print, and distribute images using standard PC hardware. Jazz Classic is intended for diagnostic and non-diagnostic purposes by dental professionals trained to provide dental care.
Device Story
Jazz Classic is a dental imaging software application operating on standard PC hardware. It captures images from intraoral x-ray sensors, panoramic/cephalometric scanners, and intraoral cameras. The software enables dental professionals to store, display, edit (resize, enhance), review, print, and distribute these images. It integrates with practice management software for scheduling, clinical notes, and billing. The device provides tools for image manipulation to assist in diagnosis and treatment planning. It is used in dental clinics by trained dental professionals. The output is displayed on a PC monitor, allowing the clinician to evaluate dental conditions. The device does not perform automated diagnosis; it serves as a tool to enhance diagnostic value for the clinician.
Clinical Evidence
Bench testing and clinical evaluation performed. Bench testing used scientific phantom targets to compare Jazz Classic against the predicate (CLIO/CLIOSOFT), demonstrating superior signal-to-noise ratio (SNR) and resolution. Clinical evaluation based on expert opinion confirmed the software meets operational and clinical needs for acquiring and evaluating dental images safely and effectively.
Technological Characteristics
Software-only medical device; operates on standard PC hardware (Windows 10 64-bit or newer). Developed on Microsoft .NET platform. Features include lossless image compression, cybersecurity detection, and pediatric-specific imaging templates. Integrates with TWAIN-compliant imaging devices and client-server practice management systems. Supports DICOM, PNG, JPEG, and TIF file formats.
Indications for Use
Indicated for general patient populations, including pediatric, requiring dental imaging. Used by trained dental professionals for acquisition, display, enhancement, review, storage, and distribution of dental radiographic and clinical images for diagnostic and non-diagnostic purposes.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K101654 — IMPROVIS IMAGING SYSTEM · American Dental Partners, Inc. · Nov 10, 2010
K250947 — VistaSoft 4.0 and VisionX 4.0 · Durr Dental SE · Aug 27, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.
November 21, 2024
Jazz Imaging % W. Edward Johansen Official Correspondent 770 Charcot Ave SAN JOSE, CA 95131
Re: K241495
Trade/Device Name: Jazz Classic Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: October 23, 2024 Received: October 23, 2024
Dear W. Edward Johansen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K241495
Device Name
Jazz Classic
### Indications for Use (Describe)
Jazz Classic is a clinical software application that captures images and data from Jazz intraoral xray sensors and various imaging sources (e.g., radiographic devices, digital still/video capture devices, and generic image devices such as scanners). Jazz Classic enables the storage of images, clinical notes, and clinical exam data. Jazz Classic is intended to be used for general populations which includes use for pediatric populations.
It is intended to acquire, display, edit (e.g., resize, enhance), review, print, and distribute images using standard PC hardware. Jazz Classic is intended for diagnostic and non-diagnostic purposes by dental professionals trained to provide dental care.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K241495 | 510(k) Summary | Prepared on: | 2024-11-19 |
|-----------|---------|----------------|--------------|------------|
|-----------|---------|----------------|--------------|------------|
Contact Details
| Applicant Name | Jazz Imaging |
|-----------------------------|-------------------------------------------------|
| Applicant Address | 770 Charcot Ave San Jose CA 95131 United States |
| Applicant Contact Telephone | (310) 795-7425 |
| Applicant Contact | Mr. W. Edward Johansen |
| Applicant Contact Email | wedjohansen@msn.com |
Device Name
| Device Trade Name | Jazz Classic |
|---------------------|------------------------------------------------|
| Common Name | Dental Imaging Software |
| Classification Name | Medical Image Management and Processing System |
| Regulation Number | 21 C.F.R. §892.2050 |
| Product Code(s) | LLZ |
Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K110886 | CLIO/CLIOSOFT | LLZ |
Device Description Summary
Jazz Classic is a dental imaging software application that equips trained dental professionals with the tools to capture, process, organize,save, and share diagnostic images of patients. The software operates on standard PC hardware and displays images on the PC's connected display or monitor. Images can be acquired from digital dental imaging devices, including intraoral x-ray sental panoramic and cephalometric scanners, and intraoral cameras. It stores images using lossless compression, and can export them as DICOM, PNG, JPEG, or TIF files. It enables dental practitioners to visualize, and manipulate patient images to assist in case diagnosis, review, and treatment planning to further enhance the diagnostic value of images.
Jazz Classic integrates with client-server practice management software, bridging patient information from the user's existing practice management solutions to be used for scheduling, clinical note-taking, and billing.
Jazz Classic neither contacts the patient nor controls any life sustaining devices. Diagnosis is not performed by the software but by dentists and other licensed dental professionals.
## Intended Use/Indications for Use
Jazz Classic is a clinical software application that captures intraoral x-ray sensors and various imaging sources (e.g., radiographic devices, digital still/video capture devices such as scanners). Jazz Classic enables the storage of images, clinical notes, and clinical exam data. Jazz Classic is intended to be used for general populations which includes use for pediatric populations.
21 CFR 807.92(a)(5)
lt is intended to acquire, display, edit (e.g., review, print, and distribute images using standard PC hardware. Jazz Classic is intended for diagnostic and non-diagnostic purposes by dental professionals trained to provide dental care.
{5}------------------------------------------------
# Indications for Use Comparison
## 21 CFR 807.92(a)(5)
The indications for use of Jazz Classic is the same as the predicate device, with few exceptions that do not constitute a new intended use or raise any new or potential risks to the user or patient. The Jazz Classic subject device shares the ClioSoft predicate device. Both devices are intended to acquire, display, edit, review, store, print, and distribute images using standard PC hardware. The subject device is intended for general patient populations whereas the indications for use of the ClioSoft predicate does not specify the intent population. Jazz Classic is intended for diagnostic and non-diagnostic use by trained dental professionals who provide dental care, while the predicate does not specify the intended user of the device.
## Technological Comparison
## 21 CFR 807.92(a)(6)
Jazz Classic and CLIO/CLIOSOFT are intended to enable dentists, and other dental professionals to capture diagnostic images.
Both software allows for the capture, storage, review, distribution, adjustment, and annotation of dental still/ video sources. Both integrate with dental imaging TWAIN device for acquiring diagnostic images.
Other similarities include that both Jazz Classic and ClioSoft both process digital images to enhance diagnostic value, both create a database of patients and exam images in patient folders on a server, both software can operate as a client-server-based system, and both offer users database management and storage features.
Some differences between the software applications include the OS versions on which the software operate, cybersecurity features, image compression features, and image templates for pediations. Jazz Classic was developed on a Microsoft .NET platform and only operates on Windows 10 64-bit OS and newer, whereas ClioSoft was developed on a Microsoft platform and operates on Windows XP OS 32-bit and 64-bit architectures. Jazz Classic is capable of lossless image compression when exporting and emailing images, whereas ClioSoft is not. Jazz Classic has cybersecurity detection features whereas ClioSoft does not. Jazz Classic has dedicated imaging templates and tooth numbering systems intended for pediatic populations, whereas ClioSoft does not.
Jazz Classic is a software only medical device. ClioSoft is a software device that is part of an X-ray system, which includes Clio X-ray sensor hardware that connects to a PC.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Jazz lmaqing has conducted design verification, bench testing validation to confirm that Jazz Classic dental imaqing software application is capable of acquiring, enhancing, reviewing, printing, and distributing diagnostic images.
Verification testing confirmed that the software design requirements have been met.
Bench testing was performed on Jazz Classic and the predicate device, Cliosoft, using images acquired from scientific phantom targets. The results confirmed that Jazz Classic produced superior signal-to-noise ratio (SNR) and resolution compared to the predicate device, validating its image processing capabilities.
Jazz Imaging also performed a clinical evaluation to assess the intended effectiveness of Jazz Classic in a clinical environment. Based on expert opinion, Jazz Classic meets operational and clinical needs to acquire and evaluate dental images safely and effectively.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.