K241094 · Turing Medical Technologies, Inc. · LLZ · Aug 13, 2024 · Radiology
Device Facts
Record ID
K241094
Device Name
Bullsai
Applicant
Turing Medical Technologies, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Bullsai is composed of a set of modules intended for analysis and processing of medical images and other healthcare data. It includes functions for image manipulation, basic measurements, and planning. Bullsai is indicated for use in image processing, registration, atlas-assisted visualization and seqmentation, and target export creation and selection of structural MRI images, where an output can be generated for use by a system capable of reading DICOM image sets. Bullsai is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts based on diffusion reconstruction methods and for the use of said maps to select and create exports. Typical users of Bullsai are medical professionals, including but not limited to surgeons, clinicians. and radiologists.
Device Story
Bullsai is a cloud-deployed, software-only image processing package for neurosurgical planning. It imports DICOM structural and diffusion-weighted MRI (DWI) images from hospital PACS. Preprocessing includes denoising, debiasing, skull stripping, susceptibility distortion, and head motion correction. The device uses generalized q-sampling imaging (GQI)—a model-free diffusion reconstruction method—to estimate voxel-level white matter microstructure and fiber orientations. An iterative heuristic tractography algorithm, emulating manual neuroanatomist seeding, generates 3D white matter tract maps. The system removes PHI, links data to an encryption key, and exports DICOM outputs (black/white and color) for viewing in external neurosurgical planning software. Used by surgeons and radiologists to support neurosurgical planning; output facilitates visualization of patient-specific anatomy and white matter tracts.
Clinical Evidence
No clinical trials; performance validated via bench testing. Clinical accuracy of outputs evaluated by experts for presurgical planning. Compatibility validated with major neuronavigation software. Repeatability and reproducibility validated across major MRI scanner manufacturer brands.
Indicated for medical professionals (surgeons, clinicians, radiologists) to process structural and diffusion-weighted MRI images for neurosurgical planning. Includes image manipulation, registration, atlas-assisted visualization/segmentation, and generation of 3D white matter tract maps for export to DICOM-compatible systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
August 13, 2024
Turing Medical Technologies, Inc. Christa Nova Lead Manager of QARA (Quality Assurance and Regulatory Affairs) 393 N Euclid Ave Suite 310 Saint Louis, Missouri 63108
Re: K241094
Trade/Device Name: Bullsai Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: July 18, 2024 Received: July 18, 2024
Dear Christa Nova:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
Bullsai
Indications for Use (Describe)
Bullsai is composed of a set of modules intended for analysis and processing of medical images and other healthcare data. It includes functions for image manipulation, basic measurements, and planning.
Bullsai is indicated for use in image processing, registration, atlas-assisted visualization and seqmentation, and target export creation and selection of structural MRI images, where an output can be generated for use by a system capable of reading DICOM image sets.
Bullsai is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts based on diffusion reconstruction methods and for the use of said maps to select and create exports.
Typical users of Bullsai are medical professionals, including but not limited to surgeons, clinicians. and radiologists.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | Bullsai |
|--------------------|-------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Turing Medical Technologies, Inc.<br>393 N EUCLID AVE STE 310,<br>SAINT LOUIS, MISSOURI 63108 UNITED STATES |
| Contact: | Christa Nova<br>Lead Manager of QARA<br>419-388-3288 |
| Date Prepared: | August 9, 2024 |
| Classification: | Medical image management and processing system, 21 CFR<br>892.2050 |
| Class: | II |
| Product Code: | LLZ |
| Primary Predicate: | Quicktome Software Suite, K222359 |
### Indications For Use:
Bullsai is composed of a set of modules intended for analysis and processing of medical images and other healthcare data. It includes functions for image manipulation, basic measurements, and planning.
Bullsai is indicated for use in image processing, registration, atlas-assisted visualization and segmentation, and target export creation and selection of structural MRI images, where an output can be generated for use by a system capable of reading DICOM image sets.
Bullsai is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts based on diffusion reconstruction methods and for the use of said maps to select and create exports.
Typical users of Bullsai are medical professionals, including but not limited to surgeons, clinicians, and radiologists.
{4}------------------------------------------------
## Device Description:
Bullsai is a software-only, cloud-deployed, image processing package which can be used to perform DICOM image processing and analysis.
Bullsai can receive ("import") DICOM images from picture archiving and communication systems (PACS), including Diffusion Weighted Imaging (DWI) and structural brain imaging.
Bullsai removes protected health information (PHI) and links the dataset to an encryption key, which is then used to relink the data back to the patient when the data is exported to a medical imaging platform such as a hospital PACS or other DICOM device.
The software provides a workflow for a clinician to:
- · Select an image for planning and visualization,
- Validate image quality,
- · Export black and white and color DICOMs for use in systems that can view DICOM images.
Bullsai uses advanced MRI processing to deliver patient-specific anatomy and tractography results to support physicians in neurosurgical planning. Bullasi preprocessing steps include denoising, debiasing, skull stripping, susceptibility distortion and head motion correction to ensure input data can support generation of tractography and segmentation results. Bullsai uses generalized q-sampling imaging (GQI) to estimate the voxel-level white matter microstructure: GQI is a model-free diffusion reconstruction method that uses the Fourier transform relationship between the diffuse MR signal and the potential diffusion displacement for resolving fiber orientations and the anisotropy of water. Patient-specific anatomical segmentation and GQI estimated fiber orientations are used in Bullsai as part of an iterative heuristic tractography algorithm that emulates the manual work of neuroanatomists who iteratively seed tracts and remove aberrant fibers. Results are shared as DICOM outputs which can readily be viewed and edited in standard neurosurgical planning software packages.
### Predicate Device:
Turing Medical Technologies, Inc. submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, Bullsai is substantially equivalent in indications, design principles, and performance to the following predicate device, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
Primary Predicate: Quicktome Software Suite (K222359)
### Performance Testing Summary:
Software verification and validation testing were conducted. Documentation and relevant standards were provided.
| Item | Predicate:<br>Quicktome Software<br>Suite<br>K222359 | Subject Device:<br>Bullsai | Comparison |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II | Class II | Identical |
| Product<br>Code | LLZ | LLZ | Identical |
| Device<br>Description | | Similar; | |
| Quicktome is a software-only, cloud- deployed, image processing package which can be used to perform DICOM image viewing, image processing, and analysis.<br><br>Quicktome can receive (“import”) DICOM images from picture archiving and communication systems (PACS), acquired with MRI, including Diffusion Weighted Imaging (DWI) sequences, T1, T2, BOLD, and FLAIR images. Quicktome can also receive Resting State functional MRI (rs-fMRI) blood- oxygen-level- dependent (BOLD) datasets.<br><br>Once received, Quicktome removes protected health information (PHI) and links the dataset to an encryption key, which is then used to relink the data back to the patient when the data is exported to hospital PACS or other DICOM device.<br><br>The software provides a workflow for a clinician to:<br>Select an image for planning and visualization, Validate image quality, Explore the available anatomical regions, network templates, tractography bundles, and parcellations, Select regions of interest, Display resting state fMRI (BOLD) correlation maps using task-analogous seeds for Motor, Vision and Language networks, and Export black and white and color DICOMs for use in systems that can view DICOM images. | Bullsai is a software-only, cloud-deployed, image processing package which can be used to perform DICOM image processing and analysis.<br><br>Bullsai can receive (“import”) DICOM images from picture archiving and communication systems (PACS), including Diffusion Weighted Imaging (DWI) and structural brain imaging. Bullsai removes protected health information (PHI) and links the dataset to an encryption key, which is then used to relink the data back to the patient when the data is exported to a medical imaging platform such as a hospital PACS or other DICOM device.<br><br>The software provides a workflow for a clinician to:<br>Select an image for planning and visualization, Validate image quality, Export black and white and color DICOMs for use in systems that can view DICOM images.<br>Bullsai uses advanced MRI processing to deliver patient-specific anatomy and tractography results to support physicians in neurosurgical planning. Bullasi preprocessing steps include denoising, debiasing, skull stripping, susceptibility distortion and head motion correction to ensure input data can support generation of tractography and segmentation results. Bullsai uses generalized q-sampling imaging (GQI) to estimate the voxel-level white matter microstructure: GQI is a model-free diffusion reconstruction method that uses the Fourier transform relationship between the diffuse MR signal and the potential diffusion displacement for resolving fiber orientations and the anisotropy of water. Patient-specific anatomical segmentation and GQI estimated fiber orientations are used in Bullsai as part of an iterative heuristic tractography algorithm that emulates the manual work of neuroanatomists who iteratively seed tracts and remove aberrant fibers. Results are shared as DICOM outputs which can readily be viewed and edited in standard neurosurgical planning software packages. | 1) Bullsai does not have a viewer and therefore does not provide display, image viewing, or clinician exploration of device outputs within the device.<br>2) Bullsai does not support the processing or visualization of resting-state MRI studies.<br>3) Description of the process and GQI included. | |
| Item | Predicate:<br>Quicktome Software Suite<br>K222359 | Subject Device:<br>Bullsai | Comparison |
| Indication<br>for Use | The Quicktome Software Suite is<br>composed of a set of modules intended<br>for display of medical images and<br>other healthcare data. It includes<br>functions for image review, image<br>manipulation, basic measurements,<br>planning, 3D visualization (MPR<br>reconstructions and 3D volume<br>rendering) and display of BOLD<br>(blood oxygen level dependent)<br>resting-state MRI scan studies.<br>Modules are available for image<br>processing, atlas-assisted visualization<br>and segmentation, resting state<br>analysis and visualization, and target<br>export creation and selection, where<br>an output<br>can be generated for use by a system<br>capable of reading DICOM image sets.<br>Quicktome is indicated for use in the<br>processing of diffusion-weighted MRI<br>sequences into 3D maps that represent<br>white matter tracts based on<br>constrained spherical deconvolution<br>methods and for the use of said maps<br>to select and create exports.<br>Quicktome can generate motor,<br>language, and vision resting state<br>fMRI correlation maps using task-<br>analogous seeds.<br>Typical users of Quicktome are medical<br>professionals, including but not limited<br>to surgeons, clinicians, and radiologists. | Bullsai is composed of a set of<br>modules intended for analysis and<br>processing of medical images and<br>other healthcare data. It includes<br>functions for image manipulation,<br>basic measurements, and planning.<br>Bullsai is indicated for use in image<br>processing, registration, atlas-assisted<br>visualization and segmentation, and<br>target export creation and selection of<br>structural MRI images, where an<br>output can be generated for use by a<br>system capable of reading DICOM<br>image sets.<br>Bullsai is indicated for use in the<br>processing of diffusion-weighted MRI<br>sequences into 3D maps that represent<br>white matter tracts based on diffusion<br>reconstruction methods and for the use<br>of said maps to select and create<br>exports.<br>Typical users of Bullsai are medical<br>professionals, including but not limited<br>to surgeons, clinicians, and radiologists. | Similar; 1)<br>Bullsai does not<br>have a viewer<br>and therefore<br>does not provide<br>display, image<br>viewing, or 3D<br>visualization. 2)<br>Bullsai does not<br>support the<br>processing or<br>visualization of<br>resting-state<br>MRI studies. 3)<br>Bullsai uses<br>diffusion<br>reconstruction<br>methods rather<br>than constrained<br>spherical<br>deconvolution<br>methods. |
### Substantial Equivalence:
{5}------------------------------------------------
{6}------------------------------------------------
Bullsai is substantially equivalent to the identified predicate (legally marketed) device. Bullsai has the same intended use and technological characteristics as the predicate device and or different intended use and technological characteristics that do not raise different questions of safety and effectiveness.
At a high level, Bullsai and predicate device are based on the following same elements:
- . Bullsai is composed of a set of modules intended for medical images and other healthcare data.
- Bullsai includes functions for image manipulation, basic measurements, and planning. ●
- . Bullsai modules are available for image processing, atlas-assisted visualization and segmentation, and target export creation and selection, where an output can be generated for use by a system capable of reading DICOM image sets.
- . Bullsai is indicated for use in the processing of diffusion-weighted MRI sequences into 3D maps that represent white matter tracts and for the use of said maps to select and create exports.
- Typical users of Bullsai are medical professionals, including but not limited to surgeons, ● clinicians, and radiologists.
- Bullsai is a software-only, cloud-deployed, image processing package which can be used ●
{7}------------------------------------------------
to perform DICOM image processing and analysis.
- . Bullsai can receive ("import") DICOM images from picture archiving and communication systems (PACS).
- Bullsai provides support to structural imaging techniques including Diffusion Weighted ● Imaging (DWI).
- Bullsai accepts standardized imaging formats for volumetric medical imaging.
- . Bullsai removes protected health information (PHI) and links the dataset to an encryption key, which is then used to relink the data back to the patient when the data is exported to a medical imaging platform such as a hospital PACS or other DICOM device.
- Bullsai provides a workflow for a clinician to select an image for planning and visualization, validate image quality, and export black and white and color DICOMs for use in systems that can view DICOM images.
At a high level, Bullsai and predicate device have the following different elements:
- Bullsai does not have a viewer and therefore does not provide display, image viewing, 3D ● visualization, or clinician exploration of device outputs within the device.
- Bullsai does not support the processing or visualization of resting-state MRI studies.
- Bullsai uses diffusion reconstruction methods rather than constrained spherical deconvolution methodologies, to process diffusion-weighted MRI sequences into 3D maps that represent white matter tracts. Generalized q-sampling imaging (GQI) is a model-free diffusion reconstruction method that uses the Fourier transform relationship between the diffuse MR signal and the potential diffusion displacement for resolving fiber orientations and the anisotropy of water.
Validation of differences compared to predicate devices:
- . Clinical accuracy of the Bullsai device outputs were evaluated and validated by clinical experts for the clinical intended uses of presurgical planning.
- Bullsai device output's were validated compatibility with major neuronavigation software ● systems.
- Bullsai device output's were validated for repeatability and reproducibility across major scanner manufacturer brands.
Differences:
- The predicate device conducted a summative usability evaluation and design validation with ● representative users; Bullsai does have a viewer and therefore a usability study was not conducted.
- The predicate device conducted performance evaluations for their BOLD processing pipeline. Bullsai does not support resting-state BOLD MRI data and therefore a similar study was not conducted.
- The predicate device compared resting state fMRI correlation maps to task-based fMRI activation maps; Bullsai does not support resting state fMRI data and therefore a similar study was not conducted.
## Conclusion:
The subject device and the predicate devices have intended use and have similar technological characteristics. The data included in this submission demonstrate substantial equivalence to the predicate device listed above. Bullsai is as safe, as effective, and performs as well as, or better, than the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.