← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K230906

# Konicaminolta DI-X1 (K230906)

_Konica Minolta, Inc. · LLZ · Apr 25, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K230906

## Device Facts

- **Applicant:** Konica Minolta, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Apr 25, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. RF Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography.

## Device Story

Standalone software installed on off-the-shelf servers/PCs; receives digital X-ray images (single/multi-frame) from imaging modalities (RF, DX, CR). Performs image processing, storage, communication, and display; supports browser-based client access. New features include TD-MODE (tracheal wall contour extraction; min/max diameter display), Position Tracking (reference point tracking with position change graphs), and Signal Value Change (graphical display of signal changes within user-specified ROI). Data redundancy provided via RAID1 mirroring. Used in clinical environments; operated by healthcare professionals to view, measure, and analyze images. Output aids clinical decision-making by providing quantitative measurements and visual analysis of X-ray data.

## Clinical Evidence

Bench testing only. Verification activities and risk analysis confirmed that all predetermined acceptance criteria were met. No clinical studies were required.

## Technological Characteristics

Standalone software for off-the-shelf servers/PCs. Supports DICOM 3.0 connectivity. Features RAID1 mirroring for data redundancy. Processing modes include FE, DM, BS, PL, PH, PH2, LM, and TD-MODE. Measurement tools include distance, angle, area, CTR, position tracking, and signal value change.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- KONICAMINOLTA DI-X1 ([K212685](/device/K212685.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

Konica Minolta, Inc. % Jan Maniscalco Executive Vice President QA/RA Konica Minolta Healthcare Americas, Inc. 411 Newark Pompton Turnpike Wayne, New Jersey 07470

Re: K230906

April 25, 2023

Trade/Device Name: Konicaminolta DI-X1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: March 31, 2023 Received: March 31, 2023

Dear Jan Maniscalco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230906

Device Name KONICAMINOLTA DI-X1

#### Indications for Use (Describe)

KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. RF Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)  </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)  </span>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Konica Minolta logo along with the text "150 YEARS". The Konica Minolta logo consists of a blue sphere with white lines inside and the company name in black letters. The "150 YEARS" text is in black, with the "150" above a blue line and the word "YEARS" below the line.

# 510(k) Summary

# K23xxxx (will be assigned)

| Company:           | KONICA MINOLTA, INC.<br>1 Sakura-machi, Hino-shi, 191-8511 Japan                                                                                                                   |  |  |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact:           | Tsutomu Fukui<br>Senior Manager of Regulatory & QMS Division<br>1 Sakura-machi, Hino-shi, 191-8511 Japan<br>Telephone: +81 42 589 8429<br>Email : tsutomu.fukui1@konicaminolta.com |  |  |
| Date Prepared:     | March 31, 2023                                                                                                                                                                     |  |  |
| Device Name:       |                                                                                                                                                                                    |  |  |
| Trade Name:        | KONICAMINOLTA DI-X1                                                                                                                                                                |  |  |
| Common Name:       | Medical Image Management and Processing System                                                                                                                                     |  |  |
| Regulation Number: | 21 CFR 892.2050                                                                                                                                                                    |  |  |
| Regulatory Class:  | Class II                                                                                                                                                                           |  |  |
| Product Code(s):   | LLZ                                                                                                                                                                                |  |  |
| Predicate Device:  | K212685 - KONICAMINOLTA DI-X1<br>KONICA MINOLTA, INC.<br>Regulation Number: 21 CFR 892.2050<br>Product Codes: LLZ                                                                  |  |  |

# Device Description

KONICAMINOLTA DI-X1 is a software device that performs image processing and display using Xray digital images (single-frame images, multi-frame images) generated by various diagnostic imaging modality consoles. It is a standalone software device intended to install onto off-the shelf Servers and PCs.

KONICAMINOLTA DI-X1 receives X-ray digital images, including serial images, processes the received images, as well as displays and sends the resulting images to PACS and other devices. In addition, KONICAMINOLTA DI-X1 can display images through the browser connection with the client that displays and process images, and instruct transmission of images.

Giving Shape to Ideas

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Konica Minolta logo and the text "150 YEARS". The Konica Minolta logo is a blue circle with white horizontal lines inside. The text "KONICA MINOLTA" is written in black letters below the logo. To the right of the logo is the number 150 in large black font with the word "YEARS" written below it.

The personal computer used in KONICAMINOLTA DI-X1 stores the same data in two hard disks in real time using RAID1 mirroring function. Thus, even if one hard disk is defective, operations can be continued with the other hard disk which has the same data.

Modifications are made to add the TD-MODE, Position Tracking and Signal Value Change. The TD-MODE is designed to extract the initial contour of the tracheal wall. The TD-MODE also displays the minimum and maximum tracheal diameter in the frame. The subject device also incorporates Position Tracking and Signal Value Change. Position Tracking is used to track the reference point specified in a frame and display a graph of the position change data. Signal Value Change graphically displays the signal value change within the ROI specified in a frame.

### Indications for Use

KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. R/F Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not in diagnostic review for mammography.

# Comparison Table

|                        | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name            | KONICAMINOLTA DI-X1                                                                                                                                                                                                                                                                                                                                                                                                             | KONICAMINOLTA DI-X1                                                                                                                                                                                                                                                                                                                                                                                                             |
| Version                | 1.30                                                                                                                                                                                                                                                                                                                                                                                                                            | 1.20                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 510(K) Number          | This Submission                                                                                                                                                                                                                                                                                                                                                                                                                 | K212685                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for<br>Use | KONICAMINOLTA DI-X1 is a software device<br>that receives digital x-ray images and data<br>from various sources (i.e. R/F Units, digital<br>radiographic devices or other imaging<br>sources). Images and data can be stored,<br>communicated, processed and displayed<br>within the system and/or across computer<br>networks at distributed locations. It is not<br>intended for use in diagnostic review for<br>mammography. | KONICAMINOLTA DI-X1 is a software device<br>that receives digital x-ray images and data<br>from various sources (i.e. R/F Units, digital<br>radiographic devices or other imaging<br>sources). Images and data can be stored,<br>communicated, processed and displayed<br>within the system and/or across computer<br>networks at distributed locations. It is not<br>intended for use in diagnostic review for<br>mammography. |
| Input image            | ● DICOM 3.0<br>● DICOM Modality (RF, DX, CR)                                                                                                                                                                                                                                                                                                                                                                                    | ● DICOM 3.0<br>● DICOM Modality (RF, DX, CR)                                                                                                                                                                                                                                                                                                                                                                                    |
| I/O Data               | ● DICOM Storage                                                                                                                                                                                                                                                                                                                                                                                                                 | ● DICOM Storage                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Image                  | ● REGIUS                                                                                                                                                                                                                                                                                                                                                                                                                        | ● REGIUS                                                                                                                                                                                                                                                                                                                                                                                                                        |

The comparison to the predicate decides was summarized in the Table below.

Giving Shape to Ideas

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Konica Minolta logo. The logo consists of a blue globe with white lines on the left and the text "KONICA MINOLTA" below it. To the right of the logo is the number "150" above a line and the word "YEARS" below the line.

|                      | Subject Device                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name          | KONICAMINOLTA DI-X1                                                                                                                                                                                                                                                                          | KONICAMINOLTA DI-X1                                                                                                                                                                                              |
| Version              | 1.30                                                                                                                                                                                                                                                                                         | 1.20                                                                                                                                                                                                             |
| 510(K) Number        | This Submission                                                                                                                                                                                                                                                                              | K212685                                                                                                                                                                                                          |
| Processing           | FE-MODE DM-MODE BS-MODE PL-MODE PH-MODE PH2-MODE LM-MODE TD-MODE                                                                                                                                                                                                                             | FE-MODE DM-MODE BS-MODE PL-MODE Pk to Pk PH-MODE PH2-MODE LM-MODE                                                                                                                                                |
| Display<br>Functions | Adjustment of density and gradation,<br>Rotation and reversal, Scaling, Panning Screen display (listing, viewer) Image display (Cine, Comparison,<br>Annotation, Overlay) Graph display (Time-series comparison) Graph display the max/min of the<br>tracheal diameter including past images | Adjustment of density and gradation,<br>Rotation and reversal, Scaling, Panning Screen display (listing, viewer) Image display (Cine, Comparison,<br>Annotation, Overlay) Graph display (Time-series comparison) |
| Measurement          | Distance Angle Area CTR (Only the image of the front of the<br>chest) Position Tracking, Signal Value Change                                                                                                                                                                                 | Distance Angle Area CTR (Only the image of the front of the<br>chest)                                                                                                                                            |
| Client               | DI-X1 client DI-X1 Server client PC client (WEB reference)                                                                                                                                                                                                                                   | DI-X1 client DI-X1 Server client PC client (WEB reference)                                                                                                                                                       |

#### Performance Data

All the verification activities required by the specification and the risk analysis for the KONICAMINOLTA DI-X1 were performed and the results showed that the predetermined acceptance criteria were met. No clinical studies were required to support the substantial equivalence.

### Conclusion

KONICAMINOLTA DI-X1 has the same intended use and indications for use, technological characteristics, and principle operations. The technological differences do not raise new issues of safety or effectiveness as compared to its predicate device (K212685). Performance tests demonstrate that the KONICAMINOLTA DI-X1 performs according to specifications and functions

Giving Shape to Ideas

- Page 3 of 4

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Konica Minolta logo along with the text "150 YEARS". The Konica Minolta logo consists of a blue globe with white horizontal lines and the text "KONICA MINOLTA" in black. The text "150" is in large black font, with a blue line underneath, and the text "YEARS" is in smaller black font below the line. The logo is likely celebrating the 150th anniversary of Konica Minolta.

as intended. All information supporting our assessment is included in the relevant sections of this submission.

Therefore, it is our conclusion that the KONICAMINOLTA DI-X1 is substantially equivalent to the predicate device and does not raise new issues of safety or effectiveness.

Giving Shape to Ideas

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K230906](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K230906)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
