← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K221405

# Neurophet SCALE PET (K221405)

_Neurophet., Inc. · LLZ · Aug 5, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K221405

## Device Facts

- **Applicant:** Neurophet., Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Aug 5, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Neurophet SCALE PET is a software for the registration, fusion, display and analysis of medical images from multiple modalities including MRI and PET. The software aids clinician in the assessment and quantification of pathologies from PET Amyloid/FDG scans of the human brain. It enables automatic analysis and visualization of amyloid protein concentration through the calculation of standard uptake volume ratios (SUVR) within target regions of interest and comparison to those within the reference regions. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with radio-tracer and disease combinations.

## Device Story

Neurophet SCALE PET is standalone medical image processing software; operates on off-the-shelf hardware (Windows 10). Inputs: DICOM/NiFTI format PET and MRI brain images. Operation: software performs automatic registration, fusion, and post-processing of images; calculates Standardized Uptake Value Ratios (SUVR) within target regions of interest compared to reference regions (e.g., pons or cerebellar white matter). Output: quantified SUVR results, visual analysis maps, and reports (.csv, pdf, nii.gz). Used in hospital settings by physicians skilled in brain MR imaging. Clinicians use output to aid in visual interpretation and diagnosis of brain pathologies. Benefits: provides automated, reliable quantification of PET tracer uptake to support clinical decision-making.

## Clinical Evidence

Bench testing only. Segmentation accuracy evaluated via Dice similarity coefficient (DSC) against expert manual segmentation, achieving 86.39±3.12% on major subcortical structures. SUVR calculation reliability assessed via intraclass correlation coefficient (ICC) against conventional PET processing tools, showing good reliability with all ROI ICC scores > 0.6.

## Technological Characteristics

Standalone software package; operates on off-the-shelf hardware (Windows 10, 64-bit). Compatible with DICOM/NiFTI data formats. Connectivity via data network or USB. Complies with ISO 14971, IEC 62304, and IEC 62366 standards. Performs automated image registration, fusion, and quantification.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Veuron-Brain-pAb ([K203142](/device/K203142.md))

## Reference Devices

- Scenium 2.0 ([K121074](/device/K121074.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 5, 2022

NEUROPHET, Inc. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave., STE 160 IRVINE CA 92612

Re: K221405

Trade/Device Name: Neurophet SCALE PET Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: May 4, 2022 Received: May 16, 2022

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Michael D. O'Hara, Ph.D. Deputy Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221405

Device Name Neurophet SCALE PET

#### Indications for Use (Describe)

Neurophet SCALE PET is a software for the registration, display and analysis of medical images from multiple modalities including MRI and PET. The software aids clinician in the assessment and quantification of pathologies from PET Amyloid/FDG scans of the human brain. It enables automatic analysis and visualization of amyloid protein concentration through the calculation of standard uptake volume ratios (SUVR) within target regions of interest and comparison to those within the reference regions.

The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with radio-tracer and disease combinations.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K221405)

This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

#### 1. Date: 7/20/2022

#### 2. Applicant / Submitter

NEUROPHET, Inc. 12F, 124, Teheran-ro, Gangnam-gu Seoul, Republic of Korea Tel : +82-2-6954-7971 Fax : +82-2-6954-7972

#### 3. U.S. Designated Agent

Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com

#### 4. Trade/Proprietary Name:

Neurophet SCALE PET

#### 5. Common Name:

Medical Image Processing Software

#### 6. Classification:

Medical Image Management and Processing System (21CFR 892.2050, Product code LLZ, Class 2, Radiology)

#### 7. Device Description:

Neurophet SCALE PET is a standalone software product that automatically calculates standardized uptake value ratios (SUVR) and provides quantified calculation results for quantitative analysis of FDG and Amyloid PET images. The calculation results are intended to aid clinicians in diagnosing patients' pathologies. Furthermore, the function provided to

{4}------------------------------------------------

visualize the results of the analysis of the image is designed to help clinicians perform accurate visual interpretation.

Functions and workflow supported by the product are as follows: The user may set specified region as reference region, however, in order to achieve reliable constant count in the reference region, FDA recommends the selection of the pons or cerebellar white matter as reference region for assessment of SUVR in FDG PET imaging.

Because this product complies with the standard DICOM medical imaging protocol, it can be used by being linked with picture archive and communications systems (PACS).

### 8. Indication for use:

Neurophet SCALE PET is a software for the registration, fusion, display and analysis of medical images from multiple modalities including MRI and PET. The software aids clinician in the assessment and quantification of pathologies from PET Amyloid/FDG scans of the human brain. It enables automatic analysis and visualization of amyloid protein concentration through the calculation of standard uptake volume ratios (SUVR) within target regions of interest and comparison to those within the reference regions.

The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with radio-tracer and disease combinations.

### 9. Predicate Device:

- Primary Predicate: . Veuron-Brain-pAb (K203142) by HEURON CO., LTD.
- . Reference Device: Scenium 2.0 (K121074) by Siemens Medical Solutions USA, Inc

{5}------------------------------------------------

## 10. Substantial Equivalence:

|                                                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name                                        | Neurophet SCALE PET                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Veuron-Brain-pAb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Scenium 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 510(k)                                             | k221405                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | k203142                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | k121074                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer                                       | NEUROPHET, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | HEURON CO., LTD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Siemens Medical Solutions<br>USA, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code                                       | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for<br>Use                             | Neurophet SCALE PET is<br>a software for the<br>registration, fusion, display<br>and analysis of medical<br>images from multiple<br>modalities including MRI<br>and PET. The software aids<br>clinician in the assessment<br>and quantification of<br>pathologies from PET<br>Amyloid/FDG scans of the<br>human brain. It enables<br>automatic analysis and<br>visualization of amyloid<br>protein concentration<br>through the calculation of<br>standard uptake volume<br>ratios (SUVR) within target<br>regions of interest and<br>comparison to those within<br>the reference regions. The<br>software is deployed via<br>medical imaging<br>workplaces and is<br>organized as a series of<br>workflows which are<br>specific to use with radio-<br>tracer and disease<br>combinations. | The Veuron-Brain-pAb is a<br>software for the registration,<br>fusion, display and analysis of<br>medical images from multiple<br>modalities including MRI and<br>PET.<br>The software aids clinician in<br>the assessment and<br>quantification of pathologies<br>from PET Amyloid scans of<br>the human brain. It enables<br>automatic analysis and<br>visualization of amyloid<br>protein concentration through<br>the calculation of standard<br>uptake volume ratios (SUVR)<br>within target regions of<br>interest and comparison to<br>those within the reference<br>regions. The software is<br>deployed via medical imaging<br>workplaces and is organized as<br>a series of workflows which<br>are specific to use with radio-<br>tracer and disease<br>combinations. | The Scenium display and<br>analysis software has been<br>developed to aid the<br>Clinician in the assessment<br>and quantification of<br>pathologies taken from PET<br>and SPECT scans.<br>The software is deploted via<br>medical imaging workplaces<br>and is organized as a series of<br>workflows which are specific<br>to use with particular drug<br>and disease combinations.<br>The software aids in the<br>assessment of human brain<br>scans enabling automated<br>analysis through<br>quantification of mean pixel<br>values located within<br>standard regions of interest. It<br>facilitates comparison with<br>existing scans derived from<br>FDG-PET, amyloid PET, and<br>SPECT studies and<br>calculation of uptake ratios<br>between regions of interest. |
| Target<br>Anatomical Sites                         | Brain                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Brain                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Brain                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Where Used                                         | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Design Features                                    | • Import DICOM data<br>Perform automatic post-<br>processing<br>• Provide the user<br>confirmation<br>• Export the resulting data<br>through data network or<br>USB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | • Import DICOM data Perform<br>automatic post-processing<br>• Provide the user confirmation<br>• Export the resulting data only<br>through USB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | • Import DICOM data<br>Perform automatic post-<br>processing<br>• Provide the user<br>confirmation<br>• Export the resulting data<br>only through data network                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Physical<br>characteristics                        | • Software package<br>• Operates on off-the-shelf<br>hardware (multiple<br>vendors)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | • Software package<br>• Operates on off-the-shelf<br>hardware (multiple vendors)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No software required<br>• Operates in a serverless<br>cloud environment<br>• User interface through<br>PACS (multiple vendors)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Operating<br>System                                | Windows 10, 64-bit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Windows 10, 64-bit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Windows 7, 64-bit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Standards                                          | • ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | • ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | • ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                    | • IEC 62304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | • IEC 62304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | • IEC 62304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                    | • IEC 62366                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | • IEC 62366                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | • IEC 62366                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Software<br>verification and<br>validation         | Tested in accordance with<br>verification and validation<br>processes and planning.<br>The testing results support<br>that all the system<br>requirements have met their<br>acceptance criteria and are<br>adequate for its intended<br>use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Tested in accordance with<br>verification and validation<br>processes and planning. The<br>testing results support that all<br>the system requirements have<br>met their acceptance criteria<br>and are adequate for its<br>intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Tested in accordance with<br>verification and validation<br>processes and planning. The<br>testing results support that all<br>the system requirements have<br>met their acceptance criteria<br>and are adequate for its<br>intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Compatible<br>Input Data<br>Format and<br>Modality | DICOM & NiFTI<br>PET, MRI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | DICOM & NiFTI<br>PET, MRI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | DICOM<br>PET, SPECT, MRI, CT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Input Patient<br>Data                              | Manual through<br>keyboard/mouse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Manual through<br>keyboard/mouse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Manual through<br>keyboard/mouse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Output Patient<br>Data                             | Pictures: nii.gz<br>Report: .csv, pdf                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Pictures: PNG<br>Report: .csv                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Pictures: JPEG<br>Report: DICOM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Study list<br>functionality                        | Search, Importing,<br>Exporting                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Search, Importing, Exporting                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Search, Importing, Exporting,<br>Deleting                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

{6}------------------------------------------------

The subject device (Neurophet SCALE PET) and the predicate device are the image analysis software for PET images scanned the Human Brain.

- Both devices have the same intended purpose to support clinicians in the assessment and quantification of pathology through PET scan images.
- Both devices automatically calculate the Standardized Uptake Value Ratio (SUVR) value using brain MRI and PET images and provide quantitative analysis results as a report.
- Both devices are DICOM compatible and operate on off-the-shelf hardware. Both devices are used by physicians skilled in brain MR imaging.
- The compatible input data format and modality are also the same.

Therefore, Neurophet SCALE PET is functionally similar and improved from a previous 510(k) market-cleared Veuron-Brain-pAb software device (K203142).

Following are the differences between Neurophet SCALE PET and the predicate device:

| ltem                            | Neurophet SCALE PET    | Veuron-Brain-pAb | Scenium 2.0                     |
|---------------------------------|------------------------|------------------|---------------------------------|
| Supported<br>PET image<br>types | Amyloid PET<br>FDG PET | Amyloid PET      | Amyloid PET<br>FDG PET<br>SPECT |
| Output                          | Pictures: nii.gz       | Pictures: PNG    | Pictures: JPEG                  |

{7}------------------------------------------------

| Patient Data       | Report: .csv, pdf                               | Report: .csv                                  | Report: DICOM                                             |
|--------------------|-------------------------------------------------|-----------------------------------------------|-----------------------------------------------------------|
| Design<br>Features | Export the resulting data<br>network or user PC | Export the resulting data<br>only through USB | Export the resulting<br>data only through<br>data network |

Although both are technically similar, there is a difference in the supported PET image type. The subject device supports Amyloid PET and FDG PET analysis, while the predicate device, Veuron-Brain-pAb(K203142). supports only Amyloid PET analysis. So, we additionally identified a reference device that covered the scope of supported PET image type for the subject device. The reference device, Scenium 2.0(K121074), is an image analysis software that has a similar purpose to the subject device and can support Amyloid PET, FDG PET, SPECT image analysis.

Also, there is a difference in output patient data between the subject device and the predicate device. However, this is only a difference in file format, and the types of output data that users can obtain are the same.

In the data export path, the predicate device can export the data only through a USB. So, we additionally identified a reference device that can export resulting data through data network. Therefore. it is possible to cover this difference by considering both the predicate device and the reference device.

### 11. Performance Data:

SW verification/validation and the measurement accuracy test were conducted to establish the performance, functionality and reliability characteristics of the subject devices. The device passed all of the tests based on pre-determined Pass/Fail criteria.

The result of the performance test is as follows.

### 1) Segmentation Accuracy

The performance of the Segmentation function was evaluated by comparing the Dice similarity coefficient (DSC) between Neurophet SCALE PET-produced segmentation and manual segmentation done by experts. For the DSC evaluation, the subject device performed a DSC of 86.39±3.12% on major subcortical brain structures.

### 2) SUVR Calculation Reliability

The reliability of the SUVR results was evaluated by statistical test using intraclass correlation coefficient between Neurophet SCALE PET-and conventional PET processing tools. For the ICC, the subject device was shown the good reliability (all ROI ICC scores > 0.6).

### 12. Conclusion:

The subject device is substantially equivalent in the areas of technical characteristics, general

{8}------------------------------------------------

function, application, and indications for use. The new device does not introduce a fundamentally new scientific technology, and the device has been validated through system level test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K221405](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K221405)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
