Retrospective study: 67 pairs of a Knee X-Ray and a CT scan of the same patient within a maximum time frame of 6 months.
1 (US Board-Certified radiologist)
Anatomical landmark positioning
—
pre-defined success criteria
Met the performance goal (measured in standard deviation of the distance in mm)
—
—
Retrospective study: 67 pairs of a Knee X-Ray and a CT scan of the same patient within a maximum time frame of 6 months.
1 (US Board-Certified radiologist)
Indications for Use
The Zebra HealthJOINT device is a software tool for 3D reconstruction of bones from a set of 2D radiographs. The device is intended for assisting clinicians in the preoperative planning of knee orthopedic surgical procedures. Zebra's HealthJOINT analyzes cases using an artificial intelligence algorithm for the 3D model reconstruction. In addition to the model, the software provides a list of anatomical landmarks with their position on the 3D model. The result is made available via a 3rd parties' software interface for further display and analysis of the 3D bone model. Clinical judgement and experience are required to properly use the models produced by this software.
Device Story
HealthJOINT is a software tool for 3D bone reconstruction from 2D knee radiographs (AP and lateral views). It uses an AI algorithm to analyze DICOM X-ray images, generating 3D bone models and identifying anatomical landmarks. The device operates as a backend component, integrating with 3rd-party software via a Web API; it is not user-facing. Clinicians use the resulting 3D models and landmark data within their existing planning workflows to assist in preoperative planning for knee orthopedic surgeries. The software does not provide clinical recommendations or treatment decisions; it requires professional clinical judgment for interpretation. Benefits include automated 3D visualization of patient anatomy to support surgical planning.
Clinical Evidence
Bench testing only. Retrospective study evaluated 3D model accuracy (RMSE) and anatomical landmark positioning against ground truth (CT scans and expert radiologist assessment). Dataset included 67 pairs of knee X-rays and CT scans. Results: Femur RMSE 1.14 (95% CI: 1.097-1.187), Tibia RMSE 1.05 (95% CI: 1.005-1.087), Fibula RMSE 0.94 (95% CI: 0.891-0.986). Landmark positioning precision met pre-defined performance goals.
Technological Characteristics
Software-only device; processes DICOM X-ray images. Uses AI-based image segmentation and processing to generate 3D bone models and anatomical landmarks. Connectivity via Web API for 3rd-party integration. Operates on standard radiological image formats. No hardware components.
Indications for Use
Indicated for clinicians assisting in preoperative planning of knee orthopedic surgical procedures. Supports 3D reconstruction of healthy bones and osteoarthritis patients (Kellgren-Lawrence grades 1-4) using 2D radiographs.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/10 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle emblem and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around it. On the right, there is the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zebra Medical Vision Ltd. % Flair Bar VP Operations Shefayim Commercial Center PO Box 25 Shefayim, 6099000 ISRAEL
December 4, 2020
## Re: K202487
Trade/Device Name: HealthJOINT Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: November 3, 2020 Received: November 6, 2020
Dear Flair Bar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202487
Device Name HealthJOINT
#### Indications for Use (Describe)
The Zebra Health.JOINT device is a software tool for 3D reconstruction of bones from a set of 2D radiographs. The device is intended for assisting clinicians in the preoperative planning of knee orthopedic surgical procedures. Zebra's HealthJOINT analyzes cases using an artificial intelligence algorithm for the 3D model reconstruction. In addition to the model, the software provides a list of anatomical landmarks with their position on the 3D model. The result is made available via a 3rd parties' software interface for further display and analysis of the 3D bone model. Clinical judgement and experience are required to properly use the models produced by this software.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Zebra Medical Vision. The logo consists of a yellow horizontal line on the left, followed by a stylized black and white icon resembling a 'Z', and the word 'zebra' in black, with 'MEDICAL.VISION' in smaller black letters underneath. The 'TM' symbol is also present next to the word 'zebra'.
# 5. 510 (k) Summary
## 510(K) Summary - HealthJOINT Zebra Medical Vision Ltd.
## 510(k) Number – K202487
| Applicant's Name: | Zebra Medical Vision Ltd.<br>Shefayim Commercial Center<br>PO Box 25<br>Shefayim, 6099000<br>ISRAEL<br>Telephone: +972-9-8827795<br>Fax: +972-9-8827795 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
- November 03, 2020 Date Prepared:
- Trade Name: HealthJOINT
## Classification Name:
LLZ - Picture Archiving and Communication System
## Classification:
Class II, Radiology
## Predicate Device:
The HealthJOINT device is substantially equivalent to the following device:
| Proprietary Name | Efficient Care 3D Planning |
|------------------------|--------------------------------------------|
| Premarket Notification | K183544 |
| Classification Name | Picture Archiving and Communication System |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | LLZ |
| Regulatory Class | II |
## Performance Standards:
No performance standards have been established for such device under Section 514 of the Federal Food,
Drug, and Cosmetic Act.
## Intended Use/Indication for Use:
The Zebra HealthJOINT device is a software tool for 3D reconstruction of bones from a set of 2D radiographs. The device is intended for assisting clinicians in the preoperative planning of knee orthopedic
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zebra Medical Vision. The logo consists of a stylized letter Z on the left and the word "zebra" on the right. Below the word "zebra" is the text "MEDICAL.VISION".
surgical procedures. Zebra's HealthJOINT analyzes cases using an artificial intelligence algorithm for the 3D model reconstruction. In addition to the model, the software provides a list of anatomical landmarks with their position on the 3D model.
The result is made available via a 3rd parties' software interface for further display and analysis of the 3D bone model.
Clinical judgement and experience are required to properly use the models produced by this software.
## Device Description:
Zebra's HealthJOINT device is a software product that uses an artificial intelligence algorithm to analyze X-ray scans. The HealthJOINT is indicated for the analysis of X-rays scans. The device receives a set of 2D radiographs and automatically provides a 3D model of the bones together with a list of anatomical landmarks with their position on the 3D model may be used by physicians for pre-operative planning of knee orthopedic surgeries. The HealthJoint supports 3D reconstructions of healthy bones, and osteoarthritis patients graded 1 to 4 based on the Kellgren-Lawrence grading system.
The HealthJOINT device functions as a component that can be used by 3rd parties via an API to generate the 3D models and provides a list of anatomical landmarks with their position on the 3D model. The software communicates with the API only, and is not user-facing. The software does not recommend clinical decisions or treatment.
The software is intended for use by clinicians in conjunction with additional patient information and professional judgment.
The following modules compose the HealthJoint software:
Data input and validation: performs validation of the input, X-ray DICOM images, assesses the input data (i.e. age, modality, view) to ensure compatibility for processing by the algorithm.
HealthJoint algorithm: Once the study has been validated the algorithm analyzes the AP (anteriorposterior) along with the LAT (lateral) knee X-ray study in order to provide 3D bone models and locations of anatomic landmarks.
IMA Integration feature: provides the capability to post studies for processing, get the study analysis status and the results of successful study analysis via a Web API.
Error codes feature: In the case of a study failure during data validation or the analysis by the algorithm, an error is provided to the calling 3rd party via the Web API.
## Performance Data:
Safety and performance of HealthJOINT has been evaluated and verified in accordance with software specifications and applicable performance standards through Software Development and Validation & Verification Process to ensure performance according to specifications, User Requirements and Federal Regulations and Guidance documents. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The HealthJOINT device performance was evaluated in a stand-alone retrospective study for accuracy of the 3D model of the Femur, Tibia, and Fibula (RMSE) and the accuracy of the three-dimensional positioning of the anatomic landmarks on the Femur and Tibia. The 3D model accuracy and the anatomical landmark positioning were evaluated comparing to the pre-defined success criteria. The data set for the 3D model accuracy included 67 pairs of a Knee X Ray and a CT scan of the same patient within a maximum
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a yellow horizontal line on the left, followed by a stylized black and white "Z" symbol. To the right of the symbol is the word "zebra" in black, with the words "MEDICAL.VISION" in smaller font below it.
time frame of 6 months. Performance accuracy of the HealthJOINT software was compared to an established Ground Truth by an experienced US Board-Certified radiologist.
The HealthJOINT software performance accuracy was determined by measuring the RMSE (Root Mean Square Error), versus the ground truth. The HealthJOINT device performance for the Femur was an RMSE of 1.14 (95% CI: [1.097,1.187]), for the Tibia an RMSE of 1.05 (95% CI: [1.005,1.087]), and for the Fibula an RMSE of 0.94 (95% CI: [0.891, 0.986]). The device performance met the pre-defined success criteria.
Additionally, HealthJOINT performance accuracy in the identification of the anatomic landmark positioning versus the ground truth was determined by measuring the distance between the software landmarks positioning and the ground truth in mm, in terms of the standard deviation of the distance. The HealthJOINT software device precision met the performance goal. In conclusion, this study demonstrated the HealthJOINT safety and effectiveness and meets its intended use statement.
## Technological Characteristics Compared to Predicate Device:
The technological characteristics, e.g., overall design, mechanism of action, mode of operation, performance characteristics, etc., and the intended use of the HealthJOINT device are substantially equivalent to the predicate device cited above.
A comparison of the technological characteristics with the predicate device is summarized below.
| Technological | Proposed Device: | Predicate Device: | Summary |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | HealthJoint | Efficient Care 3D Planning<br>(K183544) | |
| Intended<br>Use/Indications for<br>Use | The Zebra HealthJOINT device is<br>a software tool for 3D<br>reconstruction of bones from a set<br>of 2D radiographs. The device is<br>intended for assisting clinicians<br>in the preoperative planning of<br>knee orthopedic surgical<br>procedures. Zebra's | Efficient Care 3D Planning is<br>software indicated for assisting<br>orthopedic surgeons in preoperative<br>planning of knee orthopedic<br>surgeries. The software allows for<br>the overlaying of 3D/2D implant<br>models and for the visualization of<br>the radiological images and 3D | Similar. In<br>addition to the 3D<br>reconstruction and<br>3D/2D overlay for<br>the visualization<br>of radiological<br>images, the<br>predicate device |
| | HealthJOINT analyzes cases<br>using an artificial intelligence<br>algorithm for the 3D model<br>reconstruction. In addition to the<br>model, the software provides a list<br>of anatomical landmarks with<br>their position on the 3D model.<br>The result is made available via a<br>3rd parties' software interface for | reconstruction of bones, and<br>includes tools for performing<br>measurements on the images or 3D<br>model of bones, and for selecting<br>and positioning the implant model.<br>Clinical judgments and experience<br>are required to properly use the<br>software. | also provide tools<br>for performing<br>measurements and<br>for selecting the<br>implant model. |
| | further display and analysis of the<br>3D bone model.<br>Clinical judgement and<br>experience are required to<br>properly use the models produced<br>by this software. | Efficient Care 3D Planning is to be<br>used with the following fixed<br>bearing knee replacement systems<br>in accordance with their indications<br>and contraindications: NexGen®<br>CR, NexGen CR-Flex, NexGen<br>CR-Flex Gender, NexGen LPS,<br>NexGen LPS-Flex, NexGen<br>LPSFlex Gender, Persona® CR, | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Zebra Medical Vision. The logo consists of a stylized "Z" symbol on the left, followed by the word "zebra" in a lowercase, sans-serif font. Below the word "zebra" are the words "MEDICAL.VISION" in a smaller font size. The logo is simple and modern, and the use of the "Z" symbol and the word "zebra" suggests a connection to medical imaging and diagnostics.
| | | Persona PS, Vanguard® CR, and<br>Vanguard PS. | |
|-----------------------------|---------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------|
| Classification/Product code | 21 CFR 892.2050/LLZ | 21 CFR 892.2050/LLZ | Same |
| User | Prespecified clinical users<br>(clinicians) | Prespecified clinical users<br>(clinicians) | Same |
| Radiological images format | DICOM | DICOM | Same |
| Modality | X-ray | X-ray | Same |
| Body part | Knee | Knee | Same |
| Software processing | Image segmentation and<br>processing | Image segmentation and processing | Same |
| Output | 3D bone reconstruction | 3D bone reconstruction | Same |
| 2D/3D Overlay | Yes | Yes | Same |
| Anatomic Landmarks | Yes | Yes | Same |
| Measurement Tools | No | Yes | Different, but do<br>not raise<br>questions of<br>safety and<br>effectiveness. |
| Implant Predictability | No | Yes | Different, but do<br>not raise<br>questions of<br>safety and<br>effectiveness. |
## Substantial Equivalence:
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, HealthJoint device raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety and effectiveness. The subject device has the same intended use, similar indications for use, and technological characteristics as the predicate device.
The results of the performance comparison study demonstrated that the HealthJoint device performs as intended, in the specified use conditions, and thus demonstrates the safety and effectiveness of the subject device.
The HealthJoint device is therefore substantially equivalent to the predicate device and share the same intended use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.