← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K193287

# CorInsights MRI (K193287)

_Adm Diagnostics, Inc. · LLZ · Nov 20, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K193287

## Device Facts

- **Applicant:** Adm Diagnostics, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Nov 20, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

The CorInsights MRI Medical Image Processing Software is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from MRI images. Volumetric measurements are compared to reference percentile data. CorInsights MRI is for adults age 45 to 95.

## Device Story

CorInsights MRI is a software-only medical device that processes T1-weighted DICOM MRI images. It performs automated artifact correction, atlas-based segmentation of brain structures, volumetric calculation, and report generation. The software includes automated quality control checks for image headers, morphometry, tissue contrast, and value ranges. It produces a PDF report comparing measured volumes to age/gender-matched reference data and a DICOM image with color-coded anatomical overlays. Used by radiologists, neurologists, and clinical researchers in clinical or research settings to support structural MRI assessment. The output assists clinicians in evaluating brain structure volumes, potentially aiding in the assessment of neurodegenerative conditions by providing quantitative morphometric data.

## Clinical Evidence

No clinical investigation was required. Bench testing used >1,400 scans from >1,100 individuals. Accuracy validated against manual ground truth (HaRP database and neuroanatomy experts) across hippocampal, cortical, and ventricular structures. Reported ICCs ranged 0.89–0.99; DICE coefficients 81–95%. Reproducibility (test-retest) showed average ICC of 0.97 and mean absolute percentage difference of 2.3%. Validation included diverse scanner models (1.5T/3T) and clinical populations (cognitively normal, MCI, AD, FTD).

## Technological Characteristics

Software-only device; processes T1-weighted DICOM MRI. Features automated artifact correction, atlas-based segmentation, and volumetric quantification. Connectivity via DICOM (NEMA PS 3.1-3.20). Quality control includes header, morphometry, and tissue contrast checks. No specific hardware materials or energy sources.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- NeuroQuant ([K170981](/device/K170981.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

November 20, 2020

ADM Diagnostics, Inc. % Robin Martin Co-Founder, Regulatory Strategist Kinetic Compliance Solutions, LLC PO Box 2134 MILWAUKEE WI 53201

Re: K193287

Trade/Device Name: CorInsights MRI Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 15, 2020 Received: October 19, 2020

Dear Robin Martin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193287

Device Name CorInsights MRI

## Indications for Use (Describe)

The CorInsights MRI Medical Image Processing Software is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from MRI images. Volumetric measurements are compared to reference percentile data. CorInsights MRI is for adults age 45 to 95.

| Type of Use (Select one or both, as applicable)                                                                                          |
|------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <b> <span style="font-size: 16px;">☑</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="padding-right: 5px;"> <b> <span style="font-size: 16px;">☐</span> </b> </span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary                                                            | K193287                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Date:                                                          | November 18, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Submitter Information:                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Submitted By:                                                             | Dawn Matthews, CEO<br>ADM Diagnostics, Inc.<br>555 Skokie Blvd; Suite 500<br>Northbrook, IL 60062                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Secondary Contact:                                                        | Robin Martin<br>Kinetic Compliance Solutions, LLC<br>Milwaukee, WI 53201                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Information:                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:                                                               | CorInsights MRI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common Name:                                                              | Medical Image Processing Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification Name:                                                      | System, Image Processing, Radiological, Picture archiving and<br>communication system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Classification:                                                    | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate Device(s):                                                      | K170981 NeuroQuant<br>LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Description:                                                       | CorInsights MRI is a fully automated MR medical image<br>processing software intended for automatic labeling, visualization<br>and volumetric quantification of identifiable brain structures from<br>DICOM formatted magnetic resonance images. The resulting<br>output consists of a pdf report for review, which can be used in<br>research and clinical use, and a DICOM image showing the<br>anatomical structure boundaries identified by the software.<br><br>The proposed device provides morphometric measurements based<br>on T1 MRI series. The output of this software only device includes<br>morphometric reports that provide comparison of measured<br>volumes to age and gender-matched reference data and an image<br>volume that has been annotated with color overlays representing<br>each segmented region.                                                                                                                                                                                                                                                                                                                                    |
|                                                                           | The architecture has a proprietary automated internal process that<br>includes artifact correction, atlas-based segmentation, volume<br>calculation, and report generation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                           | Quality control measures include automated quality control<br>including image header checks to verify that the scan acquisition<br>protocol and provided data adhere to system requirements, an<br>image morphometry check, a tissue contrast check, and value<br>range checks.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indications for Use:                                                      | The CorInsights MRI Medical Image Processing Software is<br>intended for automatic labeling, visualization and volumetric<br>quantification of segmentable brain structures from MR images.<br>Volumetric measurements are compared to reference percentile<br>data. CorInsights MRI is intended for adults age 45 to 95.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Comparison to Predicate:                                                  | Functionally, both devices include the ability to automatically<br>label, visualize and conduct volumetric quantification of<br>segmentable brain structures from MR images. Performance<br>testing outlined below demonstrated equivalent performance in<br>segmentation accuracy and reproducibility compared to published<br>predicate values.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                           | The predicate device indications include lesion analysis. The<br>proposed CorInsights MRI does not. There is no significant impact<br>to safety or efficacy based on this.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                           | Both systems are used by medical professionals, such as<br>radiologists, neurologists and neuroradiologists, as well as by<br>clinical researchers, as a support tool in assessment of structural<br>MRIs.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                           | The proposed device employs the same fundamental scientific<br>technology and the change in indications does not impact the<br>intended use of the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Performance Testing:                                                      | No mandatory performance standards have been established for<br>this device (Section 514 of the Act).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                           | The following testing was conducted / standards complied with to<br>support the substantial equivalence of the proposed device to the<br>predicate:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Category                                                                  | Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Accuracy                                                                  | CorInsights MRI hippocampal volume accuracy was tested using 80 subjects<br>from the HaRP database as ground truth (Boccardi, et al.).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | • CorInsights MRI measured hippocampal volume with an IntraClass<br>Correlation coefficient of 0.95 and a DICE coefficient of 83% left (standard<br>deviation 2.5%) and 83% right (standard deviation 2.7%) and a mean<br>absolute percentage difference of 6.2% left (standard deviation 4.4%) and<br>5.9% right (standard deviation 5.1%) as compared to ground truth volumes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                           | CorInsights MRI cortical segmentation accuracy was tested using 80<br>subjects from a variety of databases with manual ground truth segmentation<br>generated by neuroanatomy experts.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | • CorInsights MRI measured total gray volume with an IntraClass Correlation<br>coefficient of 0.99, a DICE coefficient of 95% (standard deviation 1.6%),<br>and a mean absolute percentage difference of 4.5% (standard deviation,<br>1.8%).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                           | • CorInsights MRI measured cortical subregions with DICE coefficients<br>ranging from 81-93% and a mean absolute percentage difference range of<br>4.6% to 13.8% across all regions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                           | CorInsights MRI Intracranial Volume (ICV) and ventricular accuracy were tested<br>using an additional 70 subjects from a variety of databases with manual ground<br>truth segmentation generated by neuroanatomy experts.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                           | • CorInsights MRI measured ICV with an IntraClass Correlation coefficient<br>of 0.89, a DICE coefficient of 95% (standard deviation 1.1%) and a mean<br>absolute percentage difference of 5.2% (standard deviation of 4.4%).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                           | • CorInsights MRI measured ventricular accuracy with an IntraClass<br>Correlation coefficient of 0.98 (left and right), a DICE coefficient of 88%<br>left (standard deviation 5.2%) and 87% right (standard deviation 5.6%) and<br>a mean absolute percentage difference of 13.9% left (standard deviation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Category                                                                  | Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                           | 9.2%) and 15.2% right (standard deviation 9.9%) compared to ground truth<br>volumes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Reproducibility                                                           | In test-retest processing of two different scans from the same subject on<br>the same scanner on the same day, CorInsights MRI measured volumes<br>with an average IntraClass Correlation coefficient of 0.97, a DICE<br>coefficient of 89% (standard deviation 4.0%) and a mean absolute<br>percentage difference range of 0.7% to 5.8% with an average of 2.3%<br>(standard deviation 2.7%) across all volumes listed in its report.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Normative<br>Reference<br>Database<br>Development                         | The CorInsights MRI reference database was developed using T1 weighted MRI<br>scans from 269 male and 331 female individuals of age 42 to 95 who were<br>clinically diagnosed to be cognitively normal.<br>In addition, these individuals were confirmed to be negative for amyloid pathology,<br>and negative for a variety of other potential confounding abnormalities including<br>overt vascular disease as evidenced by white matter lesions, stroke or tumor,<br>normal pressure hydrocephalus, and with no history of traumatic brain injury or<br>severe neuropsychiatric illness as documented with the data set or provided in the<br>data set's inclusion criteria. This screening was performed to ensure that the<br>CorInsights MRI reference represented normal values without potential influence<br>by disease, even when asymptomatic.                                                                                                                                                                                                                                                                                                            |
| Validation of<br>Volume<br>Measurement<br>in Clinically<br>Relevant Cases | Testing of CorInsights MRI included scans acquired from scanners with 1.5T and<br>3T field strengths, using accelerated and non-accelerated acquisition sequences,<br>and representing a variety of scanner models from Siemens, GE, and Philips.<br>Validation testing was conducted using patient scans from ages 45 to 95, with a<br>broad spectrum of clinical diagnoses including cognitively normal, Mild Cognitive<br>Impairment, typical and atypical Alzheimer's disease, and non-Alzheimer's<br>dementias. Alzheimer's disease (AD) cases (confirmed for amyloid positivity)<br>included late onset AD, Early Onset AD, Posterior Cortical Atrophy (PCA),<br>Logopenic Progressive Aphasia, and Corticobasal Syndrome. Non-Alzheimer's<br>cases included behavioral variant (bv) Frontotemporal Dementia (FTD), semantic<br>variant (sv) FTD, Nonfluent Primary Progressive Aphasia, Primary Progressive<br>Aphasia (other), amyloid negative Corticobasal Syndrome, vascular disease,<br>moderate to severe white matter disease, and ventricular enlargement. Mild<br>Cognitive Impairment (MCI) cases included early MCI, late MCI, and persons who<br> |
| Category                                                                  | Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                           | Volumes measured by CorInsights MRI were tested using normal subject scans, as<br>well as data sets expected to have below normal gray tissue volumes or above<br>normal ventricle volumes based upon well-established literature. These data sets<br>included individuals with diagnoses of MCI, MCI who converted to a clinical<br>diagnosis of AD at 12 months post-scan, late onset AD, Early Onset AD, bvFTD,<br>svFTD, and PCA. CorInsights MRI values were compared to the percentile and z-<br>score ranges expected based upon peer reviewed published literature for these data<br>sets, and were confirmed to be in these ranges. Relationships between regions<br>based on z-score ranking were also compared to published data and were in<br>agreement with the literature. In all cases, the cognitively normal test group was<br>confirmed not to differ from the normative reference group or reference values.<br>In total, more than 1,400 scans from over 1,100 individuals were used in testing of<br>CorInsights MRI.                                                                                                                          |

{4}------------------------------------------------

{5}------------------------------------------------

- Software testing was conducted to verify performance in accordance . with FDA's guidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- In addition, the following standard was followed for DICOM . Conformance: DICOM NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set
- Animal testing was not required to demonstrate substantial . equivalence for the CorInsights MRI device.
- Performance Validation: The following performance data was . analyzed to support substantial equivalence:

{6}------------------------------------------------

{7}------------------------------------------------

Clinical:

A clinical investigation was not required to demonstrate substantial equivalence to the predicate device.

The above testing supports that the proposed device is as safe, Conclusion: effective and performs as well as the legally marketed predicate in its intended use. Therefore, the proposed device is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K193287](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K193287)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
