← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K192692

# Brainomix 360 e-CTA (K192692)

_Brainomix Limited · LLZ · May 28, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K192692

## Device Facts

- **Applicant:** Brainomix Limited
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** May 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Brainomix 360° e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "off-the-shelf" hardware (physical or virtualized) and can be used to perform image viewing, processing and analysis of images are acquired through DICOM compliant imaging devices. Brainomix 360° e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography). Brainomix 360° e-CTA is not intended for mobile diagnostic use.

## Device Story

Brainomix 360° e-CTA is a software-only medical imaging tool for clinical/hospital environments; operated by physicians or medical technicians. Input: DICOM-compliant CTA images. Processing: image registration, visualization of large cerebral vessels, and analysis of hemispheric differences via contralateral comparison (quantifying contrast-agent concentration ratios). Output: processed images and analysis results viewable on PACS workstations or web interfaces. Notification feature: sends preview images via email to clinicians. Clinical utility: assists trained professionals in interpreting CTA scans to inform clinical decision-making. Benefits: provides automated quantification of hemispheric contrast differences to aid in patient evaluation.

## Clinical Evidence

No clinical data. Bench testing performed via software verification and validation, including algorithm validation using phantom data with known properties.

## Technological Characteristics

Software-only device; runs on off-the-shelf physical or virtual servers. DICOM-compliant. No energy delivery to patients. Connectivity via hospital network/LAN. No materials, biocompatibility, or sterilization requirements.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- iSchemaView RAPID ([K172477](/device/K172477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 28, 2020

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Brainomix Limited % Gwilym Owen Head of Quality and Regulatory Suites 11-14 Suffolk House, 263 Banbury Road Oxford. Oxfordshire OX2 7HN UNITED KINGDOM

Re: K192692

Trade/Device Name: Brainomix 360° e-CTA Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: April 30, 2020 Received: May 7, 2020

Dear Gwilym Owen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K192692

Device Name Brainomix 360° e-CTA

#### Indications for Use (Describe)

Brainomix 360° e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "off-the-shel" hardware (physical or virtualized) and can be used to perform image viewing, processing and analysis of images are acquired through DICOM compliant imaging devices.

Brainomix 360° e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography).

Brainomix 360° e-CTA is not intended for mobile diagnostic use.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain graphic above the word "Brainomix". The brain graphic is divided vertically, with the left half colored green and the right half colored red. The word "Brainomix" is written in a simple, sans-serif font and is positioned directly below the brain graphic.

Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

K192692

# 510(K) Summary of Safety and Effectiveness

| Date of summary:         | 30th April 2020                                                                                     |
|--------------------------|-----------------------------------------------------------------------------------------------------|
| Submitter's name:        | Brainomix Limited                                                                                   |
| Submitter's address:     | Brainomix Limited, Suites 11-14 Suffolk House, 263 Banbury road,<br>Oxford. OX2 7HN. United Kingdom |
| Submitter's contact:     | Gwilym Owen                                                                                         |
| Telephone number:        | +44 (0)1865 582730                                                                                  |
| Device Proprietary Name: | Brainomix 360° e-CTA                                                                                |
| Device Common Name(s):   | PACS - Medical Imaging Software                                                                     |
| Classification Name:     | Class II: Picture Archiving and Communications System<br>(892.2050) Product Code: LIZ               |

Brainomix 360° e-CTA is Substantially Equivalent to the following Legally Marketed device:

#### Predicate Devices

| 510(k) Number | Trade Name        | Manufacturer    |
|---------------|-------------------|-----------------|
| K172477       | iSchemaView RAPID | iSchemaView Inc |

## Indications for Use

Brainomix 360° e-CTA is an image processing software package to be used by trained professionals, including, but not limited to physicians and medical technicians. The software runs on standard "offthe-shelf" hardware (physical or virtualized) and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.

Brainomix 360° e-CTA provides viewing and analysis capabilities for imaging datasets acquired with CTA (CT Angiography).

Brainomix 360° e-CTA is not intended for mobile diagnostic use.

## Device Description and Technological Characteristics

Brainomix 360° e-CTA is a medical image visualization and processing software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server.

Brainomix 360° e-CTA allows for the visualization, analysis and post-processing of DICOM compliant CTA images which, when interpreted by a trained physician or medical technician, may yield information useful in clinical decision making.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Brainomix. The logo consists of a stylized brain graphic above the word "Brainomix". The brain graphic is split vertically, with the left half colored green and the right half colored red. The word "Brainomix" is written in a simple, sans-serif font.

Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

Brainomix 360° e-CTA provides a wide range of basic image viewing, processing and manipulation functions, through multiple output formats. Functionality includes image registration and visualization of large cerebral vessels to provide an analysis of hemispheric difference via contralateral comparison (displayed as a relative percentage).

Brainomix 360° e-CTA can connect with other DICOM-compliant devices, for example to transfer CTA scans from a Picture Archiving and Communication System (PACS) to Brainomix 360° e-CTA software for processing.

Results and images can be sent to a PACS via DICOM transfer and can be viewed on a PACS workstation or via a web user interface on any machine contained and accessed within a hospital network and firewall and with a connection to the Brainomix 360° e-CTA software (e.g. a LAN connection)

Brainomix 360° e-CTA notification capabilities enable clinicians to preview images through via e-mail notification with result image attachments.

Images that are previewed via e-mail are compressed, are for informational purposes only, and not intended for diagnostic use beyond notification.

Brainomix 360° e-CTA is not intended for mobile diagnostic use. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.

## Substantial Equivalence Discussion

Brainomix 360° e-CTA and the predicate have substantially similar technological characteristics in that both devices are software packages used for image processing and run on standard physical and/or virtual servers. Both are intended to be used by trained physicians and provide image viewing, processing and analysis of DICOM compliant images from DICOM compliant imaging devices.

Both Brainomix 360° e-CTA and the predicate device have substantially similar intended use as both perform image processing of CTA data to visualize large cerebral vessels and analyze hemispheric difference via contralateral comparison. Both devices can quantify the presence of the contrast agent within the brain and indicate the difference (ratio) in contrast-agent concentration between the hemispheres.

The primary difference between the proposed device and the predicate device is that Brainomix 360° e-CTA offers a sub-set of the indications for use and functionality of the predicate device.

Where the predicate offers visualisation and analysis capabilities for MRI and CTP data, the proposed device does not and therefore the risks associated with this type of analysis capability for MRI and CTP data are not applicable to the proposed device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain graphic above the word "Brainomix". The brain graphic is split vertically, with the left half colored green and the right half colored red. The word "Brainomix" is written in a simple, sans-serif font.

Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

Brainomix 360° e-CTA includes similar CTA processing features and technological characteristics as compared to the predicate device. Where technological differences exist, these are predominantly where the predicate device offers more features, functionality and a broader set of imaging indications as compared to the proposed device. The differences in technological characteristics for the proposed device do not raise any new or unanswered questions of safety or effectiveness.

We therefore conclude that the proposed device is substantially equivalent to the predicate device.

| Characteristic/Parameter                                                  | Brainomix 360° e-CTA –<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | iSchemaView RAPID - Predicate<br>Device (K172477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                                                              | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulation                                                                | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use                                                       | Brainomix 360° e-CTA is an<br>image processing software<br>package to be used by trained<br>professionals, including, but<br>not limited to physicians and<br>medical technicians. The<br>software runs on standard "off-<br>the-shelf" hardware (physical<br>or virtualized) and can be used<br>to perform image viewing,<br>processing and analysis of<br>images. Data and images are<br>acquired through DICOM<br>compliant imaging devices.<br><br>Brainomix 360° e-CTA provides<br>viewing and analysis<br>capabilities for imaging<br>datasets acquired with CTA (CT<br>Angiography).<br><br>Brainomix 360° e-CTA is not<br>intended for mobile diagnostic<br>use. | iSchemaView's RAPID is an<br>image processing software<br>package to be used by trained<br>professionals, including but not<br>limited to physicians and<br>medical technicians.<br><br>The software runs on a standard<br>off-the-shelf computer or a<br>virtual platform, such as<br>VMware, and can be used to<br>perform image viewing,<br>processing and analysis of<br>images. Data and images are<br>acquired through DICOM<br>compliant imaging devices.<br><br>The iSchemaView RAPID<br>provides both viewing and<br>analysis capabilities for<br>functional and dynamic imaging<br>datasets acquired with CT<br>Perfusion (CT-P), CT<br>Angiography (CTA), and MRI<br>including a Diffusion Weighted<br>MRI (DWI) Module and a<br>Dynamic Analysis Module<br>(dynamic contrast-enhanced<br>imaging data for MRI and CT). |
| Characteristic/Parameter                                                  | Brainomix 360° e-CTA –<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | iSchemaView RAPID - Predicate<br>Device (K172477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Environment of use                                                        | Clinical/Hospital environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Clinical/Hospital environment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Energy used and/or<br>delivered                                           | None - software only<br>application. The software<br>application does not deliver or<br>depend on energy delivered to<br>or from patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | None - software only<br>application. The software<br>application does not deliver or<br>depend on energy delivered to<br>or from patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Human Factors                                                             | Designed to be used by trained<br>clinicians                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Designed to be used by trained<br>clinicians                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Design: Supported<br>Modalities for image<br>processing and visualization | CTA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | CT, MRI, CTA, CTP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Design: PACS functionality                                                | View process and analyze<br>medical images. performs<br>standard PACS functions with<br>respect to querying and listing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Design: DICOM compliance                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Design: Computer Platform                                                 | Standard off-the-shelf server or<br>virtual server                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Design: Data acquisition                                                  | Acquires medical image data<br>from DICOM compliant imaging<br>devices and modalities                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Acquisition and modalities<br>features: MRI                               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Diffusion Weighted Image (DWI)<br>Dynamic Analysis tissue flow<br>(perfusion) and tissue blood<br>volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Acquisition and modalities<br>features: CTP                               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CTP cerebral blood flow and<br>blood volume                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Characteristic/Parameter                                                  | Brainomix 360° e-CTA –<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | iSchemaView RAPID - Predicate<br>Device (K172477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Acquisition and modalities features: CTA                                  | CTA large vessel density analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | CTA large vessel density analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Additional tools                                                          | Arterial input function<br>(AIF)Venous output function<br>(VOF) – CTA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Arterial input function<br>(AIF)Venous output function<br>(VOF) - MRI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Functional overview                                                       | Brainomix 360° e-CTA is a<br>software package that provides<br>for the study of changes of<br>tissue in digital images<br>captured by CT. Brainomix 360°<br>e-CTA provides viewing and<br>quantification for CTA images.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | RAPID is a software package<br>that provides for the<br>visualization and study of<br>changes of tissue in digital<br>images captured by CT and MRI.<br>RAPID provides<br>viewing and quantification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Materials                                                                 | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Biocompatibility                                                          | N/A – Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterility                                                                 | N/A – Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Electrical Safety                                                         | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Mechanical Safety                                                         | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Chemical Safety                                                           | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Thermal Safety                                                            | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Radiation Safety                                                          | N/A - Software only device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain shape, with the left hemisphere colored green and the right hemisphere colored red. Below the brain image, the word "Brainomix" is written in a simple, sans-serif font.

Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Brainomix. The logo consists of an image of a brain that is split down the middle, with the left half colored green and the right half colored red. Below the image of the brain, the word "Brainomix" is written in a simple, sans-serif font.

Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

# Testing

Brainomix 360° e-CTA is tested against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission. Algorithm validation testing has additionally been performed using phantom data with known properties.

The results of software verification and validation and algorithmic testing demonstrate that Brainomix 360° e-CTA has met all design requirements and specifications associated with the intended use of the software.

Design, risk management, verification and Validation for Brainomix 360° e-CTA has been carried out in compliance with the requirements of CFR 21 Part 820 and in adherence to the DICOM standard.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain graphic, with the left hemisphere colored green and the right hemisphere colored red. Below the brain graphic, the word "Brainomix" is written in a simple, sans-serif font.

Brainomix Limited Suites 11-14 Suffolk House 263 Banbury Road Oxford OX2 7HN

# Performance Data

No clinical testing or non-clinical bench performance testing has been carried out, submitted or referenced for this submission. As Brainomix 360° e-CTA is a stand-alone software device, testing for Brainomix 360° e-CTA is captured within software and algorithm verification and validation.

# Conclusion

In conclusion, the predicate device has the same technological characteristics and intended use as Brainomix 360° e-CTA. Brainomix 360° e-CTA is therefore substantially equivalent to the selected predicate device and does not raise any questions of safety or effectiveness.

Software verification and validation and algorithmic testing and risk management demonstrates that Brainomix 360° e-CTA is safe and effective for use as intended and described in its indications for use.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K192692](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K192692)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
