icobrain

K180326 · Icometrix · LLZ · Mar 8, 2018 · Radiology

Device Facts

Record IDK180326
Device Nameicobrain
ApplicantIcometrix
Product CodeLLZ · Radiology
Decision DateMar 8, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Volumetric quantification of brain structuresPearson correlation coefficient 0.91; intraclass correlation coefficient 0.89Validation experiments including 463 subject datasets (healthy subjects, Alzheimer's disease patients, traumatic brain injury patients, depression patients).

Indications for Use

icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images. icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long. icobrain cross is intended to provide volumes from images acquired at a single timepoint icobrain long is intended to provide changes in volumes between two images that were acquired on the samer, with the same image acquisition protocol and with same contrast at two different timepoints The results of icobrain cross cannot be compared with the results of icobrain long.

Device Story

Input: T1-weighted and FLAIR DICOM MR images (single or multiple timepoints). Processing: Pre-processing detects scan type, converts DICOM to NIFTI; image processing performs segmentation and volumetric calculation of brain structures; longitudinal pipeline calculates volume changes between scans from same scanner/protocol. Output: Electronic reports (PDF/DICOM) containing volumetric measures and segmentations overlaid on input images. Usage: Clinical setting; used by healthcare providers to automate manual identification/quantification of brain structures. Benefit: Provides objective, automated volumetric data to assist in clinical assessment of brain structure changes.

Clinical Evidence

Bench testing only. Validation included 463 subject datasets (healthy, Alzheimer's, TBI, depression). Accuracy validated against simulated/manual ground truth; reproducibility validated via test-retest datasets. Results: Pearson correlation coefficient 0.91; intraclass correlation coefficient 0.89. Verification tests confirmed system capabilities.

Technological Characteristics

Medical image processing software; DICOM/NIFTI formats. Connectivity: Networked/DICOM. Standards: ISO 14971, IEC 62304, IEC 62366, ISO 12052, ISO 13485. Software-based segmentation and volumetric quantification. Normalization for head size via statistical model comparison to healthy population.

Indications for Use

Indicated for automatic labeling, visualization, and volumetric quantification of brain structures from MR images in patients including healthy subjects, Alzheimer's disease, traumatic brain injury, and depression patients.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 8, 2018 icometrix NV % Dirk Smeets VP Clinical Applications Kolonel Begaultlaan 1b /12 Leuven, 3012 BELGIUM Re: K180326 Trade/Device Name: icobrain Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: February 2, 2018 Received: February 6, 2018 Dear Dirk Smeets: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ ### Page 2 - Dirk Smeets and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Robert Sauer for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) #### K180326 Device Name icobrain Indications for Use (Describe) icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images. icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long. icobrain cross is intended to provide volumes from images acquired at a single timepoint icobrain long is intended to provide changes in volumes between two images that were acquired on the samer, with the same image acquisition protocol and with same contrast at two different timepoints The results of icobrain cross cannot be compared with the results of icobrain long. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # l Submitter | Name: | icometrix NV | |-------------------|-------------------------------------------------------| | Address: | Kolonel Begaultlaan 1b / 12<br>3012 Leuven<br>BELGIUM | | Contact Person: | Dirk Smeets | | Telephone number: | +32 16 369 000 | | Fax Number: | N.A. | | E-mail: | dirk.smeets@icometrix.com | | Date Prepared: | 01 Feb 2018 | # 2 Device | Device Trade Name: | icobrain | |-----------------------|----------------------------------------| | Common Name: | Medical Image Processing Software | | Classification Name: | System, Image processing, Radiological | | Number: | 892.2050 | | Product Code: | LLZ | | Classification Panel: | Radiology | Image /page/3/Figure/5 description: The image shows a fraction, specifically "2 / 6". The number 2 is above the fraction bar, and the number 6 is below the fraction bar. The background is a gradient of white to blue-green. {4}------------------------------------------------ ## 3 Predicate Device | Device | icobrain | |-----------------------|-------------------------------------------| | 510(k) Number | K161148 | | Manufacturer | icometrix NV | | Common Name: | Medical Image Processing<br>Software | | Classification Name: | System, Image processing,<br>Radiological | | Number | 892.2050 | | Product Code | LLZ | | Classification Panel: | Radiology | ## 4 Device Description icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images. icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long. - icobrain cross is intended to provide volumes from images acquired at a single timepoint - · icobrain long is intended to provide changes in volumes between two images that were acquired on the same scanner, with the same image acquisition protocol and with same contrast at two different timepoints The results of icobrain cross cannot be compared with the results of icobrain long. The following flowchart illustrates the overall architecture of icobrain. Image /page/4/Figure/9 description: This image is a flowchart that shows the steps involved in image processing. The first step is to input images in DICOM format. The next steps are preprocessing, image processing, and output generation (report/images). The final step is to generate a final report in PDF or DICOM format, as well as output images in DICOM format. As input, icobrain uses TI-weighted and a fluid-attenuated inversion recovery (FLAIR) DICOM MR images from a single or from multiple time points. In case of multiple time points, i.e. multiple MRI scans from the same subject, for each time point one FLAIR and one TI image are used as input. During the pre-processing, the scan type (TI, FLAIR) is detected for every input image before it is converted from DICOM format to NIFTI format. The image processing then performs the actual segmentation and {5}------------------------------------------------ calculates the volumes of the brain structures. In case MRI scans from the same subject on multiple time points are available, the changes in volume of the brain structures are calculated as well. Finally, the computed volumes and volume changes (in case of multiple time points) are summarized into an electronic report and (some) segmentations are overlaid on the input images. The software displays the following volumetric measures in three reports Report I: - unnormalized volume and volume changes (total, new, enlarging and shrinking) of FLAIR white matter hyperintensities, in different regions (peri-ventricular, juxta-cortical, infra-tentorial, deep white matter). - unnormalized volume and volume changes (total, new, enlarging and shrinking) of TI white matter hyperintensities and hypointensities. - · normalized volume and volume changes of the whole brain (sum of white and grey matter), - · normalized volume and volume changes of grey matter Report 2: - normalized volume and volume changes of the frontal, temporal and parietal lobe - · normalized volume and volume changes of the hippocampi - · unnormalized volume and volume changes of FLAIR white matter hyperintensities Report 3: - unnormalized volume and volume changes (total, new, enlarging and shrinking) of FLAIR white matter hyperintensities, in different regions (sub-cortical, corpus callosum, brainstem) - · normalized volume and volume changes of the cortical gray matter - · normalized volume and volume changes of the hippocampi The whole brain, gray matter, cortical gray matter, frontal lobe, temporal lobe, parietal lobe and hippocampal volumes are normalized for head size by comparing to a healthy population using a statistical model. The reported FLAIR or TI white matter hyperintensities volumes are not normalized since they are not comparable to a reference population. ### 5 Intended use icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. This software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images. icobrain consists of two distinct image processing pipelines: icobrain cross and icobrain long. - · icobrain cross is intended to provide volumes from images acquired at a single timepoint - · icobrain long is intended to provide changes in volumes between two images that were acquired on the same scanner, with the same image acquisition protocol and with same contrast at two different timepoints The results of icobrain cross cannot be compared with the results of icobrain long. This intended use of the modified device is equal to the intended use of the previously cleared predicate device [K161148]. {6}------------------------------------------------ # 6 Device modifications icobrain 3.0 is an update of the icobrain 1.3 (the predicate device). The modifications of the device mostly include new measurements of brain segmentable structures, new output reports, new output images, and new output format (json). New measurements: - unnormalized volume and volume changes (total, new, enlarging and shrinking) of FLAIR white matter hyperintensities, in different regions (peri-ventricular, juxta-cortical, infra-tentorial, deep white matter, sub-cortical, corpus callosum, brainstem), - unnormalized volume and volume changes (total, new, enlarging and shrinking) of T I white matter hyperintensities and hypointensities, - · normalized volume and volume changes of the frontal, temporal and parietal lobe, - · normalized volume and volume changes of the hippocampi, - · normalized volume and volume changes of the cortical gray matter, - · normalized volume and volume changes of the hippocampi. These new measurements, together with the original measurements, are now structured in three different report types. New output images include: - · the segmentation of FLAIR WM hyperintensities, color-coded for the region and overlaid on the input FLAIR images, - the segmentation of new and enlarging FLAIR WM hyperintensities, overlaid on the input FLAIR images, - · the segmentation of frontal, temporal and parietal lobe and the hippocampi, overlaid on the input TI images. ### 7 Performance testing ### Quality and safety icobrain has been designed, developed and tested in accordance with the following product and process standards: - · ISO 14971:2007 Medical devices Application of risk management to medical devices - · IEC 62304:2006 Medical device software Software life-cycle processes - · IEC 62366:2014 Medical devices Application of usability engineering to medical devices - · ISO 12052:2006 Digital imaging and communication in medicine (DICOM) - · CFR 21 part 820 Quality System Regulation for Medical Devices - · ISO 13485:2016 Medical devices Quality management systems The changes made in icobrain compared to the original device do not affect the safety of the device. This conclusion is based on: - · Failure mode and effects analysis on the added functionality. - · Risk category classification of new software components. ### Product performance To demonstrate the performance of icobrain, the measured volumes and volume changes of the segmentable brain structures are validated for accuracy and reproducibility. The subjects upon whom {7}------------------------------------------------ the device was tested include healthy subjects, Alzheimer's disease patients, traumatic brain injury patients, depression patients. In the accuracy experiments, the volumes / volume changes are compared to simulated and/or manually labeled ground truth volumes / volume changes; in the reproducibility experiments, the volumes / volume changes are compared on test-retest image data sets. A literature review has been performed to set relevant acceptance criteria for each type of experiments passed the acceptance criteria. The experiments encompassed 463 subject datasets in total. Averaged over all experiments, the Pearson correlation coefficient between the compared measurements was 0.91 and the intraclass correlation coefficient was 0.89. Besides the validation experiments, verification tests demonstrate the system as a whole provides all the capabilities necessary to operate according to its intended use. ## 8 Conclusions Because of the identical intended use same and technological characteristics, icobrain 3.0 is substantially equivalent to a device (icobrain 1.3) that has been approved in the United States. The performance testing demonstrates that icobrain 3.0 is as safe and effective as the predicate device. | Declarations: | • This summary includes only information that is also covered in the body of the 510(k).<br>• This summary does not contain any puffery or unsubstantiated labeling claims.<br>• This summary does not contain any raw data, i.e., contains only summary data.<br>• This summary does not contain any trade secret or confidential commercial information.<br>• This summary does not contain any patient identification information. | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| This document is reviewed and approved by Dirk Smeets, Vice President Clinical Applications of ico metrix, based on the present data and information. | Signature | | |-----------|---------------| | Date | ■ 08 Mar 2018 | 6 / 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...