SICAT Endo

K180262 · Sicat GmbH & Co. KG · LLZ · Mar 12, 2018 · Radiology

Device Facts

Record IDK180262
Device NameSICAT Endo
ApplicantSicat GmbH & Co. KG
Product CodeLLZ · Radiology
Decision DateMar 12, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Aiding diagnosis in the oral-maxillofacial region Aiding comparisons of different treatment options Aiding endodontic treatment planning Aiding treatment planning for endodontic surgical guides

Device Story

SICAT Endo is a software-only device for viewing/analyzing medical 3D volume data (CBCT/CT), intraoral images, and 3D optical surface data (optical impressions). Used by dental professionals in clinics to visualize oral-maxillofacial anatomy; identify/mark root canals; define drill canals; and plan endodontic surgical guides. The software provides tools for 3D volume rendering, orthogonal slices, panoramic views, and curved planar reformation (CPR). It allows registration of optical surface data to volume data using user-specified reference points. Output consists of visualized anatomical data and exported planning data for use in CAD or rapid prototyping systems for surgical guide fabrication. By enabling precise 3D visualization and drill path planning, the device assists clinicians in endodontic treatment planning and surgical guide design, potentially improving treatment predictability and quality.

Clinical Evidence

Bench testing only. Verification and validation activities included design reviews, unit tests, code reviews, usability tests, integration tests, and system verification. Special bench testing verified endodontic planning visualization quality and quantitative accuracy of root canal treatment planning. The device passed all verification and validation activities.

Technological Characteristics

Pure software device; runs on Windows 7/8/10 (64-bit). Inputs: DICOM (CBCT/CT), STL, proprietary SSI/SIXD containers. Visualization: 3D volume/surface rendering, orthogonal slices, panoramic views, CPR. Connectivity: Standalone or integrated as SIDEXIS XG/4 module. Software lifecycle: AAMI/ANSI/IEC 62304 compliant. Accuracy: 100 μm length, 1-degree angular measurement.

Indications for Use

Indicated for aiding diagnosis, treatment planning, and comparison of treatment options in the oral-maxillofacial region, specifically for endodontic treatment and endodontic surgical guide planning.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 12, 2018 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. SICAT GmbH & Co. KG % Frederik Kunze Quality Management and Regulatory Affairs Brunnenallee 6 Bonn, NRW 53177 GERMANY Re: K180262 Trade/Device Name: SICAT Endo Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 29, 2018 Received: January 30, 2018 Dear Frederik Kunze: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180262 Device Name SICAT Endo Indications for Use (Describe) - Aiding diagnosis in the oral-maxillofacial region - Aiding comparisons of different treatment options - · Aiding endodontic treatment planning - · Aiding treatment planning for endodontic surgical guides Type of Use (Select one or both, as applicable) | | Prescription Use (Part 21 CFR 801 Subpart D) | |--|----------------------------------------------| | | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Content and format as required by section 21 CFR 807.92 ( http://www.fda.gov/ MedicalDevices/ DeviceRequlationandGuidance/ HowtoMarket YourDevice/PremarketSubmissions/PremarketNotification510k/ucm142651.htm) ### 1. SUBMITTER/510(K) HOLDER SICAT GmbH & Co. KG Brunnenallee 6 53177 Bonn Germany | Establishment<br>Registration Number: | 3006098230 | |---------------------------------------|------------------------| | Telephone: | +49 (228) 854697-82 | | Facsimile: | +49 (228) 854697-99 | | Primary Contact: | Mr. Dr. Manfred Breuer | | Secondary Contact: | Mr. Frederik Kunze | | Date Prepared: | January 29th, 2018 | ### 2. DEVICE NAME AND DEVICE CLASSIFICATION | Proprietary Name: | SICAT Endo | |-------------------------|--------------------------------------------------------------| | Common/Usual Name: | Radiological Visualization Software for Diagnosis and Dental | | | Treatment Planning | | Classification Name: | System, Image Processing, Radiological | | Regulation Description: | Picture archiving and communications system | | Product Code: | LLZ | | Regulation Number: | 21 CFR 892.2050 | | Classification Class: | Class II Product | ### 3. PREDICATE DEVICES - SICAT Implant (K103723) (Primary predicate device) . - 3D Endo™ Software (K171115) . ### 4. DEVICE DESCRIPTION SICAT Endo is a pure software device. SICAT Endo is a software tool intended for viewing and analyzing medical information: {4}------------------------------------------------ - 5 510(k) Summary for SICAT Endo - medical 3D volume data such as volumetric X-ray data from Cone Beam CT (CBCT) and CT scanners, and - . intraoral images, and - . 3D optical surface data like optical impression data from optical scanners, and SICAT Endo provides tools for analyzing the root canal and to mark it visually. It allows to define a drill canal and ordering of a corresponding surgical quide with a drill sleeve. ### 5. Indications for Use Indications for use for SICAT Endo are: - Aiding diagnosis in the oral-maxillofacial region - . Aiding comparisons of different treatment options - . Aiding endodontic treatment planning - Aiding treatment planning for endodontic surgical quides . ### 6. Device Comparison Table The following table shows a summary of the intended use, technological characteristics, design and function of SICAT Endo and the predicate devices. The predicate device SICAT Implant is the primary predicate device. | | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|------------------------------------------------------------------------------------------------------|------------| | | Common / Usual Name | | | | | | | | Radiological<br>Visualization<br>Software for<br>Diagnosis and<br>Endodontic<br>Treatment Planning | Radiological<br>Visualization Software<br>for Diagnosis and<br>Endodontic Treatment<br>and Re-treatment<br>Planning | Radiological<br>Visualization<br>Software for<br>Diagnosis and<br>Dental Implant<br>Planning | | | | | Device Description | | | | | | | | General Description | SICAT Endo is a<br>pure software<br>device.<br>SICAT Endo is a<br>software application<br>for visualization of<br>imaging information<br>of the oral-<br>maxillofacial region.<br>The imaging data | 3D Endo™ is a CBCT<br>based software device<br>designed to improve<br>endodontic treatment<br>planning for more<br>predictability.<br>It is designed to<br>improve treatment<br>quality: to isolate the<br>tooth of interest, to | SICAT Implant<br>V1.2 is a pure<br>software device.<br>SICAT Implant<br>V1.2 is a<br>software<br>application for<br>the visualization<br>of imaging<br>information of | | | | | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | | | originates from<br>medical scanners<br>such as CT or CBCT<br>scanners. It is also<br>used as a software<br>system to aid<br>medical<br>professionals with<br>the planning, the<br>evaluation and the<br>comparison of<br>treatment options<br>and the planning of<br>endodontic accesses<br>for an endodontic<br>treatment. The<br>medical<br>professionals'<br>planning data may<br>be exported from<br>SICAT Endo and<br>used for the<br>realization of the<br>planned therapy. | clearly visualize the<br>tooth anatomy in 3D,<br>to identify all canals<br>and to anticipate<br>areas.<br>The use of 3D Endo<br>provides:<br>- Diagnosis and<br>pathology<br>- 3D tooth anatomy<br>- Canal system<br>- 3D canal anatomy<br>- Treatment plan<br>The Wizard allows to:<br>· Diagnose the case;<br>• Isolate the tooth of<br>interest;<br>· Investigate the<br>canal system;<br>• Understand the 3D<br>Canal anatomy;<br>· Create a Treatment<br>plan.<br>The use of 3D Endo<br>allows to:<br>· evaluate 3D working<br>length and cavity<br>access depth<br>• locate canal orifices<br>without opening the<br>tooth<br>• plan for an optimal<br>cavity access and your<br>final instrument<br>· investigate canals'<br>irregularities following<br>their curvatures<br>In addition, the<br>simulated Intra-Root<br>Camera View Visualize | the oral-maxillo-<br>facial region. The<br>imaging data<br>originates from<br>medical scanners<br>such as CT or<br>DVT scanners.<br>SICAT Implant<br>V1.2 is intended<br>for use as<br>planning and<br>simulation<br>software to aid<br>qualified dental<br>professionals in<br>the placement of<br>dental implants<br>and the planning<br>of surgical<br>treatments.<br>SICAT Implant<br>V1.2 allows to<br>name, position,<br>move, rotate,<br>resize and<br>visualize dental<br>implants and<br>other planning<br>objects (i.e.<br>nerve canals)<br>within the<br>visualized 3D<br>volume. Thus,<br>dental<br>professionals like<br>implantologists<br>are enabled to<br>precisely plan<br>the positions,<br>orientations,<br>types and sizes<br>of implants to be<br>placed in the<br>patient's<br>mandible/maxilla | | | | | | | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | | | | throughout the canal<br>the theoretical cutting<br>envelope of the<br>Dentsply Sirona<br>Endodontics<br>instruments. | together with the<br>related surgical<br>procedures.<br>The dental<br>professionals'<br>planning data<br>may be exported<br>from SICAT<br>Implant V1.2 and<br>used as input<br>data for CAD or<br>Rapid<br>Prototyping<br>Systems. | | | | | Indications for Use | | | | | | | | Indications for<br>Use | Aiding diagnosis in<br>the oral-maxillofacial<br>region<br>Aiding comparisons<br>of different<br>treatment options<br>Aiding endodontic<br>treatment planning<br>Aiding treatment<br>planning for<br>endodontic surgical<br>guides | 3D Endo™ Software is<br>intended to aid in the<br>visualization,<br>diagnosis and<br>planning of<br>endodontic treatment<br>and re-treatment<br>cases utilizing DICOM<br>images. | SICAT Implant is<br>a software<br>application for<br>the visualization<br>of imaging<br>information of<br>the oral-<br>maxillofacial<br>region. The<br>imaging data<br>originates from<br>medical scanners<br>such as CT or<br>DVT scanners.<br>SICAT Implant is<br>intended for use<br>as planning and<br>simulation<br>software to aid<br>qualified dental<br>professionals in<br>the placement of<br>dental implants<br>and the planning<br>of surgical<br>treatments. The<br>dental<br>professionals | | | | | | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | | | | | planning data<br>may be exported<br>from SICAT<br>Implant and<br>used as input<br>data for CAD or<br>Rapid<br>Prototyping<br>Systems. | | | | | Contraindications | None | Do not use 3D Endo™<br>Software for implants<br>or any other dental<br>procedure outside<br>endodontics.<br>There are no known<br>serious adverse<br>reactions or potential<br>safety hazards. | None | | | | | | Medical Data Viewing | | | | | | | Data types<br>visualized | 3D volume data, 3D<br>surface data (optical<br>impressions), 2D<br>radiographs | 2D and 3D visualized<br>CBCT data | 3D volume data,<br>3D surface data<br>(optical<br>impressions,<br>implant objects) | | | | | Imaging data<br>visualization<br>region | Oral-maxillofacial<br>region | Oral-maxillofacial<br>region | Oral-maxillofacial<br>region | | | | | | Viewing Modes typically used for dental procedures | | | | | | | *Orthogonal<br>slices | Axial, coronal,<br>sagittal within the<br>Endoline wizard | Axial | Axial, coronal,<br>sagittal | | | | | *Panoramic view | Panoramic view (ray<br>sum) based on a<br>panoramic curve.<br>Panoramic curve can<br>be manually<br>adjusted. | No | Panoramic view<br>(ray sum) based<br>on a panoramic<br>curve.<br>Panoramic curve<br>can be manually<br>adjusted. | | | | | Curved Planar<br>Reformation | Curved planar<br>reformation based | Yes | None | | | | | | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | | (CPR) also called<br>Multi Planar<br>Reformation<br>(MPR) that is<br>less precise | on a spline<br>positioned within the<br>volume. By<br>positioning the<br>spline the view is<br>adjusted.<br>No equidistant scale. | | | | | | | *Panoramic slice | Movable "Panoramic<br>Slicing Window"<br>embedded in<br>panoramic view | No | Movable<br>"Panoramic<br>Slicing Window"<br>embedded in<br>panoramic view | | | | | *TSA and LSA | Transversal slice<br>(TSA) and<br>longitudinal slice<br>(LSA), both with<br>respect to<br>panoramic curve | Transversal slice | Transversal slice<br>(TSA) and<br>longitudinal slice<br>(LSA), both with<br>respect to<br>panoramic curve | | | | | *3D volume<br>rendering | Yes…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...