K173467 · Philips Medical Systems Nederland B.V. · LLZ · Dec 12, 2017 · Radiology
Device Facts
Record ID
K173467
Device Name
Advanced Diffusion Analysis (ADA) application
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LLZ · Radiology
Decision Date
Dec 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The Philips Medical Systems' Advanced Diffusion Analysis (ADA) application is a post processing software application to be used by trained professionals including but not limited to physicians and medical technicians. The Philips Medical Systems' Advanced Diffusion Analysis (ADA) application can be used to perform image viewing, process and analysis of MRI Diffusion Weighted Images (DWI).
Device Story
ADA application is post-processing software for MRI Diffusion Weighted Images (DWI); used by physicians/technicians in clinical settings. Inputs: DICOM-compliant DWI MRI data. Processing: performs image registration, calculates computed DWI (cDWI) at user-selected b-values, and fits diffusion models to generate parametric maps (Perfusion fraction f, Pseudo Diffusion coefficient D*, Diffusion coefficient D, Kurtosis K). Outputs: parametric maps, 'Goodness of fit' maps/values, fitted curves, and ROI-based tabular/graphical data. Clinicians use outputs to visualize local water diffusion properties and tissue characteristics; supports diagnostic decision-making. Benefits: provides advanced visualization and quantitative analysis of diffusion MRI data.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation activities confirmed device functionality, performance claims, and compliance with ISO 14971, IEC 62304, and NEMA PS 3.1-3.20 (DICOM).
Technological Characteristics
Post-processing software application; DICOM-compliant; runs on standard workstations. Features: image registration, diffusion model fitting (D, D*, f, K), and 'Goodness of fit' analysis. Software life cycle per IEC 62304. Connectivity: DICOM standard. No hardware components.
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform image viewing, processing, and analysis of MRI Diffusion Weighted Images (DWI).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K163294 — syngo.MR General, syngo.MR Cardiology, syngo.MR Neurology · Siemens Medical Solutions USA, Inc. · Feb 6, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
December 12, 2017
Philips Medical Systems Nederland B.V. % Yoram Levy Qsite General Manager Osite 31 Haavoda Street Binyamina 30500 ISRAEL
Re: K173467
Trade/Device Name: Advanced Diffusion Analysis (ADA) application Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 2, 2017 Received: November 8, 2017
Dear Yoram Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173467
Device Name
Advanced Diffusion Analysis (ADA) application
#### Indications for Use (Describe)
The Philips Medical Systems' Advanced Diffusion Analysis (ADA) application is a post processing software application to be used by trained professionals including but not limited to physicians and medical technicians. The Philips Medical Systems' Advanced Diffusion Analysis (ADA) application can be used to perform image viewing, process and analysis of MRI Diffusion Weighted Images (DWI).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is plain white, which makes the blue letters stand out. The font is sans-serif and appears to be a standard typeface.
# 510(K) SUMMARY
# Advanced Diffusion Analysis (ADA) application
Date prepared:
November 2, 2017
## I. Submitter's name and address
| Establishment name: | Philips Medical Systems Nederland B.V. |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands |
| Establishment registration: | 3003768277 |
| Primary Contact person: | Yoram Levy, Qsite<br>QA/RA Consultant<br>31 Haavoda Street<br>Binyamina, Israel 30500<br>Tel (972)4-638-8837;<br>Fax (972)4-638-0510<br>Yoram@qsitemed.com |
| Alternative contact person | Anat Hersch<br>Regulatory Affairs Lead, ICAP<br>Philips Medical Systems Nederland B.V<br>E-mail: anat.hersch@philips.com |
| II. Device information | |
| Trade name: | Advanced Diffusion Analysis (ADA) |
| Device Classification Name | System, Image processing, Radiological |
| Device Class | Class II |
| Classification Panel | LLZ |
| Product Code | Radiological Image Processing Software |
| Regulation Description | 21 CFR 892.2050 |
{4}------------------------------------------------
## III. Device Description:
Philips Medical Systems' Advanced Diffusion Analysis (ADA) application is a post-processing software to be used as an advanced visualization application of diffusion MRI medical images. The ADA application is intended to perform image viewing, process and analysis of MRI Diffusion Weighted Images (DWI).
The ADA application can display images acquired at different b-values, where the b-value is a factor that reflects the strength and timing of the gradients used to generate Diffusion-Weighted Images. The ADA application provides advanced supportive analysis and visualization tools of diffusion MRI images and parametric maps, which can be used by the physician for further analysis.
The physician retains the ultimate responsibility for making the final diagnosis.
## Key Features
ADA application has the following key features:
- 1. Support visualization and processing of isotropic diffusion-weighted MRI data.
- 2. Calculate and display a computed Diffusion Weighted Image (cDWI) at a b-value of choice.
- 3. Support input image registration in a pre-processing step.
- 4. Present a default analysis model based on the available original DWI images and provide a selection of alternative available models.
- 5. Provide diffusion analysis models, as well as parametric maps of Perfusion fraction (f), Pseudo Diffusion coefficient (D*), Diffusion coefficient (D) and Kurtosis (K).
- 6. Provide a 'Goodness of fit' map, 'Goodness of fit' value and fitted curve showing the fitting quality of the selected model.
- 7. Display parameter values from user defined ROI's (Regions of Interest).
- 8. Display the ROI results in tabular and graphical formats.
- 9. Support export of the parametric maps as grayscale or RGB images for visualization in other viewers or PACS systems.
## [V. Intended use:
The Philips Medical Systems' Advanced Diffusion Analysis
### 2- 3 -
## Advanced Diffusion Analysis (ADA) application– Traditional 510k Submission
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is plain white, which makes the blue letters stand out.
(ADA) application is a post processing software application to be used by trained professionals including but not limited to physicians and medical technicians. The Philips Medical Systems' Advanced Diffusion Analysis (ADA) application can be used to perform image viewing, process and analysis of MRI Diffusion Weighted Images (DWI).
## V. Predicate Devices:
The Advanced Diffusion Analysis (ADA) application is substantially equivalent to the following market-cleared devices:
## Table 2-1 Predicates table
| | Device Name | Manufacturer | 510k No | Date of<br>Clearance |
|----------------------|------------------|--------------|---------|----------------------|
| Primary<br>predicate | Olea Sphere V3.0 | Olea Medical | K152602 | March 3, 2016 |
The proposed Philips Medical Systems Advanced Diffusion Analysis (ADA) application and its predicate device, Olea Sphere V3.0 (K152602) are substantially equivalent in regards to their intended uses, clinical indications, principle of operation and fundamental technology principles.
## VI. Substantial Equivalence to Predicate Devices
{6}------------------------------------------------
| Feature | The proposed device:<br>Advanced Diffusion | Primary Predicate:<br>Olea Sphere V3.0 |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Analysis (ADA) application | (K152602) |
| Device | System, Image processing, | System, Image processing, |
| Classification<br>Name | Radiological | Radiological |
| Device Class | Class II | Class II |
| Classification<br>Panel | Radiology | Radiology |
| Product Code | LLZ | LLZ |
| Regulation | Picture Archiving and | Picture Archiving and |
| Description | communication system | communication system |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Indication for<br>Use | The Philips Medical Systems'<br>Advanced Diffusion Analysis<br>(ADA) application is a post<br>processing software<br>application to be used by<br>trained professionals<br>including but not limited to<br>physicians and medical<br>technicians. The Philips<br>Medical Systems' Advanced<br>Diffusion Analysis (ADA)<br>application can be used to<br>perform image viewing,<br>processing and analysis of<br>MRI Diffusion Weighted<br>Images (DWI). | Olea Sphere V3.0 is an image<br>processing software package to be<br>used by trained professionals<br>including, but not limited to,<br>physicians and medical<br>technicians. The software runs on a<br>standard "off-the-shelf" workstation<br>and<br>can be used to perform image<br>viewing, processing, image collage<br>and analysis of medical images.<br>Data and images are acquired<br>through DICOM compliant imaging<br>devices and modalities. Olea<br>SphereV3.0 provides both viewing<br>and analysis capabilities of<br>functional and dynamic imaging<br>datasets acquired with MRI or<br>other relevant modalities,<br>including a Diffusion Weighted<br>MRI (DWI) / Fiber Tracking<br>Module and a Dynamic Analysis<br>Module (e.g., dynamic exogenous<br>or<br>endogenous contrast enhanced<br>imaging data for MRI and CT).<br>The DWI Module is used to<br>visualize local water diffusion |
| | | properties from the analysis of diffusion-weighted MRI data. The Fiber Tracking feature utilizes the directional dependency of the diffusion to display the white matter structure in the brain or more generally the central nervous system. The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast while repeating acquisitions (e.g., over time with or without variable acquisition parameters) where such techniques are useful or necessary. This functionality is referred to as: Perfusion Module – the calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Permeability Module – the calculation of parameters related to leakage of injected contrast material from intravascular to extracellular space. Arterial Spin Labeling (ASL) Module – the calculation of parameters related to tissue flow based on a MR technique using the water in arterial blood as endogenous tracer to evaluate the perfusion. Relaxometry Module – the calculation of parameters related to the MR longitudinal and transversal relaxation time and rate. Metabolic Module – the calculation of parameters related to the fat signal fraction based on a MR technique using opposed-phase imaging |
| Intended users | Trained professionals including but not limited to physicians and medical technicians. | Trained professionals including but not limited to physicians and medical technicians. |
| Intended Body part | All body | All body |
| Type of scans | Diffusion MRI scans | Multi-Modality support: CT ,MRI PET/CT and SPECT/CT.<br>The DWI Module is for Diffusion MRI scans |
| Visualization<br>MRI data<br>capabilities | Yes<br>Supports visualization and<br>processing of isotropic<br>diffusion-weighted MRI data | Yes<br>Supports visualization and process<br>isotropic diffusion-weighted MRI<br>data |
| Display images<br>capabilities | Yes<br>Display images acquired at<br>different diffusion gradient<br>factors (b-values) | Yes<br>Display images acquired at different<br>diffusion gradient factors (b-values) |
| Image<br>Registration | Yes | Yes |
| Computed<br>Diffusion<br>Weighted Image<br>(cDWI) | Yes<br>Calculate and display a<br>computed Diffusion Weighted<br>Image (cDWI) at a b-value of<br>choice. | Yes<br>Calculate and display a computed<br>Diffusion Weighted Image (cDWI)<br>at a b-value of choice. |
| Computed<br>simple ADC<br>Diffusion maps | Yes | Yes |
| Different<br>diffusion models | Yes<br>The application suggests a<br>selection of diffusion models. | Yes |
| Region of<br>Interest | Yes<br>The user can select and draw<br>the Region of Interest | Yes<br>The user can select and draw the<br>Region of Interest |
| Computation of<br>Pseudo-diffusion<br>coefficient (D*)<br>parametric map | Yes | Yes |
| Computation of<br>diffusion<br>coefficient (D)<br>parametric map | Yes | Yes |
| Computation of<br>perfusion<br>fraction (f)<br>parametric map | Yes | Yes |
| Computation of<br>Kurtosis (K)<br>parametric map | Yes | Yes |
| Computation of<br>goodness of fit<br>parametric map | Yes<br>Provide a ‘Goodness of fit’<br>map, ‘Goodness of fit’ value<br>and fitted curve showing the<br>fitting quality of the selected<br>model. | No |
| Result | Display results in tabular and<br>graphical format | Display results in tabular and<br>graphical format |
| Export image<br>Option | Yes | Yes |
| DICOM format<br>communication | Yes | Yes |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The image is a close-up of the word, and the letters are the main focus.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue letters.
The Philips Medical Systems Advanced Diffusion Analysis (ADA) and the identified predicate Olea Sphere V3.0 (K152602) are substantially equivalent in terms of indications for use and intended users, design features, principle of operation and fundamental scientific technology, and safety and/or effectiveness.
In conclusion, Philips believes that the Advanced Diffusion Analysis (ADA) does not introduce any new potential safety and/or effectiveness issues and is substantially equivalent to the identified predicate devices, Olea Sphere V3.0 (K152602).
## VII. Brief discussion of the nonclinical tests submitted, referenced or relied on
Non-clinical performance testing has been performed on ADA and demonstrates compliance with the following International and FDA-recognized consensus standards and FDA guidance document:
- ISO 14971 Medical devices Application of risk management to medical devices
- IEC 62304 Medical device software Software life cycle processes
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
- NEMA PS 3.1-3.20 standard Digital Imaging and Communications in Medicine (DICOM) Standard
Philips Medical Systems Advanced Diffusion Analysis (ADA) application was tested in accordance with Philips verification and validation processes. Verification and Validation tests have been performed to address intended use, the technological characteristics claims, requirement specifications and the risk management results.
The test results in this 510(k) premarket notification demonstrate that Advanced Diffusion
### Analysis (ADA)
- Complies with the aforementioned international and FDA-recognized consensus . standards and FDA guidance document, and
- Meets the acceptance criteria and is adequate for its intended use and specifications. .
## VIII. Brief discussion of clinical tests submitted, referenced or relied on
The subject of this premarket submission, Advanced Diffusion Analysis (ADA) application did not require clinical studies to support equivalence.
## IX. The conclusions drawn from the nonclinical and clinical tests
Verification and Validation (V&V) activities required to establish performance and functionality of Advanced Diffusion Analysis (ADA) were performed. Testing performed demonstrated the Advanced Diffusion Analysis (ADA) meets all defined functionality requirements and performance claims.
## X. Overall conclusion:
The Advanced Diffusion Analysis (ADA) is substantially equivalent to the identified predicate device, Olea Sphere V3.0 (K152602) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, verification and validation testing demonstrate the safety and efficacy of the device to meet its intended use and specifications.
Philips Medical believes that the proposed device, Advanced Diffusion Analysis (ADA)
## 2- 9 -Advanced Diffusion Analysis (ADA) application– Traditional 510k Submission
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
application, is substantially equivalent to its identified predicate device and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.