← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K173274

# Ceevra Reveal 2.0 (K173274)

_Ceevra, Inc. · LLZ · Jul 10, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K173274

## Device Facts

- **Applicant:** Ceevra, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Jul 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Ceevra Reveal 2.0 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. Ceevra Reveal 2.0 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

## Device Story

Software-only medical imaging system; processes DICOM CT/MR images into 3D/VR formats. System comprises two components: cloud-based Processing Interface (used by imaging technicians for image selection and segmentation) and mobile Viewer Interface (used by clinicians/surgeons on off-the-shelf mobile devices). VR viewing requires compatible off-the-shelf VR headset. Used in hospitals/clinics for preoperative planning and intraoperative display. Clinicians review 3D/VR models to assist in surgical decision-making; device does not integrate with surgical navigation systems. Benefits include enhanced visualization of patient anatomy to support surgical planning and intraoperative reference.

## Clinical Evidence

Bench testing only. Software verification and validation performed per IEC 62304:2006/AC:2008 and FDA guidance for software and cybersecurity. No clinical data provided.

## Technological Characteristics

Software-only device; cloud-based processing interface; mobile application viewer interface. Compatible with off-the-shelf mobile devices and VR headsets. DICOM-compliant. Software life cycle processes per IEC 62304.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Clarity Reveal 1.0 ([K171356](/device/K171356.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 10th, 2018

Ceevra, Inc. % Rory Carrillo Quality/Regulatory RAC Medical LLC 368 San Carlos St. SAN FRANCISCO, CA 94104

Re: K173274

Trade/Device Name: Ceevra Reveal 2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 9, 2017 Received: October 12, 2017

Dear Rory Carrillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jeff Ryges

for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173274

Device Name Ceevra Reveal 2.0

### Indications for Use (Describe)

Ceevra Reveal 2.0 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multidimensional digital images. Ceevra Reveal 2.0 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </div> |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CEEVRA. The logo consists of a stylized letter 'C' formed by a series of blue curved lines that resemble waves or a digital signal. Below the symbol, the word "CEEVRA" is written in a sans-serif font, also in blue, matching the color of the symbol above.

# Section 5. 510(k) Summary

#### 1. General Information

| 510(k) Sponsor        | Ceevra, Inc.               |
|-----------------------|----------------------------|
| Address               | 180 Sansome St., 2nd Floor |
|                       | San Francisco CA 94104     |
| Correspondence Person | Rory A. Carrillo           |
|                       | Quality and Regulatory     |
|                       | Ceevra, Inc.               |
| Contact Information   | Email: rcarrillo@gmail.com |
|                       | Phone: 562-533-7010        |
| Date Prepared         | June 8, 2018               |

#### 2. Proposed Device

| Proprietary Name    | Ceevra Reveal 2.0                          |
|---------------------|--------------------------------------------|
| Common Name         | Reveal 2.0                                 |
| Classification Name | System, Image Processing, Radiological     |
| Regulation Number   | 21 CFR 892.2050                            |
| Regulation Name     | Picture archiving and communication system |
| Product Code        | LLZ                                        |
| Regulatory Class    | II                                         |

#### 3. Predicate Device

| Proprietary Name       | Clarity Reveal 1.0                          |
|------------------------|---------------------------------------------|
| Premarket Notification | K171356                                     |
| Classification Name    | System, Image Processing, Radiological      |
| Regulation Number      | 21 CFR 892.2050                             |
| Regulation Name        | Picture archiving and communications system |
| Product Code           | LLZ                                         |
| Regulatory Class       | II                                          |

#### 4. Device Description

Ceevra Reveal 2.0 is a software-only device that allows clinicians to review CT and MR image data in three-dimensional (3D) format and/or stereoscopic 3D format (commonly known as virtual reality, or VR). The 3D and VR images are accessible through the Ceevra Reveal 2.0 mobile application which is used by clinicians for preoperative surgical planning and for the intraoperative display of the aforementioned 3D and VR images.

Ceevra, Inc., Traditional 510(k)

10 of 66

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CEEVRA. The logo consists of a stylized crescent shape made up of blue lines that get progressively lighter in color. Below the crescent shape, the word "CEEVRA" is written in a sans-serif font, with the letters also transitioning from dark blue to light blue.

Ceevra Reveal 2.0 includes two main software-based user interface components, the Processing Interface and Viewer Interface. The Processing Interface is hosted on a cloud-based, virtual workstation and only accessed by authorized personnel, such as an imaging technician. The Processing Interface contains a graphical user interface where an imaging technician can select DICOM-compatible medical images, segment such imitiate processing into a 3D format. The Viewer Interface is a mobile application that is accessible via a compatible, touchscreen enabled, off-the-shelf mobile device to allow for clinicians to review the medical images in 3D and/or VR formats. Only when the compatible mobile device is used in conjunction with a compatible off-the-shelf VR headset can the surgeon view medical images in the VR format.

The product is intended to be used by trained medical professionals, including imaging technicians and clinicians/surgeons, and is used to assist in clinical decision making.

The 3D images generated using Ceevra Reveal 2.0 are intended to be used in connection with surgical operations in which CT or MR images are used for preoperative planning and/or reviewed intraoperatively.

The manner in which the 3D images are viewed and used does not vary between surgery types. The 3D images are viewed solely from the clinicians' compatible mobile devices, and are not viewed through or otherwise integrated with surgical navigation systems.

#### 5. Intended Use

Ceevra Reveal 2.0 is intended as a medical imaging system that allows the processing, review, analysis. communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. Ceevra Reveal 2.0 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

#### 6. Substantial Equivalence

| Feature/<br>Function | Proposed Device:<br>Ceevra Reveal 2.0                                                                  | Predicate Device:<br>Clarity Reveal 1.0 (K171356)                                                          |
|----------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Intended Use         | Intended as a medical imaging<br>system that allows the processing,<br>review, analysis, communication | Intended as a medical imaging<br>system that allows the processing,<br>review, analysis, communication and |

Ceevra, Inc., Traditional 510(k)

11 OF 66

{5}------------------------------------------------

|                               | and media interchange of multi-<br>dimensional digital images acquired<br>from CT or MR imaging devices. It<br>is also intended as software for<br>preoperative surgical planning, and<br>as software for the intraoperative<br>display of the aforementioned multi-<br>dimensional digital images. Ceevra<br>Reveal 2.0 is designed for use by<br>health care professionals and is<br>intended to assist the clinician who<br>is responsible for making all final<br>patient management decisions. | media interchange of multi-<br>dimensional digital images acquired<br>from CT or MR imaging devices. It<br>is also intended as software for<br>preoperative surgical planning.<br>Clarity Reveal is designed for use by<br>health care professionals and is<br>intended to assist the clinician who is<br>responsible for making all final<br>patient management decisions. |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Users                | Health care Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Health care Professionals                                                                                                                                                                                                                                                                                                                                                   |
| Intended<br>Environment       | Healthcare facilities such as<br>hospitals and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                              | Healthcare facilities such as hospitals<br>and clinics                                                                                                                                                                                                                                                                                                                      |
| Device Class                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class II                                                                                                                                                                                                                                                                                                                                                                    |
| Image<br>Analysis<br>Features | Interactive<br>manipulation and<br>segment                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Interactive<br>manipulation and segment                                                                                                                                                                                                                                                                                                                                     |
| Preoperative<br>Use           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                         |
| Intraoperative<br>Use         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                                                                                                                                                                                                                                          |
| Pan image                     | Pan image in any direction                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | No                                                                                                                                                                                                                                                                                                                                                                          |

#### 7. Performance Data

Safety and performance of Ceevra Reveal 2.0 has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/AC: 2008- Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices. "

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for CEEVRA. The logo consists of a stylized letter C formed by a series of blue arcs that gradually decrease in size, creating a sense of motion or flow. Below the graphic is the word "CEEVRA" in a sans-serif font, with the letters in a teal or blue color.

#### 8. Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, Ceevra Reveal 2.0 raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.

13 of 66

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K173274](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K173274)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
