QuantX

K170195 · Quantitative Insights, Inc. · LLZ · May 17, 2017 · Radiology

Device Facts

Record IDK170195
Device NameQuantX
ApplicantQuantitative Insights, Inc.
Product CodeLLZ · Radiology
Decision DateMay 17, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor. When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis. QuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report. Limitations: QuantX is not intended for primary interpretation of digital mammography images.

Device Story

QuantX is a software-based quantitative image analysis tool for breast MR imaging; assists radiologists in lesion characterization. Inputs: MR image data (T1, DCE, T2, DWI) loaded manually or via DICOM. Operation: User selects seed point or ROI; software performs automatic image registration, segmentation, and analysis. Outputs: Volumetric/surface area analysis, QI Most Enhancing Curve, and time-intensity curves displayed in a dedicated analysis panel. Also functions as a multi-modality viewer (ultrasound, mammography) with measurement and reporting tools. Used in clinical settings by radiologists to supplement diagnostic assessment; does not replace clinical judgment. Benefits: Provides quantitative metrics to aid in lesion assessment and characterization.

Clinical Evidence

No clinical data. Evidence consists of non-clinical performance testing, including verification of MR/mammography/ultrasound image loading and manipulation, accuracy of kinetic colormaps, lesion segmentation, and BIRADS reporting. Validation testing confirmed conformance to user needs and intended use.

Technological Characteristics

MRI data post-processing software; DICOM compatible. Features: 2D/3D visualization, MIPs, reformats, image registration, subtraction, kinetic analysis, parametric image maps, ADC maps, and automated/manual segmentation. Connectivity: DICOM standard interfaces. Software-based analysis.

Indications for Use

Indicated for radiologists to assist in the assessment and characterization of breast abnormalities using MR image data. May also be used as a multi-modality image viewer for ultrasound and mammography. Not intended for primary interpretation of digital mammography images.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 17, 2017 Quantitative Insights, Inc. % Mr. Robert Tomek Chief Technology Officer 1452 E 530 Street CHICAGO IL 60615 Re: K170195 Trade/Device Name: QuantX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 21, 2017 Received: February 21, 2017 Dear Mr. Tomek: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set Image /page/0/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. {1}------------------------------------------------ forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170195 Device Name QuantX ### Indications for Use (Describe) QuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor. When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis. QuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report. Limitations: QuantX is not intended for primary interpretation of digital mammography images. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # PREMARKET NOTIFICATION 510(k) Summary As required by 21 CFR §807.92(c) # Submitter | Manufacturer: | Quantitative Insights, Inc. | |---------------|------------------------------------------| | Address: | 1452 E. 53rd Street<br>Chicago, IL 60615 | | Contact Person: | |-----------------------| | Contact Person Title: | | Telephone: | | Email: | Robert Tomek Chief Technology Officer 872.529.2239 rtomek@guantinsights.com Date Prepared: May 10, 2017 ## Device Information Trade Name: QuantX Common Name: Image Processing Software Classfication Name: Picture archiving and communications system Product Code: LLZ #### LEGALLY MARKETED PREDICATE DEVICE A. Legally marketed predicate device is: CADStream Version 5 - K081556, K092954 #### B. DEVICE DESCRIPTION QuantX is a software program that analyzes patient breast images, and is designed to aid radiologists in the characterization of lesions. After MR images are acquired from a third-party acquisition device, they can be loaded into the QuantX database manually, or automatically if connected to a DICOM-compatible device. Users then select and load the patient case to use the QuantX software tools in the examination of the images. Different types of MR sequences (T1, DCE, T2, DWI, etc.) can be viewed at the same time as mammography or ultrasound images from the same patient. A variety of viewing tools are available to users. The MR images can be examined under different image planes (axial, sagittal, and coronal) as well as different image time points and slices. Users can use keyboard shortcuts or a scrolling mechanism to navigate through MR image slices. Colored axes serve as slice location markers for ease of pinpointing regions of interest (ROI). Images can be panned, changed in contrast, zoomed in or out. and measured. The Colormap feature visualizes contrast uptake (enhancement) studies, and a time intensity curve can be viewed for any location on the MR image. {4}------------------------------------------------ QuantX includes image registration, and automated segmentation and analysis functions, based on a seed point indicated by the user. Users can select a ROI manually from the MR image, or use the automatic segmentation tool to obtain and accept a ROI, for input to the QuantX analytics. The QuantX analytics display the QI Most Enhancing Curve, volume and surface area of the specified region. A user experienced with the significance of such data will be able to view and interpret this additional information during the diagnosis of breast lesions. #### C. INDICATIONS FOR USE OuantX is a quantitative image analysis software device used to assist radiologists in the assessment and characterization of breast abnormalities using MR image data. The software automatically registers images, and segments and analyzes user-selected regions of interest (ROI). QuantX extracts image data from the ROI to provide volumetric and surface area analysis. These imaging features are then displayed to the user in a dedicated analysis panel on the display monitor. When interpreted by a skilled physician, this device provides information that may be useful for screening and diagnosis. Patient management decisions should not be made based solely on the results of QuantX analysis. OuantX may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software also includes tools that allow users to measure and document images, and output in a structured report. Limitations: QuantX is not intended for primary interpretation of digital mammography images. An assessment has been made on the differences in indications between QuantX and the predicate device, and it has been determined that these changes do not raise any new questions regarding the safety and effectiveness of the device. ### TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required by D. 807.92(a)(6) QuantX has the following same technological characteristics as its predicate device. It is an MRI data post-processing software device, and is DICOM compatible. These similar features include: - Visualization tools - standard 2D and 3D image viewing tools. MIPs. reformats - Analysis registration, subtractions, kinetic analysis, parametric image maps, ● apparent diffusion coefficient (ADC) maps, automatic and manual segmentation tools - Standardized reporting tools incorporating user selected finding and assessment ● - User interface and interaction tools that enhance workflow efficiency . {5}------------------------------------------------ - . Communication and storage functions including DICOM standard interfaces The following technological differences exist between the subject and predicate device; these differences do not affect the safety and effectiveness of the device when used as labeled: - . Subject does not support coil inhomogeneity correction - . Subject does not support interventional planning - Workflow efficiencies are implemented differently in the subject and predicate . devices - . In addition to volume computation, the subject device includes a surface area computation which provides an additional indicator of lesion size ### E. NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEWS - as required by 807.92(b)(1) QuantX has been successfully tested for its software requirements and features. Nonclinical tests included: - Verification of the proper loading, display, and manipulation of MR image data. ● - . Verification of the proper loading, display, and manipulation of Mammographic and Ultrasound image data. - Verification that the calculation and display of kinetic colormaps were accurate ● on MR image data. - Verification testing of lesion segmentation on MR image data. - Verification that all measurements and BIRADS reporting were recorded correctly. - Validation testing demonstrates the device conforms to user needs and intended ● use. #### F. CONCLUSION OuantX provides features to aid radiologists in the visualization and analysis of breast MR images. The potential hazards, including risk analysis and design considerations, have been examined and incorporated into the software development process. This 510(k) Premarket Notification provides adequate information to establish substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...