← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K163294

# syngo.MR General, syngo.MR Cardiology, syngo.MR Neurology (K163294)

_Siemens Medical Solutions USA, Inc. · LLZ · Feb 6, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K163294

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Feb 6, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

The software comprising the syngo.MR post-processing applications is post-processing software/applications to be used for viewing and evaluating the designated images provided by a magnetic resonance diagnostic device. All of the software applications comprising the syngo.MR post-processing applications have their own indications for use. syngo.MR General is a syngo based post-processing software for viewing, manipulating and evaluating MR images. syngo.MR Cardiology is a syngo based post-processing software for viewing, manipulating and evaluating MR cardiac images. syngo.MR Neurology is a syngo based post-processing software for viewing, manipulating, and evaluating MR neurological images.

## Device Story

Software suite for post-processing, viewing, and evaluating MR images; inputs include raw MR data (angiography, 4D breast, cardiac, BOLD, diffusion series); transforms data via arithmetic tools, elastic motion correction, and specialized engines (Cardiac 4D, Neuro Perfusion, BOLD/DTI processing); outputs include statistical maps, diffusion parameters (FA, ADC, etc.), and quantified lesion volumes; used in clinical settings by healthcare professionals; supports structured reporting (PI-RADS v2); aids clinical decision-making by providing quantitative metrics and enhanced visualization of MR datasets.

## Clinical Evidence

Bench testing only. No clinical data presented. Safety and effectiveness supported by software verification and validation, risk management per ISO 14971:2007, and conformance to recognized standards (IEC 62304, DICOM PS 3.1-3.20).

## Technological Characteristics

Software-based post-processing suite; operates on syngo platform; supports DICOM interoperability; utilizes arithmetic, motion correction, and quantification algorithms; conforms to AAMI/ANSI/IEC 62366-1:2015, ISO 14971:2007, IEC 62304:2006, and NEMA PS 3.1-3.20 standards.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- syngo.MR Post-Processing Software (Version SMRVA16A) ([K130749](/device/K130749.md))
- syngo.MR Post-Processing Software Version SMRVB10A ([K151353](/device/K151353.md))
- syngo MR E11C Software ([K153343](/device/K153343.md))
- syngo.via (VB10) ([K150843](/device/K150843.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 6, 2017

Siemens Medical Solutions USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Specialist 40 Liberty Blvd. Mail Code 65-1A MALVERN PA 19355

#### Re: K163294

Trade/Device Name: syngo.MR General: syngo.MR Cardiology: syngo.MR Neurology Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 21, 2016 Received: November 22, 2016

#### Dear Mr. Fields:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Michael D'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is displayed in a light blue color. The background is white, providing a clean contrast to the text.

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

k163294

Device Name

syngo.MR General; syngo.MR Cardiology; syngo.MR Neurology

Indications for Use (Describe)

The software comprising the syngo.MR post-processing applications is post-processing software/applications to be used for viewing and evaluating the designated images provided by a magnetic resonance diagnostic device. All of the software applications comprising the syngo.MR post-processing applications have their own indications for use.

syngo.MR General is a syngo based post-processing software for viewing, manipulating MR images.

syngo.MR Cardiology is a syngo based post-processing software for viewing, manipulating MR cardiac images.

syngo.MR Neurology is a syngo based post-processing software for viewing, manipulating MR neurological images.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

Traditional 510(k) Premarket Notification for: syngo.MR Post-Processing Applications Software VB20

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light blue or teal. The letters are evenly spaced and the word is horizontally oriented.

# Section 5 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.

# Date of Summary Preparation: November 16, 2016

# I. General Information

| Importer / Distributor: | Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA                                                                                                                                                           |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number:    | 2240869                                                                                                                                                                                                                                                              |
| Manufacturer:           | Siemens Healthcare GmbH<br>Henkestr. 127<br>D-91052 Erlangen, Germany                                                                                                                                                                                                |
| Registration Number:    | 3002808157                                                                                                                                                                                                                                                           |
| Contact Person:         | Cordell L. Fields, Esq.<br>Regulatory Affairs Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Pkwy<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Fax: (610) 427-3465<br>Email: cordell.fields@siemens.com |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The word is horizontally oriented and centered in the image.

# Device Name and Classification

| Data                                           | Details                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name /<br>Device<br>Proprietary<br>Name: | syngo.MR General<br>syngo.MR General contains several MR Radiology<br>workflows and MR specific Evaluation features.<br>It covers single and engine applications:<br>syngo.MR Readingsyngo.MR General Routinesyngo.MR Cardiac Readersyngo.MR Composingsyngo.MR General Enginesyngo.MR General Engine is the precondition for all other,<br>advanced MR post-processing applications and Engines. |
|                                                | syngo.MR Cardiology<br>syngo.MR Cardiology covers single and engine<br>applications:<br>syngo.MR Cardiac 4D Ventricular Functionsyngo.MR Cardiac Flowsyngo.MR Cardio Engine                                                                                                                                                                                                                      |
|                                                | syngo.MR Neurology<br>syngo.MR Neurology covers single and engine<br>applications:<br>syngo.MR Neuro Perfusionsyngo.MR Neuro Perfusion Mismatchsyngo.MR Neuro fMRIsyngo.MR Tractographysyngo.MR Neuro Perfusion Enginesyngo.MR Neuro 3D Engine                                                                                                                                                   |
| Classification<br>Name:                        | Regulation Description:<br>Picture Archiving and Communication System (PACS)                                                                                                                                                                                                                                                                                                                     |
| Classification<br>Panel:                       | Radiology                                                                                                                                                                                                                                                                                                                                                                                        |
| Device<br>Classification:                      | Class II devices                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation<br>number:                          | 21 CFR § 892.2050                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are all capitalized and are a light blue color. The background is white, which makes the word stand out.

| Data          | Details                      |
|---------------|------------------------------|
| Product Code: | Primary: LLZ, Secondary: LNH |

| Data                                           | Details                                                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name /<br>Device<br>Proprietary<br>Name: | syngo.MR General<br>syngo.MR General contains several MR Radiology<br>workflows and MR specific Evaluation features.<br>It covers single and engine applications:<br>• syngo.MR Reading<br>• syngo.MR General Routine<br>• syngo.MR Cardiac Reader<br>• syngo.MR Composing<br>• syngo.MR General Engine<br><br>syngo.MR General Engine is the precondition for all<br>other, advanced MR post-processing applications and<br>Engines. |
| Trade name /<br>Device<br>Proprietary<br>Name  | syngo.MR Cardiology<br>syngo.MR Cardiology covers single and engine<br>applications:<br>• syngo.MR Cardiac 4D Ventricular Function<br>• syngo.MR Cardiac Flow<br>• syngo.MR Cardio Engine                                                                                                                                                                                                                                             |
| Trade name /<br>Device<br>Proprietary<br>Name  | syngo.MR Neurology<br>syngo.MR Neurology covers single and engine<br>applications:<br>• syngo.MR Neuro Perfusion<br>• syngo.MR Neuro Perfusion Mismatch<br>• syngo.MR Neuro fMRI<br>• syngo.MR Tractography<br>• syngo.MR Neuro Perfusion Engine<br>• syngo.MR Neuro 3D Engine                                                                                                                                                        |

# II. Safety and Effectiveness Information Supporting Substantial Equivalence

### Intended Use

The software comprising the syngo.MR post-processing applications is postprocessing software/applications to be used for viewing and evaluating the

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are all capitalized and are a light blue color. The background is white, which makes the word stand out.

designated images provided by a magnetic resonance diagnostic device. All of the software applications comprising the syngo.MR post-processing applications have their own indications for use.

syngo.MR General is a syngo based post-processing software for viewing, manipulating and evaluating MR images.

syngo.MR Cardiology is a syngo based post-processing software for viewing, manipulating and evaluating MR cardiac images.

syngo.MR Neurology is a syngo based post-processing software for viewing, manipulating, and evaluating MR neurological images.

## Device Description

The syngo.MR post-processing applications are synqo based post-processing software/applications to be used for viewing and evaluating' MR images provided by a magnetic resonance diagnostic device and enabling structured evaluation of MR images.

With SMRVB20 there are some new features, improvements and changes within the syngo.MR post-processing applications.

Within syngo.MR General VB20 there are the following new features and improvements:

- Arithmetic tools (new): Addition, Division, Multiplication
- Motion Correction (Elastic) (new): A Motion Correction algorithm can ● be used to perform elastic motion correction for angiography series (pre/post) or within 4D Breast datasets.
- MR Combine feature (new): Composing is also available for axial series. ●
- MR Prostate workflow provides PI-RADS™ v2 reporting (improved) ●
- Harmonized MR Basic workflow (improved): Several basic workflows . for routine reading are consolidated in one MR Basic workflow.
- . MR Neurology workflow: The MR Neurology workflow merges the already cleared workflows of MR Head, MR Neuro Perfusion, and MR Neuro Dynamics.
- Easy Reading Layout in all workflows (improved): All workflows now include a viewing step with Easy Reading Layout.
- Improved result management (improved): Multiple export options for . findings in the interactive Findings details dialog

<sup>1</sup> While viewing (i.e. assessing) of images from other vendors is always possible; for advanced post-processing applications, some of the post-processing steps may depend on information contained in private DICOM tags ,therefore evaluation and processing of images can´t be guaranteed for other vendors.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the word "SIEMENS" in a sans-serif font. The word is written in a bright cyan color. The letters are evenly spaced and appear to be a part of a company logo.

Within syngo.MR Cardiology VB20 there are the following new features and improvements:

- Volume Quantification Tool (new): Volume Quant provides the . capability to evaluate lesion volumes in the myocardium.
- . Improved Result Distribution (improvement): Segmentation images can be exported as a result series.

Within syngo.MR Neurology VB20 there are the following new features, changes and improvements:

- MR Neuro 3D workflow:
	- Offline BOLD (new): offers the capability to run the GLM ● evaluations on raw BOLD data to generate fMRI statistical maps.
	- . Offline DTI (new): offers the capability to generate TENSOR data together with all other diffusion maps (including b0. ADC. TraceW, FA, AD, RD) from raw diffusion series.
	- DTI evaluation (new): offers quantitative evaluation of diffusion . parameters (FA, RD, AD, ADC, ...) using ROI, VOI, or voxels.
- . Neuro-specific Mean Curve Tool (minor improvement): The application svngo.MR Neuro Dynamics described in K151353 (and cleared August 07, 2015) is only available within syngo.MR Neuro Perfusion (neurospecific Mean Curve Tool), but no longer as a single application.

The syngo.MR Neuro Perfusion Engine Pro, described in K151353 (and cleared August 07, 2015), is therefore obsolete within SMRVB20, as all applications are already part of syngo.MR Neuro Perfusion Engine.

### General Safety and Effectiveness Concerns

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk Management is ensured via a Risk Analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a Risk Analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards. Furthermore, the operators are healthcare professionals familiar with and responsible for the evaluating and post-processing of MR images.

syngo.MR General, syngo.MR Cardiology and syngo.MR Neurology conform to the applicable FDA recognized and international IEC, ISO and NEMA

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are all capitalized and are a light blue color. The background is white, which makes the text stand out.

standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document.

| Rec.- No.  | Product Area | Title of Standard                                                                      | Ref.- No. & Date        | Standards<br>Development<br>Organization |
|------------|--------------|----------------------------------------------------------------------------------------|-------------------------|------------------------------------------|
| 5-96       | General      | Medical devices -<br>Application of usability<br>engineering to medical<br>devices     | 62366-<br>1:2015        | AAMI ANSI<br>IEC                         |
| 5-40       | General      | Medical devices - application of<br>risk management to medical<br>devices              | 14971:<br>2007          | ISO                                      |
| 13-<br>32  | Software     | Medical device software -<br>Software life cycle<br>processes                          | 62304:2006              | AAMI ANSI<br>IEC                         |
| 12-<br>300 | Radiology    | Digital Imaging and<br>Communications in<br>Medicine (DICOM) Set PS<br>3.1-3.20 (2011) | PS 3.1 -<br>3.20 (2016) | NEMA                                     |

The standards conformed to are the following:

# Substantial Equivalence

Each of the medical devices, syngo.MR General, syngo.MR Cardiology and syngo.MR Neurology, running with the herein described version SMRVB20 has the same Intended Use as the primary predicate device (see Table 1). The conclusions from the non-clinical data suggest that the additional features bear an equivalent safety and performance profile as that of the predicate device and also does not affect the Indications for use of it. Therefore syngo.MR General, syngo.MR Cardiology and syngo.MR Neurology are considered to be substantially equivalent to their primary predicate devices, which are current legally marketed devices.

Table 1: Predicate devices for syngo.MR General, syngo.MR Cardiology and syngo.MR Neurology

| Primary Predicate Device                                           | FDA Clearance                         | Product Code | for Medical<br>Device syngo.x |
|--------------------------------------------------------------------|---------------------------------------|--------------|-------------------------------|
| <i>syngo.</i> MR Post-Processing<br>Software<br>(Version SMRVA16A) | K130749<br>cleared August 20,<br>2013 | LLZ, LNH     | MR General<br>MR Cardiology   |
| <i>syngo.</i> MR Post-Processing<br>Software Version<br>SMRVB10A   | K151353<br>cleared August 07,<br>2015 | LLZ, LNH     | MR Neurology<br>MR General    |

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The background is white. 

| Secondary Predicate<br>Device                                                                                                              | FDA Clearance                        | Product<br>Code  | for Medical<br>Device syngo.x               |
|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|------------------|---------------------------------------------|
| syngo MR E11C Software for<br>Siemens MR Systems:<br>MAGNETOM Aera (1.5T),<br>MAGNETOM Skyra (3T) and<br>MAGNETOM<br>Prisma/Prismafit (3T) | K153343<br>cleared April 15,<br>2016 | LNH, LNI,<br>MOS | MR General<br>MR Neurology                  |
| syngo.via (VB10)                                                                                                                           | K150843<br>cleared April 24,<br>2015 | LLZ              | MR General<br>MR Cardiology<br>MR Neurology |

## Conclusion as to Substantial Equivalence

The syngo.MR post-processing applications are intended for similar indications as cleared in their according primary predicate devices.

In summary, Siemens is of the opinion that the syngo.MR post-processing applications do not raise new questions of safety or effectiveness and are substantially equivalent to the currently marketed primary predicate devices:

- syngo.MR Post-Processing Software, version SMRVA16A (K130749 . cleared August 20, 2013) for syngo. MR General and syngo.MR Cardiology
- syngo.MR Post-Processing Software, version SMRVB10A (K151353 . cleared August 07, 2015) for syngo.MR Neurology and syngo.MR General

There is new added functionality for the syngo.MR post-processing applications. The differences give the devices greater capabilities than the predicates, but the Intended Use, the basic technological characteristics and functionalities remain the same.

Therefore, Siemens believes that the subject device, the syngo.MR postprocessing applications syngo.MR General, syngo.MR Cardiology and syngo.MR Neurology, are substantially equivalent to their primary predicate devices listed above in Table 1.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K163294](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K163294)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
