SIMPLANT Online Case Review and SIMPLANT Editor are indicated for use as medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. These are intended for use as preoperative software programs for generating and reviewing plans for dental implant placement and surgical treatment.
Device Story
SIMPLANT Online Case Review and SIMPLANT Editor are standalone software applications for preoperative dental implant planning. Input consists of SIMPLANT project files created by DENTSPLY Implants, which merge patient image data, patient information, and implant data. SIMPLANT Online Case Review is a web application for reviewing and approving plans; SIMPLANT Editor is a desktop application for editing plans. Dental professionals use these tools to visualize 2D gray value images and 3D geometries, perform measurements, check for collisions, and approve surgical plans. The output supports the manufacturing of surgical guides or clinical evaluation during surgery. The device is used in clinical settings by dental professionals. It benefits patients by facilitating precise preoperative planning for dental implant procedures.
Clinical Evidence
Bench testing only. Verification and validation were conducted per IEC 62304, including unit testing, peer code review, integration testing, internal release testing, smoke testing, formal system testing, acceptance testing, and beta testing. Testing confirmed functionality, safety, and efficacy compared to the predicate device.
Technological Characteristics
Standalone medical software. SIMPLANT Online Case Review is a web application (C#); SIMPLANT Editor is a desktop application (C++). Both operate on Windows. Functionality includes 2D gray value image visualization, 3D viewing, measurement tools, and collision detection. No patient contact. Software life cycle processes follow IEC 62304.
Indications for Use
Indicated for use by medically trained dental professionals for visualizing gray value images and preoperative planning of dental implant placement and surgical treatment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a stacked formation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 28, 2016
Dentsply Sirona % Ms. Helen Lewis Director, Corporate Regulatory Affairs 221 West Philadelphia Street Suite 60 YORK PA 17404
Re: K161118
Trade/Device Name: SIMPLANT Online Case Review and SIMPLANT Editor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 7, 2016 Received: July 8, 2016
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161118
Device Name
SIMPLANT Online Case Review and SIMPLANT Editor
Indications for Use (Describe)
SIMPLANT Online Case Review and SIMPLANT Editor are indicated for use as medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. These are intended for use as preoperative software programs for generating and reviewing plans for dental implant placement and surgical treatment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a curved leaf or wave, rendered in gray. To the right of the shape, the words "Dentsply" appear on the top line, and "Sirona" appear on the bottom line, both in a sans-serif font and also in gray.
## SECTION 5. 510(k) SUMMARY
#### for
### SIMPLANT Online Case Review and SIMPLANT Editor
- Submitter Information: 1.
Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-487-1332 |
| Fax Number: | 717-849-4343 |
Date Prepared: April 15, 2016
- 2. Device Name:
| Proprietary Name: | SIMPLANT Online Case Review and SIMPLANT Editor |
|----------------------|-------------------------------------------------|
| Classification Name: | Picture Archiving and Communication System |
| CFR Number: | 21 CFR §892.2050 |
| Device Class: | Class II |
| Product Code: | LLZ |
#### 3. Predicate Device:
The medical device SimPlant 2011 was selected as the predicate device for this submission for the proposed devices SIMPLANT Online Case Review and SIMPLANT Editor. Simplant 2011 is designed, developed and manufactured by DENTSPLY Implants NV (formerly Materialise Dental NV) and cleared for the US market in K110300.
#### 4. Description of Device:
The proposed devices, SIMPLANT Online Case Review and SIMPLANT Editor, are software intended for pre-operative planning to generate and review plans for dental implant placement and surgical treatment, without patient contact.
SIMPLANT Online Case Review is a web application used for review and approval of dental implant plans provided by DENTSPLY Implants. SIMPLANT Online Case Review can also be used in combination with the desktop software application SIMPLANT Editor when an implant plan, provided by DENTSPLY Implants, is edited by the dental professional.
The software SIMPLANT Online Case Review and SIMPLANT Editor are used by dental professionals with clinical experience in implant surgery as well as training in medical image review.
{4}------------------------------------------------
Implant plans, in the format of SIMPLANT project files are created by DENTSPLY Implants and made available to the dental professional. The SIMPLANT project is created by DENTSPLY Implants using dental professional supplied patient image data, patient information and implants data developed by DENTSPLY Implants which are merged together. The SIMPLANT project is the basis for implant surgery planning by dental professionals.
The dental plan, which is the result of the dental implant planning process, can be used for manufacturing a surgical guide or for evaluation of treatment options by the dental professional during the implant surgery procedure.
- ડ. Indications for Use:
SIMPLANT Online Case Review and SIMPLANT Editor are indicated for use as medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. These are intended for use as preoperative software programs for generating and reviewing plans for dental implant placement and surgical treatment.
- 6. Substantial Equivalence:
Technological characteristics
The proposed devices, SIMPLANT Online Case Review and SIMPLANT Editor, are stand-alone medical device software used without patient contact.
| Element | Proposed Devices | | Predicate Device | Difference |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | SIMPLANT<br>Online Case<br>Review | SIMPLANT<br>Editor | Simplant 2011<br>(K11300) | |
| Indications<br>for Use | SIMPLANT Online Case Review<br>and SIMPLANT Editor are<br>indicated for use as medical front-<br>end software that can be used by<br>medically trained people for the<br>purpose of visualizing gray value<br>images. These are intended for use<br>as pre-operative software programs<br>for generating and reviewing plans<br>for dental implant placement and<br>surgical treatment. | | Simplant 2011 is intended for<br>use as a software interface and<br>image segmentation system for<br>the transfer of imaging<br>information from a medical<br>scanner such as a CT scanner or<br>a Magnetic Resonance scanner.<br>It is also intended as pre-<br>planning software for dental<br>implant placement and surgical<br>treatment. | Yes |
| Intended<br>Purpose | Review and<br>approve implant<br>plan | Edit implant<br>plan for<br>approval in<br>Online Case<br>Review. | Create, edit and approve implant<br>plan. | Yes |
| Function | | | Volume Rendering | Yes |
| | | | Segmentation Wizard | Yes |
| | | | Reorient axial images to occlusal plane | Yes |
| | | | Advanced virtual teeth | Yes |
| | | | Advanced grafts and Volumes | Yes |
| | | | Dual scan registration | Yes |
| Element | Proposed Devices | | Predicate Device | Difference |
| | SIMPLANT<br>Online Case<br>Review | SIMPLANT<br>Editor | Simplant 2011<br>(K11300) | |
| | | | Optical scan registration | Yes |
| | | | Occlusion tool | Yes |
| | | | Virtual occludator* | Yes |
| | | | Advanced soft tissue simulation | Yes |
| | | | Surgical guide wizard | Yes |
| | Open project | Open project | Open project | No |
| | | Save project | Save project | No |
| | Load project | | Load project | No |
| | 3D viewing | 3D viewing | 3D viewing | No |
| | 2D gray value images | 2D gray value images | 2D gray value images | No |
| | Navigation | Navigation | Navigation | No |
| | | | Review and accept nerve | Yes |
| | Distance | Distance | Distance | No |
| | Geometry<br>presentation | Geometry<br>presentation | Geometry presentation | No |
| | Measurements | Measurements | Measurements | No |
| | Review implants | Review implants | Review implants | No |
| | | Edit implants | Edit implants | No |
| | | Change implant | Change implant | No |
| | Collision detection | Collision detection | Collision detection | No |
| | Approve plan | | Approve plan | No |
| Media for<br>Delivery | Web application | Software-File<br>for download | Software-Magnetic media | Yes |
| Principles of<br>operation | Web application | Desktop<br>software<br>application | Desktop software application | Yes |
| Program<br>language | C# | C++ | C++ | Yes |
| Operating<br>System | Windows | Windows | Windows | No |
Table 5-1 Substantial equivalence comparison table
{5}------------------------------------------------
*Occludator means a function to visualize how the upper and lower jaw will come together when the mouth is closed.
# Analysis of differences
The proposed devices compared to the predicate have similar indications for use and partly the same basic functionalities. The first 11 functions listed in Table 5-1 for the predicate device are used during creation of an implant plan, which will not be performed with any of the proposed devices. The differences here are due to the proposed devices', SIMPLANT Online Case Review and SIMPLANT Editor's dependency to a pre-created implant plan made by DENTPLY Implants.
The 'Review and accept nerve' function is no longer included in the software applications as a separate function. For the proposed devices, SIMPLANT Online Case Review and SIMPLANT Editor, the nerve is accepted implicitly when the complete dental implant plan is approved by the dental professional.
{6}------------------------------------------------
The differences in principles of operation and media for delivery are due to the simplification of the review process for the dental professional allowing them to view implant plans in the web application SIMPLANT Online Case Review. This does not require transfer of all the data in the implant plan to a local computer nor is a local installation of a software application required.
The differences in programming languages are also due to principles of operation and the fact that SIMPLANT Online Case Review is a web application. For this reason. C# is selected for its suitability to support development of web applications.
- 7. Software testing
Software verification and validation was conducted to ensure the functionality and compatibility of all system components and to support the safety and effectiveness of the proposed devices.
The software testing has been compared to the predicate device, Simplant 2011 (K110300), image accuracy testing and tests applied in accordance with the software life cycle processes, as defined in IEC 62304.
The verification and validation testing consists of the following activities:
- Unit test
- Peer Code Review
- Integration test ●
- Internal release test ●
- Smoke test
- Formal system test
- Acceptance test ●
- Beta test ●
Verification and validation testing confirms that all user needs and performance requirements according to the design input are fulfilled. The comparison tests confirm the functionality, safety and efficacy of the proposed devices.
#### 8. Conclusion Regarding Substantial Equivalence
The SIMPLANT Online Case Review and SIMPLANT Editor are dental software which are intended for use as a medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. The SIMPLANT Online Case Review and SIMPLANT Editor are intended for use as pre-operative software programs for generating and reviewing plans for dental implant placement and surgical treatment.
The SIMPLANT Online Case Review and SIMPLANT Editor have similar indications for use and incorporate the same fundamental functions as the predicate device SimPlant 2011 cleared under premarket notification K110300. Test data to verify the performance of the SIMPLANT Online Case Review and SIMPLANT Editor have been provided, where the results of this testing, combined with the comparison to the predicate device, support substantial equivalence and do not raise any new issues of safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.