← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K153705

# PROcedure Rehearsal Studio (K153705)

_3Dsystems, Simbionix Products · LLZ · Sep 2, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K153705

## Device Facts

- **Applicant:** 3Dsystems, Simbionix Products
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Sep 2, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

The PROcedure Rehearsal Studio software is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner to an output file. It is also intended as pre-operative software for surgical planning

## Device Story

PROcedure Rehearsal Studio is a software-based radiological image processing system used by clinicians for pre-operative surgical planning. It takes patient CT scan data as input and performs 3D anatomical segmentation to create patient-specific 3D models. The software supports thoracic (TEVAR), abdominal (EVAR), carotid, and neurological (cerebral circulation) anatomies. The generated 3D models assist physicians in visualizing patient anatomy for surgical planning; models may also be exported to the Simbionix ANGIO Mentor Simulator for training and post-operative debriefing purposes. The device facilitates clinical decision-making by providing detailed anatomical representations derived from patient imaging.

## Clinical Evidence

No clinical data. Evidence consists of software verification and validation testing using four patient datasets (Carotid, EVAR, TEVAR, and Neuro Intervention types). Testing included segmentation quality, phantom testing, and usability testing. All tests passed successfully.

## Technological Characteristics

Software-based radiological image processing system. Functions as an interface and segmentation tool for CT imaging data. Utilizes 3D segmentation algorithms to generate anatomical models. Operates as a standalone software application. Verification conducted per FDA software guidance.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- PROcedure Rehearsal Studio ([K123269](/device/K123269.md))
- PROcedure Rehearsal Studio ([K112387](/device/K112387.md))
- PROcedure Rehearsal Studio ([K093269](/device/K093269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 2, 2016

3DSystems, Simbionix Products % Ms. Kim Torluemke Vice President, Ouality and Regulatory Affairs, Healthcare 3D Systems, Inc. 5381 South Alkire Circle LITTLETON CO 80127

Re: K153705

Trade/Device Name: PROcedure Rehearsal Studio Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 12, 2016 Received: August 15, 2016

Dear Ms. Torluemke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153705

Device Name PROcedure Rehearsal Studio

Indications for Use (Describe)

The PROcedure Rehearsal Studio software is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner to an output file. It is also intended as pre-operative software for surgical planning

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                               |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY [as required by section 807.92(c)] PROcedure Rehearsal Studio™ 510(k) Number K

#### 1. SUBMITTER

## Applicant's Name:

3DSystems, Simbionix Products Beit Golan Corner of Golan and Hanegev St. Airport City, 70151, Israel Tel: +972-3-9114444 Fax: +972-3-9114455

## Contact Person:

Kim Torluemke Vice President, Quality and Regulatory Affairs, Healthcare 5381 South Alkire Circle Littleton, Colorado 80127 Phone: 720-643-1001; Fax: 720-643-1009; kim.torluemke@3dsystems.com Date Prepared: December 22, 2015

#### 2. DEVICE

Trade Name: PROcedure Rehearsal Studio™

## Common or Usual Name:

Radiological Image Processing System

Classification: Name: System, Image Processing, Radiological Product Code: LLZ Regulation No: 892.2050 Class: 2 Classification Panel: Radiology

#### 3. PREDICATE DEVICES

## Main predicate:

PROcedure Rehearsal Studio™ cleared for marketing under K123269, K112387 and K093269

{4}------------------------------------------------

#### 4. DEVICE DESCRIPTION

The PROcedure Rehearsal Studio software allows clinicians to create a patient specific 3D anatomical model based on a patient's CT for the purpose of preoperative surgical planning.

The 3D segmentation model produced by the PROcedure Rehearsal Studio may be exported to the Simbionix ANGIO Mentor Simulator Practice Environment and allow the physician to create a library of modules for training and post-operative debriefing. The ANGIO Mentor Simulator Practice Environment is not meant for clinical purposes and is intended to be used for training purposes only.

The modifications subject to this Special 510(k) submission are: (1) Graphic User Interface changes in various locations; (2) functional change in various locations which enable the addition of a Neuro module that allows the software to create 3D models for neurological scans in addition to the thoracic (TEVAR), abdominal (EVAR) and Carotid options that were previously cleared.

The Neuro module builds upon the previously cleared PROcedure Rehearsal Studio modules by adding the support for segmentation of the cerebral circulation system in addition to existing carotid, thoracic, and abdominal anatomies.

#### ട്. INDICATIONS FOR USE

The PROcedure Rehearsal Studio software is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner to an output file. It is also intended as preoperative software for surgical planning.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6. PREDICATE DEVICE

The software uses the same imaging inputs and substantially equivalent segmentation methods as the predicate device to create a 3D model which can be used for pre-operative surgical planning.

#### 7. PERFORMANCE DATA

Below is a list of the tests that have been performed and successfully completed for the PROcedure Rehearsal Studio.

## Software Verification and Validation Testing:

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

{5}------------------------------------------------

The testing activities were conducted according to the following phases of the user work flow: Importing Patient Data, Segmentation and Centerlines. Verification was performed using four patients' datasets (Dataset A through Dataset D). Dataset A was of Carotid Type, Dataset B was of EVAR type, Dataset C was of TEVAR type and Dataset D was of Neuro Intervention type.

In addition to the verification testing above, the following tests were completed:

- Segmentation quality testing ●
- Phantom testing
- Usability testing ●

All the verification and validation tests passed successfully.

#### 8. CONCLUSION:

The PROcedure Rehearsal Studio has the same intended use and technological characteristics as the previously cleared version of the same device (K123269, K112387 and K093269). Consequently, it is clear that the PROcedure Rehearsal Studio is as safe and effective as its predicate devices without raising any new safety and/or effectiveness concerns. Any differences have been addressed by extensive testing and validations and therefore negate any safety or effectiveness concerns.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K153705](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K153705)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
