K140587 · Medis Medical Imaging Systems BV · LLZ · Jan 6, 2015 · Radiology
Device Facts
Record ID
K140587
Device Name
MR-CT VVA
Applicant
Medis Medical Imaging Systems BV
Product Code
LLZ · Radiology
Decision Date
Jan 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
MR-CT VVA is indicated for use in clinical settings where more reproducible than manually derived quantified results are needed to support the visualization and analysis of MR and CT images of the heart and blood vessels for use on individual patients with cardiovascular disease. Further, MR-CT VVA allows the quantification of T2* in MR images of the heart and the liver. Finally, MR-CT VVA can be used for the quantification of cerebral spinal fluid in MR velocity-encoded flow images. When the quantified results provided by MR-CT VVA are used in a clinical setting on MR and CT images of an individual patient, they can be used to support the clinical decision making for the patient. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians.
Device Story
MR-CT VVA is standalone image post-processing software for viewing/quantifying MR and CT images of heart, blood vessels, liver, and cerebral spinal fluid. Input: MR/CT image data via PACS, network, or CD/DVD. Operation: Clinician/technician uses semi-automatic contour detection to segment anatomy; software performs cardiac function quantification, flow quantification, tissue segmentation, and parametric mapping (T1, T2, T2*). Output: On-screen visualization (cine, 3D MIP, volume rendering) and exported reports. Used in clinical settings by cardiologists/technicians to support clinical decision-making; results are not sole basis for diagnosis. Benefits: Provides more reproducible quantification than manual methods.
Clinical Evidence
Bench testing only. Validation performed against design input requirements. No clinical data provided.
Technological Characteristics
Standalone Windows-based software. Modular architecture integrating previously cleared Medis software components. Features semi-automatic contour detection, 3D volume rendering, MIP, and parametric mapping. Connectivity via PACS/network/CD/DVD. Complies with ISO 13485, ISO 14971, and IEC 62304.
Indications for Use
Indicated for patients with cardiovascular disease requiring visualization and analysis of MR and CT images of the heart and blood vessels. Also indicated for T2* quantification in heart/liver MR images and cerebral spinal fluid quantification in MR velocity-encoded flow images. Intended for use by clinicians to support clinical decision-making.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medis medical imaging systems, b.v. % Bob Goedhart, Ph.D. VP Corporate R&D Schuttersveld 9 2316 XG Leiden THE NETHERLANDS
January 6, 2015
Re: K140587
Trade/Device Name: MR-CT VVA Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 7, 2014 Received: November 10, 2014
Dear Dr. Goedhart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140587
Device Name MR-CT VVA
### Indications for Use (Describe)
MR-CT VVA is indicated for use in clinical settings where more reproducible than manually derived quantified results are needed to support the visualization and analysis of MR and CT images of the heart and blood vessels for use on individual patients with cardiovascular disease. Further, MR-CT VVA allows the quantification of T2* in MR images of the heart and the liver. Finally, MR-CT VVA can be used for the quantification of cerebral spinal fluid in MR velocity-encoded flow images.
When the quantified results provided by MR-CT VVA are used in a clinical setting on MR and CT images of an individual patient, they can be used to support the clinical decision making for the patient. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 5. 510(k) Summary of Safety and Effectiveness
Submission in accordance with the requirements of 21 CFR Part 807.87(h).
| Submitter | : Medis medical imaging systems bv |
|------------------------|-----------------------------------------------|
| Address | : Schuttersveld 9 |
| | : 2316 XG Leiden |
| | : The Netherlands |
| Contact Person | : B. Goedhart, PhD, VP Corporate R&D |
| Email | : bgoedhart@medis.nl |
| Telephone | : +31 71 522 3244 |
| Fax | : +31 71 521 5617 |
| Prepared | : January 5, 2015 |
| Trade / Device Name | : MR-CT VVA |
| Common Name | : Radiological Image Processing Software |
| Regulatory Class | : II |
| Regulation Description | : Picture Archiving and Communications System |
| Regulation / Procode | : 21 CFR 892.2050 / LLZ |
#### Predicate Devices
- · The Cardiovascular Imaging Solutions Ltd .: CMRtoolsTM (K073194)
- · The GE Healthcare: CardiacVXTM (K121762)
### Device Description
MR-CT VVA (MR-CT Vessel and Ventricular Analysis) is image post-processing software for the viewing and quantification of MR and CT images of blood vessels, of the heart and MR images of the liver and cerebral spinal fluid. Semi-automatic contour detection forms the basis for the analyses. Its functionality is independent of the type of vendor acquisition equipment. The analysis results are available on screen and can be exported in various electronic formats.
MR-CT VVA has been developed as a standalone application to run on a Windows based operating system. The import of images and the export of analysis results are via CD / DVD, a PACS or network environment.
MR-CT VVA has a modular structure that consists of its previously cleared predicate devices: MRI-MASS, CT-MASS, MRI-FLOW, CMS-VIEW and MRA-CMS. MR-CT VVA comprises their respective functionalities for analyzing the blood vessels and the heart. In addition, MR-CT VVA includes new functionality for the 3D review of MR volumetric data.
### Intended Use
MR-CT VVA is software intended to be used for the visualization and analysis of MR and CT images of the heart and blood vessels.
MR-CT VVA is intended to support the following visualization functionalities:
- cine loop and 2D review -
- double oblique review -
- 3D review by means of MIP and volume rendering -
- -3D reformatting
- performing caliper measurements -
{5}------------------------------------------------
MR-CT VVA is also intended to support the following analyses:
- cardiac function quantification -
- MR velocity-encoded flow quantification .
- anatomy and tissue segmentation .
- signal intensity analysis for the myocardium and infarct sizing -
- MR parametric maps (such as T1, T2, T2* relaxation) -
MR-CT VVA is also intended to be used for:
- quantification of T2* results in MR images that can be used to characterize iron loading in the heart and the liver
- MR velocity-encoded flow quantification of cerebral spinal fluid
These analyses are based on contours that are either manually drawn by the clinician or trained medical technician who is operating the software, or automatically detected by the software and subsequently presented for review and manual editing. The results obtained are displayed on top of the images and provided in reports.
The analysis results obtained with MR-CT VVA are intended for use by cardiologists to support clinical decisions concerning the heart and vessels.
#### Indications for Use
-
MR-CT VVA is indicated for use in clinical settings where more reproducible than manually derived quantified results are needed to support the visualization and analysis of MR and CT images of the heart and blood vessels for use on individual patients with cardiovascular disease. Further, MR-CT VVA allows the quantification of T2* in MR images of the heart and the liver. Finally, MR-CT VVA can be used for the quantification of cerebral spinal fluid in MR velocity-encoded flow images.
When the quantified results provided by MR-CT VVA are used in a clinical setting on MR and CT images of an individual patient, they can be used to support the clinical decision making for the diagnosis of the patient. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians.
### Substantial Equivalence Information
Two devices have been selected as the predicate devices for MR-CT VVA: CMRtools™ and Cardiac VX™.
CMRtools ("a software package for the visualization and analysis of cardiovascular MR images") and Cardiac VX ("a post-processing analytical software application, which provides tools for review and reporting of cardiac MR datasets") have technological features and characteristics comparable to MR-CT VVA.
MR-CT VVA provides (see Intended Use description above) visualization functionalities for cine loop and 2D review and caliper measurements, which are also supported by CardiacVX ("Multi-phase sequences of images can be displayed in cine mode" and "Available tools include point, distance, area and volume measurement"). Further, MR-CT VVA provides visualization functionalities for 3D review and reformatting, which are also supported by CMRtools ("It allows high performance image manipulation, 3D visualization and advanced image processing").
MR-CT VVA provides (see Intended Use description above) analysis functionalities for cardiac function quantification, flow quantification, which are also supported by CardiacVX ("Semi-automatic tools are available for left ventricular contour detection, valve plane detection and vessel contour detection for flow quantification"). Further, MR-CT VVA provides analysis functionalities for tissue segmentation and MR parametric maps (including T2* for the liver), which are also supported by CMRtools ("cardiac assessment for left ventricular assessment and Thalassaemia tools for image quantification ... to view and analyze human cardiac and liver images").
Next to this, the following devices from Medis medical imaging system bv are predecessors of MR-CT VVA And can be used as reference devices: MRI-MASS (K994283), CT-MASS (K033774), MRI-FLOW (K994282), CMS-VIEW (K993761), and MRA-CMS (K040746).
{6}------------------------------------------------
#### Conclusions
MR-CT VVA has the same intended use as the predicate devices.
MR-CT VVA complies with international process standards (ISO 13485, ISO 14971 and IEC 62304).
Testing and validation have produced results consistent with design input requirements. MR-CT VVA is a software-only device for which there no applicable mandatory performance standards.
During the development, potential hazards were controlled by a risk management plan, including risk analysis, risk mitigation, verification and evaluation.
Medis concludes that MR-CT VVA is a safe and effective medical device, and is at least as safe and effective as its predicate devices. The use of MR-CT VVA does not change the intended use of MRI or CT scanners, nor does the use of this software result in any new potential hazards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.