K131158 · Marrek, Inc. · LLZ · Aug 21, 2013 · Radiology
Device Facts
Record ID
K131158
Device Name
CORVIEW
Applicant
Marrek, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Left Atrial Enhancement (LAE) quantification
—
—
—
—
—
—
—
Indications for Use
The Corview software application is indicated for viewing and post-processing of cardiovascular magnetic resonance images (MRIs) to obtain Left Atrial Enhancement (LAE) quantification and visualization on a 3D model. It provides measurements of LAE within the left atrial wall. It enables: Importing Cardiac MRIs in DICOM format Qualitative analysis of the cardiac MRIs using display functionality such as panning, windowing, zooming, and navigation through series/slices and phases Quantitative measurement in cardiac MRIs, specifically LAE It is intended for use by qualified medical professionals experienced in examining and evaluating cardiovascular MRIs, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. The target population of Corview is not restricted. However image acquisition by an MRI scanner with cardiac capability may limit the use of the device for certain sectors of the general public. Corview is intended to view only cardiac magnetic resonance images acquired from an MRI scanner with cardiac capabilities. The CORVIEW software application is intended for use in the clinical setting.
Device Story
CORVIEW is a standalone software application for processing digitized cardiac MRI datasets. It imports DICOM-formatted images from commercial MRI scanners. The software allows clinicians to perform qualitative analysis (panning, windowing, zooming, slice navigation) and quantitative analysis of the left atrial wall. Using a feature classification algorithm on 3D-acquired datasets, the device identifies endo- and epi-cardial borders to construct a 3D model of the left atrium and calculate the total percentage of left atrial enhancement (LAE). The output is visualized for the clinician to assist in diagnostic decision-making. It is intended for use in clinical settings by qualified medical professionals experienced in cardiovascular MRI evaluation.
Clinical Evidence
Bench testing only. Software verification and validation were performed at the code and system level according to pre-established protocols. All software requirements were verified, and results met pass/fail criteria. No clinical data provided.
Technological Characteristics
Standalone software application for PC or Apple computers. Processes DICOM-formatted cardiac MRI images. Features include 3D modeling, image segmentation, and feature classification algorithms for LAE quantification. Software verification and validation conducted per internal protocols.
Indications for Use
Indicated for viewing and post-processing cardiovascular MRI images to quantify and visualize Left Atrial Enhancement (LAE) in a 3D model. Intended for use by qualified medical professionals in clinical settings. No specific age or gender restrictions; use limited by availability of cardiac-capable MRI image acquisition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K13/1158
Page 1 of 4
#### SECTION 5: 510(k) SUMMARY
#### 510(k) Summary
| Date Prepared: | July 16, 2013 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Marrek, Inc.<br>3293 Niblick Drive<br>Park City, UT 84098<br>Phone: 1 (512) 582-2453<br>Facsimile: 1 (713) 589-7964 |
| Contact: | Jeremy M. Fotheringham, President & CEO<br>Mobile: 1 (801) 870-4323 |
| Trade/Proprietary<br>Name of Device: | CORVIEW |
| Common Name<br>of Device: | System, Image Processing, Radiological |
| Classification: | Class II per 21 CFR 892.2050, Picture archiving and<br>communications system, Product Code LLZ |
| Legally Marketed<br>Predicate<br>Device: | 1. 2D Cardiac Performance Analysis MR 1.0 device<br>(K120135), manufactured by TomTec Imaging<br>Systems<br>2. Virtue (CORE Module) (K111833), manufactured by<br>Diagnosoft, Inc. |
#### Description of New CORVIEW Device:
CORVIEW is an independent software solution for the segmentation and quantitative analysis of digitized cardiac magnetic resonance images (MRIs). CORVIEW operates on higher-end off-the-shelf PC or Apple computers and is compatible with Cardiac MRI digital images in DICOM format from a range of commercial MRI machine vendors. The CORVIEW software platform is also designed to process MRIs received from various modes of image exchange.
CORVIEW presents a state-of-the-art method to process and quantify the extent of left atrial enhancement (LAE) within the left atrial wall by contouring and identifying the endo- and epi-cardial borders from contrast enhanced MR image sequences that have been acquired as part of a conventional cardiac MRI examination.
{1}------------------------------------------------
CORVIEW is designed for the processing of 3D image datasets allowing users to produce contour models and visualizations of left atrial structure and contrast enhancement. Based on 3D-acquired datasets, a feature classification algorithm supports the calculation of LAE. From these corresponding algorithms, total percentage of enhancement can be derived. In addition, the corresponding shape of the left atrium can also be obtained.
#### Indications for Use of the New Device:
Indications for Use:
The Corview software application is indicated for viewing and post-processing of cardiovascular magnetic resonance images (MRIs) to obtain Left Atrial Enhancement (LAE) quantification and visualization on a 3D model. It provides measurements of LAE within the left atrial wall.
#### It enables:
- . Importing Cardiac MRIs in DICOM format
- Qualitative analysis of the cardiac MRIs using display functionality such as . panning, windowing, zooming, and navigation through series/slices and phases
- . Quantitative measurement in cardiac MRIs, specifically LAE
It is intended for use by qualified medical professionals experienced in examining and evaluating cardiovascular MRIs, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process.
The target population of Corview is not restricted. However image acquisition by an MRI scanner with cardiac capability may limit the use of the device for certain sectors of the general public.
Corview is intended to view only cardiac magnetic resonance images acquired from an MRI scanner with cardiac capabilities.
The CORVIEW software application is intended for use in the clinical setting.
## Comparison of the Technological Features of the New Device and Predicate Devices:
The new CORVIEW device, the lawfully marketed predicate Tom Tec Imaging Systems 2D Cardiac Performance Analysis MR 1.0 device (K120135), and the Diagnosoft Virtue-CORE device (K111833) are all stand-alone software
{2}------------------------------------------------
applications using cardiac data obtained from standard magnetic resonance images. Features of CORVIEW are comparable to those of the predicate devices. The main differences include:
- 1. The new CORVIEW quantifies the amount of contrast enhancement within the left atrial wall. The predicate Tom Tec Imaging Systems 2-D Cardiac Performance Analysis MR 1.0 device quantifies the myocardial deformation and movement of the entire heart muscle.
- 2. The new CORVIEW constructs and visualizes a 3D model of the left atrium.
- 3. The new CORVIEW quantifies the amount of contrast enhancement within the left atrial wall. The predicate Diagnosoft Virtue - CORE device quantifies the amount of contrast enhancement within the right and left ventricles as well as the surrounding vessels.
### Testing:
#### Non-Clinical Performance Data Testing:
Software testing and validation were done at the code and system level according to written test protocols established before testing was conducted.
The test procedure was performed and documented. The test results were reviewed by designated technical professionals to ensure acceptability criteria were satisfied prior to the release of the software.
The results are summarized in the test summary report. The conclusion states that:
- Verification strategies and test procedures used are appropriate
- Software system test procedures trace to software requirements
- All software requirements are tested or otherwise verified
- Test results meet the required pass/fail criteria
#### Adverse Effects on Health:
The potential risks are identified in a risk management analysis and report and are controlled by the software development team, risk management processes, documented verification and validation processes to ensure performance to specifications, Federal Regulations, user requirements, and adherence to industry and international standards.
Risk analysis aspects were treated in the risk management report. Based on this document, the applied methods in the literature and the techniques of the product (which are considered in the risk analysis) were evaluated. No further risks were identified.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows handwritten text. The text at the top reads "K131158". Below that, the text reads "Page 4 of 4". The handwriting appears to be cursive.
### Conclusion:
The conclusions drawn from the specifications and performance testing of the new CORVIEW device demonstrate that the new CORVIEW device is at least as safe and as effective and performs as well as or better than the predicate TomTec Imaging Systems 2D Cardiac Performance Analysis MR 1.0 device, (K120135) and the Diagnosoft Virtue - CORE device (K111833). For these reasons, we believe the new CORVIEW device is substantially equivalent to the predicate devices.
Signed,
Jeremy M. Fotheringham President & CEO Marrek, Inc.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, with three curved lines representing its wings and body. The logo is rendered in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2013
Marrek, Inc. % Mr. Jeremy M. Fotheringham President & CEO 3293 Niblick Drive PARK CITY UT 84098
Re: K131158 Trade/Device Name: Corview Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 16, 2013 Received: August 19, 2013
Dear Mr. Fotheringham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Fotheringham
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K131158
Device Name: Corview
Indications for Use:
The Corview software application is indicated for viewing and post-processing of cardiovascular magnetic resonance intris) to obtain Left Arrial Enhancement (LAE) quantification and visualization on a 3D model. It provides measurements of LAE within the lett atrial wall.
It enables:
- . Importing Cardiac MRIs in DICOM format
- . Qualitative analysis of the cardiac MRIs using display functionality snch as paming. windowing. and navigation through senes shees and phases
- . Quantitative measurement in cardiae MRIs. specifically LAE
It is intended for use by qualified medical professionals experienced in examining cardiovascular MRIs. for the purpose of obtaining diamanon as part of a comprehensve diagnostic decision-making process.
The target population of Corview is not restricted. However image acquisition by an MRI scardiac capability may linit the use of the device for certain sectors of the general public.
Corview is intended to view only cardine resonance intages acquired from an MRI scamer with cardiac capabilities.
The CORVIEW sothware application is intended for use in the clinical setting.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostics and Radiological Health (OIR)
| (min) |
|-------|
|-------|
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 510(k) K131158
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.