The ZMED VP4000 Video Processor is intended for use in any application where a viewing device (fluoroscope, endoscope, laparoscope, etc.) and monitor are incorporated to aid in diagnosis and treatment of a disease such as an arthroscopy or cholecystectomy.
Device Story
ZMED VP4000 is a real-time video processing system; connects between video source (e.g., camera) and display. Inputs raw video signals (DVI-D, HD-SDI up to 1080p); applies locally-adaptive contrast enhancement via large-kernel convolution and histogram-based operations to improve clarity, contrast, and detail; outputs enhanced video. Used in clinical settings (e.g., OR) by physicians to aid visualization through smoke, haze, or poor lighting. Enhancements assist in diagnosis and treatment; real-time processing ensures minimal latency. Device is non-patient contact; housed in aluminum chassis.
Clinical Evidence
Bench testing only. Device tested for safety per IEC60601-1 Second Edition and electromagnetic compatibility per FCC Part 15 Subpart B. No clinical data provided.
Technological Characteristics
Aluminum chassis; 15.7" x 12.0" x 1.7"; 4.95 lbs. Compatible with DVI-D and HD-SDI signals up to 1080p. Internal medical-grade power supply (100-240VAC). Real-time video enhancement via locally-adaptive contrast enhancement (large-kernel convolution) and histogram-based operations. Non-patient contact.
Indications for Use
Indicated for use in clinical applications requiring visualization via fluoroscope, endoscope, or laparoscope to aid in the diagnosis and treatment of diseases such as arthroscopy or cholecystectomy.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K130929
Page 1 of 3
APR 1 2 2013
# APPENDIX 1 510(k) SUMMARY
| APPLICANT: | ZMED, Inc.<br>9820 Summers Ridge Road<br>San Diego, CA 92121<br>858.831.7000 |
|-------------------|------------------------------------------------------------------------------|
| CONTACT: | Randall S. Millar<br>858.831.7059 |
| DATE OF SUMMARY: | 20 MARCH 2013 |
| DEVICE NAME: | ZMED VP4000 Video Processor |
| PREDICATE DEVICE: | DigiVision CE3000<br>FDA 510(k) file #: K962260<br>Year approved: 1996 |
### DEVICE DESCRIPTION
The ZMED VP4000 is a simple-to install, easy to use video processing system which provides powerful real-time enhancement of input video signals. The enhanced output video provides increased clarity, contrast, and detail, aiding in the ability to see through smoke, in poor lighting conditions, and other suboptimal conditions.
The VP4000 is connected between a video source and a display. An example of a typical installation is shown below:
Image /page/0/Figure/7 description: The image shows a diagram of a ZMED VP4000 connected to a video source and a video display. The video source is labeled as "VIDEO SOURCE (camera, etc.)" and the video display is labeled as "VIDEO DISPLAY". Arrows indicate the direction of the video signal from the video source to the ZMED VP4000 and from the ZMED VP4000 to the video display.
{1}------------------------------------------------
K130928
Page 2 of 3
The system uses standard electronic components to implement patented algorithms which are designed to improve the ability of the human eye to perceive fine details in the image which might otherwise be obscured by smoke, haze, poor lighting, etc. This is accomplished via a combination of locallyadaptive contrast enhancement using a very large-kernel convolution operation (identical to that used in the CE3000), and histogram-based operations. Like the CE3000, a key feature of the VP4000 is its ability to perform these complex operations in real time, adding virtually imperceptible system latency from input of raw video to output of enhanced video. The effectiveness of the algorithms combined with this real-time capability makes the VP4000 ideally suited for use in applications which live video is required to perform a given function.
The VP4000 chassis is constructed of lightweight aluminum, measures 15.7" wide by 12.0" deep by 1.7" high, and weighs 4.95 pounds. The device is compatible with DVI-D and HD-SDI video signals in a variety of standard resolutions, up to 1080p. Internally, the VP4000 is constructed using UL- and IEC-approved components, circuit boards, and harnessing. The VP4000 has been successfully tested to meet IEC60601-1 Second Edition safety standards, and FCC Part 15 Subpart B. These tests were performed by:
> TÜV SÜD America Inc., 10040 Mesa Rim Road. San Diego, CA 92121 Tel: (858) 678-1400
No adaptations to or deviations from these Standards were employed during testing.
### STATEMENT OF INTENDED USE
" The ZMED VP4000 Video Processor is intended for use in any application where a viewing device (fluoroscope, endoscope, laparoscope, etc.) and monitor are incorporated to aid in diagnosis and treatment of a disease such as an arthroscopy or cholecystectomy."
This statement is virtually identical to that of the predicate device (CE3000). Both the subject and predicate devices have the same function and purpose and the differences in intended use do not create new safety or effectiveness issues.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The VP4000 is an evolutionary development of the CE3000. Whereas the CE3000 is compatible with outdated analog video signals, the VP4000 has been designed to work with modern high-definition digital video formats. Additionally, advancements in electronic component capability since the CE3000 was designed have allowed for a higher degree of integration (fewer components required) when implementing the enhancement algorithms.
Both systems operate on standard AC power (100-240VAC). Both systems utilize an internal medical-grade power supply to convert the input AC power to low-voltage DC supplies for use by the electronics. Refer to the table below for a
{2}------------------------------------------------
KB0929
Page 3 of 3
summary comparison of changes between the unmodified (CE3000) and proposed (VP4000) devices.
| Features | VP4000 | CE3000 (unmodified) | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Similarities | | | |
| Intended Use | The ZMED VP4000 Video<br>Processor is intended for use<br>in any application where a<br>viewing device<br>(fluoroscope, endoscope,<br>laparoscope, etc.) and<br>monitor are incorporated to<br>aid in diagnosis and<br>treatment of a disease such<br>as an arthroscopy or<br>cholecystectomy. | The CE3000 Contrast<br>Enhancer like the<br>FluoroVision, may be used in<br>any application where a<br>viewing device (fluoroscope,<br>endoscope, laparoscope, etc.)<br>and monitor is incorporated to<br>aid in diagnosis and treatment<br>of a disease such as an<br>arthroscopy or<br>cholecystectomy. | |
| Device | Non-patient contact | Non-patient contact | |
| Category | Pre-market notification<br>510(k) | Approved via 510(k)<br>submission | |
| Technology | Real-time video<br>enhancement | Real-time video enhancement | |
| Differences | | | |
| User Interface | Simplified, selectable<br>presets | Individual controls for<br>parametric settings | |
| Supported<br>video formats | Digital, high-definition | Analog, standard-definition | |
(
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 12, 2013
ZMed, Inc. % Mr. Randall Millar VP of Engineering 9820 Summers Ridge Road SAN DIEGO CA 92121
Re: K130929
Trade/Device Name: ZMED VP4000 Video Processor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 27, 2013 Received: April 3, 2013
Dear Mr. Millar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Millar
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/delault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K130929 510(k) Number (if known):
ZMED VP4000 Video Processor Device Name:
Indications for Use:
The ZMED VP4000 Video Processor is intended for use in any application where a viewing device (fluoroscope, endoscope, laparoscope, etc.) and monitor are incorporated to aid in diagnosis and treatment of a disease such as an arthroscopy or cholecystectomy.
Prescription Use X ____________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) # K130929
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.