K123922 · Siemens Medical Solutions USA, Inc. · LLZ · Jan 16, 2013 · Radiology
Device Facts
Record ID
K123922
Device Name
SYNGO DYNAMICS
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
syngo® Dynamics is an image and information system intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting. As a Cardiology PACS and information system, syngo® Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology, Obstetrics and Gynecology or other departments. syngo® Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.
Device Story
syngo® Dynamics Version VA10A is a software-only Picture Archiving and Communication System (PACS) for managing digital medical images (ultrasound, CT, MR, X-ray, nuclear medicine, PET). It functions as a DICOM server to receive, store, distribute, and archive images and structured reports. Used in clinical settings by physicians and authorized personnel, the system provides workstations for image review, manipulation, quantitative analysis, and diagnostic reporting. It integrates with HIS/RIS/CIS to support clinical workflows. The system enables collaborative reporting and provides non-diagnostic, read-only access via mobile devices (iPhone/iPad). By centralizing cardiovascular and other imaging data, it assists clinicians in diagnosis and treatment planning, facilitating efficient access to patient information and documentation.
Clinical Evidence
Bench testing only. Verification and validation activities were performed to confirm system performance against design requirements. Compliance with standards including DICOM 3.x, IEC/ISO 10918-1, IEC 62366, and SMPTE was demonstrated. No clinical data was required or provided.
Technological Characteristics
Software-only system installed on standard IT hardware. Backend: Windows 2008 Server. Clients: Windows XP/7. Connectivity: DICOM, HL7, TCP/IP, HTTP. Image processing: Window/Leveling, Edge Enhancement, Digital Subtraction. Quantitative: Pixel Size Evaluation, vascular/cardiac measurements. Data compression: Lossless (factor 2-3) and lossy. Standards: DICOM 3.x, IEC/ISO 10918-1, IEC 62366, SMPTE.
Indications for Use
Indicated for use by physicians in healthcare institutions for the interpretation and evaluation of digital medical images, specifically in cardiology, obstetrics, gynecology, and other departments. Not indicated for the display of digital mammography images for diagnosis in the U.S.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K123922
# SIEMENS
Healthcare
JAN 1 6 2013
# 510(k) Summary
## svngo® Dvnamics (Version VA10A)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
- General Information l.
Date of Summary Preparation: December 5, 2012
#### Establishment:
- Address: .
Siemens Medical Solutions USA. Inc. 400 W. Morgan Road Ann Arbor, MI 48108
- Registration Number: ●
1836549
Contact Person: ●
Yuri Ikeda Quality Engineer, Quality & Regulatory Phone: (734) 205-2442 Fax: (734) 205-2683 Email: yuri.ikeda@siemens.com
#### Device Name and Classification:
- Trade Name: syngo® Dynamics Version VA10A
- Picture Archiving and Communications System Classification Name: .
- Classification Panel: Radiology
- 21 CFR §892.2050 CFR Number: ●
- Class II Device Class: .
- . Product Code:
Safety and Effectiveness Information Supporting the Substantial ll. Equivalence Determination
LLZ
#### Intended Use:
.
syngo® Dynamics is an image and information system intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting.
As a Cardiology PACS and information system, syngo® Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology, Obstetrics and Gynecology or other
Siemens Medical Solutions USA, Inc.
400 West Morgan Road Ann Arbor, MI 48108 USA
Tel .: +1-734-205-2400 www.usa.siemens.com/healthcare
{2}------------------------------------------------
#### departments.
syngo® Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.
#### Device Description:
This premarket notification addresses the Siemens syngo® Dynamics Version VA10A Picture Archiving and Communication System.
The system is a "software only" medical device. It defines recommended requirements to the hardware it runs on. The hardware itself is not considered a medical device and not in the scope of this 510(k) submission.
syngo® Dynamics is a system that includes a DICOM Server which receives, stores, distributes, and archives images from digital image acquisition devices such as ultrasound, computer tomography, magnetic resonance and x-ray angiography machines. The system has workplaces that can be used to review, edit, and manipulate image data, as well as to generate quantitative data, qualitative data, and diagnostic reports.
syngo® Dynamics supports the physician in diagnosis and treatment planning. It also supports storage and archiving of DICOM Structured Reports. In a comprehensive imaging suite syngo® Dynamics integrates Hospital / Radiology / Cardiology Information Systems (HIS/RIS/CIS) to enable customer specific workflows.
syngo® Dynamics Version VA10A provides advanced reporting support for cardiology, OB/GYN, MFM (maternal fetal medicine), vascular ultrasound, including specific echo and cath lab oriented features for documentation support in the cardiology department.
#### Data Management:
syngo® Dynamics allows all authorized personnel fast and continuous access to data such as cardiovascular images and information. Its functionality ranges from availability of images with regard to data security, open interfaces, storage media and central system administration, to provide a flexible storage hierarchy.
#### Technological Characteristics:
syngo® Dynamics is a "software only" system, which will be delivered on CD-ROM / DVD to be installed on common IT hardware. Hardware must meet the defined requirements. Any special needs such as integration in a specific environment and updates / upgrades will be covered by individual service contract and fulfilled by special trained service technicians.
The backend communication and storage solution is based on Windows 2008 operating system. The client machines are based on Windows XP, and Windows 7. Any hardware platform, which meets the specified recommended hardware and software requirements and with successful installation verification and
{3}------------------------------------------------
validation activities can be supported. syngo® Dynamics supports DICOM formatted images and objects.
syngo® Dynamics Version VA10A will be used to display, process, read, report, communicate, distribute and store digital medical images, much like its predicate, syngo® Dynamics Version 9.0 (K102150).
The difference between the syngo® Dynamics Version VA10A and the predicate device syngo® Dynamics Version 9.0 are to give the subject device greater capabilities than the predicate device. syngo® Dynamics Version VA10A has similar technological characteristics as the predicate device and is similar to the functionalities of the predicate device. The table below summarizes the similarities and the differences between the two devices.
| Functionality | syngo® Dynamics<br>Version VA10A | syngo® Dynamics<br>Version 9.0 |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Siemens Medical Solutions<br>USA, Inc. | Siemens Medical Solutions<br>USA, Inc. |
| Intended Use | syngo® Dynamics is an image<br>and information system<br>intended for acceptance,<br>transfer, display, storage,<br>archive and manipulation of<br>digital medical images,<br>including review, analysis,<br>quantification and reporting.<br>As a Cardiology PACS and<br>information system, syngo®<br>Dynamics supports the | syngo® Dynamics is a Picture<br>Archiving and Communication<br>System (PACS) intended for<br>acceptance, transfer, display,<br>storage, archive and<br>manipulation of digital medical<br>images, including quantification<br>and report generation.<br>syngo® Dynamics is not<br>intended to be used for reading<br>of mammography images. |
| | physician in interpretation and<br>evaluation of examinations<br>within healthcare institutions, in<br>particular, in Cardiology,<br>Obstetrics and Gynecology or<br>other departments. | |
| | syngo Dynamics® is not<br>intended to be used for<br>displaying of digital<br>mammography images for<br>diagnosis in the U.S. | |
| Functionality | syngo® Dynamics<br>Version VA10A | syngo® Dynamics<br>Version 9.0 |
| Operating<br>Systems | Server<br>Windows 2008 R2 Server<br>Standard Edition R2 SP1 or<br>SP2 (64-bit)<br><br>Workplace<br>Microsoft Windows XP SP2 or<br>higher<br>32-bit or 64-bit, orMicrosoft Windows 7 or<br>Windows 7 SP1 or higher32-bit or 64-bitUltimate, Professional,<br>Enterprise, Ultimate N,<br>Professional N, or<br>Enterprise N<br>Portal Website Host<br>Windows Server 2008 R1 32-<br>bit or greater | Server<br>Windows Server 2008 R2 SP1<br>Standard Edition (64-bit)<br><br>Workplace<br>Windows 2000, 2003, XP,<br>Vista, 7 (32 or 64-bit)<br><br>Portal Website Host<br>Windows Server 2008 R1 32-<br>bit or greater |
| Image Source | DICOM Ultrasound, XA, CT,<br>MR, DX, DR and Nuclear<br>Medicine, including PET. | DICOM Ultrasound, XA, CT,<br>MR, DX, DR and Nuclear<br>Medicine, including PET. |
| Image Display | Ultrasound, XA, CT, MR, DX,<br>DR, PET and Nuclear Medicine<br>through Corridor4DM | Ultrasound, XA, CT, MR, DX,<br>DR, PET and Nuclear Medicine<br>through MI Mobile or<br>Corridor4DM |
| Data Export<br>Image<br>Communication | DICOM, bmp, avi<br>Within the network, the<br>following communication<br>protocols are used:<br>TCP/IP: for communication<br>and transportDICOM and HL7 at<br>application levelHTTP for communication<br>and transport of thumbnails | DICOM, bmp, avi<br>Within the network, the<br>following communication<br>protocols are used:<br>TCP/IP: for communication<br>and transportDICOM and HL7 at<br>application levelHTTP for communication<br>and transport of thumbnails |
| Image Data<br>Compression | Lossless compression with<br>compression factor 2 to 3 and<br>lossy compression with higher<br>compression rate. | Lossless compression with<br>compression factor 2 to 3 and<br>lossy compression with higher<br>compression rate. |
| Imaging<br>Algorithms | Window/Leveling, Edge<br>Enhancement, and Digital<br>Subtraction | Window/Leveling, Edge<br>Enhancement, and Digital<br>Subtraction |
| Quantitative<br>Algorithms | Pixel Size Evaluation | Pixel Size Evaluation |
| Network Access | Yes | Yes |
| Analysis | Yes | Yes |
| Reporting | Yes | Yes |
| Multimodality<br>storage and<br>review | Yes | Yes |
| Functionality | syngo® Dynamics<br>Version VA10A | syngo® Dynamics<br>Version 9.0 |
| Web Server for<br>images and clips | Yes | Yes |
| Report upload to<br>Information<br>Systems | Yes, through broker or<br>interface engine | Yes, through broker or<br>interface engine |
| DICOM Structured<br>Reporting | Yes | Yes |
| Export/Import<br>Data Sets via<br>removable media<br>or network means | Yes | Yes |
| Vascular<br>Quantification | Yes, measurements and<br>calculations | Yes, measurements and<br>calculations |
| Data Mining | Yes | Yes |
| Discrete Data<br>Export | Yes | Yes |
| Cardiac<br>Measurements | Yes | Yes |
| Interactive<br>graphical<br>documentation for<br>reporting | Yes, through Soarian<br>Cardiology, version 2.0 utilizing<br>patient/user context sharing | Yes, through Soarian<br>Cardiology, version 2.0 utilizing<br>patient/user context sharing |
| Hemodynamic<br>data import (third<br>party vendor) | Yes | Yes |
| Web Reporting | Yes | Yes |
| Hardware | Software-only option for server<br><br>Workstation: software only<br>(HW is not part of the medical<br>device, but needs to meet<br>recommended requirements as<br>specified by syngo Dynamics) | Software-only option for server<br><br>Workstation: software only<br>(HW is not part of the medical<br>device, but needs to meet<br>recommended requirements as<br>specified by syngo Dynamics) |
| Collaborative<br>Reporting | Yes | No |
| Mobile Device<br>Support | Yes - Non-diagnostic, and<br>read-only. Support Apple<br>iPhone and iPad | No |
{4}------------------------------------------------
:
・
.
{5}------------------------------------------------
:
# Summary of Non-Clinical Tests:
Integration and System Testing were performed for verification and validation of The device. Siemens Medical Solutions USA, Inc. complies with voluntary standards DICOM Version 3.x (2011), IEC/ISO 10918-1:1994 + TC 1:2005 Standards DrOOm Version 6.X (2017) 1207 Version 2.3.1, IEC 62366:2007, and SMPTE.
# General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device.
{6}------------------------------------------------
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify potential hazards. These potential hazards are controlled via software development, labeling, and verification and validation testing.
The device has no patient contacting materials and is used only by trained professionals. The output of the device is evaluated by trained professionals allowing sufficient review for identification and intervention in the event of a malfunction.
Siemens believes that syngo® Dynamics Version VA10A is as safe and effective as its predicate device as it does not, raise new types of safety and effectiveness concerns, or introduce new technology.
#### Substantial Equivalence
The syngo® Dynamics Version VA10A, addressed in this premarket modification is substantially equivalent to the following commercially available device:
| Manufacturer | Predicate Device Name | FDA Clearance<br>Number |
|----------------------------------------|----------------------------|-------------------------|
| Siemens Medical Solutions<br>USA, Inc. | syngo Dynamics Version 9.0 | K102150 |
The potential hazards of modifications to the device have been evaluated and controlled as part of the product development process, including risk analysis and design considerations. Siemens conducts testing to verify the design output met the design input requirements and to validate the device conformance to the intended use. Predefined acceptance criteria was met and demonstrated that the device is as safe and effective as the predicate device.
In summary, Siemens is of the opinion that syngo® Dynamics Version VA10A does not introduce any new significant potential safety risks and is substantially equivalent to and performs as well as the predicate device.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### January 16. 2013
Siemens Medical Solutions USA, Inc. % Mr. Yuri Ikeda Ouality Engineer, Ouality & Regulatory TUV SUD America, Inc. 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891
Re: K123922
Trade/Device Name: syngo Dynamics (Version VA10A) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 17, 2012 Received: December 20, 2012
Dear Mr. Ikeda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance, of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## Sean M.ABoyd -S for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123922
Device Name: syngo® Dynamics (Version VA10A)
Indications for Use:
syngo® Dynamics is an image and information system intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including review, analysis, quantification and reporting.
As a Cardiology PACS and information system, syngo® Dynamics supports the physician in interpretation and evaluation of examinations within healthcare institutions, in particular, in Cardiology, Obstetrics and Gynecology or other departments.
syngo® Dynamics is not intended to be used for displaying of digital mammography images for diagnosis in the U.S.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
# Sean M. Boyd -S
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K123922 510(k)
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.