The Tractus TissueMapper Reviewer Application provides two and threedimensional image review, manipulation, and analysis tools to assist users in screening, diagnosis, planning and performing image-guided interventional procedures. The supported imaging modality is Ultrasound (US). Images and data are received from various imaging systems and other sources such as calibrated spatial positioning devices. This device provides the capability to overlay annotations on 2D or 3D medical image displays. These annotations may represent the position of instruments including but not limited to imaging probes or other tracked devices. This device is intended to assist skilled medical professionals in clinical screening and interventions, for anatomical structures where imaging is currently used for visualizing such structures, including head and neck, breast, thoracic, and abdominal applications.
Device Story
Software application for review/analysis of previously recorded ultrasound examinations. Inputs: digital ultrasound images and spatial data from calibrated positioning devices via USB storage. Operation: software maps images to user-identified anatomical landmarks (e.g., nipple, navel) and overlays annotations representing tracked instruments (e.g., imaging probes). Features: 2D/3D display, brightness/contrast/magnification adjustment, video playback, measurement of regions of interest (ROI) on two axes, and electronic report generation. Used in clinical settings by medical professionals for procedure planning and comparison with other imaging modalities. Output: visual display of mapped images/annotations and electronic reports. Benefits: assists in anatomical localization and procedure planning.
Clinical Evidence
Bench testing only. Validation performed using two breast phantoms containing 11 randomly positioned masses. All 11 masses were successfully identified. Registration accuracy tests confirmed performance within specified tolerances, comparable to the predicate device.
Technological Characteristics
Software-based image review system. Operates on Windows 7 PC (min 2.5 GHz, 6GB RAM, 750GB HDD). Displays images via 1920x1080 monitor (≥140 DPI). Uses electromagnetic sensors for spatial registration. Software class B per ANSI/AAMI/IEC 62304:2006. Moderate level of concern.
Indications for Use
Indicated for skilled medical professionals to assist in clinical screening and image-guided interventions for anatomical structures (head, neck, breast, thoracic, abdominal) using ultrasound imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Sentinelle Medical Aegis Navigation Application (K093672)
Submission Summary (Full Text)
{1}------------------------------------------------
#### 510(k) SUMMARY
#### Tractus TissueMapper Reviewer Application
JAN 2 9 2013
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Tractus Corporation
981 El Cajon Drive, Danville, CA 94526
Phone: (503) 758-4964 Facsimile: (866) 397-4732
Contact Person: Eric A. Eggers.
Date Prepared: January 23, 2013
Name of Device and Name/Address of Sponsor
TissueMapper Reviewer Application
Tractus Corporation
981 El Cajon Drive
Danville, CA 94526
Common or Usual Name: Picture Archiving and Communications System
Regulation Number: 21 CFR 892.2050
Product Code: LLZ
Device Class: Class II
Predicate Devices
Sentinelle Medical Aegis Navigation Application (K093672)
Intended Use / Indications for Use:
The Tractus TissueMapper Reviewer Application provides two and threedimensional image review, manipulation, and analysis tools to assist users in screening, diagnosis, planning and performing image-guided interventional procedures. The supported imaging modality is Ultrasound (US). Images and data are received from various imaging systems and other sources such as calibrated spatial positioning devices.
{2}------------------------------------------------
This device provides the capability to overlay annotations on 2D or 3D medical image displays. These annotations may represent the position of instruments including but not limited to imaging probes or other tracked devices.
This device is intended to assist skilled medical professionals in clinical screening and interventions, for anatomical structures where imaging is currently used for visualizing such structures, including head and neck, breast, thoracic, and abdominal applications.
#### Technological Characteristics
The TissueMapper Reviewer Application consists of:
- . the Tractus TissueMapper Reviewer Application software
The TissueMapper Reviewer Application is an electronic image review and reporting software program intended to operate on a Windows Operating System (OS) computer. The device allows the review of previously recorded ultrasound examinations which are performed using standard ultrasound systems and other sources such as calibrated spatial positioning devices, the images of which were recorded digitally.
The images are displayed on a computer monitor. The images may be reviewed individually or as a self-playing sequence (video clip). Certain image functions may be adjusted by the software, such as image brightness, image contrast and image size magnification or reduction. The speed of the video presentation may be adjusted by the software.
The device software presents sequences of mapped ultrasound images to review previously-recorded hand-held ultrasound examinations, which can be transferred to a computer using a media storage device (USB). Each image is mapped anatomically with respect to a user-identified anatomical landmark, such as the nipple or navel. The image mapping information allows anatomic location of useridentified regions of interest for comparison with other imaging techniques or for procedure planning purposes.
The device software allows the user to measure the size, on two user-defined axes, of the region of interest. The device software allows the user to create and store electronic reports on each region of interest.
The TissueMapper Reviewer Application requires two off-the-shelf accessories:
- . Off-the-shelf computer (PC) to run the Tractus TissueMapper Reviewer Application software, which meets the following requirements: o Display
■≥140 DPI ± 5%
{3}------------------------------------------------
· Minimum Resolution of 1920x1080
- o Keyboard and Pointing device
- o Processor speed of ≥ 2.5 GHz
- o Hard drive ≥750 GB
- o RAM≥6GB
- o Ethernet
- o ≥4 USB 2.0 ports or better
- o UL60950 ITE Compliant
- o Operating System: Windows 7
- Off-the-shelf media storage (USB) to move image files from standard legally marketed ultrasound machines to the off-the-shelf computer noted above
Pursuant to 809.92(a)(6), basically, both the applicant's device ("Tractus TissueMapper Reviewer Application") and the predicate device ("Sentinelle Medical Aegis Navigation Application"), present previously recorded images that are spatially mapped with calibrated spatial positioning devices for subsequent review, employ electromagnetic sensors to detect the location of an ultrasound transducer for a registration of transducer location with respect to the region of body scanned with ultrasound, allow for review of ultrasound images collected in the course an ultrasound scan and are software products that run on accessory PC computers. The only technological differences between the Tractus TissueMapper Reviewer Application and its predicate ("Sentinelle Medical Aegis Navigation Application") are: (1) the TissueMapper Reviewer Application is only
used with the imaging modality of Ultrasound; (2) the TissueMapper Reviewer Application is used to identify and map regions of interest for planning subsequent non-invasive diagnostic examinations, such as a diagnostic X-Rav, diagnostic ultrasound, or MRI, where the Sentinelle Medical Aegis Navigation Application may be also used to overlay instrument templates for planning interventional procedures. such as biopsy. Both the applicant's device ("Tractus TissueMapper Reviewer Application") and the predicate device ("Sentinelle Medical Aegis Navigation Application") are software devices that have a Moderate Level of Concern.
The Tractus TissueMapper Reviewer Application software is Safety Class B according to ANSI/AAMI/IEC 62304:2006. Determination of the LOC and Safety Class is the result of risk assessment activities per ISO 14971.
3
{4}------------------------------------------------
### Tractus corporation
### Tissuemapper Reviewer application
## Substantial Equivalence Chart
| | [Tractus TissueMapper Reviewer<br>Application, k123901] | [Sentinelle Medical Aegis<br>Navigation Application,<br>k093672] |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Tractus TissueMapper Application<br>Software provides two and three-<br>dimensional image review, manipulation,<br>and analysis tools to assist users in<br>screening, diagnosis, planning and<br>performing image-guided interventional<br>procedures. The supported<br>imaging modality is Ultrasound<br>(US). Images and data are received from<br>various imaging systems and other<br>sources such as calibrated spatial<br>positioning devices.<br>This device provides the capability to<br>overlay annotations on 2D or 3D medical<br>image displays. These annotations may<br>represent the position of instruments<br>including but not limited<br>to imaging probes or other tracked<br>devices.<br>This device is intended to assist skilled<br>medical professionals in clinical screening<br>and interventions, for anatomical<br>structures where imaging is currently<br>used for visualizing such structures,<br>including head and neck, breast, thoracic,<br>and abdominal applications. | This device provides two and<br>three-dimensional image review,<br>manipulation, and analysis tools to<br>assist users in screening,<br>diagnosis, planning and<br>performing image-guided<br>interventional<br>procedures. Supported imaging<br>modalities include Magnetic<br>Resonance (MR), Ultrasound<br>(US),<br>Single Photon Emission Computed<br>Tomography (SPECT), Computed<br>Tomography (CT), Positron<br>Emission Tomography (PET),<br>Fluoroscopy and Endoscopy.<br>Images and data are received from<br>various imaging systems and other<br>sources such as calibrated spatial<br>positioning devices.<br>This device provides the capability<br>to overlay annotations on 2D or 3D<br>medical image displays.<br>These annotations may represent<br>the position of instruments<br>including but not limited to biopsy<br>needles, guidance wires, imaging<br>probes or other tracked devices.<br>This device is intended to assist<br>skilled medical professionals in<br>clinical screening and<br>interventions, for anatomical<br>structures where imaging is<br>currently used for visualizing such<br>structures, including head and<br>neck, breast, thoracic, and<br>abdominal applications (including<br>pelvis).<br>Lossy compressed mammographic<br>images and digitized film screen<br>images must not be<br>reviewed for primary image |
| | | interpretations. Mammographic<br>images may only be interpreted<br>using a FDA approved monitor that<br>offers at least 5 MPixel resolution<br>and meets other technical<br>specifications reviewed and<br>accepted by the FDA. |
| Indications for Use | Same as above | Same as above |
| User Population | Skilled medical professionals | Skilled medical professionals |
| Technological<br>Characteristics | System, Image Processing,<br>Radiological; Picture Archiving and<br>Communications System | System, Image Processing,<br>Radiological; Picture Archiving<br>and Communications System |
| Primary<br>component(s) | Software: TissueMapper Reviewer<br>Application | Software: Aegis Navigation<br>Application |
| Accessories (if<br>any) | Computer (PC) and storage media<br>(USB) | Not specified in 510(k), assumed<br>to include computer to run<br>software and some type of media<br>or method to transmit files to<br>computer |
| Safety Features | Software risk analysis developed in<br>accordance with ISO 14971 | Not available in 510(k) |
| Software | Yes | Yes |
| 3-D Rendering | Yes | Yes |
| Regions of Interest<br>- Location | Yes | Yes |
| Regions of Interest<br>- Annotation | Yes | Yes |
| Supported Imaging<br>Modalities | Ultrasound | MRI, Ultrasound, SPECT, CT,<br>Fluoroscopy, and Endoscopy. |
| Software level of<br>concern | Moderate level of concern | Moderate level of concern |
| Standards with<br>which the Device<br>Complies | None required | Not available in 510(k) |
.
{5}------------------------------------------------
:
.
.
. . . . .
·
,
1. The state of the state of the states of the states of the states of the states of the states of the states of the states of the states of the states of the states of the s
:
{6}------------------------------------------------
#### Performance Data
Testing for TissueMapper Reviewer Application was performed to ensure that all functional requirements have been met, and that core functions execute as expected. Testing was conducted in-house by trained personnel in a simulated work-environment using phantoms to obtain the functional and accuracy test results. Registration accuracy tests were performed to ensure that the registration and correspondence between ultrasound meets or exceeds specified criteria. The test methodology employed was identical to that of the Aegis predicate device, conducted by targeting locations within a phantom and confirming that the selected target location based on the registration calculation is within the same tolerance range or better than the Aegis predicate device. System Validation Testing of the Tractus TissueMapper Reviewer Application was performed on a two breast phantoms with 11 masses total randomly positioned. All 11 of the masses were successfully identified.
The results of these tests demonstrate that the TissueMapper Reviewer Application validation is within specification.
As such, TissueMapper Reviewer is as safe and effective as the predicate devices and is substantially equivalent to existing products on the market today. The software performs as well as, or better than legally marketed predicate devices.
Tractus' TissueMapper Reviewer Applications for use are drawn from the indications for use of a legally marketed predicate device: Sentinelle's Aegis Navigation Application. Tractus TissueMapper Reviewer Application draws from features of this predicate device and does not provide novel functionality. As such, the features provided by Tractus TissueMapper Reviewer Application do not in themselves raise new concerns of safety or effectiveness.
In all instances, the TissueMapper Reviewer Application functioned as intended and the operation observed was as expected.
#### Substantial Equivalence
The TissueMapper Reviewer Application is as safe and effective as the Sentinelle Medical Aegis Navigation Application. The TissueMapper Reviewer Application has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the TissueMapper Reviewer Application and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the TissueMapper Reviewer is as safe and effective as Sentinelle Medical Aegis Navigation Application Thus, the TissueMapper Reviewer Application is substantially equivalent.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an abstract symbol, possibly representing a human figure or a flame.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 29, 2013
Tractus Corporation %Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K123901
Trade/Device Name: Tractus TissueMapper Reviewer Application Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 14, 2013 Received: January 15, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2-Mr. Job
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. ·
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Sean MyBoyd -S for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): . k123901
Tractus TissueMapper Reviewer Application Device Name:
Indications for Use:
The Tractus TissueMapper Reviewer Application provides two and three-dimensional image review, manipulation, and analysis tools to assist users in screening, diagnosis, planning and performing image-guided interventional procedures. The supported imaging modality is Ultrasound (US). Images and data are received from various imaging systems and other sources such as calibrated spatial positioning devices.
This device provides the capability to overlay annotations on 2D or 3D medical image displays. These annotations may represent the position of instruments including but not limited to imaging probes or other tracked devices.
This device is intended to assist skilled medical professionals in clinical screening and interventions, for anatomical structures where imaging is currently used for visualizing such structures; including head and neck, breast, thoracic, and abdominal applications
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
# Sean M. Boyd -S
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K123901 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.