MEDICVOD, MEDICAST, MEDISHOW, MEDISEARCH, CENTRAL STORAGE SYSTEM - CSS

K123426 · Orpheus Medical , Ltd. · LLZ · Mar 1, 2013 · Radiology

Device Facts

Record IDK123426
Device NameMEDICVOD, MEDICAST, MEDISHOW, MEDISEARCH, CENTRAL STORAGE SYSTEM - CSS
ApplicantOrpheus Medical , Ltd.
Product CodeLLZ · Radiology
Decision DateMar 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Indications for Use

The MedicVOD System is a picture archival and communications systems (PACS) that is intended for image capturing, archiving, displaying and recording of audio/video sequences and patient data during a procedure. The system allows for capture and annotation of the surgical procedure for documentation purposes and for viewing at a later time either locally or on a secure distributed network. The captured audio/video sequences can be broadcasted in real-time with video-conferencing capability during the broadcast. Information captured and stored by the system are for viewing and reference purposes only and are not intended for primary diagnosis. The MedicVOD system is intended as a radiological image processing system that allows for the capture, transfer, display, storage, and digital processing of medical images.

Device Story

MedicVOD System is a PACS for medical procedure documentation; captures, archives, displays, and records audio/video and patient data. Components: mediCAST station (in OR, captures/streams video/images), mediShow (real-time monitoring), mediSearch (archive search/edit), and MedicVOD Server (central storage). Operates by streaming video from imaging devices to server for storage and retrieval; supports real-time broadcast with video-conferencing. Used by medical teams in clinical settings. Output viewed by clinicians for reference/documentation; does not support primary diagnosis. Benefits include improved procedure documentation, remote viewing, and real-time collaboration.

Clinical Evidence

Bench testing only. No clinical data. Testing included hardware verification (first article inspection, functional verification, electrical safety, EMC), software verification/validation (unit, functional, and operational tests), and system-level performance testing (integrated system performance and simulated use in a clinical environment). All tests passed acceptance criteria.

Technological Characteristics

PACS for medical image/video management. Components: Panel PC unit, central storage system. Connectivity: Networked (local/distributed). Standards: DICOM PS 3.3-2004, IEC/EN 60601-1 (2nd ed.), IEC/EN 60601-1-2, IEC 62304, IEC/EN 60601-1-4, ISO 15223-1. Software: Proprietary applications (mediCAST, mediShow, mediSearch).

Indications for Use

Indicated for image capturing, archiving, displaying, and recording of audio/video sequences and patient data during medical procedures for documentation and reference. Not intended for primary diagnosis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Orpheus Medical, Ltd. % Clay Anselmo President Reglera, LLC 11925 West I-70 Frontage Road, Suite 900 Wheat Ridge, CO 80033 JUL 2 7 2015 Re: K132426 Trade/Device Name: Medic VOD System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated (Date on orig SE Itr): January 25, 2013 Received (Date on orig SE ltr): January 31, 2013 Dear Clay Anselmo, This letter corrects our substantially equivalent letter of March 1, 2013. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {2}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K123426 Device Name: MedicVOD System Indications for Use: The MedicVOD System is a picture archival and communications systems (PACS) that is intended for image capturing, archiving, displaying and recording of audio/video sequences and patient data during a procedure. The system allows for capture and annotation of the surgical procedure for documentation purposes and for viewing at a later time either locally or on a secure distributed network. The captured audio/video sequences can be broadcasted in realtime with video-conferencing capability during the broadcast. Information captured and stored by the system are for viewing and reference purposes only and are not intended for primary diagnosis. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/2/Picture/10 description: The image shows the name "Long Chen-A" in a stylized font. To the right of the name is a digital signature block. The signature block includes the text "Digitally signed by Long H. Chen-A" and a date stamp of "2013.02.28 15:28:41-05'00". (Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number_ K123426 Page 1 of 1 {4}------------------------------------------------ K123426 PAGE 1 of 5 # 510(k) Summary: MedicVOD System # MAR 0 1 2013 #### INTRODUCTION 1 - This document contains the 510(k) Summary for the MedicVOD system. The content of 1.1 this document is based on the requirements of 21 CFR 807,92 ## 2 APPLICANT NAME AND ADDRESS - 2.1 Name: Orpheus Medical Ltd. - 2.2 Address: 5 Haetgar St. POB 15140, MATAM Industrial Zone HAIFA 31905, Israel - 2.3 Official Contact: David Rainis, CEO - 2.4 Summary Preparation Date: October 29, 2012 #### 3 DEVICE NAME AND CLASSIFICATION - Name of Device: MedicVOD System 3.1 - 3.2 Common Name: Picture Archiving and Communication System - 3.3 Classification Name: Picture Archiving and Communication System - 3.4 Classification: II - 3.5 Classification Panel: Radiology - 3.6 Regulation Number: 21 CFR 892.2050, 21 CFR 876.1500 - 3.7 Product Code: LLZ, KOG #### 4 PREDICATE DEVICE: - 4.1 AIDA with DICOM and HL7 interface, by Storz, cleared under K043324 - 4.2 660 HD Image Management System, by Smith & Nephew, cleared under K060777 - 4.3 SwitchPoint Infinity®3 (SPI3) Control System, by Stryker Communications, cleared under K100852 #### DEVICE DESCRIPTION: 5 - 5.1 The MedicVOD System is a video system enabling medical teams to document. record and broadcast live medical procedures in real time. Its four components are: - 5.1.1 The mediCAST station, attached to the imaging device in an operating theater - 5.1.2 The mediShow application, a real-time monitoring application, displaying sessions from all currently active mediCAST stations - 5.1.3 The mediSearch video archive search and edit application - 5.1.4 The MedicVOD Server which is a central storage system (CSS) for archival of all mediCAST sessions - 5.2 The MedicVOD key features include: - 5.2.1 Digital recording of medical procedures streamed directly to on-site server for storage and retrieval. - 5.2.2 Live broadcast of a medical procedure to an unlimited number of viewers. {5}------------------------------------------------ K123426 PAGE 2 OF 5 - 5.2.3 Archiving of video assets to a controlled storage resource, - 5.2.4 Data fetching with dynamic categories which may be used by other 300 party applications for purposes such as statistical analysis of data usage and/or user operation and report production, and - 5.2.5 Ability to view/review videos from any mediShow location at any time - 5.3 The mediCAST station is capable of capturing and broadcasting a video stream in real time. MediCAST station captures and stores video and still images locally, and in parallel, uploads the capture images and video streams to the MedicVOD server. The mediCAST station uploads the videos and images when there is communication with the MedicVOD server. - 5.4 The mediSearch application provides the ability to search within the MedicVOD archive. The mediSearch application allows users to perform searches for video recordings within a captured mediCAST session. Each session could be a combination of several recordings of various kinds. For example, a session could be a combination of several video recordings, several still images (snapshot taken) and several DICOM images that have been stored. MediSearch provides the ability to view the assets which were found, and provides certain editing capabilities. - The mediShow application automatically connects into the MedicVOD server and 5.5 displays currently active sessions or procedures captured by any mediCAST station on the local network. #### INDICATIONS FOR USE: 6 - The MedicVOD System is a picture archival and communications systems (PACS) that is 6.1 intended for image capturing, archiving, displaying and recording of audio/video sequences and patient data during a procedure. The system allows for capture and annotation of the surgical procedure for documentation purposes and for viewing at a later time either locally or on a secure distributed network. The captured audio/video sequences can be broadcasted in real-time with video-conferencing capability during the broadcast. Information captured and stored by the system are for viewing and reference purposes only and are not intended for primary diagnosis. #### INTENDED USE: 7 - 7.1 The MedicVOD system is intended as a radiological image processing system that allows for the capture, transfer, display, storage, and digital processing of medical images. #### 8 PERFORMANCE STANDARDS: - 8.1 This 510(k) submission was written in accordance with the FDA guidance document "Guidance for the Submission of Premarket Notifications for Medical Image Management Devices- July 27, 2000", "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005", and "General Principles for Software Validation; January 2002". The design of the MedicVOD System conforms to the following voluntary standards: - 8.1.1 Digital Imaging and Communications in Medicine (DICOM): PS 3.3-2004, National Electrical Manufacturers Association. - 8.1.2 IEC/EN 60601-1-4:1997, General Requirements for Programmable Electrical Medical System. - 8.1.3 IEC 62304:2006, Medical Device Software Software Lifecycle Processes. {6}------------------------------------------------ K123426. PAGE 3 of 5 - 8.1.4 IEC/EN-60601-1: 1988 (2nd ed.), Medical Electrical Equipment; Part 1: General Requirements for Safety. Second edition , including amendments #1(1993), #2(1995) and #3(1996). - 8.1.5 IEC/EN 60601-1-2:2001, Medical Electrical Equipment; Part 1-2: Collateral Standard: Electromagnetic Compatibility- Requirements and Tests - 8.1.6 ISO 15223-1:2012, Medical devices Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements #### TEST DATA: 9 - The MedicVOD System has been subjected to extensive safety and performance testing, 9.1 and verification / validation before release. Final testing of the MedicVOD System included various performance tests and software validation tests designed to ensure that the device meet all of its functional specifications. - 9.2 Tests have been performed to ensure the device complies with industry and safety standards and to demonstrate substantial equivalence to the identified predicate devices. All performance testing showed results consistent with the intended uses and performance specifications of the device. - ਰ.3 The following list and table below summarizes the testing performed on the device; - 9.3.1 Hardware Verification / Validation - 9.3.1.1 Component First Article Inspection - 9.3.1.2 Component Functional Verification - 9.3.1.3 Electrical Safety - 9.3.1.4 Electromagnetic Compatibility - 9.3.2 Software Verification / Validation - 9.3.2.1 Software unit test - 9.3.2.2 Software verification tests - 9.3.2.3 Software validation tests - 9.3.3 System Level Performance - 9.3.3.1 Integrated System Performance Testing - 9.3.3.2 Simulated Use | Verification / Validation<br>Description | Unit or<br>Configuration<br>Tested | Method | Acceptance Criteria | Results | |------------------------------------------|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------| | Component First<br>Article Inspection | Panel PC unit<br>CSS | Visual inspection,<br>installation of proprietary<br>software and testing<br>using designated test<br>software that tests each<br>component and function | Passing results<br>defined within the<br>test software | Passed all<br>applicable tests | | Verification / Validation<br>Description | Unit or<br>Configuration<br>Tested | Method | Acceptance Criteria | Results | | Component Functional<br>Verification | Panel PC unit<br>CSS | Visual inspection,<br>installation of proprietary<br>software and testing<br>using designated test<br>software that tests each<br>component and function | Passing results<br>defined within the<br>test software | Passed all<br>applicable tests | | Electrical safety | Panel PC unit<br>CSS | Applicable tests in a<br>certified lab per IEC<br>60601-1<br>Applicable tests in a<br>certified lab per IEC<br>60950 | As required by the<br>applicable<br>standards | Passed all<br>applicable tests | | Electromagnetic<br>compatibility | Panel PC unit<br>CSS | Applicable tests in a<br>certified lab per IEC<br>60601-1-2<br>Applicable tests in a<br>certified lab per IEC<br>60950 | As required by the<br>applicable<br>standards | Passed all<br>applicable tests | | Software unit test | mediCast<br>mediShow<br>mediSearch<br>CSS | Bench verification of unit<br>functionality against its<br>requirements using<br>simulated input signals | Compliance with<br>defined<br>requirements,<br>absence of bugs | Passed all tests | | Software verification<br>test | mediCast<br>mediShow<br>mediSearch<br>CSS | Complete bench<br>functional and operational<br>tests derived from the<br>software requirements<br>specifications and<br>additional tests for<br>regression, destructive<br>tests, system debugging,<br>white-box and black box<br>tests | Compliance with<br>defined<br>requirements,<br>absence of bugs | Passed all tests | | Software validation<br>test | mediCast<br>mediShow<br>mediSearch<br>CSS | Complete bench<br>functional and operational<br>tests derived from the<br>software requirements<br>specifications | As defined in the<br>software test<br>description (see<br>section 16) | Passed all tests | | Verification / Validation<br>Description | Unit or<br>Configuration<br>Tested | Method | Acceptance Criteria | Results | | Integrated System<br>Performance Testing | MedicVOD<br>integrated<br>system | Functional bench testing<br>using a designated test<br>software that tests each<br>and every system<br>components and function,<br>for each system prior to<br>its release | Passing all<br>defined tests<br>defined by the test<br>protocol and<br>embedded within<br>the test software | Passed all tests | | Simulated use | MedicVOD<br>integrated<br>system | Functional testing in a<br>clinical environment by<br>potential user/s | Full functionality<br>and satisfactory<br>usability per a<br>defined user<br>questionnaire | Passed all tests | {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows handwritten text and numbers. The text at the top reads "K123426". Below that, the text reads "Page 4 of 5". The numbers are written in a casual, handwritten style. · . {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the text "K123426 page 5 of 5". The text appears to be handwritten. The words "page" and "of" are in lowercase letters, while the other text is in uppercase letters and numbers. # 10 SUBSTANTIAL EQUIVALENCE: - 10.1 The intended use of the MedicVOD System is the same as identified predicate devices. - 10.2 The operating principles and technological characteristics of the MedicVOD System are the same or substantially equivalent to those of the AIDA System, 660 HD System and also to those of the SP13 Control System. The algorithms used by the MedicVOD System are similar in principle-of-operation and specifications to the algorithms used by the identified predicate devices. - 10.3 The physical and performance specifications of the MedicVOD System are the same or similar to the specifications of the AIDA System, 660 HD System and also to those of the SPI3 Control System. ## 11 CONCLUSION: - 11.1 The MedicVOD System is substantially equivalent with respect to the indications for use. technological characteristics and performance characteristics to the identified legally marketed predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...