K121076 · Toshiba America Medical Systems, In.C · LLZ · Oct 9, 2012 · Radiology
Device Facts
Record ID
K121076
Device Name
ULTRAEXTEND FX, ULTRASOUND WORKSTATION PACKAGE
Applicant
Toshiba America Medical Systems, In.C
Product Code
LLZ · Radiology
Decision Date
Oct 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The UltraExtend FX (TUW-U001S) is designed to allow the user to observe images and perform analysis using the examination data acquired with specified diagnostic ultrasound systems Aplio 500, Aplio 400 and Aplio 300.
Device Story
UltraExtend FX is post-processing software installed on a general-purpose PC; enables loading and analysis of diagnostic ultrasound data acquired from Aplio 300, 400, and 500 systems. Provides seamless image reading environment; implements functionality equivalent to diagnostic ultrasound systems. Features include basic measurement, cardiac measurement, vascular measurement, stress echo, 4D review, Elastography, Auto-IMT, and 2D wall motion tracking. Used by clinicians in a reading environment to observe images and perform analysis; output facilitates clinical decision-making by extending diagnostic capabilities from the ultrasound system to a PC workstation.
Clinical Evidence
Bench testing only. Verification and validation tests were conducted to confirm device safety and effectiveness. No clinical data provided.
Technological Characteristics
PC-based software package. Implements functionality equivalent to diagnostic ultrasound systems. Supports 2D, M-mode, spectral Doppler, 4D review, Elastography, Auto-IMT, and 2D wall motion tracking. Software development follows IEC 62304 standards.
Indications for Use
Indicated for users to observe and analyze diagnostic ultrasound examination data acquired from Aplio 300, 400, and 500 systems. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
UltraExtend (Model USWS-900A)
Reference Devices
Aplio 500/400/300 Diagnostic Ultrasound System (K110870)
Submission Summary (Full Text)
{1}------------------------------------------------
K121076
Toshiba America Medical Syst Pre-Market Notification 510(k) UltraExtend FX, TUW-U001S
#### 510(k) SUMMARY AND EFFECTIVENESS
OCT
- 1. DEVICE NAME:
System, Image Processing, Radiological Generic Name: Ultrasound Workstation Package, UltraExtend FX Trade/ Proprietary Name: TUW-U001S Model Name:
#### 2. ESTABLISHMENT REGISTRATION: 2020563
# 3. CONTACT PERSON AND U.S AGENT INFORMATION:
Charlemagne Chua Contact Person: Manager, Regulatory Affairs (714) 730-5000
U.S. Agent Name: Paul Biggins Director, Regulatory Affairs (714) 730-5000 Establishment Name and Address: Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA. 92780
Toshiba Medical Systems Corporation 4. MANUFACTURING SITE: 1385 Shimoishigami Otawara-shi, Tochigi 324-8550 Japan
5. DATE OF SUBMISSION: April 3, 2012 (revised August 3, 2012)
#### 6. DEVICE DESCRIPTION:
UltraExtend FX is a software package that can be installed in a general-purpose personal computer (PC) to enable data acquired from Aplio diagnostic ultrasound system (Aplio 300, Aplio 400 and Aplio 500), to be loaded onto the PC for image processing with other application software product. UltraExtend FX is a post-processing software that implements functionality and operability equivalent to that of the diagnostic ultrasound system, providing a seamless image reading environment from examination using the diagnostic ultrasound system to diagnosis using the PC.
9 202
{2}------------------------------------------------
#### 7. INDICATION FOR USE
The UltraExtend FX (TUW-U001S) is designed to allow the use to observe images and perform analysis using the examination data acquired with specified diagnostic ultrasound systems Aplio 300, Aplio 400 and Aplio 500.
#### DESIGN CHANGE 8.
The change allows the ability for the UltraExtend FX Workstation to process images from the Aplio Ultrasound System.
### 9. SUMMARY OF DESIGN CONTROL ACTIVITIES
PS Risk List for software is attached. Risk Analysis and Design Reviews were conducted. Verification and validations tests were conducted on the subject device through bench testing to confirm device safety and effectiveness. IEC 62304 processes was implemented in the development of the subject device. A declaration of conformity with design controls is included in this submission.
#### 10. TRUTHFUL AND ACCURACY CERTIFICATION
A certification of the truthfulness and accuracy of the UltraExtend FX Workstation described in this submission is provided in this submission.
#### 11. SUBSTANTIAL EQUIVALENCE
UltraExtend (Predicate) is a software that can be installed in a general-purpose personal computer (PC) to enable data acquired using diagnostic ultrasound systems (Aplio Artida SSH-880CV, Aplio XG SSA-790A, or Xario XG SSA-680A) to be loaded onto the PC for various application software programs. This software (UltraExtend) implements functions and operability of the basic measurement, cardiac application measurements using 2D, M and spectral Doppler, and report generation.
UltraExtend FX (Subiect) is designed to allow the user to observe images and perform analysis using the examination data acquired from an Aplio diagnostic ultrasound system (Aplio 300, Aplio 400 and Aplio 500), It has the application of basic measurement, cardiac measurement, vasicular measurement, stress echo, 4D review, Elastography, Auto-IMT and 2D wall motion tracking.
UltraExtend (Model USWS-900A) and UltraExtend FX (Model TUW-U001S) Ultrasound Workstation Package are capable of acquiring images from a specific ultrasound diagnostic systems and perform analysis. Their applications are not the same but the intended use of the software is the same. The UltraExtend FX has been found substantially equivalent to the previously cleared UltraExtend Workstation (Model USWS-900A) and Aplio 500/400/300 Diagnostic Ultrasound System (K110870) referenced in this submission.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
OCT
Toshiba Medical Systems Corporation, Japan % Ms. Charlemagne Chua Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K121076
Trade/Device Name: UltraExtend FX, TUW-U0001S, v2.02 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, IYN, and IYL Dated: September 28, 2012 Received: October 1, 2012
#### Dear Ms. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
9
_012
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K121076
Device Name:
UltraExtend FX, TUW-U001S, v2.02
Indications for Use:
The UltraExtend FX (TUW-U001S) is designed to allow the user to observe images and perform analysis using the examination data acquired with specified diagnostic ultrasound systems Aplio 500, Aplio 400 and Aplio 300.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Signature
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K121076
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.