OrthoCAD is an option that provides the morphometry of the lumbo-sacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
Device Story
OrthoCAD is a software option for the Esaote G-scan MR system. It processes sagittal FSE T2 MR images of the lumbosacral spine (L1-S1). Users perform manual or semi-automatic (wizard) segmentation of vertebral bodies and the spinal canal to construct 3D models. The system calculates clinical parameters including vertebral wedging, listhesis index, intervertebral translation/angles, vertebral collapse index, spinal curvature, spinal canal thickness/section, and foramen area. It supports endoscopic virtual navigation within the segmented canal and allows comparison of images acquired in clinostatic and orthostatic positions. Used by physicians in clinical settings to assist in diagnosis by providing quantitative geometric data.
Clinical Evidence
Bench testing only. Testing included comparison between manual and semi-automated segmentation of lumbar/sacral vertebrae and spinal canal; evaluation of variability, repeatability, and reproducibility of measurements; and validation of software correctness in measuring MRI images of the lumbar spine.
Technological Characteristics
Software-based image processing package. Operates on MR images acquired via G-scan (K111803). Features include semi-automatic segmentation, 3D model construction, and geometric measurement tools. Connectivity via database integration for patient management. No change to fundamental scientific technology of the host MR system.
Indications for Use
Indicated for morphometric analysis of the lumbosacral spine (L1-S1) using MR images to assist trained physicians in diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K120288
SEP
1
1 9 2012
510(k) Summary OrthoCAD software option for G-scan Esaote, S.p.A.
# 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(a).
# Submitter Information
Donatella Ragghianti Esaote S.p.A. Via Siffredi 58 Genova 16153 Italy
Contact Person: Allison Scott, RAC (317) 569-9500 x 106 Phone: Facsimile: (317) 569-9520 ascott@ansongroup.com
| Date: | August 30, 2012 |
|-------------------------|---------------------------------------------|
| Trade Name: | OrthoCAD software option for G-scan |
| Common Name: | System, Image Processing, Radiological |
| Classification Name(s): | Picture archiving and communications system |
| Classification Number: | LLZ |
## Predicate Device(s)
| Trade name | Common name | Class | Product code | Manufacturer | K number |
|---------------|--------------------------------------------------|-------|--------------|-----------------|----------|
| G-scan | System, nuclear<br>magnetic resonance<br>imaging | II | LNH | Esaote S.p.A. | K111803 |
| SpineAnalyzer | Spine Analysis<br>Software | II | LLZ | OPTASIA MEDICAL | K103475 |
{2}------------------------------------------------
510(k) Summarv OrthoCAD software option for G-scan Esaote, S.p.A.
## Device Description
The OrthoCAD software option is a software package intended to be used with Esaote Gscan system cleared via K111803. OrthoCAD provides the morphometry of the lumbosacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
G-scan is a Magnetic Resonance (MR) system, which produces images of the internal structures of the patient's limbs and joints.
The OrthoCAD system allows you to visualize, analyse and compare Magnetic Resonance images. The system is connected to a database that enables the physician to keep track of all the patients examined over time.
The images are acquired by running FSE T2 Rel sequences on the G-scan of the lumbosacral tract of the vertebral spine, in the sagittal plane, and are transferred to the OrthoCAD database following acquisition.
When the MR images are stored on the OrthoCAD database, the user can proceed with a manual or semi-automatic (wizard) segmentation of the vertebral bodies (from L1 up to S1) and of the spinal canal. During the segmentation of the anatomical elements, 3D models are constructed based on the segmented structures.
When the segmentation procedure has been terminated, the user can proceed with the evaluation of the following significant clinical parameters:
- . � Vertebral wedging
- Listhesis index .
- Intervertebral translation index .
- Intervertebral angles .
- Vertebral collapse index .
- . Spinal curvature
- . Spinal canal thickness
- . Spinal canal section
- Foramen area . .
Following this process, the endoscopic virtual navigation within the segmented spinal canal is enabled.
{3}------------------------------------------------
Finally, if the user has worked on MR images acquired both in the clinostatic and orthostatic mode, the measures calculated and the virtual navigation of the two the positions can be compared, and a report containing all the information is produced.
OrthoCAD is made up of six environments:
- . Patient Management: contains the functions required for the display and management of patients stored in the database associated with the system.
- Home: keeps track of the procedures executed overtime for the selected patient . (analyses present, status of examinations associated with the various analyses, etc.).
- . Segmentation: carries out the functions used for the segmentation and those related to the construction of 3D models of anatomic elements
- Measurements: includes all tools required to measure the clinical parameters . used for the analysis of the currently selected exam.
- Navigation: enables endoscopic virtual navigation within the segmented . anatomical structures by means of the definition of anatomic points, in order to construct one (or more) navigation routes.
- . Comparison: enables the comparison of two different examinations within the same analysis or within different analyses provided they are the same type. This environment enables:
- . The simultaneous display, or superimposed display when required, of anatomical elements which belong to the two volumes being compared.
- 미 The simultaneous display of the different measurements, with an indication of the main differences between these values.
### Intended Use(s)
OrthoCAD is an option that provides the morphometry of the lumbo-sacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
#### Technological Characteristics
The addition of the OrthoCAD Software Option, reflected in this 510(k), does not alter the fundamental scientific technology of the G-scan system, cleared via K111803.
#### Performance Data
Non-clinical testing of the G-scan system with the addition of the OrthoCAD Software Option demonstrated that it met performance requirements and is as safe and effective as the predicate devices.
{4}------------------------------------------------
Tests performed:
- Comparison between the manual and semi-automated segmentation on lumbar ● and first sacral vertebrae
- . Comparison between the manual and semi-automated segmentation on spinal canal
- Comparison of the manual and semi-automated measurements and evaluation of . variability, repeatability and reproducibility
4
- Validation of new software OrthoCAD in its correctness in measuring MRI . images f the Lumbar spine
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is facing to the right. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Esaote S.p.A. % Ms. Allison Scott RAC The Anson Group 9001 Wesleyan Road, Suite 200 INDIANAPOLIS IN 46268
Re: K120288
Trade/Device Name: OrthoCAD Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 30, 2012 Received: August 31, 2012
### Dear Ms. Scott:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 910(t) presidentially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the chorosury to regars in medical Device Amendments, or to conninered print to May 20, 1978, in accordance with the provisions of the Federal Food, DNA devices mat have occh recassinou in asses approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may, mercere, market the act include requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (see above) into enass fr (or your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination costs of the at Please be advised that I DA 3 issumed on a second is with other requirements of the Act
that FDA has made a determination that your device complies with other requirements of that IDA has made a determination and regulations administered by other Federal agencies. You must of any I edelal statutes and regularions and including, but not limited to: registration and listing (21 Comply with an the rece requirems, 801 and 809); medical device reporting (reporting of
SEP 19 2012
{6}------------------------------------------------
### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device felated du rerse evality systems (QS) regulation (21 CFR Part 820). This letter requirements as bet form in the quarny device as described in your Section 510(k) premarket with anow you to began maing of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you don't opening an receive of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket s 150. Theose not 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may obtain other gefacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
K120288 510(k) Number (if known):
Device Name: OrthoCAD
Indications for Use:
OrthoCAD is an option that provides the morphometry of the lumbo-sacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mundoff
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safe 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.