← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K113701

# VISIA NEURO (K113701)

_MeVis Medical Solutions AG · LLZ · Feb 16, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113701

## Device Facts

- **Applicant:** MeVis Medical Solutions AG
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Feb 16, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Visia™ Neuro is a medical image processing software application intended for the visualization of images from various sources such as Magnetic Resonance Imaging systems or from image archives. The system provides viewing, quantification, manipulation, and printing of medical images. Visia™ Neuro provides both analysis and viewing capabilities for anatomical and physiologic/functional imaging datasets, including blood oxygen dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic review, and vessel visualization. Data can be visualized in both 2D and 3D views.

## Device Story

Visia™ Neuro is a medical image processing software application for MRI data. It receives images from MRI systems or archives via network. It provides tools for 2D/3D visualization, quantification, and manipulation of anatomical and functional datasets, including BOLD fMRI, diffusion, fiber tracking, dynamic studies, and vessel visualization. The software runs on off-the-shelf PCs as a standalone workstation or in a client-server configuration. Clinicians use the software to annotate, measure, and generate reports. The output assists physicians in identifying susceptibility changes, visualizing water diffusion, tracking white matter, analyzing contrast changes over time, and visualizing vascular structures. It does not perform automated diagnosis; it supports clinical decision-making by providing processed images for physician interpretation.

## Clinical Evidence

No clinical data. Bench testing only. Validation testing evaluated all input/output functions and software actions in all operational modes. Results demonstrated that predetermined acceptance criteria were met.

## Technological Characteristics

Software-only device; no patient contact. Operates on off-the-shelf PC hardware. Supports standalone or client-server network configurations. Provides 2D/3D image processing, registration, and visualization modules. Does not control life-sustaining equipment. Diagnosis performed by human clinicians.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- DC Neuro ([K081262](/device/K081262.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for MeVis MEDICAL SOLUTIONS. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent a medical company.

FEB 1 6 2012

KII3701 P/4

# 510(k) Summary

| Submitted By:                         | MeVis Medical Solutions AG<br>Universitaetsallee 29<br>28359 Bremen, Germany                                                                                                            |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Name:                | MeVis Medical Solutions AG                                                                                                                                                              |
| Establishment<br>Registration Number: | Applied for and awaiting assignment by FDA                                                                                                                                              |
| Contact Person:                       | Thomas E. Tynes<br>President & CEO<br>MeVis Medical Solutions, Inc.<br>N27 W24075 Paul Court - Suite 100<br>Pewaukee, WI 53072<br>USA<br>Phone: +1-262-691-9530<br>Fax: +1-262-691-9531 |
| Date Prepared:                        | 12/14/2011                                                                                                                                                                              |
| Trade Name:                           | Visia™ Neuro                                                                                                                                                                            |
| Common Name:                          | Medical Image Processing Software                                                                                                                                                       |
| Classification Name:                  | Image Processing System                                                                                                                                                                 |
| Classification<br>Regulation Number:  | 892.2050                                                                                                                                                                                |
| Class:                                | II                                                                                                                                                                                      |
| Panel:                                | Radiology                                                                                                                                                                               |
| Product Code:                         | LLZ                                                                                                                                                                                     |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for MeVis MEDICAL SOLUTIONS. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent a medical company.

## Device Description

Visia™ Neuro is a medical image processing software application intended for the visualization of images from various sources such as Magnetic Resonance Imaging systems or from image archives. The system provides viewing, quantification, manipulation, and printing of medical images.

Visia™ Neuro integrates within typical clinical workflow patterns through receiving and transferring medical images over a computer network. The software can be loaded on a standard off-the-shelf personal computer (PC) and can operate as a stand-alone workstation or in a distributed server-client configuration across a computer network.

The software provides functionality for processing and analyzing both anatomical and physiologic/functional imaging datasets. Specifically, the software includes user defined processing modules for image registration, blood oxygen level dependent (BOLD) functional magnetic resonance imaging (fMRI), diffusion imaging, fiber tracking, dynamic imaging, and vessel imaging. Processed images are stored as separate files from the original data such that the original data is preserved.

Images may be displayed based on physician preferences using configurable layouts, or hangings. Visia™ Neuro provides the clinician with a broad set of viewing and analysis tools in both 2D and 3D. The software includes tools to annotate, measure, and output selected image views or user defined reports.

## Intended Use

Visia ™ Neuro is a medical image processing software application intended for the visualization of images from various sources such as Magnetic Resonance Imaging systems or from image archives. The system provides viewing, quantification, manipulation, and printing of medical images. Visia™ Neuro provides both analysis and viewing capabilities for anatomical and physiologic/functional imaging datasets, including blood oxygen dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic review, and vessel visualization. Data can be visualized in both 2D and 3D views.

## BOLD fMRI Review

The BOLD fMRI feature is useful in identifying small susceptibility changes arising from neuronal activity during performance of a specific task.

## Diffusion Review

The diffusion review feature is intended for visualization and analysis of the diffusion of water molecules through brain tissue.

## Fiber Tracking Review

The fiber tracking feature uses the directional portion of the diffusion vector to track and visualize white matter structures within the brain.

> Visia™ Neuro 510(k) 510(k) Summary - 12/14/2011 Page 2 of 4

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo consists of the word "MeVis" in large, bold, sans-serif font, with the letters slightly overlapping. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in a smaller, sans-serif font.

## Dynamic Review

Dynamic review feature is intended for visualization and analysis of MRI dynamic studies, showing changes in contrast over time, where such techniques are useful or necessary.

#### Vessel Visualization

The vessel feature is used to identify and visualize the vascular structures of the brain.

#### 3D Visualization

The 3D visualization feature allows image data to be reconstructed as 3D objects that are visualized and manipulated on a 2D screen.

#### Predicate Device Information

| Product      | Predicate Device Name | Predicate 510(k)<br>Submission Reference |
|--------------|-----------------------|------------------------------------------|
| Visia™ Neuro | DC Neuro              | K081262                                  |

## Safety and Effectiveness

The Visia™ Neuro labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via MeVis Medical Solution AG's Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via software development and verification & validation testing.

## Nonclinical Testing and Performance Information

The complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the software in each operational mode and followed the process documented in the Validation Test Plan.

Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification activities and that the results demonstrated that the predetermined acceptance criteria were met.

## Technological Characteristics

Visia™ Neuro is a software device that does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians.

A physician, providing ample opportunity for competent human interprets images and information being displayed and printed.

> Visia™ Neuro 510(k) 510(k) Summary - 12/14/2011 Page 3 of 4

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MeVis MEDICAL SOLUTIONS. The logo is in black and white. The word "MeVis" is in large, bold letters, and the words "MEDICAL SOLUTIONS" are in smaller letters below it.

K113701

P 4/4

The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.

## Conclusion

The 510(k) Pre-Market Notification for Visia™ Neuro contains adequate information, data, and nonclinical test results to enable FDA – CDRH to determine substantial equivalence to the predicate device. MeVis Medical Solutions has determined that its device, Visia™ Neuro, is substantially equivalent to the identified predicate device listed above. A comparison with the legally marketed predicate device indicates that it is substantially equivalent to this device, and that it does not raise any new safety or efficacy concerns. Nonclinical tests demonstrate that the device is safe, effective, and is substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved, overlapping shapes.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

MeVis Medical Solutions AG % Mr. Thomas E. Tynes President and CEO MeVis Medical Solutions, Inc. N27 W24075 Paul Court, Suite 100 PEWAUKEE WI 28359

FEB 1 6 2012

Re: K113701

Trade/Device Name: Visia™ Neuro Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 14, 2011 Received: December 16, 2011

Dear Mr. Tynes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{5}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and consists of the word "MeVis" in large, bold letters. Below the word "MeVis" are the words "MEDICAL SOLUTIONS" in smaller, less bold letters.

# Indications for Use Statement

510(k) Number (if known): K113701

Device Name: Visia™ Neuro

Indications for Use:

Visia™ Neuro is a medical image processing software application intended for the visualization of images from various sources such as Magnetic Resonance Imaging systems or from image archives. The system provides viewing, quantification, manipulation, and printing of medical images. Visia™ Neuro provides both analysis and viewing capabilities for anatomical and physiologic/functional imaging datasets, including blood oxygen dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic review, and vessel visualization. Data can be visualized in both 2D and 3D views.

#### BOLD fMRi Review

The BOLD MRI feature is useful in identifying small susceptibility changes arising from neuronal activity during performance of a specific task.

#### Diffusion Review

The diffusion review feature is intended for visualization and analysis of the diffusion of water molecules through brain tissue.

#### Fiber Tracking Review

The fiber tracking feature uses the directional portion of the diffusion vector to track and visualize white matter structures within the brain.

#### Dynamic Review

Dynamic review feature is intended for visualization and analysis of MRI dynamic studies, showing changes in contrast over time, where such techniques are useful or necessary.

#### Vessel Visualization

The vessel feature is used to identify and visualize the vascular structures of the brain.

#### 3D Visualization

The 3D visualization feature allows image data to be reconstructed as 3D objects that are visualized and manipulated on a 2D screen.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary Spoth

Division Sign Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)

Visia™ Neuro 510(k) Indications for Use Statement - 12/14/2011 Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113701](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113701)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
