← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K113404

# DENTAL 3D PLANNING SOFTWARE (K113404)

_Symbyo Dental, LLC · LLZ · Dec 6, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113404

## Device Facts

- **Applicant:** Symbyo Dental, LLC
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Dec 6, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Intended Use

IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.

## Device Story

IPS is a dental 3D planning software; inputs DICOM-compliant radiological data (CT/Cone beam CT); transforms data into 3D models of oral maxillofacial region; provides visualization, segmentation, and measurement tools (distance, angle, bone density); used by dentists and dental lab technicians in clinical settings; output compatible with CAD and rapid prototyping machines; aids in surgical treatment planning and implant placement; facilitates clinical decision-making through anatomical visualization and measurement.

## Clinical Evidence

No clinical data. Bench testing only; validation performed to demonstrate DICOM conformance, including importing, sorting, and grouping of DICOM data.

## Technological Characteristics

Pure software device; PC-based; DICOM standard compliant; compatible with Windows XP, Vista, 7, and Intel-based Mac (booted in Windows); outputs compatible with CAD and rapid prototyping systems.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Simplant system ([K033849](/device/K033849.md))
- Sicat implant ([K090119](/device/K090119.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510k Summary

DEC - 6 2011

Device Owner Address:

Telephone: Fax:

Official Correspondent

Telephone: Email:

Symbvo Dental, LLC, 15310 Amberly Drive suite#250 Tampa FL 33647 (877) 479-6296 (866) 697-8559

Amberry Drive suite#250
FL 33647
479-6296
697-8559

Parul Chansoria 87 Hulme Ct Apt. 216 Stanford CA 94305 877-479-6296 parul@symbyo.com

## Device Name and Classification

| Trade Name/Product Name:            | IPS                                                                                                     | Areas of<br>Comparison                         | Simplant system<br>K033849                                                                                                                       | Sicat implant<br>K090119                                                                                                                         | IPS                                                                                                                                                |
|-------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:                  | Dental 3D Planning Software                                                                             | Description                                    | Pure software device                                                                                                                             | Pure software device                                                                                                                             | Pure software device                                                                                                                               |
| Classification Name:                | System, Image Processing, Radiological; 21CFR892.2050                                                   | Input source                                   | CT scanners, MRI<br>machines                                                                                                                     | CT scanners, DVT<br>scanners                                                                                                                     | Cone beam CT<br>machines, CT<br>scanners                                                                                                           |
| Product Code:                       | LLZ                                                                                                     | Data conformance                               | DICOM                                                                                                                                            | DICOM                                                                                                                                            | DICOM                                                                                                                                              |
| Regulation Number:                  | Picture Archiving and Communication System, 21CFR892.2050                                               | Regulation<br>description                      | Picture archiving and<br>communication system                                                                                                    | Picture archiving and<br>communication system                                                                                                    | Picture archiving and<br>communication system                                                                                                      |
| Panel:                              | Radiology                                                                                               | Intended use                                   | Visualization and<br>planning simulation that<br>aids the user in the<br>placement of dental<br>implants and planning<br>for surgical treatments | Visualization and planning<br>simulation that aids the<br>user in the placement of<br>dental implants and<br>planning for surgical<br>treatments | Visualization and<br>planning simulation<br>that aids the user in the<br>placement of dental<br>implants and planning<br>for surgical treatments   |
| Class                               | 2                                                                                                       | Intended user                                  | Well trained dentist \<br>dental technicians                                                                                                     | Well trained dentist                                                                                                                             | Dentists, dental<br>technician                                                                                                                     |
| Substantial Equivalence claimed to: | MATERIALISE N.V. Simplant system (K033849)                                                              | Main tools                                     | Visualization, implant<br>placement,<br>measurement of<br>distance, angle<br>measurement, density<br>determination,<br>segmentation tool         | Visualization, implant<br>placement, measurement<br>of distance, angle<br>measurement, density<br>determination                                  | Visualization, implant<br>and nerve placement,<br>measurement of<br>distance, angle<br>measurement, density<br>determination,<br>segmentation tool |
|                                     | siCAT GmbH & Co. KG Sicat implant (K090119)                                                             | Host platform                                  | PC                                                                                                                                               | PC                                                                                                                                               | PC                                                                                                                                                 |
| Operating system                    | Not specified                                                                                           | Not specified                                  | Windows XP, vista, 7                                                                                                                             |                                                                                                                                                  |                                                                                                                                                    |
| Output<br>compatibility             | CAD, rapid prototyping<br>machines                                                                      | CAD, rapid prototyping<br>machines             | CAD, rapid<br>prototyping machines                                                                                                               |                                                                                                                                                  |                                                                                                                                                    |
| Operating system<br>compatibility   | - Windows XP or vista<br>- Intel based MAC<br>(computer should be<br>booted from windows<br>not MAC OS) | - Windows XP<br>- Windows Vista<br>- WINDOWS 7 | - Windows XP<br>- Windows Vista<br>- WINDOWS 7                                                                                                   |                                                                                                                                                  |                                                                                                                                                    |

### Device Description

IPS is a Dental 3D Planning Software that aids dentists and dental lab technicians in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants. The data used for planning the treatment course is obtained from radiological devices (e.g., CT) complying with DICOM format. IPS provides dentists with various visualization and measurement tools which facilitate in generating dental panoramic view, segmenting of the dental anatomical structures and measuring bone density and distance between the dental implants.

#### Indications for Use

IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.

## Substantial Equivalence

Symbyo Dental LLC. wishes to use the following devices as a predicates: MATERIALISE N.V. Simplant system (K033849) siCAT GmbH & Co. KG Sicat implant (K090119)

The claim of substantial equivalence to the predicate devices is based on both intended use and technology used.

{1}------------------------------------------------

.

Predicate Comparison Table

{2}------------------------------------------------

## Based Upon Intended Used:

IPS and its predicate devices have intended use; which is Visualization and planning simulation that aids the user in the placement of dental implants and planning for dental surgical treatment.

#### Based Upon Technology Used:

IPS, like the predicate devices, is a pure software device with DICOM data conformance.

The technological characteristics of IPS vs. the predicate devices are summarized below.

Our software conforms to DICOM standard and it has been verified and validated for it.

Validation and verification was performed to demonstrate DICOM conformance

### Following test cases were used to validate DICOM conformance:

- Importing DICOM Data.

- Sorting DICOM data.
- Grouping multiple DICOM case

These test cases have been adopted to perform a validation process according to the IPS DICOM Conformance statement. These test cases are performed on the patient's DICOM images to ensure that the IPS application's modules are fully compliant with the DICOM conformance.

#### Conclusion

The results of all testing demonstrate that IPS does not raise any new issues of safety, effectiveness or performance when compared to the existing predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room – WO66-G609 Silver Spring, MD 20993-0002

Symbyo Dental, LLC. % Mr. Mark Job -- Official Correspondent 1394 25th Street NW Buffalo MN 94305

Re: K113404 Trade/Device Name: Dental 3D Planning Software IPS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 17, 2011 Received: November 18, 2011

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

DEC - 6 2011

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Device Name: Dental 3D Planning Software IPS

#### Indications for Use:

IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

.l.O.D.R.

ostic Device

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113404](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K113404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
