← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K111052

# SYNGO NEURO PBV-IR (K111052)

_Siemens Medical Solutions USA, Inc. · LLZ · May 20, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K111052

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** May 20, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format. Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis. This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)

## Device Story

Software application extending InSpace 3D; processes 2D angiographic C-arm X-ray images into 3D volumes. Uses contrast agent to enhance blood flow visualization; applies color-coded relative values to represent contrast-filled blood volume in brain vasculature. Used in interventional radiography suites by physicians; assists in diagnosing/treating vessel malformations (aneurysms, AVMs, stenoses). Provides CT-like perfusion visualization without patient transport. Output supports clinical decision-making for surgical planning and treatment follow-up.

## Clinical Evidence

Bench testing only. Hazard analysis and software verification/validation testing performed to ensure safety and effectiveness. No clinical data provided.

## Technological Characteristics

Software-only application; extension of InSpace 3D. Operates on angiographic C-arm hardware. Features post-processing algorithms for 3D reconstruction and color-coded blood volume visualization. Integrated into existing InSpace 3D task card interface.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- InSpace 3D Soft ([K011447](/device/K011447.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 2 0 2011

K1110S2

510(k) Summary For syngo Neuro PBV IR

Submitted by: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355

April 08, 2011

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

#### General Information 1.

Importer / Distributor Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, G-01 Malvern, PA 19355 Establishment Registration Number 2240869

# Manufacturing Site

SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany

### 2. Contact Person

Mr. Gary Johnson Sr. Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Vallev Stream Parkway G-01 Malvem, PA 19355-1406 Fax: (610) 448-1787 Phone: (610) 448 1778 Email: garyjohnson@siemens.com

### 3. Device Name and Classification

Trade Name: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:

syngo Neuro PBV IR Accessory to Angiographic X-Ray System Radiology 21 CFR §892.1600 Class II 90JAA

{1}------------------------------------------------

#### বঁ.
বাংলাদেশের বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্বকাপ বিশ্ববিদ্যালয় বিশ্বকাপ বিশ্ববিদ্যালয় বিশ্বকাপ করে বিশ্ববিদ্যালয় বিশ্বকাপ করে বিশ্ববিদ্যালয় বি Legally Marketed Device

| Trade Name:           | InSpace 3D Soft |
|-----------------------|-----------------|
| 510(k) Clearance      | K011447         |
| Clearance Date        | August 3, 2001  |
| Classification Name:  | Accessory to An |
| Classification Panel: | Radiology       |
| CFR Section:          | 21 CFR §892.16  |
| Device Class:         | Class II        |
| Product Code:         | 90JAA           |

ware Option giographic X-Ray System 00

#### ട്. Device Description

syngo Neuro-PBV IR is a post-processing software application designed to provide the physician with images similar to CT cerebral perfusion in the Interventional Radiography suite. The physician can use these images to visualize blood volume in the vasculature and blood-brain barrier without moving the patient to a CT system. A contrast agent is used to enhance the visualization of the blood flow. As with CT perfusion, the image is further enhanced using color as a reference of the amount of contrast filled blood in an area of the brain.

#### ર્ભ. Intended Use

Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format.

Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design fro the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis.

This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)

### 7. Substantial Equivalence

syngo Neuro-PBV IR Software Application is substantially equivalent to the commercially available Siemens ssoftware application, Inspace 3D. The Inspace 3D software option was described in premarket notification K011447 which received FDA Clearance on August 03, 2001.

The syngo Neuro-PBV IR software is an extension to InSpace 3D and uses the same hardware and software components as the InSpace 3D software option.

### Summary of Technological Characteristics of the Principal Device as Compared with 8. the Predicate Device

syngo Neuro PBV IR is a software extension to Inspace 3D. The principal device syngo Neuro PBV IR features the same Postprocessing software, user interface, archiving and

Siemens Medical Systems, Inc.

{2}------------------------------------------------

communication like the predicate Inspace 3D. The design of the syngo Neuro PBV IR is the same as it is integrated into the Inspace 3D task card.

#### 9. General Safety and Effectiveness Concerns

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.

#### 10. Conclusion as to Substantial Equivalence

syngo Neuro PBV IR is intended for the same indications for use as the predicate Inspace 3D. syngo Neuro PBV IR Software Application is a further post processing and enhanced visualization technique within Inspace 3D (K011447).

Most of the functionality remains the same as used with the predicate device Inspace 3D (K011447). The functionality is the same or similar to the predicate device with enhanced algorithm to display syngo Neuro PBV IR images.

It is Siemens opinion, that the syngo Neuro PBV IR is substantially equivalent to the Inspace 3D (K011447).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Gary L. Johnson Senior Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway G-01 MALVERN PA 19355-1406

MAY 20 2011

Re: K111052

Trade/Device Name: syngo Neuro-PBV IR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 9, 2011 Received: April 15, 2011

Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## SECTION 7

### INDICATIONS FOR USE

510(k) Number (if known): __

Device Name:

Indications for Use:

Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format.

Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis.

This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)

(Please do not write below this line - continue on another page if needed)

Concurrence of the CDRH, Office of Device Evaluation (ODE)

Prescription Use X (per 21 CFR 801.109)

OR Over-The-Counter Use _

Mary Shatt

Division Sign-Off Division of Radiological Devic Office of In Vitro Diagnostic Device Evaluation and Safety

510K K11052

Page 13 of 60 510(k) for Siemens 'syngo Neuro PBV IR'

Siemens Medical Systems, Inc.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K111052](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K111052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
