K111052 · Siemens Medical Solutions USA, Inc. · LLZ · May 20, 2011 · Radiology
Device Facts
Record ID
K111052
Device Name
SYNGO NEURO PBV-IR
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
May 20, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format. Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis. This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)
Device Story
Software application extending InSpace 3D; processes 2D angiographic C-arm X-ray images into 3D volumes. Uses contrast agent to enhance blood flow visualization; applies color-coded relative values to represent contrast-filled blood volume in brain vasculature. Used in interventional radiography suites by physicians; assists in diagnosing/treating vessel malformations (aneurysms, AVMs, stenoses). Provides CT-like perfusion visualization without patient transport. Output supports clinical decision-making for surgical planning and treatment follow-up.
Clinical Evidence
Bench testing only. Hazard analysis and software verification/validation testing performed to ensure safety and effectiveness. No clinical data provided.
Technological Characteristics
Software-only application; extension of InSpace 3D. Operates on angiographic C-arm hardware. Features post-processing algorithms for 3D reconstruction and color-coded blood volume visualization. Integrated into existing InSpace 3D task card interface.
Indications for Use
Indicated for imaging soft tissue in the head to visualize contrast-enhanced blood distribution in arterial and venous vessels. Assists physicians in diagnosis, surgical planning, interventional procedures, and treatment follow-up for vessel malformations including aneurysms, AVMs, and stenoses.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
MAY 2 0 2011
K1110S2
510(k) Summary For syngo Neuro PBV IR
Submitted by: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
April 08, 2011
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### General Information 1.
Importer / Distributor Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, G-01 Malvern, PA 19355 Establishment Registration Number 2240869
# Manufacturing Site
SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany
### 2. Contact Person
Mr. Gary Johnson Sr. Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Vallev Stream Parkway G-01 Malvem, PA 19355-1406 Fax: (610) 448-1787 Phone: (610) 448 1778 Email: garyjohnson@siemens.com
### 3. Device Name and Classification
Trade Name: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
syngo Neuro PBV IR Accessory to Angiographic X-Ray System Radiology 21 CFR §892.1600 Class II 90JAA
{1}------------------------------------------------
#### বঁ.
বাংলাদেশের বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্ববিদ্যালয় বিশ্বকাপ বিশ্ববিদ্যালয় বিশ্বকাপ বিশ্ববিদ্যালয় বিশ্বকাপ করে বিশ্ববিদ্যালয় বিশ্বকাপ করে বিশ্ববিদ্যালয় বি Legally Marketed Device
| Trade Name: | InSpace 3D Soft |
|-----------------------|-----------------|
| 510(k) Clearance | K011447 |
| Clearance Date | August 3, 2001 |
| Classification Name: | Accessory to An |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.16 |
| Device Class: | Class II |
| Product Code: | 90JAA |
ware Option giographic X-Ray System 00
#### ട്. Device Description
syngo Neuro-PBV IR is a post-processing software application designed to provide the physician with images similar to CT cerebral perfusion in the Interventional Radiography suite. The physician can use these images to visualize blood volume in the vasculature and blood-brain barrier without moving the patient to a CT system. A contrast agent is used to enhance the visualization of the blood flow. As with CT perfusion, the image is further enhanced using color as a reference of the amount of contrast filled blood in an area of the brain.
#### ર્ભ. Intended Use
Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format.
Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design fro the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis.
This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)
### 7. Substantial Equivalence
syngo Neuro-PBV IR Software Application is substantially equivalent to the commercially available Siemens ssoftware application, Inspace 3D. The Inspace 3D software option was described in premarket notification K011447 which received FDA Clearance on August 03, 2001.
The syngo Neuro-PBV IR software is an extension to InSpace 3D and uses the same hardware and software components as the InSpace 3D software option.
### Summary of Technological Characteristics of the Principal Device as Compared with 8. the Predicate Device
syngo Neuro PBV IR is a software extension to Inspace 3D. The principal device syngo Neuro PBV IR features the same Postprocessing software, user interface, archiving and
Siemens Medical Systems, Inc.
{2}------------------------------------------------
communication like the predicate Inspace 3D. The design of the syngo Neuro PBV IR is the same as it is integrated into the Inspace 3D task card.
#### 9. General Safety and Effectiveness Concerns
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
#### 10. Conclusion as to Substantial Equivalence
syngo Neuro PBV IR is intended for the same indications for use as the predicate Inspace 3D. syngo Neuro PBV IR Software Application is a further post processing and enhanced visualization technique within Inspace 3D (K011447).
Most of the functionality remains the same as used with the predicate device Inspace 3D (K011447). The functionality is the same or similar to the predicate device with enhanced algorithm to display syngo Neuro PBV IR images.
It is Siemens opinion, that the syngo Neuro PBV IR is substantially equivalent to the Inspace 3D (K011447).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Gary L. Johnson Senior Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway G-01 MALVERN PA 19355-1406
MAY 20 2011
Re: K111052
Trade/Device Name: syngo Neuro-PBV IR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 9, 2011 Received: April 15, 2011
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 7
### INDICATIONS FOR USE
510(k) Number (if known): __
Device Name:
Indications for Use:
Syngo Neuro-PBV IR is an extended software application to the InSpace 3D software option which allows the reconstruction of two-dimentional images acquired with a standard angiographic C-arm device into a three-dimentional image format.
Syngo Neuro-PBV-IR is intended for imaging primarily soft tissue for diagnosis, surgical planning, interventional procedures and treatment follow-up. It is design for the visualization of contrast enhanced blood distribution in the arterial and venous vessels in the head using color coded relative values for diagnosis.
This software is designed to visually assist physicians in the diagnosis and treatment of vessel malformations (i.e. Aneurysms, AVM's and Stenoses)
(Please do not write below this line - continue on another page if needed)
Concurrence of the CDRH, Office of Device Evaluation (ODE)
Prescription Use X (per 21 CFR 801.109)
OR Over-The-Counter Use _
Mary Shatt
Division Sign-Off Division of Radiological Devic Office of In Vitro Diagnostic Device Evaluation and Safety
510K K11052
Page 13 of 60 510(k) for Siemens 'syngo Neuro PBV IR'
Siemens Medical Systems, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.